<article-meta>
<article-id pub-id-type="doi">10.3389/fphys.2017.00659</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Physiology</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>T Wave Safety Margin during the Process of ICD Implantation As a Novel Predictor of T Wave Oversensing</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Sun</surname> <given-names>Ya-Xun</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="author-notes" rid="fn004"><sup>&#x02020;</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/458938/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Gao</surname> <given-names>Jing</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="author-notes" rid="fn004"><sup>&#x02020;</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/470773/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Jiang</surname> <given-names>Chen-Yang</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Xue</surname> <given-names>Yu-Mei</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Xu</surname> <given-names>Yi-Zhou</given-names></name>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Liu</surname> <given-names>Gang</given-names></name>
<xref ref-type="aff" rid="aff4"><sup>4</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/451874/overview"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Guo</surname> <given-names>Ji-Hong</given-names></name>
<xref ref-type="aff" rid="aff5"><sup>5</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Sheng</surname> <given-names>Xia</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Ye</surname> <given-names>Yang</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>He</surname> <given-names>Hong</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Zhao</surname> <given-names>Yun-Tao</given-names></name>
<xref ref-type="aff" rid="aff6"><sup>6</sup></xref>
</contrib>
<contrib contrib-type="author">
<name><surname>Barajas-Martinez</surname> <given-names>Hector</given-names></name>
<xref ref-type="aff" rid="aff7"><sup>7</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/186696/overview"/>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name><surname>Fu</surname> <given-names>Guo-Sheng</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="author-notes" rid="fn001"><sup>&#x0002A;</sup></xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name><surname>Hu</surname> <given-names>Dan</given-names></name>
<xref ref-type="aff" rid="aff7"><sup>7</sup></xref>
<xref ref-type="aff" rid="aff8"><sup>8</sup></xref>
<xref ref-type="author-notes" rid="fn002"><sup>&#x0002A;</sup></xref>
<uri xlink:href="http://loop.frontiersin.org/people/29301/overview"/>
</contrib>
</contrib-group>
<aff id="aff1"><sup>1</sup><institution>Department of Cardiology, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine</institution> <country>Hangzhou, China</country></aff>
<aff id="aff2"><sup>2</sup><institution>Department of Cardiology, Guangdong Cardiovascular Institute, Guangdong General Hospital</institution> <country>Guangzhou, China</country></aff>
<aff id="aff3"><sup>3</sup><institution>Department of Cardiology, Hangzhou First People&#x00027;s Hospital</institution> <country>Hangzhou, China</country></aff>
<aff id="aff4"><sup>4</sup><institution>Department of Cardiology, First Hospital of Hebei Medical University</institution> <country>Shijiazhuang, China</country></aff>
<aff id="aff5"><sup>5</sup><institution>Division of Cardiology, Peking University People&#x00027;s Hospital</institution> <country>Beijing, China</country></aff>
<aff id="aff6"><sup>6</sup><institution>Department of Cardiology, Aerospace Center Hospital</institution> <country>Beijing, China</country></aff>
<aff id="aff7"><sup>7</sup><institution>Masonic Medical Research Laboratory</institution> <country>Utica, NY, United States</country></aff>
<aff id="aff8"><sup>8</sup><institution>Department of Cardiology and Cardiovascular Research Institute, Renmin Hospital of Wuhan University</institution> <country>Wuhan, China</country></aff>
<author-notes>
<fn fn-type="edited-by"><p>Edited by: Jiashin Wu, University of South Florida, United States</p></fn>
<fn fn-type="edited-by"><p>Reviewed by: Lukas R. C. Dekker, Catharina Ziekenhuis, Netherlands; Jonas S. De Jong, Onze Lieve Vrouwe Gasthuis, Netherlands</p></fn>
