<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article article-type="research-article" dtd-version="2.3" xml:lang="EN" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Physiol.</journal-id>
<journal-title>Frontiers in Physiology</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Physiol.</abbrev-journal-title>
<issn pub-type="epub">1664-042X</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="publisher-id">1248899</article-id>
<article-id pub-id-type="doi">10.3389/fphys.2023.1248899</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Physiology</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>PreEpiSeizures: description and outcomes of physiological data acquisition using wearable devices during video-EEG monitoring in people with epilepsy</article-title>
<alt-title alt-title-type="left-running-head">Abreu et al.</alt-title>
<alt-title alt-title-type="right-running-head">
<ext-link ext-link-type="uri" xlink:href="https://doi.org/10.3389/fphys.2023.1248899">10.3389/fphys.2023.1248899</ext-link>
</alt-title>
</title-group>
<contrib-group>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Abreu</surname>
<given-names>Mariana</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<xref ref-type="corresp" rid="c001">&#x2a;</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1702910/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Carmo</surname>
<given-names>Ana Sofia</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1601603/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Peralta</surname>
<given-names>Ana Rita</given-names>
</name>
<xref ref-type="aff" rid="aff3">
<sup>3</sup>
</xref>
</contrib>
<contrib contrib-type="author">
<name>
<surname>S&#xe1;</surname>
<given-names>Francisca</given-names>
</name>
<xref ref-type="aff" rid="aff4">
<sup>4</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/46052/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Pl&#xe1;cido da Silva</surname>
<given-names>Hugo</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<xref ref-type="aff" rid="aff5">
<sup>5</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/586182/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Bentes</surname>
<given-names>Carla</given-names>
</name>
<xref ref-type="aff" rid="aff3">
<sup>3</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1551422/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Fred</surname>
<given-names>Ana Lu&#xed;sa</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/305620/overview"/>
</contrib>
</contrib-group>
<aff id="aff1">
<sup>1</sup>
<institution>Instituto de Telecomunica&#xe7;&#xf5;es</institution>, <addr-line>Lisboa</addr-line>, <country>Portugal</country>
</aff>
<aff id="aff2">
<sup>2</sup>
<institution>Departamento de Bioengenharia</institution>, <institution>Instituto Superior T&#xe9;cnico</institution>, <institution>Universidade de Lisboa</institution>, <addr-line>Lisboa</addr-line>, <country>Portugal</country>
</aff>
<aff id="aff3">
<sup>3</sup>
<institution>Lab EEG-Sono</institution>, <institution>Centro Hospitalar Universit&#xe1;rio Lisboa Norte</institution>, <institution>Hospital de Santa Maria</institution>, <addr-line>Lisboa</addr-line>, <country>Portugal</country>
</aff>
<aff id="aff4">
<sup>4</sup>
<institution>Departamento Neurologia</institution>, <institution>Centro Hospitalar Lisboa Ocidental</institution>, <institution>Hospital Egas Moniz</institution>, <addr-line>Lisboa</addr-line>, <country>Portugal</country>
</aff>
<aff id="aff5">
<sup>5</sup>
<institution>Lisbon Unit for Learning and Intelligent Systems (LUMLIS), A Unit of the European Laboratory for Learning and Intelligent Systems (ELLIS)</institution>, <addr-line>Lisboa</addr-line>, <country>Portugal</country>
</aff>
<author-notes>
<fn fn-type="edited-by">
<p>
<bold>Edited by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/290344/overview">Gin&#xe9;s Viscor</ext-link>, University of Barcelona, Spain</p>
</fn>
<fn fn-type="edited-by">
<p>
<bold>Reviewed by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/2364246/overview">Petroula Laiou</ext-link>, King&#x2019;s College London, United Kingdom</p>
<p>
<ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/890163/overview">Cesar Alexandre Teixeira</ext-link>, University of Coimbra, Portugal</p>
</fn>
<corresp id="c001">&#x2a;Correspondence: Mariana Abreu, <email>mariana.abreu@tecnico.ulisboa.pt</email>
</corresp>
</author-notes>
<pub-date pub-type="epub">
<day>10</day>
<month>10</month>
<year>2023</year>
</pub-date>
<pub-date pub-type="collection">
<year>2023</year>
</pub-date>
<volume>14</volume>
<elocation-id>1248899</elocation-id>
<history>
<date date-type="received">
<day>27</day>
<month>06</month>
<year>2023</year>
</date>
<date date-type="accepted">
<day>04</day>
<month>09</month>
<year>2023</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#xa9; 2023 Abreu, Carmo, Peralta, S&#xe1;, Pl&#xe1;cido da Silva, Bentes and Fred.</copyright-statement>
<copyright-year>2023</copyright-year>
<copyright-holder>Abreu, Carmo, Peralta, S&#xe1;, Pl&#xe1;cido da Silva, Bentes and Fred</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/">
<p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p>
</license>
</permissions>
<abstract>
<p>The PreEpiSeizures project was created to better understand epilepsy and seizures through wearable technologies. The motivation was to capture physiological information related to epileptic seizures, besides Electroencephalography (EEG) during video-EEG monitorings. If other physiological signals have reliable information of epileptic seizures, unobtrusive wearable technology could be used to monitor epilepsy in daily life. The development of wearable solutions for epilepsy is limited by the nonexistence of datasets which could validate these solutions. Three different form factors were developed and deployed, and the signal quality was assessed for all acquired biosignals. The wearable data acquisition was performed during the video-EEG of patients with epilepsy. The results achieved so far include 59 patients from 2 hospitals totaling 2,721 h of wearable data and 348 seizures. Besides the wearable data, the Electrocardiogram of the hospital is also useable, totalling 5,838 h of hospital data. The quality ECG signals collected with the proposed wearable is equated with the hospital system, and all other biosignals also achieved state-of-the-art quality. During the data acquisition, 18 challenges were identified, and are presented alongside their possible solutions. Though this is an ongoing work, there were many lessons learned which could help to predict possible problems in wearable data collections and also contribute to the epilepsy community with new physiological information. This work contributes with original wearable data and results relevant to epilepsy research, and discusses relevant challenges that impact wearable health monitoring.</p>
</abstract>
<kwd-group>
<kwd>wearable devices</kwd>
<kwd>physiological signals</kwd>
<kwd>protocol design</kwd>
<kwd>epilepsy dataset</kwd>
<kwd>epilepsy monitoring</kwd>
<kwd>cardiorespiratory function</kwd>
<kwd>physiological data acquisition</kwd>
</kwd-group>
<custom-meta-wrap>
<custom-meta>
<meta-name>section-at-acceptance</meta-name>
<meta-value>Integrative Physiology</meta-value>
</custom-meta>
</custom-meta-wrap>
</article-meta>
</front>
<body>
<sec id="s1">
<title>1 Introduction</title>
<p>Over the past years, research in epilepsy has been integrating more physiological information (i.e. biosignals) besides the electrical brain activity, captured by the Electroencephalography (EEG) (<xref ref-type="bibr" rid="B35">Ranjan et al., 2019</xref>). In fact, some devices with non-EEG biosignals have already been approved as medical devices by the Federal Drugs Administration (FDA), for their ability to detect of Generalised Tonic-clonic Seizures (GTCS) (<xref ref-type="bibr" rid="B7">Bruno et al., 2020</xref>). Examples include the Embrace Smartwatch by Empatica (<xref ref-type="bibr" rid="B27">McCarthy et al., 2016</xref>; <xref ref-type="bibr" rid="B36">Regalia et al., 2019</xref>), or the SPEAC device by Brain Sentinel (<xref ref-type="bibr" rid="B52">Whitmire et al., 2019</xref>). Additionally, non-EEG biosignals are also being used in epilepsy research for ambulatory daily monitoring of people with epilepsy (PWE), and for the development of seizure forecast and prediction models, where the most frequently mentioned device is the smartwatch E4 by Empatica (<xref ref-type="bibr" rid="B28">Meisel et al., 2020</xref>; <xref ref-type="bibr" rid="B6">Brinkmann et al., 2021</xref>; <xref ref-type="bibr" rid="B49">Vieluf et al., 2022</xref>). One essential aspect for the development of end-user solutions is the access to large datasets of multimodal quality data. This has been the focus of multiple international initiatives, such as the Remote Assessment of Disease and Relapse-Central Nervous System (RADAR-CNS), an initiative focused on ambulatory data collection is based on an open-source platform (RADAR-Base) to integrate data from different sources including smartphone and wearables (<xref ref-type="bibr" rid="B35">Ranjan et al., 2019</xref>).</p>
<p>The use of biosignals in epilepsy is especially relevant for the prevention of Sudden Unexpected Death in Epilepsy (SUDEP) (<xref ref-type="bibr" rid="B46">Van de Vel et al., 2013</xref>), which is thought to be a consequence of severe cardio-respiratory dysfunctions (<xref ref-type="bibr" rid="B40">Sivathamboo et al., 2020</xref>; <xref ref-type="bibr" rid="B38">2021</xref>). With this in mind, our research group has given special focus to peripheral physiological signals which have shown to be of interest for these pathological changes, such as Electrocardiography (ECG), Respiration (RESP) and Chest Motion.</p>
<p>The PreEpiSeizures project was motivated by the collection of multimodal non-EEG data from PWE through wearables, in order to be leveraged towards the development of methods for early seizure detection, as well as expanding the understanding of epilepsy in different lenses. Although the use of biosignals in epilepsy is common, their contribution to different epilepsy types is still yet to prove, thus motivating the collection of comprehensive data to sustain and improve non-EEG based solutions. The rest of this document is structured as follows: <xref ref-type="sec" rid="s2">Section 2</xref> provides the current landscape of epilepsy-related data collection as well as commercial wearables which serve that purpose; <xref ref-type="sec" rid="s3">Section 3</xref> describes the experimental protocol employed in this project, along with the characteristics of the wearable devices used, together with the data quality assessment metrics; <xref ref-type="sec" rid="s4">Section 4</xref> details the main findings extracted from the data collected so far, including the analysis of the signal quality and patients&#x2019; metadata; <xref ref-type="sec" rid="s5">Section 5</xref> overviews several challenges faced in the context of wearable data collection, possible solutions, and presents a discussion of the results achieved. Lastly, <xref ref-type="sec" rid="s6">Section 6</xref> summarises the main outcomes and conclusions, followed by some future work directions for the PreEpiSeizures project.</p>
</sec>
<sec id="s2">
<title>2 Background</title>
<sec id="s2-1">
<title>2.1 Datasets in epilepsy</title>
<p>Throughout the years, the epilepsy research community has been working in multi-centre collaborations. The joint efforts of stakeholders from different backgrounds, such as industry, healthcare and academia, as well as different countries, is essential to provide better and faster solutions to epilepsy. Some of the collaborations found were <ext-link ext-link-type="uri" xlink:href="http://iEEG.org">iEEG.org</ext-link>, SeizeIT, Epilepsy Ecosystem and EPILEPSIAE. Additionally, other publicly available datasets were found in the literature, which also comprise data from epileptic patients. The description of these datasets is summarised in <xref ref-type="table" rid="T1">Table 1</xref> alongside the one proposed in this project. This includes the number of patients, their age, the modalities acquired, the number of seizures, the seizure types, the total duration and the setting.</p>
<table-wrap id="T1" position="float">
<label>TABLE 1</label>
<caption>
<p>Comparison of public datasets with PreEpiSeizures, regarding number of patients, age, signal modalities, number of seizures, seizure types and total number of recording hours, and monitoring settings. If the number of seizures also includes subclinical, the number of subclinical seizures is specified in brackets. The numbers in brackets in the patients and total hours for the PreEpiSeizures are related to the wearable data, where the leftside numbers correspond to the hospital data. (&#x2a; The 8 seizure types in the TUH EEG Corpus are: focal non specific seizure, generalised non-specific seizure, FAS, FIAS, absence, myoclonic, tonic and tonic-clonic seizures.)</p>
</caption>
<table>
<thead valign="top">
<tr>
<th align="left">Dataset name</th>
<th align="center">&#x23; Patients</th>
<th align="center">Age</th>
<th align="center">Modalities</th>
<th align="center">&#x23; Seizures</th>
<th align="center">Seizure types</th>
<th align="center">Total hours</th>
<th align="center">Setting</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">PreEpiSeizures</td>
<td align="center">59 (37)</td>
<td align="center">[16,60]</td>
<td align="center">ECG, Resp, ACC, EDA, PPG, EMG</td>
<td align="center">348 (89)</td>
<td align="center">FAS, FIAS, FUAS, FBTCS, E</td>
<td align="center">5,838 (2,721)</td>
<td align="center">Video-EEG</td>
</tr>
<tr>
<td align="left">My Seizure Gauge Wearable (footnote 4)</td>
<td align="center">27</td>
<td align="center">-</td>
<td align="center">EDA, PPG, ACC, EMG, GYR, EEG</td>
<td align="center">226 (85)</td>
<td align="center">GTCS, FIAS, FUAS, FAS, E</td>
<td align="center">2640</td>
<td align="center">Unclear</td>
</tr>
<tr>
<td align="left">Siena Scalp EEG <xref ref-type="bibr" rid="B12">Detti (2020)</xref>
</td>
<td align="center">14</td>
<td align="center">44 &#xb1; 14</td>
<td align="center">EEG, ECG</td>
<td align="center">47</td>
<td align="center">FAS, FIAS and FBTCS</td>
<td align="center">128</td>
<td align="center">Hospital long-term</td>
</tr>
<tr>
<td align="left">EPILEPSIAE <xref ref-type="bibr" rid="B20">Klatt et al. (2012)</xref>
</td>
<td align="center">30</td>
<td align="center">41 &#xb1; 14</td>
<td align="center">EEG, ECG</td>
<td align="center">276</td>
<td align="center">-</td>
<td align="center">2881</td>
<td align="center">Hospital long-term</td>
</tr>
<tr>
<td align="left">NeuroVista Trial <xref ref-type="bibr" rid="B10">Cook et al. (2013)</xref>
</td>
<td align="center">15</td>
<td align="center">44 &#xb1; 13</td>
<td align="center">iEEG</td>
<td align="center">2&#x2013;12 monthly</td>
<td align="center">Focal</td>
<td align="center">
<inline-formula id="inf1">
<mml:math id="m1">
<mml:mo>&#x3e;</mml:mo>
<mml:mn>1</mml:mn>
<mml:mi>y</mml:mi>
</mml:math>
</inline-formula>
</td>
<td align="center">Ambulatory</td>
</tr>
<tr>
<td align="left">Bern-Barcelona <xref ref-type="bibr" rid="B3">Andrzejak et al. (2012)</xref>
</td>
<td align="center">5</td>
<td align="center">-</td>
<td align="center">EEG and iEEG</td>
<td align="center">-</td>
<td align="center">-</td>
<td align="center">100 &#xd7; 23.6s</td>
<td align="center">Hospital</td>
</tr>
<tr>
<td align="left">Bonn-Barcelona micro-, and macro- EEG database <xref ref-type="bibr" rid="B26">Mart&#xed;nez et al. (2020)</xref>
</td>
<td align="center">3</td>
<td align="center">[20,39,47]</td>
<td align="center">iEEG</td>
<td align="center">0</td>
<td align="center">-</td>
<td align="center">66</td>
<td align="center">Hospital long-term</td>
</tr>
<tr>
<td align="left">TUH EEG Corpus <xref ref-type="bibr" rid="B32">Obeid and Picone (2016)</xref>
</td>
<td align="center">
<inline-formula id="inf2">
<mml:math id="m2">
<mml:mo>&#x3e;</mml:mo>
<mml:mn>10</mml:mn>
<mml:mi>k</mml:mi>
</mml:math>
</inline-formula>
</td>
<td align="center">51 &#xb1; 56</td>
<td align="center">EEG</td>
<td align="center">
<inline-formula id="inf3">
<mml:math id="m3">
<mml:mo>&#x2265;</mml:mo>
<mml:mn>3050</mml:mn>
</mml:math>
</inline-formula>
</td>
<td align="center">8 seizure types&#x2a;</td>
<td align="center">29.1<italic>y</italic>
</td>
<td align="center">Hospital long and short-term</td>
</tr>
<tr>
<td align="left">New Delhi EEG <xref ref-type="bibr" rid="B42">Swami et al. (2016)</xref>
</td>
<td align="center">10</td>
<td align="center">-</td>
<td align="center">EEG</td>
<td align="center">-</td>
<td align="center">-</td>
<td align="center">10 &#xd7; 50 &#xd7; 5.12s</td>
<td align="center">Hospital</td>
</tr>
<tr>
<td align="left">Epilepsy Seizure Recognition <xref ref-type="bibr" rid="B2">Andrzejak et al. (2001)</xref>
</td>
<td align="center">500</td>
<td align="center">-</td>
<td align="center">EEG</td>
<td align="center">-</td>
<td align="center">-</td>
<td align="center">500 &#xd7; 23.6s</td>
<td align="center">Hospital short-term</td>
</tr>
<tr>
<td align="left">CHB MIT <xref ref-type="bibr" rid="B37">Shoeb (2009)</xref>
</td>
<td align="center">23</td>
<td align="center">10 &#xb1; 5.6</td>
<td align="center">EEG</td>