<fn fn-type="corresp" id="fn001"><p>&#x0002A;Correspondence: Guo-Sheng Fu <email>fugs&#x00040;zju.edu.cn</email></p></fn>
<fn fn-type="corresp" id="fn002"><p>Dan Hu <email>dianah&#x00040;mmrl.edu</email></p></fn>
<fn fn-type="other" id="fn003"><p>This article was submitted to Cardiac Electrophysiology, a section of the journal Frontiers in Physiology</p></fn>
<fn fn-type="other" id="fn004"><p>&#x02020;These authors have contributed equally to this work.</p></fn></author-notes>
<pub-date pub-type="epub">
<day>01</day>
<month>09</month>
<year>2017</year>
</pub-date>
<pub-date pub-type="collection">
<year>2017</year>
</pub-date>
<volume>8</volume>
<elocation-id>659</elocation-id>
<history>
<date date-type="received">
<day>17</day>
<month>06</month>
<year>2017</year>
</date>
<date date-type="accepted">
<day>18</day>
<month>08</month>
<year>2017</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#x000A9; 2017 Sun, Gao, Jiang, Xue, Xu, Liu, Guo, Sheng, Ye, He, Zhao, Barajas-Martinez, Fu and Hu.</copyright-statement>
<copyright-year>2017</copyright-year>
<copyright-holder>Sun, Gao, Jiang, Xue, Xu, Liu, Guo, Sheng, Ye, He, Zhao, Barajas-Martinez, Fu and Hu</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/"><p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p></license>
</permissions>
<abstract><p><bold>Introduction:</bold> T wave oversensing (TWOS) is a major drawback of implantable cardioverter defibrillator (ICD) and data on predictors of TWOS in ICD is limited. We aimed to calculate a novel index of T wave safety margin (TWSM) and assess its potential for evaluating TWOS during the procedure of ICD implantation.</p>
<p><bold>Methods and Results:</bold> Thirty-two consecutive patients with ICD implantation were enrolled. During each procedure of ICD implantation, different ICD generators were connected to implanted sensing lead through active-fixation leads and bridging cables. R and T wave amplitudes were measured on ICD printouts according to the gain. The ICDs were programed to the most sensitive settings to reveal possible TWOS. A novel index TWSM was calculated according to the corresponding sensing algorithm of ICD. There was discrepancy of R wave amplitudes measured by different ICDs (<italic>P</italic> &#x0003C; 0.01). In Fortify and Teligen ICDs, T wave amplitudes showed no difference (<italic>P</italic> &#x0003E; 0.05) and TWSMs were sufficiently high (post sensing: 13.0 &#x000B1; 7.6 and 28.3 &#x000B1; 16.5, respectively, post pacing: 5.0 &#x000B1; 2.2 and 4.6 &#x000B1; 0.9, respectively). In nine patients with 10 TWOS episodes detected during the procedure of ICD implantation, generators with the highest TWSM were chosen. Only one TWOS episode during pacing was recorded during the 25 &#x000B1; 7 mo follow-up period.</p>
<p><bold>Conclusions:</bold> We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM might help to reduce TWOS episodes in patients with high risk of TWOS. Prospective studies are warranted to validate this index and its potential to reduce TWOS episodes.</p></abstract>
<kwd-group>
<kwd>implantable cardioverter-defibrillator</kwd>
<kwd>T wave oversensing</kwd>
<kwd>safety margin</kwd>
<kwd>inappropriate therapy</kwd>
<kwd>intracardiac electrocardiogram</kwd>
</kwd-group>
<counts>
<fig-count count="5"/>
<table-count count="2"/>
<equation-count count="0"/>
<ref-count count="29"/>
<page-count count="9"/>
<word-count count="5320"/>
</counts>
</article-meta>
</front>
<body>
<sec sec-type="intro" id="s1">
<title>Introduction</title>
<p>Implantable cardioverter-defibrillators (ICDs), either alone or with cardiac resynchronization therapy (CRT), reduce mortality in patients at high risk for cardiac death (Moss et al., <xref ref-type="bibr" rid="B12">1996</xref>, <xref ref-type="bibr" rid="B14">2002</xref>, <xref ref-type="bibr" rid="B13">2009</xref>). However, inappropriate ICD therapy can result in impaired quality of life, psychiatric disturbance, and life-threatening ventricular arrhythmias (Mark et al., <xref ref-type="bibr" rid="B11">2008</xref>; Juan and Pollack, <xref ref-type="bibr" rid="B4">2010</xref>; van Rees et al., <xref ref-type="bibr" rid="B27">2011</xref>; Powell et al., <xref ref-type="bibr" rid="B19">2013</xref>). How to reduce inappropriate therapy is a significant challenge in ICD patients.</p>