<td align="center">198</td>
<td align="center">Not mentioned</td>
<td align="center">969</td>
<td align="center">Hospital long-term</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>The <italic>iEEG.org</italic> online platform<xref ref-type="fn" rid="fn1">
<sup>1</sup>
</xref>, supported by the National Institutes of Neurological Disorders and Stroke, provides access to over 900 datasets on human and animal epilepsy, as well as analytic and visualisation tools to aid data exploration. By inserting the corresponding keywords, the datasets are filtered to a desired EEG montage, population and seizure types, just to name a few. Nonetheless, the source of the signal is always assumed to be EEG and other physiological signals are not an option when filtering.</p>
<p>SeizeIT<xref ref-type="fn" rid="fn2">
<sup>2</sup>
</xref> is another collaborative project within epilepsy, which is supported by the European Institute of Innovation and Technology (EIT Health). This consortium aims to tackle the continuous ambulatory monitoring of epilepsy, in order to accurately register seizures, thereby enabling the optimisation of treatment plans.</p>
<p>The Epilepsy Ecosystem<xref ref-type="fn" rid="fn3">
<sup>3</sup>
</xref> (<xref ref-type="bibr" rid="B23">Kuhlmann et al., 2018</xref>) is a crowd-sourcing environment supported through the My Seizure Gauge project, supported by the Epilepsy Foundation of America, the Aikenhead Centre for Medical Discovery at St. Vincent&#x2019;s Hospital Melbourne, the University of Melbourne, Monash University and Seer Medical. Its goal is to improve the performance of seizure prediction algorithms with the purpose of making them a viable option for PWE. They are responsible for gathering and maintaining two public datasets: NeuroVista Trial (<xref ref-type="bibr" rid="B10">Cook et al., 2013</xref>) and My Seizure Gauge Wearable<xref ref-type="fn" rid="fn4">
<sup>4</sup>
</xref>. Both datasets are available in the Seer platform requiring only a login setup. The NeuroVista trial (<xref ref-type="bibr" rid="B10">Cook et al., 2013</xref>) contains 15 patients that were submitted to an ambulatory intracranial EEG (iEEG) monitoring for more than 80 days. This dataset consists of a set of carefully selected patients whom experienced 2 to 12 monthly seizures, while still having a level of independence compatible with using the device for daily management. Another dataset made available in the Seer platform is the My Seizure Gauge Wearable, which is a collection of wearable data using 3 wearable devices: Empatica E4, Epilog (<xref ref-type="bibr" rid="B16">Frankel et al., 2021</xref>) and Byteflies (<xref ref-type="bibr" rid="B43">Swinnen et al., 2021</xref>). It comprises 27 patients that were monitored for approximately 4.4 &#xb1; 2.4 days. Although this dataset provides unprecedented wearable data in the context of epilepsy, it does not provide any information regarding the positioning of the devices nor some relevant clinical symptoms such as sleep/vigilance state and loss of awareness during the seizure events. Since no information was found regarding the acquisition conditions, we could not assume whether this dataset was acquired during video-EEG or in ambulatory.</p>
<p>The EPILEPSIAE database (<xref ref-type="bibr" rid="B20">Klatt et al., 2012</xref>) was created under a joint European project between Epilepsy Center, University Hospital of Freiburg; Center for Informatics and Systems, University of Coimbra; and Epilepsy Unit, CHU Pitie-Salpetriere, Paris. This database contains more than 275 patients that performed Video-EEG monitoring, focusing on the EEG signal. Still, since the ECG is also acquired by the hospital system, it is also present in this database. Several works have used this database both for EEG and ECG analysis (<xref ref-type="bibr" rid="B45">Truong et al., 2019</xref>; <xref ref-type="bibr" rid="B17">G&#xf3;mez et al., 2020</xref>; <xref ref-type="bibr" rid="B24">Leal et al., 2021</xref>). However, EPILEPSIAE only makes available 30 patients for external researchers, which is always the same subset, without disclosing the type of seizures, or the clinical semiology. Moreover, it does not provide wearable data, only hospital data.</p>
<p>The Siena Scalp EEG Epilepsy Database (<xref ref-type="bibr" rid="B13">Detti et al., 2020</xref>) contains the data of 14 patients that performed the video-EEG monitoring in the Unit of Neurology and Neurophysiology of the University of Siena. This dataset is curated to the seizure episodes, hence for some patients, only the peri-ictal period is available to the public. The signal sources present in this database include EEG and ECG, both acquired by the hospital system.</p>
<p>Other epilepsy datasets also available in the literature were included in <xref ref-type="table" rid="T1">Table 1</xref> for dimensionality comparison, however, they are out of the scope of this research, since they provide neither non-EEG data or wearable data. For more information in EEG datasets (<xref ref-type="bibr" rid="B53">Wong et al., 2023</xref>), provides a comprehensive review.</p>
<p>The majority of available datasets is focused on EEG analysis, and only a few also include the ECG signal acquired by the hospital system. Regarding the acquisition settings, four datasets contain data from long-term or video-EEG monitorings, and four datasets contain short-term EEG acquisitions. The EEG was intracranial in three datasets (indicated with iEEG), and the NeuroVista Trial is the only ambulatory dataset, to the best of our knowledge. The only multimodal wearable dataset found was My Seizure Gauge Wearable data, however it does not provide clear information on seizure types, the acquisition conditions, or the clinical manifestations of seizures. Most events annotated in this dataset have little seizure information, and for some of them it is unclear if they are even seizures or not. Larger datasets are needed to sustain the use of non-EEG data as lens for epileptic activity. Recently, projects such as My Seizure Gauge and Radar-CNS started to create multicenter datasets with hundreds of patients acquired in ambulatory scenarios (<xref ref-type="bibr" rid="B5">B&#xf6;ttcher et al., 2022</xref>). However, it was not yet proven the ability of these devices to capture pathophysiological manifestations in non-EEG modalities related to non-convulsive seizures. Hence more evidence based on data collected in a semi-controlled scenario (e.g. as the video-EEG monitoring), is essential towards the use of non-EEG physiological modalities in the prediction and forecast of epileptic seizures, and in the detection of non-convulsive epileptic seizures.</p>
</sec>
<sec id="s2-2">
<title>2.2 Peripheral multimodal devices</title>
<p>The ever-decreasing size of devices with built-in sensors, internal battery and powerful processors, allow continuous wearable long-term monitoring of movement patterns and physiological variables. A wearable device similar to a smartwatch can be decomposed in four main layers: top cover, electronics board, lithium battery and bottom cover<xref ref-type="fn" rid="fn5">
<sup>5</sup>
</xref>. The top cover serves as a protective shield for the other components; the electronics board contains sensors, communication antenna (Bluetooth or WiFi), memory, and MCU; the lithium battery provides a power supply; finally the bottom cover often contains the interface with the patient when necessary (e.g., dry electrodes). This example of a wearable device unfolds the physical constraint associated with each layer: on the one hand, multiple embedded sensors lead to more information, and larger battery sizes lead to longer continuous acquisitions; on the other hand, devices should be small and unobtrusive. Thus, there is a trade-off between the device&#x2019;s size and its modus operandi.</p>
<p>The devices that can simultaneously record data from multiple sensors are more interesting for epilepsy monitoring since they can increase the chances of capturing manifestations from different seizure types, thus enhancing the performance of detection or early prediction algorithms (<xref ref-type="bibr" rid="B48">Verdru and Van Paesschen, 2020</xref>). Some devices are commercially available for the purpose of physiological signal acquisition, which could be used for research purposes. In <xref ref-type="table" rid="T2">Table 2</xref>, relevant devices found in the literature are described, which were validated and can record multimodal physiological signals. In this table, the devices are compared based on their target use, validation, modalities available, body location, battery duration, sampling frequency (FS) and data accessibility. For comparison, the bottom part of <xref ref-type="table" rid="T2">Table 2</xref>, contains the characteristics of the devices used in this project.</p>
<table-wrap id="T2" position="float">
<label>TABLE 2</label>
<caption>
<p>Devices with multiple sensors, characterised with respect to their applicability to research, validation status, form factor or body positioning, battery duration, method to access the data, measured signals and sampling frequency.</p>
</caption>
<table>
<thead valign="top">
<tr>
<th align="center">Product</th>
<th align="center">Research use</th>
<th align="center">Validation</th>
<th align="center">Modalities</th>
<th align="center">Site</th>
<th align="center">Battery</th>
<th align="center">FS(Hz)</th>
<th align="center">Data accessibility</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="center">EQ02<xref ref-type="table-fn" rid="Tfn9">
<sup>9</sup>
</xref>
</td>
<td align="center">Ambulatory <xref ref-type="bibr" rid="B34">Orphanidou and Drobnjak (2017)</xref>, Athletes <xref ref-type="bibr" rid="B19">Kenny et al. (2017)</xref>
</td>
<td align="center">With Holter <xref ref-type="bibr" rid="B1">Akintola et al. (2016)</xref>
</td>
<td align="center">ECG, Resp, Temp, ACC</td>
<td align="center">Vest</td>
<td align="center">48 h</td>
<td align="center">256</td>
<td align="center">BT transmission</td>
</tr>
<tr>
<td align="center">E4 footnote 5</td>
<td align="center">Migraines <xref ref-type="bibr" rid="B22">Koskim&#xe4;ki et al. (2017)</xref>, Stress <xref ref-type="bibr" rid="B33">Ollander et al. (2016)</xref>
</td>
<td align="center">CE <xref ref-type="bibr" rid="B36">Regalia et al. (2019)</xref>
</td>
<td align="center">EDA, PPG, Temp, ACC</td>
<td align="center">Bracelet</td>
<td align="center">&#x2b;32 h</td>
<td align="center">128</td>
<td align="center">BLE to Cloud Storage</td>
</tr>
<tr>
<td align="center">VisiMobile<xref ref-type="table-fn" rid="Tfn10">
<sup>10</sup>
</xref>
</td>
<td align="center">Ambulatory <xref ref-type="bibr" rid="B51">Weenk et al. (2017)</xref>
</td>
<td align="center">FDA</td>
<td align="center">ECG, Resp, SpO<sub>2</sub>, Temp</td>
<td align="center">Wristband</td>
<td align="center">&#x2b;12 h</td>
<td align="center">-</td>
<td align="center">WiFi transmission</td>
</tr>
<tr>
<td align="center">Bioharness3 <xref ref-type="bibr" rid="B18">Johnstone et al. (2012)</xref>
</td>
<td align="center">Sports</td>
<td align="center">HR validation <xref ref-type="bibr" rid="B31">Nepi et al. (2016)</xref>
</td>
<td align="center">ECG, Resp, Temp, ACC</td>
<td align="center">Chest Strap</td>
<td align="center">12&#x2013;24 h</td>
<td align="center">-</td>
<td align="center">BLE transmission</td>
</tr>
<tr>
<td align="center">Hexoskin<xref ref-type="table-fn" rid="Tfn11">
<sup>11</sup>
</xref>
</td>
<td align="center">Remote</td>
<td align="center">With gold-standard <xref ref-type="bibr" rid="B41">Smith et al. (2019)</xref>
</td>
<td align="center">ECG, Resp, ACC</td>
<td align="center">Smart Shirt</td>
<td align="center">12 h</td>
<td align="center">256</td>
<td align="center">Cloud Storage or BLE</td>
</tr>
<tr>
<td align="center">Vivosmart4<xref ref-type="table-fn" rid="Tfn12">
<sup>12</sup>
</xref>
</td>
<td align="center">Stress</td>
<td align="center">With gold-standard <xref ref-type="bibr" rid="B44">Tedesco et al. (2019)</xref>
</td>
<td align="center">Optical HRV, SpO<sub>2</sub>, Light, ACC</td>
<td align="center">Smartwatch</td>
<td align="center">5 days</td>
<td align="center">-</td>
<td align="center">BLE transmission</td>
</tr>
<tr>
<td align="center">Byteflies<xref ref-type="table-fn" rid="Tfn13">
<sup>13</sup>
</xref>
</td>
<td align="center">Epilepsy</td>
<td align="center">CE</td>
<td align="center">EEG, ACC, Resp, HR</td>
<td align="center">Behind Ear</td>
<td align="center">24 h</td>
<td align="center">250</td>
<td align="center">Cloud Storage</td>
</tr>
<tr>
<td align="center">ArmBIT &#x2b; ForearmBIT</td>
<td align="left"/>
<td align="left"/>
<td align="center">ECG, EDA, EMG, PPG, ACC</td>
<td align="center">Arm &#x2b; Wrist</td>
<td align="center">12 h</td>
<td align="center">1,000</td>
<td align="center">BT transmission</td>
</tr>
<tr>
<td align="center">WristBIT</td>
<td align="left"/>
<td align="left"/>
<td align="center">EDA, PPG, ACC</td>
<td align="center">Wrist</td>
<td align="center">24 h</td>
<td align="center">1,000</td>
<td align="center">BT transmission</td>
</tr>
<tr>
<td align="center">ChestBIT</td>
<td align="left"/>
<td align="left"/>
<td align="center">ECG, PZT, ACC</td>
<td align="center">Chest</td>
<td align="center">24 h</td>
<td align="center">1,000</td>
<td align="center">BLE transmission</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<fn id="Tfn9">
<label>9</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://equivital.com/">https://equivital.com/</ext-link>
</p>
</fn>
<fn id="Tfn10">
<label>10</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://soteradigitalhealth.com/">https://soteradigitalhealth.com/</ext-link>
</p>
</fn>
<fn id="Tfn11">
<label>11</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.hexoskin.com/">https://www.hexoskin.com/</ext-link>
</p>
</fn>
<fn id="Tfn12">
<label>12</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.garmin.com/en-US/p/605739">https://www.garmin.com/en-US/p/605739</ext-link>
</p>
</fn>
<fn id="Tfn13">
<label>13</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://byteflies.com/">https://byteflies.com/</ext-link>
</p>
</fn>
</table-wrap-foot>
</table-wrap>
<p>
<xref ref-type="table" rid="T2">Table 2</xref> illustrates the characteristics that prove to be the most relevant in wearable devices applied to epilepsy. However, it also uncovers some limitations of current options: from restrictive form factors, to short battery life, lack of validation, limited storage options, or insufficient physiological information. In fact, this last point accurately illustrates one pressing issue in epilepsy research: the extent of the applicability of seizure detection devices is inherently related to their location/signals they acquire, for example, an ACC- or EMG-based device will not be able to detect non-motor seizures. Even though the research community&#x2019;s intention is in detecting all seizure types, there is an imbalance, in both research and evidence, towards an overrepresentation of convulsive motor seizures (<xref ref-type="bibr" rid="B7">Bruno et al., 2020</xref>). Therefore, research efforts should be endorsed for other seizure types. Moreover, additional research is needed towards the applicability of multimodal wearables in epilepsy in ambulatory scenarios, especially combined with seizure risk forecasting.</p>
<p>The benchmark device for epilepsy research is the Empatica E4 smartwatch, since it is discreet, it captures four biosignals and its battery lasts for 32 h. However, the data management is performed in a cloud-based platform, which compromises our privacy protection clauses. Moreover, (<xref ref-type="bibr" rid="B47">Vandecasteele et al., 2017</xref>), have reported the higher reliability of ECG, comparatively to PPG, to obtain the heart rate and heart rate variability, which are considered significant autonomic change predictors. These two factors motivated the development of our own devices, designed to ensure the privacy preservation and the recording of relevant physiological modalities, such as the ECG, PZT, PPG, EDA and ACC (for chest motion).</p>
</sec>
</sec>
<sec sec-type="methods" id="s3">
<title>3 Methodology</title>
<p>Data collection within the PreEpiSeizures project started as a collaboration from Instituto Superior T&#xe9;cnico (IST) with the EEG-Sleep Lab of Hospital de Santa Maria - Centro Hospitalar Universit&#xe1;rio Lisboa Norte (CHULN), that is responsible for supervising video-EEG monitorings. This project was approved in 2015 by the ethics committee of CHULN. This collaboration was later extended to Centro de Epilepsia Refrat&#x00E1;ria of Hospital Egas Moniz&#x2014;Centro Hospitalar Lisboa Ocidental towards continuous monitoring in the scope of the PreEpiSeizures project underwent multiple iterations with the purpose of achieving a friendlier and more autonomous solution for patients. The approach is divided into the development of the wearable device and in the acquisition system for data collection and storage.</p>
<sec id="s3-1">
<title>3.1 Wearable devices</title>
<p>The devices developed in this scope are based on BITalino (<xref ref-type="bibr" rid="B11">Da Silva et al., 2014</xref>), which allowed for the experimentation around form-factors and sensors modalities. The BITalino device consists of a core block with a microprocessor (MCU) and a Bluetooth (BT) module for communication, sensors for each physiological modality, and a lithium-ion polymer (LiPo) battery. BITalino has the ability of recording simultaneously up to 6 analog channels (with a FS up to 1000Hz). When all channels are being acquired, the first four channels also allow for 10-bit resolution, while the last two only record with 6-bit resolution.</p>
<p>The first prototypes developed within the PreEpiSeizures project were ArmBIT and ForearmBIT (shown in <xref ref-type="fig" rid="F1">Figure 1</xref>), both consisting of one BITalino core block (with MCU and BT module), and respective sensors, all enclosed within a 3D printed case with an adjustable strap. The ArmBIT, which is displayed in left side of <xref ref-type="fig" rid="F1">Figure 1</xref>, collected the equivalent of lead I ECG and the triaxial ACC of the upper arm. The ForearmBIT, represented in right side of <xref ref-type="fig" rid="F1">Figure 1</xref>, was design to record the bicep&#x2019;s EMG, EDA on the palm of the hand, fingertip&#x2019;s PPG, and forearm&#x2019;s triaxial ACC. Interface with the patient was performed with the use of pre-gelled Ag/AgCl electrodes. These two devices were used simultaneously and were connected by a cable, to enable synchronisation (not shown in <xref ref-type="fig" rid="F1">Figure 1</xref>). Although this montage is acceptable for short-term data acquisitions, it becomes impractical for longer sessions. This is due to the prolonged use of pre-gelled electrodes and connection cables, which can cause discomfort, associated with the fact that wearables are secured to the arm, inhibiting patients from adopting certain sleeping positions. Since these two devices are both on the arm and physically connected by a cable, the remainder of this work uses ArmBIT to refer to the ensemble with both devices.</p>
<fig id="F1" position="float">
<label>FIGURE 1</label>
<caption>
<p>Prototypes of ArmBIT <bold>(A)</bold> and ForearmBIT <bold>(B)</bold>, the first wearable devices used to record multimodal data during video-EEG acquisitions.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g001.tif"/>
</fig>
<p>Given the higher resolution in the first four channels, these were usually used for the more relevant physiological signals (e.g., ECG, EEG, EMG and PPG). Since the ACC possesses three axis, and the human motion has frequencies lower than 5 Hz (<xref ref-type="bibr" rid="B21">Korohoda et al., 2013</xref>), two of its axis were usually acquired in the channels with lower resolution.</p>
<p>The second iteration of the PreEpiSeizures wearable devices, consisted on rejecting gel electrodes and finding other ways to collect the same signals. The ForearmBIT was replaced by a BITalino assembled in a wristband (WristBIT), whereas the ArmBIT was replaced by a BITalino mounted in a chestband (ChestBIT). The WristBIT device (shown in <xref ref-type="fig" rid="F2">Figure 2</xref>), is composed of an elastic wristband with a pocket to enclose the BITalino core block, as well as EDA and ACC sensors. The EDA signal was being recorded on the wrist, using two metallic electrodes encapsulated in the strechable fabric. Additionally, the finger strap enclosing the PPG sensor was connected to the wristband.</p>
<fig id="F2" position="float">
<label>FIGURE 2</label>
<caption>
<p>WristBIT prototype used in the PreEpiSeizures project. In the <bold>(A)</bold>, the WristBIT is being worn in the left hand, where the EDA and ACC are recorded in the wrist and the PPG in the finger. The <bold>(B)</bold> shows the interior of the wristband, with the BITalino main unit attached to the Bluetooth module and sensors.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g002.tif"/>
</fig>
<p>The ChestBIT is shown in <xref ref-type="fig" rid="F3">Figure 3</xref>, and it was previously applied in other contexts within the research group (<xref ref-type="bibr" rid="B25">Manso et al., 2018</xref>). The chestband used was a Polar chest strap<xref ref-type="fn" rid="fn6">
<sup>6</sup>
</xref>, with the ECG sensor connected to the metal buttons of the chest strap. The BITalino MCU was glued to the chest strap, alongside the ACC, the Bluetooth module, and a piezoelectric PZT sensor used to record the RESP signal. The entire electronics was protected with a soft sponge and covered with a synthetic leather textile, to improve endurance. The battery remained outside, in a pocket, and a small opening allowed the battery connector to reach the BITalino power module; this enabled easier natter charging and replacement to ensure near-continuous operation. The on/off button was accessible through the side, since the left side of the textile was not fully closed.</p>
<fig id="F3" position="float">
<label>FIGURE 3</label>
<caption>
<p>Current prototype of ChestBIT in the PreEpiSeizures project. On the <bold>(A,B)</bold> the ChestBIT components are shown, namely, the BItalino mini, and the mounted sensors ECG and ACC. On the <bold>(C)</bold> the connection between the battery and the BITalino Mini is shown, and in the <bold>(D)</bold>, the components are covered by a comfortable textile.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g003.tif"/>
</fig>
<p>To ensure greater battery life, the batteries used in ArmBIT and ForearmBIT were LiPo with 750 mAh and lasted for 12 h, which required a bi-diary battery exchange. However, the new wearables ChestBIT and WristBIT reduce the physical constrain of battery size, reason for which LiPo batteries of 1,600 mAh were used, requiring a single daily battery change.</p>
</sec>
<sec id="s3-2">
<title>3.2 Data quality assessment</title>
<p>The quality of all data recorded was assessed both for the hospital system and for the wearable devices, following the approach found in (<xref ref-type="bibr" rid="B5">B&#xf6;ttcher et al., 2022</xref>) for data quality evaluation. In this approach, the quality of the acquired data is assessed globally for each dataset, through statistical metrics (mean, median, standard deviation, minimum value and maximum value). These metrics are applied to the patient data variables, namely: duration, data completeness, and each biosignal quality.</p>
<p>The duration variable consists on the time span from the first timestamp marking the beginning of the acquisition until the last timestamp. It is presented in hours, minutes and seconds (H:M:S). The percentage of usable data (i.e., not loss) is measured by the data completeness. This variable is the ratio between the recorded duration and aforementioned duration variable. The recorded duration is given by all the timestamps associated to a data value between the first and last data points.</p>
<p>The ECG quality was evaluated in segments of 10 s and in three levels: low, medium and high. Low quality corresponds to an unusable segment, whether due to the presence of an artifact, or due to the loss of connection with the body. Medium quality (ECG MQ) corresponds to a segment where it is possible to extract rhythm information, in particular the heart rate and heart rate variability. In this scenario, the R peak detection is reliable, but the rest of the ECG waveform can be indistinguishable. Lastly, the high quality (ECG HQ) corresponds to a reliable and consistent depiction of the ECG wave throughout the segment. This qualitative stratification was also described in previous works and gives a perception of possible applications of the acquired data. To reach these levels, each segment is processed according to the following method: the R peaks locations are found and the instantaneous heart rate is calculated. If the heart rate is between 40 and 200 beats per minute, the low quality level is discarded (the low quality is chosen otherwise). Then, the ECG wave cycles are cross-correlated through Pearson correlation (between 0 and 1). If the cross correlation is above 0.8, then the segment is classified as high quality (and medium quality if below the threshold).</p>
<p>For the ACC, (<xref ref-type="bibr" rid="B30">Nasseri et al., 2020</xref>), evaluated the quality of Empatica E4 in the context of video EEG-monitoring, proposing as SQI the ratio between human motion activity frequency and the entire frequency range, as given by Eq. <xref ref-type="disp-formula" rid="e1">1</xref>, where <italic>p</italic> refers to the power&#x2019;s spectrum in the indicated frequency band and FN is the Nyquist frequency (half of the sampling frequency). In this paper, the metric was calculated for 4-s non-overlapping segments and averaged across 10-min intervals.<disp-formula id="e1">
<mml:math id="m4">
<mml:msub>
<mml:mrow>
<mml:mi>S</mml:mi>
<mml:mi>Q</mml:mi>
<mml:mi>I</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>C</mml:mi>
<mml:mi>C</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mo>&#x3d;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:msub>
<mml:mrow>
<mml:mi>P</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="[" close="]">
<mml:mrow>
<mml:mn>0.5</mml:mn>
<mml:mo>,</mml:mo>
<mml:mn>8</mml:mn>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:msub>
</mml:mrow>
<mml:mrow>
<mml:msub>
<mml:mrow>
<mml:mi>P</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="[" close="]">
<mml:mrow>
<mml:mn>0.8</mml:mn>
<mml:mo>,</mml:mo>
<mml:msub>
<mml:mrow>
<mml:mi>F</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mi>N</mml:mi>
</mml:mrow>
</mml:msub>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:msub>
</mml:mrow>
</mml:mfrac>
</mml:math>
<label>(1)</label>
</disp-formula>
</p>
<p>The signal quality analysis of the RESP signal was calculated following the metric proposed by (<xref ref-type="bibr" rid="B9">Charlton et al., 2021</xref>) for a respiratory signal acquired through impedance pneumography. This metric consists on the analysis of valid breaths from segments with 32 s of length. For each segment, the peaks are extracted and used to find breath cycles. The segment is valid if the breaths respect some predefined conditions proposed by the authors. This metric results in a binary separation between high quality and low quality.</p>
<p>The EDA quality was retrieved from (<xref ref-type="bibr" rid="B5">B&#xf6;ttcher et al., 2022</xref>), where the binary decision high/low quality is based on two factors. The first is the signal amplitude, which has to be above 0.05 <italic>&#x3bc;S</italic>; then, the signal amplitude change rate is calculated as the ratio between the min-max amplitude and the first of these two extremes to appear. This rate should be below 0.2 in the assessment of EDA for windows with 2 s.</p>
<p>The aforementioned work, evaluates the quality of the PPG also in a binary low/high quality distinction, using the spectral entropy in windows of 4 s. A spectral entropy value should be below 0.8 for the segment to be considered high quality.</p>
<p>The decision to follow the quality assessment approaches taken by (<xref ref-type="bibr" rid="B30">Nasseri et al., 2020</xref>; <xref ref-type="bibr" rid="B5">B&#xf6;ttcher et al., 2022</xref>) enabled the comparison between our wearable devices and other datasets and data acquired in similar conditions. Both these works acquired wearable data using the Empatica E4.</p>
</sec>
<sec id="s3-3">
<title>3.3 Experimental protocol</title>
<p>When a patient is admitted to the hospital stay, they sign an informed consent from the hospital so that their data can be analysed and used for research purposes. After the hospital system setup, the patient is approached to participate in the wearable data collection as well. If accepted, a new informed consent is given for the use of wearable devices during the hospital stay. The proposed acquisition system collects data unobtrusively and only requires a daily short-time interruption for battery replacement. This section describes the process of accessing the wearable and the hospital data, the structure of the dataset, as well as the several formats in which each data source was saved.</p>
<sec id="s3-3-1">
<title>3.3.1 Hospital data extraction</title>
<p>The hospital data of the video-EEG monitoring consisted on the patient&#x2019;s reports and hospital admission notes, along with the data recorded by the monitoring unit. The physiological data is always reviewed by an EEG technician and a neurologist resulting in a video-EEG monitoring report with all the relevant events annotated. This is usually saved as Word and PDF documents. The reports were only consulted with the purpose of retrieving the seizures&#x2019; onsets and details, being dissociated from the biosignals&#x2019; records, which are anonymised thus preserving patients&#x2019; and data privacy. The physiological data is stored in the hospital servers and accessed through a secure computer in the hospital&#x2019;s facility. All physiological data files are retrieved with all acquired channels, varying from 24 (22 EEG channels and 2 ECG channels) up to 77 (72 EEG channels, 2 ECG, 1 SpO2, 1 EMG, and 1 EOG). However, most of the time, EOG and SpO2 are disconnected, and EMG is only connected in some motor seizures, usually on the deltoid or tibia.</p>
<p>Concerning the data extraction, we came across different systems in the two hospitals included in this study, which led to different extraction processes and file formats. The HSM video-EEG data is managed by the Nihon Kohden DMS system<xref ref-type="fn" rid="fn7">
<sup>7</sup>
</xref>; all HSM files are extracted in the European Data Format (EDF), resulting in files with 2 h of data, sampled at 1,000 Hz. Through the mne python package, it is also possible to use the raw Nihon Kohden files (.EEG format) (<xref ref-type="bibr" rid="B54">Gramfort et al., 2013</xref>). The HEM video-EEG data is managed by Micromed software<xref ref-type="fn" rid="fn8">
<sup>8</sup>
</xref>. All HEM files are extracted in the micromed raw file format (TRC), resulting in files with 2 h of data, sampled at 256 Hz.</p>
<p>In both hospital systems, two ECG channels are recorded, corresponding to the positive and negative channels. The difference between these channels results in the approximation of the Lead I derivation, sometimes requiring inversion of the signal (i.e., when the positive and negative electrodes are placed on the wrong side).</p>
</sec>
<sec id="s3-3-2">
<title>3.3.2 Wearable data extraction</title>
<p>The acquisition setup relied on a processing unit to receive BITalino data in real-time and store it appropriately. This was initially performed using a laptop and Python script, but was later replaced by a more user-friendly acquisition system developed within our research group, named EpiBOX (<xref ref-type="bibr" rid="B8">Carmo et al., 2022</xref>). Nonetheless, both systems stored the data in the same structure.</p>
<p>The data collected by the wearables was saved every time the system disconnected, or after 1 h in a stabilised connection. These files are in the TXT format, comprising the data of one or two devices (if they are acquired simultaneously), with a list-like header (with one dictionary per device, where the key corresponds to the device MAC address) on the first few rows and the values over the next rows, in a table-like structure. All the header fields are detailed with examples in the <xref ref-type="sec" rid="s13">Supplementary Material</xref>. Besides the header, each subsequent row corresponds to a new time instant (1/<italic>FS</italic>), where FS corresponds to the sampling frequency of 1,000 Hz. Each TXT file is named after its starting time (e.g., 2021-04-02 18-09-05.TXT). The table produced with two devices is equal to one device with the only difference being the addition of the second device&#x2019;s columns. The order of first and second is known by the order in which they appear in the header. When retrieving data from the second device, it is necessary to recognise the presence of the first, and use its number of columns as an offset. This structure is easily scalable to additional devices.</p>
</sec>
</sec>
</sec>
<sec sec-type="results" id="s4">
<title>4 Results</title>
<p>The results detailed in this section are divided into the signal quality analysis and the characteristics of the dataset acquired so far.</p>
<sec id="s4-1">
<title>4.1 PreEpiSeizures data quality</title>
<p>The quality of biosignals was studied for all data acquired by the hospital system, the ChestBIT, the WristBIT and the ArmBIT and these results are summarised in <xref ref-type="table" rid="T3">Table 3</xref>. The hospital data was acquired for 59 patients totalling almost 6 k hours of ECG data. Data completeness is fairly high with a median of 99.21%. The majority of the signal was classified, on average, as high quality (around 70%), however, on average, 15% of the data was of poor quality. Regarding the ChestBIT, 37 patients used this device, with some sessions being more successful than others (this is observable by the minimum and maximum metrics). The data completeness is lower (in mean and median) when compared to the hospital system, which is expected since the system communicates through Bluetooth, hence being more susceptible to connection losses. Nonetheless, the discrepancy between the mean and the median indicates that the majority of acquisitions was mostly successful in terms of data completeness. In ChestBIT, more segments were discarded as poor quality, however, the median of low quality segments is also around 15% (100 - (ECG HQ &#x2b; ECG MQ)). Interestingly, the median percentage of ECG HQ is higher in the ChestBIT than in the hospital system, and very few segments were classified as medium quality. The quality of RESP signals is on average 60%, which is reasonable.</p>