<p>T wave oversensing (TWOS) is one of the major reasons of inappropriate ICD therapy. In some subgroups of patients, such as those with Brugada syndrome, long QT syndrome, short QT syndrome, hypertrophic cardiomyopathy, and younger patients (Schimpf et al., <xref ref-type="bibr" rid="B23">2003</xref>; Weretka et al., <xref ref-type="bibr" rid="B28">2003</xref>; Sacher et al., <xref ref-type="bibr" rid="B20">2006</xref>, <xref ref-type="bibr" rid="B21">2013</xref>; Magnusson et al., <xref ref-type="bibr" rid="B10">2015</xref>; Olde Nordkamp et al., <xref ref-type="bibr" rid="B17">2016</xref>), TWOS after ICD implantation has been reported in 3 &#x0007E; 14% of these patients. However, the studies focusing on the predictors of TWOS in ICD are limited (Maesato et al., <xref ref-type="bibr" rid="B9">2011</xref>). In these patients, evaluating the risk of TWOS during the procedure of ICD implantation might help in reducing TWOS episodes.</p>
<p>In this study, we introduced a novel TWSM index and evaluated its potential of reducing TWOS episodes in mid-term follow up.</p>
</sec>
<sec sec-type="methods" id="s2">
<title>Methods</title>
<p>This study was approved by the Ethics Committee of the Sir Run Run Shaw Hospital at Zhejiang University. Informed written consents were obtained from all study participants.</p>
<sec>
<title>Clinical and device characteristics</title>
<p>From April 2013 to July 2014, 32 consecutive patients with ICD implantation indices were enrolled. The average age of the 32 patients (59.4% male) was 54.9 &#x000B1; 12 y. The indications for ICD implantation and clinical characteristic of the patients are summarized in Table <xref ref-type="table" rid="T1">1</xref>. The case with short QT syndrome was reported previously (Sun et al., <xref ref-type="bibr" rid="B25">2010</xref>, <xref ref-type="bibr" rid="B24">2011</xref>). Six ICDs (CRT/D) from three manufacturers (St. Jude Medical, Inc.; Medtronic, Inc.; and Boston Scientific, Inc.) were applied as testing ICDs in this study; all testing ICDs were functionally normal.</p>
<table-wrap position="float" id="T1">
<label>Table 1</label>
<caption><p>Demographic and clinical characteristics of patients.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th/>
<th valign="top" align="center"><bold>Patients (<italic>n</italic> &#x0003D; 32)</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Age, y/o (Mean &#x000B1;<italic>SD</italic>)</td>
<td valign="top" align="center">54.9 &#x000B1; 12.3</td>
</tr>
<tr>
<td valign="top" align="left">Male, <italic>n</italic> (%)</td>
<td valign="top" align="center">19 (59%)</td>
</tr>
<tr>
<td valign="top" align="left">Primary prevention, <italic>n</italic> (%)</td>
<td valign="top" align="center">23 (72%)</td>
</tr>
<tr>
<td valign="top" align="left">Secondary prevention, <italic>n</italic> (%)</td>
<td valign="top" align="center">9 (28%)</td>
</tr>
<tr>
<td valign="top" align="left" colspan="2" style="background-color:#bdbec1"><bold>INDICATIONS FOR ICD IMPLANTATION, <italic>n</italic> (%)</bold></td>
</tr>
<tr>
<td valign="top" align="left">Dilated cardiomyopathy</td>
<td valign="top" align="center">10 (31%)</td>
</tr>
<tr>
<td valign="top" align="left">Reduced LVEF post-myocardial infarction</td>
<td valign="top" align="center">6 (19%)</td>
</tr>
<tr>
<td valign="top" align="left">Survived idiopathic cardiac arrest</td>
<td valign="top" align="center">5 (16%)</td>
</tr>
<tr>
<td valign="top" align="left">Brugada syndrome</td>
<td valign="top" align="center">4 (13%)</td>
</tr>
<tr>
<td valign="top" align="left">Hypertrophic cardiomyopathy</td>
<td valign="top" align="center">3 (9%)</td>
</tr>
<tr>
<td valign="top" align="left">Long-QT syndrome</td>
<td valign="top" align="center">2 (6%)</td>
</tr>
<tr>
<td valign="top" align="left">Alcoholic cardiomyopathy</td>
<td valign="top" align="center">1 (3%)</td>
</tr>
<tr>
<td valign="top" align="left">Short-QT syndrome</td>
<td valign="top" align="center">1 (3%)</td>
</tr>
<tr>
<td valign="top" align="left">LVEF, % (<italic>n</italic>)</td>
<td valign="top" align="center">48 &#x000B1; 19 (32)</td>
</tr>