<table-wrap id="T3" position="float">
<label>TABLE 3</label>
<caption>
<p>Signal quality analysis for the data acquired by the hospital system and the devices ChestBIT and WristBIT. The N indicates the number of patients and the total duration is in H:M:S. The first column indicates the statistical metric, followed by the data completeness (in %), the quality of the biosignals, where the ECG is divided in high and medium quality (in %) and the acquisition duration (in H:M:S).</p>
</caption>
<table>
<thead valign="top">
<tr>
<th colspan="2" align="left">Hospital data, <italic>N</italic> &#x3d; 59</th>
<th align="left"/>
<th align="left"/>
<th align="left">Total duration: 5,838:19:12</th>
</tr>
<tr>
<th align="left">
<italic>p</italic>
</th>
<th align="left">Data Completeness (%)</th>
<th align="left">ECG HQ (%)</th>
<th align="left">ECG MQ (%)</th>
<th align="left">Duration</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">Mean</td>
<td align="left">92.33</td>
<td align="left">68.58</td>
<td align="left">17.15</td>
<td align="left">98:57:16</td>
</tr>
<tr>
<td align="left">Median</td>
<td align="left">99.21</td>
<td align="left">71.83</td>
<td align="left">16.11</td>
<td align="left">94:33:40</td>
</tr>
<tr>
<td align="left">SD</td>
<td align="left">15.09</td>
<td align="left">20.29</td>
<td align="left">11.33</td>
<td align="left">33:51:50</td>
</tr>
<tr>
<td align="left">Min</td>
<td align="left">43.40</td>
<td align="left">4.10</td>
<td align="left">0.88</td>
<td align="left">10:43</td>
</tr>
<tr>
<td align="left">Max</td>
<td align="left">100</td>
<td align="left">96.60</td>
<td align="left">52.23</td>
<td align="left">165:47:15</td>
</tr>
</tbody>
</table>
<table>
<thead valign="top">
<tr>
<th colspan="2" align="left">WristBIT, <italic>N</italic> &#x3d; 9</th>
<th align="left"/>
<th align="left"/>
<th colspan="2" align="left">Total Duration: 510:04:07</th>
</tr>
<tr>
<th align="left">
<italic>p</italic>
</th>
<th align="left">Data Completeness (%)</th>
<th align="left">EDA (%)</th>
<th align="left">BVP (%)</th>
<th align="left">ACC</th>
<th align="left">Duration</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">Mean</td>
<td align="left">70.64</td>
<td align="left">92.84</td>
<td align="left">64.33</td>
<td align="left">0.44</td>
<td align="left">56:39:46</td>
</tr>
<tr>
<td align="left">Median</td>
<td align="left">69.95</td>
<td align="left">98.23</td>
<td align="left">71.72</td>
<td align="left">0.39</td>
<td align="left">60:33:25</td>
</tr>
<tr>
<td align="left">SD</td>
<td align="left">21.11</td>
<td align="left">13.73</td>
<td align="left">25.31</td>
<td align="left">0.19</td>
<td align="left">15:21:54</td>
</tr>
<tr>
<td align="left">Min</td>
<td align="left">31.82</td>
<td align="left">57.55</td>
<td align="left">20.39</td>
<td align="left">0.22</td>
<td align="left">26:31:28</td>
</tr>
<tr>
<td align="left">Max</td>
<td align="left">93.38</td>
<td align="left">99.86</td>
<td align="left">89.72</td>
<td align="left">0.80</td>
<td align="left">76:25:50</td>
</tr>
</tbody>
</table>
<table>
<thead valign="top">
<tr>
<th colspan="2" align="left">ChestBIT, <italic>N</italic> &#x3d; 37</th>
<th align="left"/>
<th align="left"/>
<th align="left"/>
<th colspan="2" align="left">Total Duration: 2,211:53:29</th>
</tr>
<tr>
<th align="left">
<italic>p</italic>
</th>
<th align="left">Data Completeness (%)</th>
<th align="left">ECG HQ (%)</th>
<th align="left">ECG MQ (%)</th>
<th align="left">PZT (%)</th>
<th align="left">ACC</th>
<th align="left">Duration</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">Mean</td>
<td align="left">69.64</td>
<td align="left">67.33</td>
<td align="left">4.44</td>
<td align="left">58.52</td>
<td align="left">0.35</td>
<td align="left">59:46:51</td>
</tr>
<tr>
<td align="left">Median</td>
<td align="left">85.98</td>
<td align="left">82.03</td>
<td align="left">3.02</td>
<td align="left">61.63</td>
<td align="left">0.34</td>
<td align="left">61:20:44</td>
</tr>
<tr>
<td align="left">SD</td>
<td align="left">31.84</td>
<td align="left">34.59</td>
<td align="left">6.07</td>
<td align="left">21.16</td>
<td align="left">0.22</td>
<td align="left">34:18:16</td>
</tr>
<tr>
<td align="left">Min</td>
<td align="left">1.15</td>
<td align="left">0.0</td>
<td align="left">0.0</td>
<td align="left">1.81</td>
<td align="left">0.0</td>
<td align="left">10:20</td>
</tr>
<tr>
<td align="left">Max</td>
<td align="left">100</td>
<td align="left">98.65</td>
<td align="left">30.89</td>
<td align="left">89.47</td>
<td align="left">0.82</td>
<td align="left">192:38:58</td>
</tr>
</tbody>
</table>
<table>
<thead valign="top">
<tr>
<th colspan="2" align="left">ArmBIT, <italic>N</italic> &#x3d; 4</th>
<th align="left"/>
<th align="left"/>
<th align="left"/>
<th colspan="2" align="left">Total Duration: 156:14:54</th>
</tr>
<tr>
<th align="left">
<italic>p</italic>
</th>
<th align="left">Data Completeness (%)</th>
<th align="left">ECG HQ (%)</th>
<th align="left">ECG MQ (%)</th>
<th align="left">EDA (%)</th>
<th align="left">BVP (%)</th>
<th align="left">Duration</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="left">Mean</td>
<td align="left">60.09</td>
<td align="left">8.18&#x2a;10<sup>&#x2013;3</sup>
</td>
<td align="left">0.84</td>
<td align="left">15.50</td>
<td align="left">58.52</td>
<td align="left">39:03:43</td>
</tr>
<tr>
<td align="left">Median</td>
<td align="left">59.19</td>
<td align="left">0.0</td>
<td align="left">0.64</td>
<td align="left">2.03&#x2a;10<sup>&#x2013;2</sup>
</td>
<td align="left">61.63</td>
<td align="left">7:27:19</td>
</tr>
<tr>
<td align="left">SD</td>
<td align="left">28.56</td>
<td align="left">1.64&#x2a;10<sup>&#x2013;2</sup>
</td>
<td align="left">0.97</td>
<td align="left">30.97</td>
<td align="left">21.16</td>
<td align="left">67:27:35</td>
</tr>
<tr>
<td align="left">Min</td>
<td align="left">26.35</td>
<td align="left">0.0</td>
<td align="left">0.0</td>
<td align="left">0.0</td>
<td align="left">1.81</td>
<td align="left">1:26</td>
</tr>
<tr>
<td align="left">Max</td>
<td align="left">95.65</td>
<td align="left">3.27&#x2a;10<sup>&#x2013;2</sup>
</td>
<td align="left">2.05</td>
<td align="left">61.95</td>
<td align="left">89.47</td>
<td align="left">139:54:13</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>The WristBIT was only acquired in 9 patients, still, it comprises 500 h of data, with data completeness percentage comparable to the ChestBIT. The quality of the EDA signal was significantly high, however the PPG quality was significantly lower, since on average, only half of the segments have high quality. This was expected, since the PPG is highly susceptible to noise. Moreover, this is comparable to the results reported by (<xref ref-type="bibr" rid="B5">B&#xf6;ttcher et al., 2022</xref>) with the Empatica E4, where the mean and median quality of PPG data was in the range [51.5, 63.3]% for the datasets acquired in hospital settings. The <italic>SQI</italic>
<sub>
<italic>ACC</italic>
</sub> values of the WristBIT are in agreement with those reported by the study of (<xref ref-type="bibr" rid="B30">Nasseri et al., 2020</xref>).</p>
</sec>
<sec id="s4-2">
<title>4.2 PreEpiSeizures dataset</title>
<p>The total number of patients in this dataset is 59. From this cohort, 36 performed the video-EEG in Hospital de Santa Maria (HSM), while 23 performed the video-EEG in Hospital Egas Moniz (HEM). This dataset includes biosignals recorded by our wearable devices, as well as records from the hospital. The classification of seizure types follows the guidelines of the 2017 ILAE&#x2019;s instruction manual (<xref ref-type="bibr" rid="B14">Fisher et al., 2017a</xref>; <xref ref-type="bibr" rid="B15">b</xref>). Focal seizures were classified as FUAS, whenever the awareness was not registered or tested. The ictal semiology described in this dataset, follows the 2022 ILAE&#x2019;s glossary of terms (<xref ref-type="bibr" rid="B4">Beniczky et al., 2022</xref>).</p>
<sec id="s4-2-1">
<title>4.2.1 Hospital de Santa Maria</title>
<p>
<xref ref-type="fig" rid="F4">Figure 4</xref> provides a visual depiction of the wearable data acquired so far in HSM. The horizontal axis represents the days of the week, whereas the vertical axis represents the different patients enrolled. The label of vertical axis contain the information of the wearable device and the acquisition system. The blue lines represent the wearable data, while the pink shade corresponds to the missing data. The red and grey diamonds represent clinical and subclinical seizures, respectively.</p>
<fig id="F4" position="float">
<label>FIGURE 4</label>
<caption>
<p>HSM Data Acquisition Temporal Overview. Patients with wearable and hospital data. The horizontal axis represents the temporal domain, from the beginning of the video-EEG monitoring, until its end. Each row consists in a single patient.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g004.tif"/>
</fig>
<p>The first 4 patients enrolled in this study used the wearables ArmBIT and ForearmBIT during their video-EEG monitoring: in <xref ref-type="fig" rid="F4">Figure 4</xref>, these are represented in the four bottom rows. These acquisitions lasted for 5 days and 19 h, 13 h, and 1 h and 30 min for the remaining two, respectively. However, only one of these acquisitions captured a seizure event (the patient with the longest acquisition did not experienced any seizure during their hospital stay, whilst the other patients&#x2019; seizures occurred outside the wearables&#x2019; recorded time spans). There were 10 patients that used the ChestBIT and WristBIT during their video-EEG monitoring, recording over 102 seizures. Their distribution according to seizure type is: subclinical &#x3d; 63; FBTCS &#x3d; 8; FAS &#x3d; 5; FIAS &#x3d; 17, and unknown focus &#x3d; 9. The percentage of recorded data by these wearable devices was on average 53% &#xb1; 18% of the entire video-EEG monitoring stay (4 full days &#x3d; 96 h). The number of actual recorded seizures by the wearable device is 19, from a total of 39 clinical seizures, which occurred during the monitorings.</p>
<p>The EpiBOX deployment as the acquisition system was mostly successful, with only 4 out of 13 patients having less than a day recorded (their recorded duration was: 13 h; 3 h; 12 min and 9 min). The ChestBIT enabled the recording of 30 out of 60 clinical seizures. Besides the mentioned patients so far, there was the case of patients who were enrolled in the study and only have hospital data. Since the hospital monitoring equipment provides ECG recordings, they were still included in the dataset, despite the lack of wearable data. 9 patients are included in this group with a total of 47 clinical seizures, with a distribution in seizure types of FBTCS &#x3d; 6; FAS &#x3d; 19; FIAS &#x3d; 16 and subclinical &#x3d; 6.</p>
</sec>
<sec id="s4-2-2">
<title>4.2.2 Hospital Egas Moniz</title>
<p>The expansion of the data acquisition to another centre was possible due to the reproducibility of EpiBOX. <xref ref-type="fig" rid="F5">Figure 5</xref> shows the time domain analysis of wearable data recorded for all patients enrolled in this hospital. This set is composed of 15 patients, who performed the wearable acquisition with the EpiBOX setup, along with the ChestBIT device. From this group, one patient did not register any data during the acquisition, hence being excluded from this plot. The time domain analysis of the remaining 14 patients ranges from 10 min up to 3 days and 9 h. The average percentage of actual recorded time domain is 64.2%, with a standard deviation of 34.0%. Nonetheless, the median recorded time percentage rises to 83.6% (the median of time loss is only 16.4%). The total of seizures in this cohort is 45 clinical seizures (FIAS &#x3d; 11, FAS &#x3d; 34) and 20 subclinical seizures. As it is observed in <xref ref-type="fig" rid="F5">Figures 5</xref>, <xref ref-type="fig" rid="F6">6</xref> patients do not have reports on clinical seizures: 2 only experienced subclinical seizures; and 4 did not experienced any seizure during the exam. Nonetheless, the wearable device was able to capture 15 clinical seizures in this cohort.</p>
<fig id="F5" position="float">
<label>FIGURE 5</label>
<caption>
<p>HEM Data Acquisition Temporal Overview. Patients with wearable and hospital data. The <italic>X</italic>-axis represents the temporal domain, from the beginning of the video-EEG monitoring, until its end. Each row consists in a single patient.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g005.tif"/>
</fig>
<fig id="F6" position="float">
<label>FIGURE 6</label>
<caption>
<p>Distribution of seizure types from the cohort. <italic>E</italic>: Subclinical; <italic>U</italic>: Unknown; FIAS, Focal Impaired Awareness Seizure; FBTCS, Focal to Bilateral Tonic-Clonic Seizure; FAS, Focal Aware Seizure; FUAS, Focal Unknown Awareness Seizures.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g006.tif"/>
</fig>
<p>Besides the data acquired in the scope of our project, the hospital also provided additional retrospective data, from other patients with temporal lobe epilepsy, who underwent the same monitoring exam. This set consisted on 9 patients, who experienced 61 clinical seizures (FBTCS &#x3d; 1, FIAS &#x3d; 36, FAS &#x3d; 20 and FUAS &#x3d; 4). This retrospective data contains the hospital system&#x2019;s measurements, which includes the ECG. This cohort was used to initiate the ECG analysis and assess the validation of ECG-based seizure manifestations.</p>
</sec>
</sec>
<sec id="s4-3">
<title>4.3 Overview</title>
<p>The patients collected in this database were performing the video-EEG monitoring for presurgical assessment. This section will detail the distribution of onset localisation and lateralisation, seizure types and ictal semiology reported by the clinical team.</p>
<p>In this cohort (59 patients), 64.41% (<italic>N</italic> &#x3d; 38) showed temporal lobe activation during the onset of ictal period. The frontal lobe ictal onset activation showed a prevalence of 25.42% (<italic>N</italic> &#x3d; 15), closely followed by the parietal lobe with 18.64% (<italic>N</italic> &#x3d; 11). Lastly, the occipital lobe ictal onset was reported in 6.78% (<italic>N</italic> &#x3d; 4). The ictal onset activation reported in more than one lobe region (e.g., fronto-temporal) was included in both categories. Concerning lateralisation, 38% had a left hemisphere onset seizure, whereas the epileptogenic zone was in the right hemisphere for 44% of patients. Bilateral ictal onsets were described in 4% of patients from the cohort.</p>
<p>In <xref ref-type="fig" rid="F6">Figure 6</xref>, the distribution of seizures per seizure types is shown, regarding the seizure types which were reported in the video-EEG reports. The total number of seizures in the PreEpiSeizures dataset is 348, distributed a across six seizure types (FBTCS &#x3d; 27, FIAS &#x3d; 95, FUAS &#x3d; 35, FAS &#x3d; 82, subclinical &#x3d; 89 and unknown &#x3d; 20). These values are illustrated as a percentage in <xref ref-type="fig" rid="F6">Figure 6</xref>, where the predominance of focal seizures is notable. The classification of seizures as Unknown was chosen when there was no information regarding the type of seizure, whereas FUAS correspond to focal seizures, in which only the awareness was not tested.</p>
<p>Besides the classification of seizure types, the video-EEG reports also contained some details regarding the events occurring during the seizure. In <xref ref-type="fig" rid="F7">Figure 7</xref>, the reported ictal semiology from the dataset is displayed in 6 categories. The Elementary Motor Behaviour was found in 120 seizures and includes Dystonic (<italic>N</italic> &#x3d; 23), head version or head orientation (<italic>N</italic> &#x3d; 45), eye blinking (<italic>N</italic> &#x3d; 12), myoclonic/clonic (<italic>N</italic> &#x3d; 14), tonic (<italic>N</italic> &#x3d; 6) and tonic-clonic (<italic>N</italic> &#x3d; 27). The Complex Motor Behaviour (<italic>N</italic> &#x3d; 183) included automatisms and hyperkinetic behaviour (<italic>N</italic> &#x3d; 3). The most common automatisms were distal (<italic>N</italic> &#x3d; 67) followed by oroalimentary (<italic>N</italic> &#x3d; 59) and proximal (<italic>N</italic> &#x3d; 16). Other reported automatisms were: verbal (<italic>N</italic> &#x3d; 11); vocal (<italic>N</italic> &#x3d; 6); mimic (<italic>N</italic> &#x3d; 7); nose-wiping (<italic>N</italic> &#x3d; 9) and Rinch (<italic>N</italic> &#x3d; 5). Sensory phenomena (<italic>N</italic> &#x3d; 33) was mostly populated by the somatosensory symptoms (<italic>N</italic> &#x3d; 15), followed by cephalic aura (<italic>N</italic> &#x3d; 5). Other reported sensory phenomena was gustatory (<italic>N</italic> &#x3d; 1) and olfactory (<italic>N</italic> &#x3d; 2). The most reported autonomic semiology was epigastric aura (<italic>N</italic> &#x3d; 20), followed by hyperventilation (<italic>N</italic> &#x3d; 12), tachycardia (<italic>N</italic> &#x3d; 11) and piloerection (<italic>N</italic> &#x3d; 1). Few seizures reported cognitive ictal phenomena (<italic>N</italic> &#x3d; 27), where the most common symptom was autoscopy (<italic>N</italic> &#x3d; 12), followed by dej&#xe1; vu (<italic>N</italic> &#x3d; 9), ictal aphasia (<italic>N</italic> &#x3d; 3) and dysarthria (<italic>N</italic> &#x3d; 3). Only 2 seizures reported emotional phenomena and the symptom was fear.</p>
<fig id="F7" position="float">
<label>FIGURE 7</label>
<caption>
<p>Overview of all ictal semiology phenomena reported in the video-EEG, grouped according to the nature of the phenomena following the 2022 ILAE&#x2019;s glossary (<xref ref-type="bibr" rid="B4">Beniczky et al., 2022</xref>).</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g007.tif"/>
</fig>
<p>In <xref ref-type="table" rid="T3">Table 3</xref>, the total duration of ArmBIT, ChestBIT and WristBIT is shown separately. <xref ref-type="fig" rid="F8">Figure 8</xref> gives a visual depiction of the number of recorded hours by each modality. The different colours represent the device responsible for each measurement. The least represented modality is EMG with 156 h, 14 min and 54 s of data. Since this measurement required gel electrodes, it was discarded in the new devices. The EDA totals 666 h, 19 min and 2 s of data, which is decomposed in 156 h of using pre-gelled electrodes (with ForearmBIT) and 510 h, 4 min and 7 s of metallic electrodes in the WristBIT. The chest-ACC corresponds to 2,211 h of data, whereas the wrist-ACC totals 510 h. Additionally, the ArmBIT and ForearmBIT add more 312 h to the ACC data. Regarding RESP, 2,211 h were recorded so far, using only the ChestBIT. The ECG was captured using pre-gelled electrodes in 156 h (ArmBIT) and the ChestBIT has so far recorded 2,211 h. The total number of recorded hours by all wearables is 2,721 h, 57 min and 36 s. The broader use of ChestBIT instead of WristBIT was based on three arguments: an easier reproducibility; a higher reliability of ECG instead of PPG to record cardiac activity; and being more discreet, since it is placed under the clothes. This last point is aligned with previously surveyed preferences of PWE (<xref ref-type="bibr" rid="B39">Simblett et al., 2020</xref>).</p>
<fig id="F8" position="float">
<label>FIGURE 8</label>
<caption>
<p>Number of hours acquired per modality. Each colour represents one form factor.</p>
</caption>
<graphic xlink:href="fphys-14-1248899-g008.tif"/>
</fig>
<p>To conclude the overview of the PreEpiSeizures dataset, the aforementioned <xref ref-type="table" rid="T1">Table 1</xref> provides a side-by-side characterisation of this dataset with those available within the state-of-the-art. Three of the public databases (Bern-Barcelona (<xref ref-type="bibr" rid="B3">Andrzejak et al., 2012</xref>), New Delhi EEG (<xref ref-type="bibr" rid="B42">Swami et al., 2016</xref>) and Epilepsy Seizure Recognition (<xref ref-type="bibr" rid="B2">Andrzejak et al., 2001</xref>)) only presented segments of a few seconds duration. Bonn-Barcelona micro- and macro-EEG database (<xref ref-type="bibr" rid="B26">Mart&#xed;nez et al., 2020</xref>), comprises data from epileptic patients without epileptic seizures.</p>
<p>The comparison of our dataset with EPILEPSIAE (<xref ref-type="bibr" rid="B20">Klatt et al., 2012</xref>) and Siena Scalp EEG (<xref ref-type="bibr" rid="B12">Detti, 2020</xref>) is performed with regards to the ECG recorded by the hospital system (this is the number in the left side of columns patient and total hours). Our dataset size, both in terms of patients, and in total recorded hours, surpasses its peers.</p>
<p>Regarding wearable data, the only other dataset found with wearable data was the My Seizure Gauge Wearable, which does not contain ECG, PZT or Chest Motion data. Both datasets duration is quite similar (our wearable duration is the right number of column Total Hours in <xref ref-type="table" rid="T1">Table 1</xref>). Concerning the number of seizures and the variety of seizure types, PreEpiSeizures is similar to EPILEPSIAE and My Seizure Gauge Wearable. Regarding the number of recorded hours, the proposed dataset also closely follows those of the literature.</p>
</sec>
</sec>
<sec sec-type="discussion" id="s5">
<title>5 Discussion</title>
<p>The patients which are admitted for video-EEG monitoring are usually patients that either have refractory epilepsy and require a differential diagnosis, or are included in the presurgical evaluation program. In HSM, a new patient comes every 2 weeks, with some exceptions like vacations or other unforeseen events, which can change this schedule. Then, the patient stays for a work week (Monday to Friday) and sometimes also spends the weekend if the hospital team finds it necessary, for example, if no seizure occurred until Friday. Although this is a very good pace, allowing for up to 26 patients per year, not all these patients meet the eligibility criteria: patients need to have at least 16 years old; the EEG monitoring should be performed with scalp electrodes, instead of intracranial; patients also require a high level of independence, compatible with using an extra device continuously throughout the hospital stay. In HEM, the EEG technicians are all day with the patients (from 8AM to 11PM), hence the majority does not need accompaniment. The cadence in this ward is of 1&#x2013;2 patients per week, however only one is asked to participate in the study. This is related also to the duration of the monitoring, since one patients stays in the hospital for the work week, whereas the other only stays for 48 h (the case of differential diagnosis) or less (in the case of other exams).</p>
<p>Despite having a great number of seizures, the proposed dataset suffers from one core issue also found in its peers: the great variability of seizure types. In both hospitals, many types of seizures are monitored (from tonic-clonic and myoclonic to subclinical seizures), and often the epileptogenic zone is not even known prior to the monitoring. As a result, it is difficult to predict what will be the outcome of the data collection. So far, the majority of seizures encountered have a focal onset, nevertheless they vary in duration, awareness or semiology. Even the broader classifications of seizure types (such as FAS and FIAS) enclose seizures that have motor and non-motor manifestations. Although patients in video-EEG monitoring have refractory epilepsy, their seizure occurrence is often highly reduced in daily life by ASM. During the monitoring, the ASM is slowly suspended to allow for resurgence of epileptic activity. For some patients, the complete withdraw of ASM could result in bilateral tonic-clonic seizures, which would hamper the identification of seizure onset. The goal in video-EEG monitoring is to achieve the balance that allows to clearly identify the ictal onset of seizures, with the minimum number of seizures. However, for some patients, no seizure occurs during the hospital admission, despite their frequent seizure rate outside the hospital.</p>
<p>The wearable data acquisition itself was an eventful process, filled with a great variety of challenges. Hereinafter, this section is divided into an overview of the problems faced and proposed solutions, followed by a discussion of the quality between the wearable device and the hospital system, and finally the quality assessment of the current acquisition setup and its comparison with the previous setup.</p>
<p>In <xref ref-type="table" rid="T4">Table 4</xref>, the various issues faced during the wearable data acquisition process are described, being divided into four categories according to their nature: real-world acquisition; wearable design; processing; and noise interference. Though this data collection was specific to a singular disease and a strict protocol, the challenges herein faced are general to wearable and even some non-wearable data collection experiments.</p>
<table-wrap id="T4" position="float">
<label>TABLE 4</label>
<caption>
<p>Common concerns of wearable data acquisition.</p>
</caption>
<table>
<thead valign="top">
<tr>
<th align="center">Name</th>
<th align="left">Description</th>
<th align="center">Nature</th>
<th align="left">Solution</th>
</tr>
</thead>
<tbody valign="top">
<tr>
<td align="center">Forget battery exchange</td>
<td align="left">Uncontrolled scenarios can lead to forgetting battery exchanges</td>
<td align="center">Real-world acquisition</td>
<td align="left">Mobile app</td>
</tr>
<tr>
<td align="center">User Compliance</td>
<td align="left">The acceptance to participate and the proper realisation of the experiment</td>
<td align="center">Real-world acquisition</td>
<td align="left">Reward system</td>
</tr>
<tr>
<td align="center">Incorrect placement</td>
<td align="left">Uncontrolled scenarios can lead to incorrect device placement by non-experts</td>
<td align="center">Real-world acquisition</td>
<td align="left">Mobile app</td>
</tr>
<tr>
<td align="center">Cohort&#x2019;s variability</td>
<td align="left">Real-world settings might translate into an unexpected variability to the cohort</td>
<td align="center">Real-world acquisition</td>
<td align="left">Multi-centre</td>
</tr>
<tr>
<td align="center">Battery size</td>
<td align="left">Battery shortage could cause more connection losses</td>
<td align="center">Wearable design</td>
<td align="left">Reduce FS</td>
</tr>
<tr>
<td align="center">Design constraints</td>
<td align="left">The signals to record are restricted by the sensors&#x2019; requirements</td>
<td align="center">Wearable design</td>
<td align="left">Adaptative factor</td>
</tr>
<tr>
<td align="center">Device resistance</td>
<td align="left">Resistance to sweat and to the continuous use over multiple acquisitions</td>
<td align="center">Wearable design</td>
<td align="left">Detachable electronics</td>
</tr>
<tr>
<td align="center">Materials Hygienisation</td>
<td align="left">Materials properly clean before and after each new acquisition, cleanable materials</td>
<td align="center">Wearable design</td>
<td align="left">Resistant materials</td>
</tr>
<tr>
<td align="center">Signal quality</td>
<td align="left">Uncontrolled scenarios can lead to noise presence in the signals of a variety of natures</td>
<td align="center">Processing</td>
<td align="left">Multimodal denoising</td>
</tr>
<tr>
<td align="center">Temporal synchronisation</td>
<td align="left">Signals acquired in different sources may present a relative temporal shift</td>
<td align="center">Processing</td>
<td align="left">HR synchronisation</td>
</tr>
<tr>
<td align="center">Big data</td>
<td align="left">High sampling frequencies and long hours leads to great amounts of data</td>
<td align="center">Processing</td>
<td align="left">Batch processing</td>
</tr>
<tr>
<td align="center">Fragmented files</td>
<td align="left">File fragmentation can lead to contiguous files with different signal offsets or amplitudes, sometimes time loss is unknown</td>
<td align="center">Processing</td>
<td align="left">Discrete analysis</td>
</tr>
<tr>
<td align="center">Device drift</td>
<td align="left">Devices can drift and originate signals with slightly different characteristics</td>
<td align="center">Processing</td>
<td align="left">Constant calibration</td>
</tr>
<tr>
<td align="center">Motion artifacts</td>
<td align="left">The presence of motion artifacts affects the proper measurements of other signals</td>
<td align="center">Noise Interference</td>
<td align="left">ACC denoising</td>
</tr>
<tr>
<td align="center">Sensor interference</td>
<td align="left">The continuous use of a device might degrade its sensors, and contamination can occur</td>
<td align="center">Noise interference</td>
<td align="left">Automatic check</td>
</tr>
<tr>
<td align="center">Contact loss</td>
<td align="left">Contact with the body can be lost during acquisition</td>
<td align="center">Noise interference</td>
<td align="left">Automatic check</td>
</tr>
<tr>
<td align="center">Network interference</td>
<td align="left">The surroundings can cause electrical disturbances in the physiological signal</td>
<td align="center">Noise interference</td>
<td align="left">Low-pass filtering</td>
</tr>
<tr>
<td align="center">Baseline wander</td>
<td align="left">Fluctuations on the ECG signal are common and might result in incorrect measurements</td>
<td align="center">Noise interference</td>
<td align="left">Baseline removal techniques</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>Four challenges were identified in relation to real-world acquisition: forget battery exchange (1.1); user compliance (1.2); incorrect placement (1.3) and cohort&#x2019;s variability (1.4). Although all data was collected in the hospital ward, there was no constant supervision by an expert, leading to several challenges, that can also be expected in an ambulatory scenario (hence their &#x201c;real-world&#x201d; nature). (1.1) Most wearable devices require a quasi-daily change of battery, as seen in <xref ref-type="table" rid="T2">Table 2</xref>. Hence, it could be expected that users forget to charge the device or to have a charged battery at-hand. To help in this process, the proposed solution is a mobile application with reminders to charge the device or change the battery. (1.2) User compliance is important for any experiment, especially those in which the patient plays a major role in the acquisition control. Mental and emotional availability is required to perform a rigorous experiment control; despite the relative good rate of potential patients, not all will meet the inclusion criteria to participate in the data collection. Moreover, this is amplified in ambulatory data collection, and strategies should be established to increase patient&#x2019;s compliance. As a suggestion, a more immediate reward system could be added to the protocol, such as their data overview, or daily statistics. (1.3) Unsupervised experiments such as this, could lead to incorrect placement of the devices, especially if more than one device is used simultaneously. For example, the ChestBIT could be inverted from its original orientation, or shifted to one side or the other. Moreover, the ChestBIT could fall during the night or even detach itself. In order to solve these issues, the device should be reinforced especially in the attachment area. Moreover, in an ambulatory setting, the mobile application could aid the patient to properly adjust the device, for example, via a signal quality assessment algorithm. (1.4) The variability of the cohort is something to account for when dealing with such specific diseases. Potential patients could have different ages, unique disease characteristics, or even specific comorbidities. In this case it could be useful to design the protocol upfront as a multi-centre data collection. The increase of potential patients could enable the encounter of similar cases and better deal with the great cohort&#x2019;s variability.</p>
<p>The category of wearable design included four challenges: battery size (2.1); design constraints (2.2); device&#x2019;s resistance (2.3); and materials hygienisation (2.4). The nature of these challenges comprises issues which conditioned the wearable design, and should be accounted for when proposing new designs for continuous wearable data collection. (2.1) The battery size is a constraint on wearable devices, especially in preliminary versions, based on lithium ion polymer batteries, hence their energy consumption should be reduced to the minimum, to ensure the maximum operability. In the proposed devices, the FS is 1,000 Hz, which corresponds to a battery drain rate of approximately 62 mAh (i.e., milliamp hour). As a result, 1,500 mAh are required for lasting 24 h. One solution could be to decrease the FS to 100 Hz for the ChestBIT (or less in the case of the WristBIT), to enable a slower rate of battery drain, while still capturing the major physiological manifestations. (2.2) The design of a wearable device is conditioned to its embedded sensors. For example, the ECG sensor requires at least two contact points with the body, one in each side of the heart, whereas the EDA has better quality in places with more eccrine glands (<xref ref-type="bibr" rid="B29">Nagai et al., 2019</xref>). Thus, when designing a new wearable it is necessary to understand which modalities are more relevant to the task, since not all are possible to acquire with the same form factor. (2.3) The device should be resistant to ensure long-term use without quality loss. This could be achieved by choosing appropriate materials, by attaching securely the electronics to the textiles, or even by adopting bendable electronics. (2.4) The materials disinfection is important to remove bacteria and fluids, especially during patient exchange. The choice of materials could also play an important part, not only comfort-wise but also for proper hygienisation.</p>