<tr>
<td valign="top" align="left">Dilated cardiomyopathy</td>
<td valign="top" align="center">31 &#x000B1; 3(10)</td>
</tr>
<tr>
<td valign="top" align="left">Ischemic cardiomyopathy</td>
<td valign="top" align="center">33 &#x000B1; 2 (6)</td>
</tr>
<tr>
<td valign="top" align="left">Hypertrophic cardiomyopathy</td>
<td valign="top" align="center">76 &#x000B1; 5 (3)</td>
</tr>
<tr>
<td valign="top" align="left">Arrhythmias related</td>
<td valign="top" align="center">62 &#x000B1; 10 (13)</td>
</tr>
<tr>
<td valign="top" align="left">Serum potassium (mmol/L)</td>
<td valign="top" align="center">3.9 &#x000B1; 0.3</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p><italic>ICD, Implantable cardioverter-defibrillator; LVEF, left ventricular ejection fraction</italic>.</p>
</table-wrap-foot>
</table-wrap>
</sec>
<sec>
<title><italic>Ex vivo</italic> ICD connection</title>
<p><italic>Ex vivo</italic> connection was designed to enable evaluation of R and T wave amplitudes across all six testing ICDs in each patient. ICD implantation was performed under local anesthesia. All the shock leads were fixed in the lower ventricular septal. All the 32 defibrillation leads were dedicated bipolar leads. A bridging cable was connected to the connector pins of sensing lead by bipolar mode at 10 min after lead fixation to minimize the influence of acute injury current (Saxonhouse et al., <xref ref-type="bibr" rid="B22">2005</xref>). According to the reported study and our experience, 10 min is long enough for normalizing acute injury current, and diminishing its influence on ST-T amplitude. The other end of the bridging cable was clipped onto both the tip and the ring electrode of an active-fixation lead (Guidant Fineline II EZ 4471). The active-fixation lead was then connected to the ventricular sensing channel of testing ICD. The testing ICD was programed by the corresponding ICD programmer (Figure <xref ref-type="fig" rid="F1">1A</xref>). The gain of the printed intracardiac electrocardiogram (IEGM) was set to an optimal value to avoid signal clipping and be eligible for T wave measurement.</p>
<fig id="F1" position="float">
<label>Figure 1</label>
<caption><p><bold>(A)</bold> <italic>Ex vivo</italic> ICD connection during the procedure of ICD implantation using the active-fixation lead. <bold>(B)</bold> TWSM measurements. The green line indicates the sensing threshold curve of ICD. <bold>(C)</bold> The timeline for the evaluation of TWSM.</p></caption>
<graphic xlink:href="fphys-08-00659-g0001.tif"/>
</fig>
<p>After being connected to testing ICDs, ventricular IEGMs were printed by ICD programmers. The ventricular amplitude was measured by two methods: R wave sensing test by ICD programming, and signal measurement on ICD printouts. Three consecutive R waves were measured for an average value. T wave amplitudes post sensing and post pacing were measured by the same method (Figures <xref ref-type="fig" rid="F1">1A,B</xref>).</p>
</sec>
<sec>
<title>TWOS and TWSM tests</title>
<p>After connected to the active-fixation lead, each testing ICD was programed to the most sensitive setting to reveal possible TWOS. TWOS episode was evaluated by two physicians with expertise in electrophysiology. If TWOS occurred, ventricular sensitivity was reduced and retested till TWOS was resolved. If TWOS occurred in certain type of ICD, another type of ICD generator with highest TWSM was implanted to avoid possible TWOS in the future. After the implantation, the ICD parameters were programed to the nominal setting unless changes were necessary. The timeline for TWSM is shown in Figure <xref ref-type="fig" rid="F1">1C</xref>. During each follow-up, the ICDs were temporarily programed to the most sensitive setting to reveal possible TWOS.</p>
<p>The TWSM was defined as the ratio of the ventricular sensing threshold at the T wave peak on IEGM to T wave amplitude (Figure <xref ref-type="fig" rid="F1">1B</xref>). The details to calculate TWSM in different ICDs were shown in Supplementary Table <xref ref-type="supplementary-material" rid="SM1">1</xref>.</p>
</sec>
<sec>
<title>Statistical analyses</title>