<p>The next category of challenges is processing, and it includes: signal quality (3.1); temporal synchronisation (3.2); big data (3.3), fragmented files (3.4); and device drift (3.5). Processing concerns the first steps towards the extraction of meaningful information from the acquired data. The challenges herein described could be detrimental and lead to misinformation, thus they should be carefully addressed, preferably during the actual data collection. (3.1) As it was previously mentioned, the device could be incorrectly placed due to various reasons. Moreover, other factors (which will be addressed in the next category) could contribute to poor quality signals. Quality control should be assured automatically through an algorithm, specific for each modality acquired. (3.2) When acquiring data from more than one device simultaneously, or when comparing two signals acquired simultaneously (such as the hospital&#x2019;s ECG against the device&#x2019;s ECG), or even when using the hospital&#x2019;s annotations with the wearables, it is relevant to confirm whether the temporal line is properly defined. It could be the case that the wearable does not save data with the correct time. To solve this issue, it should be implemented a temporal synchronisation. The temporal synchronisation between the same modality time series is possible using cross-correlation (e.g., in ECG/PPG, one can use the instantaneous heart rate from both sources). This could be used between the hospital&#x2019;s signal and the device&#x2019;s signal, and the devices time should be adjusted according to this temporal offset. In our case the EpiBOX smartphone&#x2019;s application allows to save the device&#x2019;s time alongside the real time (fetched by the smartphone). This time difference could also be used to adjust the time to a better estimation prior to signal synchronisation.</p>
<p>(3.3) The FS of 1,000 Hz, associated to several days of acquisition, originates several gigabytes of data per patient. This poses a problem for opening such files during the processing phase. One solution could be batch processing; through this technique, the time is divided into intervals of a few hours (or even less than 1 h), and the signals are extracted from the TXT format to a more compressed format, such as parquet (a columnar file optimized for processing big data). This could allow to save space and time in future operations. (3.4) There is always the case of ending one file and starting another in the middle of the acquisition. This also occurs for hospital files. While it is expected that the next file will be in the same conditions as the previous, changes can occur in both signal&#x2019;s amplitude and/or offset. Moreover, battery exchange can lead to a slight change in the devices position. During processing, the timeseries should be considered discontiguous if they come from fragmented files, and processing should occur separately for each fragment. (3.5) As it was previously mentioned, a drift in device position during the course of the acquisition is inevitable, due to several reasons. The timeseries should be seen as an evolving signal, and not static according to an initial defined baseline. The normalisation or even signal shape correctedness should be assessed with nearby points and not the entire signal. This does not invalidate standard references of quality for the entire signal, but simply whether signal changes are a result of drifts or they contain meaningful information towards the final purpose.</p>
<p>Noise interference in timeseries is especially prominent in wearable acquisitions, with a multitude of noise types (<xref ref-type="bibr" rid="B50">Vijayan et al., 2021</xref>). Noises can be difficult to remove when they fall into the frequency bands of the acquired signal. The following noises found in the data acquisition were: motion artifacts (4.1); sensor interference (4.2); contact loss (4.3); network interference (4.4); and baseline wander (4.5). The proper identification of each noise type, could be the first step for their automatic removal during the processing phase. (4.1) Motion artifacts fall into most signals frequencies of interest, thus their removal is not trivial. Since acceleration is acquired in most wearable devices, one solution could be to use the ACC signal to remove motion from the other signal modalities. This would mostly work when the sensors are placed in proximity. For example, the ACC from WristBIT could have common motion with the EDA, however motion captured in the PPG acquired at the finger might not have been captured by the ACC. (4.2) The continuous long-term use of the device could deteriorate the electronic board contacts and originate improper signals, which should be tested regularly. (4.3) Loss of contact between the device and the skin is something to account for, since the connection is still maintained but the desired signal is no longer there. This should also be assessed automatically by identifying the types of signal shapes induced by such issue, and seeking them in the original signal prior to meaningful extraction. (4.4) The presence of other electronic devices in the surrounding space can influence the quantity of noise in the signal. The video-EEG room contains a great amount of electrical devices acquiring simultaneously and constantly, and one should be prepared for a small signal-to-noise ratio. (4.5) The presence of baseline wander, especially in the ECG signal, is common in both the wearable device and the hospital&#x2019;s system. Similarly to motion artifacts, baseline wander can also fall in the signal frequency range and affect meaningful information extraction. There are several baseline wander techniques proposed in the literature which could be implemented during the processing step.</p>
<p>Given this last set of challenges, one potentially valuable addition to the PreEpiSeizures dataset would be to provide the quality assessment within the metadata of each file. The developed approach should be based on the assessment of quality against the physiological basis in an agnostic manner. This will be addressed in the future work.</p>
</sec>
<sec sec-type="conclusion" id="s6">
<title>6 Conclusion</title>
<p>The PreEpiSeizures project was motivated by the limited number of available wearable data in epilepsy; this collaboration has enabled the collection of data from 59 patients over the past few years (where 37 patients also wore a chest device, nine wore a wrist device, and four wore an arm device). The data collection resulted in 5,838 h of hospital data and 2,721 h of wearable data.</p>
<p>So far, three different form factors of wearable devices were worn by the patients: ArmBIT, WristBIT and ChestBIT. The ChestBIT&#x2019;s ability to record the ECG signal using non-gelled conductive materials, is an advantage compared with the ArmBIT. The quality assessment of ChestBIT was at the level of the hospital system (in the case of the ECG), and WristBIT quality was comparable to the quality of Empatica E4 used in similar conditions, reported in previous works. The advantage of using our own devices, allowed to have the liberty to experiment with form-factors and physiological modalities. Although not all patients have used the wearable devices, the availability of the hospital&#x2019;s ECG signal allows to draw similar conclusions regarding the physiological manifestations of epilepsy in cardiac activity, which is also aligned with the literature.</p>
<p>The video-EEG monitoring in the hospital significantly reduces the variability of the real-world while allowing the access of ground truth annotations. The ambulatory setting may reveal additional challenges which are not being considered yet, nevertheless it was possible to identify 18 challenges that impacted the wearable data collection and may also appear in real-world scenarios. Further research on these challenges should be a priority to avoid data and time losses. Moreover, the concerns and challenges raised in the discussion affect overall continuous health monitoring studies and are not exclusively epilepsy monitoring.</p>
<p>The future of the PreEpiSeizures project will include the maintenance of the data collection process, as well as the curation of the dataset and the expansion to ambulatory monitoring is also envisioned. Additionally, the project will address some of the acquisition challenges mentioned, such as: the battery size, the real-time signal quality assessment, and noise reduction.</p>
</sec>
</body>
<back>
<sec sec-type="data-availability" id="s7">
<title>Data availability statement</title>
<p>The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.</p>
</sec>
<sec id="s8">
<title>Ethics statement</title>
<p>The studies involving humans were approved by the Comiss&#xe3;o de &#xc9;tica do Centro Acad&#xe9;mico de Medicina de Lisboa&#x2014;CAML. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.</p>
</sec>
<sec id="s9">
<title>Author contributions</title>
<p>HP, AF, MA, and AC were responsible for the wearable design and function. MA performed the data quality analysis. MA, CB, and FS were responsible for the data collection and availability. MA wrote the article. AC, AP, FS, CB, HP, and AF reviewed the article. All authors contributed to the article and approved the submitted version.</p>
</sec>
<sec id="s10">
<title>Funding</title>
<p>This work was partially supported by the Funda&#x00E7;&#x00E3;o para a Ci&#x00EA;ncia e (FCT), Portugal, under the grants 2021.08297.BD and 2022.12369.BD, under the Scientific Employment Stimulus&#x2014;Individual Call&#x2014;2022.04901.CEECIND, and project PCIF/SSO/0163/2019 &#x201c;SafeFire&#x201d;. The work has also been partially supported by Centro Hospitalar Universit&#x00E1;rio Lisboa Norte, EPE, under the project &#x201c;Pre_EpiSeizures&#x201d;, by the FCT/Minist&#x00E9;rio da Ci&#x00EA;ncia, Tecnologia e Ensino Superior (MCTES) through national funds and when applicable co-funded by EU funds under the project UIDB/50008/2020, and by the Instituto de Telecomunica&#xe7;&#xf5;es (IT).</p>
</sec>
<ack>
<p>The authors would like to thank Instituto de Telecomunica&#xe7;&#xf5;es for the materials and help with the acquisition setup, namely, the technical engineer Jos&#xe9; Gouveia. In Hospital de Santa Maria, the authors thank the EEG technician Rosa Santos and their team for clinical support and supervision. In Hospital Egas Moniz, the authors acknowledge the contribution of the EEG technician Oct&#xe1;via Br&#xe1;s and their team for their involvement during the acquisitions and annotations. The authors would also like to express their gratitude to all the volunteers that participated in the study.</p>
</ack>
<sec sec-type="COI-statement" id="s11">
<title>Conflict of interest</title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="disclaimer" id="s12">
<title>Publisher&#x2019;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<sec id="s13">
<title>Supplementary Material</title>
<p>The Supplementary Material for this article can be found online at: <ext-link ext-link-type="uri" xlink:href="https://www.frontiersin.org/articles/10.3389/fphys.2023.1248899/full#supplementary-material">https://www.frontiersin.org/articles/10.3389/fphys.2023.1248899/full&#x23;supplementary-material</ext-link>
</p>
<supplementary-material xlink:href="Presentation1.pdf" id="SM1" mimetype="application/pdf" xmlns:xlink="http://www.w3.org/1999/xlink"/>
</sec>
<fn-group>
<fn id="fn1">
<label>1</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.ieeg.org/">https://www.ieeg.org/</ext-link>
</p>
</fn>
<fn id="fn2">
<label>2</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://eithealth.eu/spotlight-story/seizeit/">https://eithealth.eu/spotlight-story/seizeit/</ext-link>
</p>
</fn>
<fn id="fn3">
<label>3</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.epilepsyecosystem.org/">https://www.epilepsyecosystem.org/</ext-link>
</p>
</fn>
<fn id="fn4">
<label>4</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.epilepsyecosystem.org/my-seizure-gauge-1">https://www.epilepsyecosystem.org/my-seizure-gauge-1</ext-link>
</p>
</fn>
<fn id="fn5">
<label>5</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.empatica.com/en-eu/research/e4/">https://www.empatica.com/en-eu/research/e4/</ext-link>
</p>
</fn>
<fn id="fn6">
<label>6</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://www.polar.com/us-en/products/accessories/polar-soft-strap">https://www.polar.com/us-en/products/accessories/polar-soft-strap</ext-link>
</p>
</fn>
<fn id="fn7">
<label>7</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://us.nihonkohden.com/">https://us.nihonkohden.com/</ext-link>
</p>
</fn>
<fn id="fn8">
<label>8</label>
<p>
<ext-link ext-link-type="uri" xlink:href="https://micromedgroup.com/">https://micromedgroup.com/</ext-link>
</p>
</fn>
</fn-group>
<ref-list>
<title>References</title>
<ref id="B1">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Akintola</surname>
<given-names>A. A.</given-names>
</name>
<name>
<surname>van de Pol</surname>
<given-names>V.</given-names>
</name>
<name>
<surname>Bimmel</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Maan</surname>
<given-names>A. C.</given-names>
</name>
<name>
<surname>van Heemst</surname>
<given-names>D.</given-names>
</name>
</person-group> (<year>2016</year>). <article-title>Comparative analysis of the equivital EQ02 lifemonitor with holter ambulatory ECG device for continuous measurement of ECG, heart rate, and heart rate variability: A validation study for precision and accuracy</article-title>. <source>Front. Physiology</source> <volume>7</volume>, <fpage>391</fpage>. <pub-id pub-id-type="doi">10.3389/fphys.2016.00391</pub-id>
</citation>
</ref>
<ref id="B2">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andrzejak</surname>
<given-names>R. G.</given-names>
</name>
<name>
<surname>Lehnertz</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Mormann</surname>
<given-names>F.</given-names>
</name>
<name>
<surname>Rieke</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>David</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Elger</surname>
<given-names>C. E.</given-names>
</name>
</person-group> (<year>2001</year>). <article-title>Indications of nonlinear deterministic and finite-dimensional structures in time series of brain electrical activity: dependence on recording region and brain state</article-title>. <source>Phys. Rev. E</source> <volume>64</volume>, <fpage>061907</fpage>. <pub-id pub-id-type="doi">10.1103/PhysRevE.64.061907</pub-id>
</citation>
</ref>
<ref id="B3">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Andrzejak</surname>
<given-names>R. G.</given-names>
</name>
<name>
<surname>Schindler</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Rummel</surname>
<given-names>C.</given-names>
</name>
</person-group> (<year>2012</year>). <article-title>Nonrandomness, nonlinear dependence, and nonstationarity of electroencephalographic recordings from epilepsy patients</article-title>. <source>Phys. Rev. E</source> <volume>86</volume>, <fpage>046206</fpage>. <pub-id pub-id-type="doi">10.1103/PhysRevE.86.046206</pub-id>
</citation>
</ref>
<ref id="B4">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Beniczky</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Tatum</surname>
<given-names>W. O.</given-names>
</name>
<name>
<surname>Blumenfeld</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Stefan</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Mani</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Maillard</surname>
<given-names>L.</given-names>
</name>
<etal/>
</person-group> (<year>2022</year>). <article-title>Seizure semiology: ILAE glossary of terms and their significance</article-title>. <source>Epileptic Disord.</source> <volume>24</volume>, <fpage>447</fpage>&#x2013;<lpage>495</lpage>. <pub-id pub-id-type="doi">10.1684/epd.2022.1430</pub-id>
</citation>
</ref>
<ref id="B5">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>B&#xf6;ttcher</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Vieluf</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Bruno</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Joseph</surname>
<given-names>B.</given-names>
</name>
<name>
<surname>Epitashvili</surname>
<given-names>N.</given-names>
</name>
<name>
<surname>Biondi</surname>
<given-names>A.</given-names>
</name>
<etal/>
</person-group> (<year>2022</year>). <article-title>Data quality evaluation in wearable monitoring</article-title>. <source>Sci. Rep.</source> <volume>12</volume>, <fpage>21412</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-022-25949-x</pub-id>
</citation>
</ref>
<ref id="B6">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Brinkmann</surname>
<given-names>B. H.</given-names>
</name>
<name>
<surname>Karoly</surname>
<given-names>P. J.</given-names>
</name>
<name>
<surname>Nurse</surname>
<given-names>E. S.</given-names>
</name>
<name>
<surname>Dumanis</surname>
<given-names>S. B.</given-names>
</name>
<name>
<surname>Nasseri</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Viana</surname>
<given-names>P. F.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>Seizure diaries and forecasting with wearables: epilepsy monitoring outside the clinic</article-title>. <source>Front. Neurology</source> <volume>12</volume>, <fpage>690404</fpage>. <pub-id pub-id-type="doi">10.3389/fneur.2021.690404</pub-id>
</citation>
</ref>
<ref id="B7">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Bruno</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Viana</surname>
<given-names>P. F.</given-names>
</name>
<name>
<surname>Sperling</surname>
<given-names>M. R.</given-names>
</name>
<name>
<surname>Richardson</surname>