<p>All analyses were conducted using SPSS software (SPSS Inc., Chicago, IL, USA). Conformity to the normal distribution was evaluated for continuous variables with the Kolmogorov&#x02013;Smirnov test. Comparisons between groups were made using a paired <italic>t</italic>-test, and One way repeated measure ANOVA was conducted to test paired data among three or more groups. R wave amplitudes measured by ICD programmers and on ICD printouts were compared using the Bland-Altman method as well as with linear regression and Pearson correlation analyses. Continuous variables are expressed as mean &#x000B1; standard deviation. A two-tailed <italic>P</italic> &#x0003C; 0.05 was considered statistically significant.</p>
</sec>
</sec>
<sec sec-type="results" id="s3">
<title>Results</title>
<sec>
<title>Basic characteristic</title>
<p>Patient characteristics are summarized in Table <xref ref-type="table" rid="T1">1</xref>. Briefly, 59% of patients were men, with a mean age of 54.9 &#x000B1; 12.3 years, mostly for primary prevention (72%) and with a mean serum potassium of 3.9 &#x000B1; 0.3 mmol/L. Mean follow-up duration was 25 &#x000B1; 7 months.</p>
</sec>
<sec>
<title>R wave and T wave amplitudes</title>
<p>The R wave amplitude measured by ICD was consistent with that measured on &#x0201C;Ventricular Amplifier&#x0201D; channel in St. Jude ICDs and that on the &#x0201C;V&#x0201D; channel in the Teligen ICD, but was not consistent with &#x0201C;Pace/sense (bipolar)&#x0201D; channels of other four ICDs (Figure <xref ref-type="fig" rid="F2">2</xref>).</p>
<fig id="F2" position="float">
<label>Figure 2</label>
<caption><p>Bland-Altman analysis comparing R wave amplitudes measured from implantable cardioverter-defibrillator printouts. <bold>(A)</bold> Measurement of R wave amplitude by Fortify ICD. Noted that IEGM of ICD automatically switched to &#x0201C;Ventricular amplifier&#x0201D; channel to measure R wave amplitude. <bold>(B,C)</bold> R wave amplitude measured from by Fortify ICD. <bold>(D,E)</bold> R wave amplitude measured by Virtuoso ICD. Inner dashed line is the mean difference; outer dashed lines are mean &#x000B1; 1.96 SD.</p></caption>
<graphic xlink:href="fphys-08-00659-g0002.tif"/>
</fig>
<p>R wave amplitudes measured by five testing ICDs (Fortify, Virtuoso, Consulta, Vitality, and Teligen) were different (<italic>P</italic> &#x0003C; 0.01, <italic>n</italic> &#x0003D; 28, Figure <xref ref-type="fig" rid="F3">3A</xref>). Four patients were not included in R wave amplitude analysis, because their R wave amplitudes measured by Fortify ICDs exceeded 12 mV and could not be measured by ICD programmer accurately (only &#x0201C;&#x0003E;12 mV&#x0201D; was shown).</p>
<fig id="F3" position="float">
<label>Figure 3</label>
<caption><p>Comparison of R wave amplitude, T wave amplitudes and TWSMs by different ICDs. <bold>(A)</bold>. R wave amplitudes by different ICDs; <bold>(B,C)</bold> T wave amplitudes post sensing intrinsic rhythm and post pacing at 90/min in Fortify and Teligen ICDs. <bold>(D)</bold> TWSM after sensing intrinsic rhythm and pacing in Fortify and Teligen. Red line indicates safety margin of 2 (200%). <sup>&#x0002A;</sup><italic>P</italic> &#x0003C; 0.05; <sup>&#x0002A;&#x0002A;</sup><italic>P</italic> &#x0003C; 0.01.</p></caption>
<graphic xlink:href="fphys-08-00659-g0003.tif"/>
</fig>
<p>T wave amplitudes post pacing were higher than those post intrinsic rhythm (Fortify: 0.39 &#x000B1; 0.13 vs. 0.24 &#x000B1; 0.11 mV; <italic>P</italic> &#x0003C; 0.001, <italic>n</italic> &#x0003D; 21; Teligen: 0.36 &#x000B1; 0.11 vs. 0.24 &#x000B1; 0.10 mV; <italic>P</italic> &#x0003C; 0.01, <italic>n</italic> &#x0003D; 23. Figure <xref ref-type="fig" rid="F3">3B</xref>). There were no difference between T wave amplitudes measured by Fortify and Teligen ICDs (post-sensing: 0.26 &#x000B1; 0.15 vs. 0.27 &#x000B1; 0.19 mV; <italic>P</italic> &#x0003D; 0.81, <italic>n</italic> &#x0003D; 32; Post-pacing: 0.39 &#x000B1; 0.11 vs. 0.38 &#x000B1; 0.10 mV; <italic>P</italic> &#x0003D; 0.72, <italic>n</italic> &#x0003D; 19; Figure <xref ref-type="fig" rid="F3">3C</xref>).</p>
</sec>
<sec>
<title>TWOS testing</title>
<p>In the TWOS test, nine patients were positive (28%). There were two sensing TWOS events and eight pacing TWOS events under the maximum sensitivity during the procedure of ICD implantation (Table <xref ref-type="table" rid="T2">2</xref>). In the short QT syndrome patient, Virtuoso and Consulta ICDs showed TWOS at the maximum sensitivity of 0.15 mV at R-T interval of 200 ms (Figures <xref ref-type="fig" rid="F4">4A,B</xref>). No TWOS event was recorded at the sensitivity of 0.3 mV. The Epic ICD showed intermittent TWOS at the interval of 172 ms between R wave and T wave sensing at decay delay of 0 ms and threshold start of 50%. No TWOS event occurred at decay delay of 0 ms and threshold start of 62.5% (Figure <xref ref-type="fig" rid="F4">4C</xref>). Fortify, Vitality, and Teligen ICDs showed no sensing or pacing TWOS under maximum sensitivity (Figures <xref ref-type="fig" rid="F4">4D&#x02013;F</xref>). The other sensing TWOS occurred in a patient with HCM, who also showed post pacing TWOS under the maximum sensitivity during the procedure of implantation.</p>
<table-wrap position="float" id="T2">
<label>Table 2</label>
<caption><p>Details of patients with TWOS during the procedure of ICD implantation.</p></caption>
<table frame="hsides" rules="groups">
<thead><tr>
<th valign="top" align="left"><bold>&#x02018;Patient</bold></th>
<th valign="top" align="left"><bold>Age, sex</bold></th>
<th valign="top" align="left"><bold>Patho-genesis</bold></th>
<th valign="top" align="center"><bold>Indication</bold></th>
<th valign="top" align="center" colspan="6" style="border-bottom: thin solid #000000;"><bold>Sensitivity</bold></th>
<th valign="top" align="left"><bold>ICD implanted</bold></th>
<th valign="top" align="left"><bold>Defibrillation lead</bold></th>
</tr>
<tr>
<th/>
<th/>
<th/>
<th/>
<th valign="top" align="center" colspan="3" style="border-bottom: thin solid #000000;"><bold>0.15 mV</bold></th>
<th valign="top" align="center" colspan="3" style="border-bottom: thin solid #000000;"><bold>0.3 mV</bold></th>
<th/>
<th/>
</tr>
<tr>
<th/>
<th/>
<th/>
<th/>
<th valign="top" align="center"><bold>VS</bold></th>
<th valign="top" align="center"><bold>VP</bold></th>
<th valign="top" align="center"><bold>PSVP</bold></th>
<th valign="top" align="center"><bold>VS</bold></th>
<th valign="top" align="center"><bold>VP</bold></th>
<th valign="top" align="left"><bold>PSVP</bold></th>
<th/>
<th/>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">1</td>
<td valign="top" align="left">54, M</td>
<td valign="top" align="left">SQTS</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Vitality</td>
<td valign="top" align="left">Riata 1570</td>
</tr>
<tr>
<td valign="top" align="left">2</td>
<td valign="top" align="left">56, M</td>
<td valign="top" align="left">Brugada</td>
<td valign="top" align="center">Secondary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x000B1;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Fortify</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">3</td>
<td valign="top" align="left">42, F</td>
<td valign="top" align="left">HCM</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Fortify</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">4</td>
<td valign="top" align="left">65, M</td>
<td valign="top" align="left">DCM</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x000B1;</td>
<td valign="top" align="center">&#x000B1;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Current</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">5</td>
<td valign="top" align="left">28, M</td>
<td valign="top" align="left">Brugada</td>
<td valign="top" align="center">Secondary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x000B1;</td>
<td valign="top" align="left">Teligen</td>
<td valign="top" align="left">RELIANCE G</td>
</tr>
<tr>
<td valign="top" align="left">6</td>
<td valign="top" align="left">52, F</td>
<td valign="top" align="left">SCD</td>
<td valign="top" align="center">Secondary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Fortify</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">7</td>
<td valign="top" align="left">44, F</td>
<td valign="top" align="left">HCM</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x000B1;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Teligen</td>
<td valign="top" align="left">RELIANCE G</td>
</tr>
<tr>
<td valign="top" align="left">8</td>