<given-names>M. P.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Seizure detection at home: do devices on the market match the needs of people living with epilepsy and their caregivers?</article-title> <source>Epilepsia</source> <volume>61</volume>, <fpage>S11</fpage>&#x2013;<lpage>S24</lpage>. <pub-id pub-id-type="doi">10.1111/epi.16521</pub-id>
</citation>
</ref>
<ref id="B8">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Carmo</surname>
<given-names>A. S.</given-names>
</name>
<name>
<surname>Abreu</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Fred</surname>
<given-names>A. L. N.</given-names>
</name>
<name>
<surname>da Silva</surname>
<given-names>H. P.</given-names>
</name>
</person-group> (<year>2022</year>). <article-title>EpiBOX: an automated platform for long-term biosignal collection</article-title>. <source>Front. Neuroinformatics</source> <volume>16</volume>, <fpage>837278</fpage>. <pub-id pub-id-type="doi">10.3389/fninf.2022.837278</pub-id>
</citation>
</ref>
<ref id="B9">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Charlton</surname>
<given-names>P. H.</given-names>
</name>
<name>
<surname>Bonnici</surname>
<given-names>T.</given-names>
</name>
<name>
<surname>Tarassenko</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Clifton</surname>
<given-names>D. A.</given-names>
</name>
<name>
<surname>Beale</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Watkinson</surname>
<given-names>P. J.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>An impedance pneumography signal quality index: design, assessment and application to respiratory rate monitoring</article-title>. <source>Biomed. Signal Process. Control</source> <volume>65</volume>, <fpage>102339</fpage>. <pub-id pub-id-type="doi">10.1016/j.bspc.2020.102339</pub-id>
</citation>
</ref>
<ref id="B10">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cook</surname>
<given-names>M. J.</given-names>
</name>
<name>
<surname>O&#x2019;Brien</surname>
<given-names>T. J.</given-names>
</name>
<name>
<surname>Berkovic</surname>
<given-names>S. F.</given-names>
</name>
<name>
<surname>Murphy</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Morokoff</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Fabinyi</surname>
<given-names>G.</given-names>
</name>
<etal/>
</person-group> (<year>2013</year>). <article-title>Prediction of seizure likelihood with a long-term, implanted seizure advisory system in patients with drug-resistant epilepsy: a first-in-man study</article-title>. <source>Lancet Neurology</source> <volume>12</volume>, <fpage>563</fpage>&#x2013;<lpage>571</lpage>. <pub-id pub-id-type="doi">10.1016/S1474-4422(13)70075-9</pub-id>
</citation>
</ref>
<ref id="B11">
<citation citation-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Da Silva</surname>
<given-names>H. P.</given-names>
</name>
<name>
<surname>Guerreiro</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Louren&#xe7;o</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Fred</surname>
<given-names>A. L.</given-names>
</name>
<name>
<surname>Martins</surname>
<given-names>R.</given-names>
</name>
</person-group> (<year>2014</year>). &#x201c;<article-title>BITalino: A novel hardware framework for physiological computing</article-title>,&#x201d; in <conf-name>International Conference on Physiological Computing Systems</conf-name>, <conf-loc>Lisbon, Portugal</conf-loc>, <conf-date>January 7-9, 2014</conf-date>, <fpage>246</fpage>&#x2013;<lpage>253</lpage>.</citation>
</ref>
<ref id="B12">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Detti</surname>
<given-names>P.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Siena scalp EEG database</article-title>. <comment>Version Number: 1.0.0 Type: dataset</comment>. <pub-id pub-id-type="doi">10.13026/5D4A-J060</pub-id>
</citation>
</ref>
<ref id="B13">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Detti</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Vatti</surname>
<given-names>G.</given-names>
</name>
<name>
<surname>Zabalo Manrique de Lara</surname>
<given-names>G.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>EEG synchronization analysis for seizure prediction: A study on data of noninvasive recordings</article-title>. <source>Processes</source> <volume>8</volume>, <fpage>846</fpage>. <pub-id pub-id-type="doi">10.3390/pr8070846</pub-id>
</citation>
</ref>
<ref id="B14">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Fisher</surname>
<given-names>R. S.</given-names>
</name>
<name>
<surname>Cross</surname>
<given-names>J. H.</given-names>
</name>
<name>
<surname>D&#x2019;Souza</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>French</surname>
<given-names>J. A.</given-names>
</name>
<name>
<surname>Haut</surname>
<given-names>S. R.</given-names>
</name>
<name>
<surname>Higurashi</surname>
<given-names>N.</given-names>
</name>
<etal/>
</person-group> (<year>2017a</year>). <article-title>Instruction manual for the ILAE 2017 operational classification of seizure types</article-title>. <source>Epilepsia</source> <volume>58</volume>, <fpage>531</fpage>&#x2013;<lpage>542</lpage>. <pub-id pub-id-type="doi">10.1111/epi.13671</pub-id>
</citation>
</ref>
<ref id="B15">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Fisher</surname>
<given-names>R. S.</given-names>
</name>
<name>
<surname>Cross</surname>
<given-names>J. H.</given-names>
</name>
<name>
<surname>French</surname>
<given-names>J. A.</given-names>
</name>
<name>
<surname>Higurashi</surname>
<given-names>N.</given-names>
</name>
<name>
<surname>Hirsch</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Jansen</surname>
<given-names>F. E.</given-names>
</name>
<etal/>
</person-group> (<year>2017b</year>). <article-title>Operational classification of seizure types by the international league against epilepsy: position paper of the ILAE commission for classification and terminology</article-title>. <source>Epilepsia</source> <volume>58</volume>, <fpage>522</fpage>&#x2013;<lpage>530</lpage>. <pub-id pub-id-type="doi">10.1111/epi.13670</pub-id>
</citation>
</ref>
<ref id="B16">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Frankel</surname>
<given-names>M. A.</given-names>
</name>
<name>
<surname>Lehmkuhle</surname>
<given-names>M. J.</given-names>
</name>
<name>
<surname>Watson</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Fetrow</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Frey</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Drees</surname>
<given-names>C.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>Electrographic seizure monitoring with a novel, wireless, single-channel EEG sensor</article-title>. <source>Clin. Neurophysiol. Pract.</source> <volume>6</volume>, <fpage>172</fpage>&#x2013;<lpage>178</lpage>. <pub-id pub-id-type="doi">10.1016/j.cnp.2021.04.003</pub-id>
</citation>
</ref>
<ref id="B54">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Gramfort</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Luessi</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Larson</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Engemann</surname>
<given-names>D. A.</given-names>
</name>
<name>
<surname>Strohmeier</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Brodbeck</surname>
<given-names>C.</given-names>
</name>
<etal/>
</person-group> (<year>2013</year>). <article-title>MEG and EEG data analysis with MNE-Python</article-title>. <source>Front. Neurosci.</source> <volume>7</volume>, <fpage>267</fpage>. <pub-id pub-id-type="doi">10.3389/fnins.2013.00267</pub-id>
</citation>
</ref>
<ref id="B17">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>G&#xf3;mez</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Arbel&#xe1;ez</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Navarrete</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Alvarado-Rojas</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Le Van Quyen</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Valderrama</surname>
<given-names>M.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Automatic seizure detection based on imaged-EEG signals through fully convolutional networks</article-title>. <source>Sci. Rep.</source> <volume>10</volume>, <fpage>21833</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-020-78784-3</pub-id>
</citation>
</ref>
<ref id="B18">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Johnstone</surname>
<given-names>J. A.</given-names>
</name>
<name>
<surname>Ford</surname>
<given-names>P. A.</given-names>
</name>
<name>
<surname>Hughes</surname>
<given-names>G.</given-names>
</name>
<name>
<surname>Watson</surname>
<given-names>T.</given-names>
</name>
<name>
<surname>Garrett</surname>
<given-names>A. T.</given-names>
</name>
</person-group> (<year>2012</year>). <article-title>Bioharness(<sup>TM</sup>) multivariable monitoring device: part. I: validity</article-title>. <source>J. Sports Sci. Med.</source> <volume>11</volume>, <fpage>400</fpage>&#x2013;<lpage>408</lpage>.</citation>
</ref>
<ref id="B19">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kenny</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Cullen</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Warrington</surname>
<given-names>G. D.</given-names>
</name>
</person-group> (<year>2017</year>). <article-title>The &#x201c;ice-mile&#x201d;: case study of 2 swimmers&#x2019; selected physiological responses and performance</article-title>. <source>Int. J. Sports Physiology Perform.</source> <volume>12</volume>, <fpage>711</fpage>&#x2013;<lpage>714</lpage>. <pub-id pub-id-type="doi">10.1123/ijspp.2016-0323</pub-id>
</citation>
</ref>
<ref id="B20">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Klatt</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Feldwisch-Drentrup</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Ihle</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Navarro</surname>
<given-names>V.</given-names>
</name>
<name>
<surname>Neufang</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Teixeira</surname>
<given-names>C.</given-names>
</name>
<etal/>
</person-group> (<year>2012</year>). <article-title>The epilepsiae database: an extensive electroencephalography database of epilepsy patients</article-title>. <source>Epilepsia</source> <volume>53</volume>, <fpage>1669</fpage>&#x2013;<lpage>1676</lpage>. <pub-id pub-id-type="doi">10.1111/j.1528-1167.2012.03564.x</pub-id>
</citation>
</ref>
<ref id="B21">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Korohoda</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Kolodziej</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Stepien</surname>
<given-names>J.</given-names>
</name>
</person-group> (<year>2013</year>). <article-title>Time-frequency analysis of accelerometry data for seizure detection</article-title>. <source>Bio-Algorithms Med-Systems</source> <volume>9</volume>, <fpage>65</fpage>&#x2013;<lpage>71</lpage>. <pub-id pub-id-type="doi">10.1515/bams-2013-0010</pub-id>
</citation>
</ref>
<ref id="B22">
<citation citation-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Koskim&#xe4;ki</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>M&#xf6;nttinen</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Siirtola</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Huttunen</surname>
<given-names>H.-L.</given-names>
</name>
<name>
<surname>Halonen</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>R&#xf6;ning</surname>
<given-names>J.</given-names>
</name>
</person-group> (<year>2017</year>). &#x201c;<article-title>Early detection of migraine attacks based on wearable sensors: experiences of data collection using Empatica E4</article-title>,&#x201d; in <conf-name>Proceedings of the 2017 ACM International Joint Conference on Pervasive and Ubiquitous Computing and Proceedings of the 2017 ACM International Symposium on Wearable Computers</conf-name>, <conf-loc>Maui Hawaii</conf-loc>, <conf-date>September 11 - 15, 2017</conf-date> (<publisher-name>ACM</publisher-name>), <fpage>506</fpage>&#x2013;<lpage>511</lpage>. <pub-id pub-id-type="doi">10.1145/3123024.3124434</pub-id>
</citation>
</ref>
<ref id="B23">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kuhlmann</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Karoly</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Freestone</surname>
<given-names>D. R.</given-names>
</name>
<name>
<surname>Brinkmann</surname>
<given-names>B. H.</given-names>
</name>
<name>
<surname>Temko</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Barachant</surname>
<given-names>A.</given-names>
</name>
<etal/>
</person-group> (<year>2018</year>). <article-title>Epilepsyecosystem.org: crowd-sourcing reproducible seizure prediction with long-term human intracranial EEG</article-title>. <source>Brain</source> <volume>141</volume>, <fpage>2619</fpage>&#x2013;<lpage>2630</lpage>. <pub-id pub-id-type="doi">10.1093/brain/awy210</pub-id>
</citation>
</ref>
<ref id="B24">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Leal</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Pinto</surname>
<given-names>M. F.</given-names>
</name>
<name>
<surname>Lopes</surname>
<given-names>F.</given-names>
</name>
<name>
<surname>Bianchi</surname>
<given-names>A. M.</given-names>
</name>
<name>
<surname>Henriques</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Ruano</surname>
<given-names>M. G.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>Heart rate variability analysis for the identification of the preictal interval in patients with drug-resistant epilepsy</article-title>. <source>Sci. Rep.</source> <volume>11</volume>, <fpage>5987</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-021-85350-y</pub-id>
</citation>
</ref>
<ref id="B25">
<citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname>Manso</surname>
<given-names>A. F.</given-names>
</name>
<name>
<surname>Fred</surname>
<given-names>A. L. N.</given-names>
</name>
<name>
<surname>Neves</surname>
<given-names>R. C.</given-names>
</name>
<name>
<surname>Ferreira</surname>
<given-names>R. C.</given-names>
</name>
</person-group> (<year>2018</year>). &#x201c;<article-title>Smart-wearables and heart-rate assessment accuracy</article-title>,&#x201d; in <source>2018 IEEE international symposium on medical measurements and applications (MeMeA)</source> (<publisher-loc>Rome</publisher-loc>: <publisher-name>IEEE</publisher-name>), <fpage>1</fpage>&#x2013;<lpage>5</lpage>. <pub-id pub-id-type="doi">10.1109/MeMeA.2018.8438812</pub-id>
</citation>
</ref>
<ref id="B26">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Mart&#xed;nez</surname>
<given-names>C. G. B.</given-names>
</name>
<name>
<surname>Niediek</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Mormann</surname>
<given-names>F.</given-names>
</name>
<name>
<surname>Andrzejak</surname>
<given-names>R. G.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Seizure onset zone lateralization using a non-linear analysis of micro vs. Macro electroencephalographic recordings during seizure-free stages of the sleep-wake cycle from epilepsy patients</article-title>. <source>Front. Neurology</source> <volume>11</volume>, <fpage>553885</fpage>. <pub-id pub-id-type="doi">10.3389/fneur.2020.553885</pub-id>
</citation>
</ref>
<ref id="B27">
<citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname>McCarthy</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Pradhan</surname>
<given-names>N.</given-names>
</name>
<name>
<surname>Redpath</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Adler</surname>
<given-names>A.</given-names>
</name>
</person-group> (<year>2016</year>). &#x201c;<article-title>Validation of the Empatica E4 wristband</article-title>,&#x201d; in <source>2016 IEEE EMBS international student conference (ISC)</source> (<publisher-loc>Ottawa, ON, Canada</publisher-loc>: <publisher-name>IEEE</publisher-name>), <fpage>1</fpage>&#x2013;<lpage>4</lpage>. <pub-id pub-id-type="doi">10.1109/EMBSISC.2016.7508621</pub-id>
</citation>
</ref>
<ref id="B28">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Meisel</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>El Atrache</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Jackson</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Schubach</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Ufongene</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Loddenkemper</surname>
<given-names>T.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Machine learning from wristband sensor data for wearable, noninvasive seizure forecasting</article-title>. <source>Epilepsia</source> <volume>61</volume>, <fpage>2653</fpage>&#x2013;<lpage>2666</lpage>. <pub-id pub-id-type="doi">10.1111/epi.16719</pub-id>
</citation>
</ref>
<ref id="B29">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Nagai</surname>
<given-names>Y.</given-names>
</name>
<name>
<surname>Jones</surname>
<given-names>C. I.</given-names>
</name>
<name>
<surname>Sen</surname>
<given-names>A.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Galvanic skin response (GSR)/Electrodermal/Skin conductance biofeedback on epilepsy: A systematic review and meta-analysis</article-title>. <source>Front. Neurology</source> <volume>10</volume>, <fpage>377</fpage>. <pub-id pub-id-type="doi">10.3389/fneur.2019.00377</pub-id>
</citation>
</ref>
<ref id="B30">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Nasseri</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Nurse</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Glasstetter</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>B&#xf6;ttcher</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Gregg</surname>
<given-names>N. M.</given-names>
</name>
<name>
<surname>Laks Nandakumar</surname>
<given-names>A.</given-names>
</name>
<etal/>
</person-group> (<year>2020</year>). <article-title>Signal quality and patient experience with wearable devices for epilepsy management</article-title>. <source>Epilepsia</source> <volume>61</volume>, <fpage>S25</fpage>&#x2013;<lpage>S35</lpage>. <pub-id pub-id-type="doi">10.1111/epi.16527</pub-id>
</citation>
</ref>
<ref id="B31">
<citation citation-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Nepi</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Sbrollini</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Agostinelli</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Maranesi</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Di Nardo</surname>
<given-names>F.</given-names>
</name>
<name>
<surname>Fioretti</surname>
<given-names>S.</given-names>
</name>
<etal/>
</person-group> (<year>2016</year>). &#x201c;<article-title>Validation of the heart:rate signal provided by the zephyr BioHarness 3.0</article-title>,&#x201d; in <conf-name>Proc. of the 43rd Computing in Cardiology Conference (CINC)</conf-name>, <conf-loc>Vancouver, BC, Canada</conf-loc>, <conf-date>11-14 September 2016</conf-date> (<publisher-name>IEEE</publisher-name>), <fpage>361</fpage>&#x2013;<lpage>364</lpage>. <pub-id pub-id-type="doi">10.22489/CinC.2016.106-358</pub-id>
</citation>
</ref>
<ref id="B32">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Obeid</surname>
<given-names>I.</given-names>
</name>
<name>
<surname>Picone</surname>
<given-names>J.</given-names>
</name>
</person-group> (<year>2016</year>). <article-title>The temple university hospital EEG data Corpus</article-title>. <source>Front. Neurosci.</source> <volume>10</volume>, <fpage>196</fpage>. <pub-id pub-id-type="doi">10.3389/fnins.2016.00196</pub-id>
</citation>
</ref>
<ref id="B33">
<citation citation-type="confproc">
<person-group person-group-type="author">
<name>
<surname>Ollander</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Godin</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Campagne</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Charbonnier</surname>
<given-names>S.</given-names>
</name>
</person-group> (<year>2016</year>). &#x201c;<article-title>A comparison of wearable and stationary sensors for stress detection</article-title>,&#x201d; in <conf-name>2016 IEEE International Conference on Systems, Man, and Cybernetics (SMC)</conf-name>, <conf-loc>Budapest, Hungary</conf-loc>, <conf-date>09-12 October 2016</conf-date> (<publisher-name>IEEE</publisher-name>), <fpage>004362</fpage>&#x2013;<lpage>004366</lpage>. <pub-id pub-id-type="doi">10.1109/SMC.2016.7844917</pub-id>
</citation>
</ref>
<ref id="B34">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Orphanidou</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Drobnjak</surname>
<given-names>I.</given-names>
</name>
</person-group> (<year>2017</year>). <article-title>Quality assessment of ambulatory ECG using wavelet entropy of the HRV signal</article-title>. <source>IEEE J. Biomed. Health Inf.</source> <volume>21</volume>, <fpage>1216</fpage>&#x2013;<lpage>1223</lpage>. <pub-id pub-id-type="doi">10.1109/JBHI.2016.2615316</pub-id>
</citation>
</ref>
<ref id="B35">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ranjan</surname>
<given-names>Y.</given-names>
</name>
<name>
<surname>Rashid</surname>
<given-names>Z.</given-names>
</name>
<name>
<surname>Stewart</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Conde</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Begale</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Verbeeck</surname>
<given-names>D.</given-names>
</name>
<etal/>
</person-group> (<year>2019</year>). <article-title>RADAR-Base: open source mobile health platform for collecting, monitoring, and analyzing data using sensors, wearables, and mobile devices</article-title>. <source>JMIR mHealth uHealth</source> <volume>7</volume>, <fpage>e11734</fpage>. <pub-id pub-id-type="doi">10.2196/11734</pub-id>
</citation>
</ref>
<ref id="B36">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Regalia</surname>
<given-names>G.</given-names>
</name>
<name>
<surname>Onorati</surname>
<given-names>F.</given-names>
</name>
<name>
<surname>Lai</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Caborni</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Picard</surname>
<given-names>R. W.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Multimodal wrist-worn devices for seizure detection and advancing research: focus on the Empatica wristbands</article-title>. <source>Epilepsy Res.</source> <volume>153</volume>, <fpage>79</fpage>&#x2013;<lpage>82</lpage>. <pub-id pub-id-type="doi">10.1016/j.eplepsyres.2019.02.007</pub-id>
</citation>
</ref>
<ref id="B37">
<citation citation-type="thesis">
<person-group person-group-type="author">
<name>
<surname>Shoeb</surname>
<given-names>A. H.</given-names>
</name>
</person-group> (<year>2009</year>). &#x201c;<article-title>Application of machine learning to epilepsy seizure onset detection and treatment</article-title>,&#x201d;. <comment>Doutoramento, Harvard</comment> (<publisher-name>MIT</publisher-name>).</citation>
</ref>
<ref id="B38">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sivathamboo</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Friedman</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Laze</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Nightscales</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>Z.</given-names>
</name>
<name>
<surname>Kuhlmann</surname>
<given-names>L.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>Association of short-term heart rate variability and sudden unexpected Death in epilepsy</article-title>. <source>Neurology</source> <volume>97</volume>, <fpage>e2357</fpage>&#x2013;<lpage>e2367</lpage>. <pub-id pub-id-type="doi">10.1212/WNL.0000000000012946</pub-id>
</citation>
</ref>
<ref id="B39">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Simblett</surname>
<given-names>S. K.</given-names>
</name>
<name>
<surname>Biondi</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Bruno</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Ballard</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Stoneman</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Lees</surname>
<given-names>S.</given-names>
</name>
<etal/>
</person-group> (<year>2020</year>). <article-title>Patients&#x2019; experience of wearing multimodal sensor devices intended to detect epileptic seizures: A qualitative analysis</article-title>. <source>Epilepsy &#x26; Behav.</source> <volume>102</volume>, <fpage>106717</fpage>. <pub-id pub-id-type="doi">10.1016/j.yebeh.2019.106717</pub-id>
</citation>
</ref>
<ref id="B40">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sivathamboo</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Constantino</surname>
<given-names>T. N.</given-names>
</name>
<name>
<surname>Chen</surname>
<given-names>Z.</given-names>
</name>
<name>
<surname>Sparks</surname>
<given-names>P. B.</given-names>
</name>
<name>
<surname>Goldin</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Velakoulis</surname>
<given-names>D.</given-names>
</name>
<etal/>
</person-group> (<year>2020</year>). <article-title>Cardiorespiratory and autonomic function in epileptic seizures: A video-EEG monitoring study</article-title>. <source>Epilepsy &#x26; Behav.</source> <volume>111</volume>, <fpage>107271</fpage>. <pub-id pub-id-type="doi">10.1016/j.yebeh.2020.107271</pub-id>
</citation>
</ref>
<ref id="B41">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Smith</surname>
<given-names>C. M.</given-names>
</name>
<name>
<surname>Chillrud</surname>
<given-names>S. N.</given-names>
</name>
<name>
<surname>Jack</surname>
<given-names>D. W.</given-names>
</name>
<name>
<surname>Kinney</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Yang</surname>
<given-names>Q.</given-names>
</name>
<name>
<surname>Layton</surname>
<given-names>A. M.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Laboratory validation of hexoskin biometric shirt at rest, submaximal exercise, and maximal exercise while riding a stationary bicycle</article-title>. <source>J. Occup. Environ. Med.</source> <volume>61</volume>, <fpage>e104</fpage>&#x2013;<lpage>e111</lpage>. <pub-id pub-id-type="doi">10.1097/JOM.0000000000001537</pub-id>
</citation>
</ref>
<ref id="B42">
<citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname>Swami</surname>
<given-names>P.</given-names>
</name>
<name>
<surname>Panigrahi</surname>
<given-names>B. K.</given-names>
</name>
<name>
<surname>Nara</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Bhatia</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Gandhi</surname>
<given-names>T.</given-names>
</name>
</person-group> (<year>2016</year>). <source>EEG epilepsy datasets</source>. <comment>Publisher: Unpublished</comment>. <pub-id pub-id-type="doi">10.13140/RG.2.2.14280.32006</pub-id>
</citation>
</ref>
<ref id="B43">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Swinnen</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Chatzichristos</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Jansen</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Lagae</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Depondt</surname>
<given-names>C.</given-names>
</name>
<name>
<surname>Seynaeve</surname>
<given-names>L.</given-names>
</name>
<etal/>
</person-group> (<year>2021</year>). <article-title>Accurate detection of typical absence seizures in adults and children using a two-channel electroencephalographic wearable behind the ears</article-title>. <source>Epilepsia</source> <volume>62</volume>, <fpage>2741</fpage>&#x2013;<lpage>2752</lpage>. <pub-id pub-id-type="doi">10.1111/epi.17061</pub-id>
</citation>
</ref>
<ref id="B44">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Tedesco</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Sica</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Ancillao</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Timmons</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Barton</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>O&#x2019;Flynn</surname>
<given-names>B.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Validity evaluation of the fitbit Charge2 and the garmin vivosmart HR&#x2b; in free-living environments in an older adult cohort</article-title>. <source>JMIR mHealth uHealth</source> <volume>7</volume>, <fpage>e13084</fpage>. <pub-id pub-id-type="doi">10.2196/13084</pub-id>
</citation>
</ref>
<ref id="B45">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Truong</surname>
<given-names>N. D.</given-names>
</name>
<name>
<surname>Kuhlmann</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Bonyadi</surname>
<given-names>M. R.</given-names>
</name>
<name>
<surname>Querlioz</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Zhou</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Kavehei</surname>
<given-names>O.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Epileptic seizure forecasting with generative adversarial networks</article-title>. <source>IEEE Access</source> <volume>7</volume>, <fpage>143999</fpage>&#x2013;<lpage>144009</lpage>. <pub-id pub-id-type="doi">10.1109/ACCESS.2019.2944691</pub-id>
</citation>
</ref>
<ref id="B46">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Van de Vel</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Cuppens</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Bonroy</surname>
<given-names>B.</given-names>
</name>
<name>
<surname>Milosevic</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Jansen</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Van Huffel</surname>
<given-names>S.</given-names>
</name>
<etal/>
</person-group> (<year>2013</year>). <article-title>Non-EEG seizure-detection systems and potential SUDEP prevention: state of the art</article-title>. <source>Seizure</source> <volume>22</volume>, <fpage>345</fpage>&#x2013;<lpage>355</lpage>. <pub-id pub-id-type="doi">10.1016/j.seizure.2013.02.012</pub-id>
</citation>
</ref>
<ref id="B47">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vandecasteele</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>De Cooman</surname>
<given-names>T.</given-names>
</name>
<name>
<surname>Gu</surname>
<given-names>Y.</given-names>
</name>
<name>
<surname>Cleeren</surname>
<given-names>E.</given-names>
</name>
<name>
<surname>Claes</surname>
<given-names>K.</given-names>
</name>
<name>
<surname>Paesschen</surname>
<given-names>W.</given-names>
</name>
<etal/>
</person-group> (<year>2017</year>). <article-title>Automated epileptic seizure detection based on wearable ECG and PPG in a hospital environment</article-title>. <source>Sensors</source> <volume>17</volume>, <fpage>2338</fpage>. <pub-id pub-id-type="doi">10.3390/s1710-2338</pub-id>
</citation>
</ref>
<ref id="B48">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Verdru</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Van Paesschen</surname>
<given-names>W.</given-names>
</name>
</person-group> (<year>2020</year>). <article-title>Wearable seizure detection devices in refractory epilepsy</article-title>. <source>Acta Neurol. Belg.</source> <volume>120</volume>, <fpage>1271</fpage>&#x2013;<lpage>1281</lpage>. <pub-id pub-id-type="doi">10.1007/s13760-020-01417-z</pub-id>
</citation>
</ref>
<ref id="B49">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vieluf</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>El Atrache</surname>
<given-names>R.</given-names>
</name>
<name>
<surname>Cantley</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Jackson</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>Clark</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Sheehan</surname>
<given-names>T.</given-names>
</name>
<etal/>
</person-group> (<year>2022</year>). <article-title>Seizure-related differences in biosignal 24-h modulation patterns</article-title>. <source>Sci. Rep.</source> <volume>12</volume>, <fpage>15070</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-022-18271-z</pub-id>
</citation>
</ref>
<ref id="B50">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vijayan</surname>
<given-names>V.</given-names>
</name>
<name>
<surname>Connolly</surname>
<given-names>J. P.</given-names>
</name>
<name>
<surname>Condell</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>McKelvey</surname>
<given-names>N.</given-names>
</name>
<name>
<surname>Gardiner</surname>
<given-names>P.</given-names>
</name>
</person-group> (<year>2021</year>). <article-title>Review of wearable devices and data collection considerations for connected health</article-title>. <source>Sensors</source> <volume>21</volume>, <fpage>5589</fpage>. <pub-id pub-id-type="doi">10.3390/s21165589</pub-id>
</citation>
</ref>
<ref id="B51">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Weenk</surname>
<given-names>M.</given-names>
</name>
<name>
<surname>van Goor</surname>
<given-names>H.</given-names>
</name>
<name>
<surname>Frietman</surname>
<given-names>B.</given-names>
</name>
<name>
<surname>Engelen</surname>
<given-names>L. J.</given-names>
</name>
<name>
<surname>van Laarhoven</surname>
<given-names>C. J.</given-names>
</name>
<name>
<surname>Smit</surname>
<given-names>J.</given-names>
</name>
<etal/>
</person-group> (<year>2017</year>). <article-title>Continuous monitoring of vital signs using wearable devices on the general ward: pilot study</article-title>. <source>JMIR mHealth uHealth</source> <volume>5</volume>, <fpage>e91</fpage>. <pub-id pub-id-type="doi">10.2196/mhealth.7208</pub-id>
</citation>
</ref>
<ref id="B52">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Whitmire</surname>
<given-names>L.</given-names>
</name>
<name>
<surname>Voyles</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Cardenas</surname>
<given-names>D.</given-names>
</name>
<name>
<surname>Cavazos</surname>
<given-names>J.</given-names>
</name>
</person-group> (<year>2019</year>). <article-title>Diagnostic utility of continuous sEMG monitoring in a home setting-real-world use of the speac&#xae; system (p4. 5-012)</article-title>. <source>Neurology</source> <volume>92</volume>.</citation>
</ref>
<ref id="B53">
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Wong</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Simmons</surname>
<given-names>A.</given-names>
</name>
<name>
<surname>Rivera-Villicana</surname>
<given-names>J.</given-names>
</name>
<name>
<surname>Barnett</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Sivathamboo</surname>
<given-names>S.</given-names>
</name>
<name>
<surname>Perucca</surname>
<given-names>P.</given-names>
</name>
<etal/>
</person-group> (<year>2023</year>). <article-title>Eeg datasets for seizure detection and prediction&#x2014;A review</article-title>. <source>Epilepsia Open</source> <volume>8</volume>, <fpage>252</fpage>&#x2013;<lpage>267</lpage>. <pub-id pub-id-type="doi">10.1002/epi4.12704</pub-id>
</citation>
</ref>
</ref-list>
</back>
</article>