<td valign="top" align="left">58, M</td>
<td valign="top" align="left">Brugada</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Current</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">9</td>
<td valign="top" align="left">54, M</td>
<td valign="top" align="left">Post MI</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Fortify</td>
<td valign="top" align="left">Durata 7122</td>
</tr>
<tr>
<td valign="top" align="left">10<xref ref-type="table-fn" rid="TN1"><sup>&#x0002A;</sup></xref></td>
<td valign="top" align="left">47, F</td>
<td valign="top" align="left">DCM</td>
<td valign="top" align="center">Primary</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x0002B; (BV)</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="center">&#x02212;</td>
<td valign="top" align="left">&#x02212;</td>
<td valign="top" align="left">Consulta</td>
<td valign="top" align="left">Sprint 6947</td>
</tr>
</tbody>
</table>
<table-wrap-foot>
<p><italic>BV, Bioventricular pacing; ICD, implantable cardioverter-defibrillator; TWOS, T-wave oversensing; VP, ventricular pacing; VS, ventricular sensing; PSVP, post-sensing ventricular pacing. &#x0002B;, persistent TWOS; &#x02212;, no TWOS; &#x000B1;, intermittent TWOS;</italic></p>
<fn id="TN1">
<label>&#x0002A;</label>
<p><italic>during routine follow-up</italic>.</p></fn>
</table-wrap-foot>
</table-wrap>
<fig id="F4" position="float">
<label>Figure 4</label>
<caption><p>TWOS episodes during the procedure of ICD implantation and sensing threshold curves with real proportion. <bold>(A)</bold> 12-lead surface electrocardiogram of the short QT syndrome patient. <bold>(B)</bold> TWOS under the maximum sensitivity in Virtuoso ICD. Red line shows the blank period after sensing; green line shows the sensing curve with minimum sensitivity; blue lines show sensing curves with sensitivity of 0.3 and 0.15 mV; red dot shows the possible range of T wave amplitudes. <bold>(C)</bold> Intermittent TWOS under maximum sensitivity in Epic ICD. Green and blue lines show the sensing curves with minimum and default sensitivity, respectively. <bold>(D)</bold> Measurement of TWSM in Fortify ICD. <bold>(E,F)</bold> Measurement of TWSM in Vitality and Teligen ICDs. Blue lines show the sensing curve with default sensitivity.</p></caption>
<graphic xlink:href="fphys-08-00659-g0004.tif"/>
</fig>
<p>Virtuoso and Consulta ICDs showed post pacing TWOS events in eight patients at a sensitivity of 0.15 mV, and in two patients under the maximum sensitivity of 0.3 mV (Figures <xref ref-type="fig" rid="F5">5A&#x02013;E</xref>). No TWOS was recorded under the maximum sensitivity of 0.45 mV. The eight pacing TWOS patients included three with Brugada syndrome, two with hypertrophic cardiomyopathy, one with dilated cardiomyopathy, one who had survived cardiac arrest, and one post-myocardial infarction with low left ventricular ejection fraction. The two patients that had pacing TWOS events under the sensitivity of 0.3 mV were Brugada syndrome and hypertrophic cardiomyopathy, respectively.</p>
<fig id="F5" position="float">
<label>Figure 5</label>
<caption><p>Post pacing TWOS revealed during the procedure of ICD implantation in Brugada syndrome patient, hypertrophic cardiomyopathy patient and dialated cardiomyopathy patient. <bold>(A)</bold> 12-lead surface EKG of a Brugada syndrome patient. <bold>(B,C)</bold> Post pacing TWOS episodes by a Consulta ICD under sensitivity of 0.15 mV. No TWOS was recorded under 0.3 mV. <bold>(D)</bold> 12-lead surface EKG of a hypertrophic cardiomyopathy patient. The implantable cardioverter-defibrillator working mode is VAT. <bold>(E)</bold> Post pacing TWOS episodes in Virtuoso ICD under the sensitivity of 0.15 mV. <bold>(F)</bold> 12-lead EKG of a CRT-D patient with dilated cardiomyopathy. <bold>(G)</bold> After the episode of TWOS (&#x02460;), PVARP was prolonged by PVC response and sinus P wave was recognized as a premature beat (&#x02461;) and not traced by ventricle (&#x02462;).</p></caption>
<graphic xlink:href="fphys-08-00659-g0005.tif"/>
</fig>
</sec>
<sec>
<title>TWSM</title>
<p>With the default parameter setting, TWSMs post intrinsic rhythm sensing by Fortify and Teligen ICDs were 13.0 &#x000B1; 7.6 and 28.3 &#x000B1; 16.5, respectively (<italic>n</italic> &#x0003D; 32, Figure <xref ref-type="fig" rid="F3">3D</xref>). TWSM post sensing was not calculated in Epic, Virtuoso, Consulta or Vitality ICDs because T wave amplitudes were not measurable on these ICD printouts.</p>
<p>TWSMs post pacing by Fortify and Teligen ICDs were 5.0 &#x000B1; 2.2 and 4.6 &#x000B1; 0.9 respectively (<italic>n</italic> &#x0003D; 23, Figure <xref ref-type="fig" rid="F3">3D</xref>). TWSMs post sensing and pacing were not evaluated in Epic, Virtuoso, Consulta, or Vitality ICDs because T wave amplitudes were not measurable on these ICD printouts.</p>
</sec>
<sec>
<title>Follow-up</title>
<p>The 32 patients were followed up for 25 &#x000B1; 7 months. One DCM patients died of refractory heart failure. One Brugada patient was shocked three times for ventricular tachycardia and ventricular fibrillation. One short QT syndrome patient was shocked twice for fracture of defibrillation lead (Riata, St. Jude Medical, Inc.). Two patients experienced inappropriate ICD shocks for atrial fibrillation. One patient with a Consulta ICD showed TWOS after biventricular pacing in the default setting at the 3-month follow-up visit (Figures <xref ref-type="fig" rid="F5">5F,G</xref>, Table <xref ref-type="table" rid="T2">2</xref>). No inappropriate therapy related to TWOS occurred during follow-up.</p>
</sec>
</sec>
<sec sec-type="discussion" id="s4">
<title>Discussion</title>
<p>To the best of our knowledge, this is the first study that evaluates the T wave amplitude and TWSM of different ICDs during the procedure of ICD implantation. We first propose the index of TWSM and validated that ventricular signal amplitude could be measured on printouts in certain ICD models. <italic>Ex vivo</italic> connection could evaluate T wave amplitude and TWSM, and might reduce TWOS episodes at mid-term follow-up.</p>
<p>Currently, there is no reliable method to identify patients at high risk for TWOS. In clinical practice, if the R wave amplitude measured by the ICD programmer is high enough (&#x0003E;5&#x0007E;8 mV), the risk of TWOS is considered to be low. Recently, T/R ratio during AAI pacing using bipolar electrograms was reported as an excellent predictor of TWOS (Maesato et al., <xref ref-type="bibr" rid="B9">2011</xref>). However, according to the sensing principle of ICDs, the devices sense the filtered ventricular signals when these signals are higher than the real-time sensing threshold. As T wave was concerned, it is TWSM rather than R wave amplitude or T/R ratio that determines the risk of TWOS when T wave amplitude varies. In this study, R wave amplitudes in all 32 patients were sufficiently high for clinical practice. But TWOS still occurred in some patients under the maximum sensitivity, which means insufficient TWSM in these patients under default parameters setting. In certain clinical situations, such as hyperkalemia, exercise, emotional fluctuation, and myocardial ischemia, T wave amplitude might increase and cause TWOS (Koul et al., <xref ref-type="bibr" rid="B8">2004</xref>; Otmani et al., <xref ref-type="bibr" rid="B18">2008</xref>). Evaluating T wave amplitude and TWSM during the procedure of ICD implantation might help clinicians to select optimal ICDs to reduce risk of TWOS by the following means: (1) When TWSM under default setting is &#x0003C;1 in one type of ICD, physician could implant another type with higher TWSM, or re-locate shock lead to avoid TWOS; (2) When TWOS occurs after ICD implantation and could not be solved by non-invasive method, replacing ICD generator with higher TWSM might resolve the problem and minimize the risk in the future; (3) When TWOS occurs during the follow-up period and could be solved by non-invasive method, TWSM could help to program appropriate sensing parameters.</p>
<p>To evaluate TWSM, two values are needed: T wave amplitude and real-time sensing threshold. The sensing threshold could be calculated according to the sensing algorithm of ICD (Supplementary Table <xref ref-type="supplementary-material" rid="SM1">1</xref>). However, T wave amplitude could not be measured on an ICD analyzer or programmer, so we developed a novel method called <italic>ex vivo</italic> connection to measure T wave amplitude.</p>
<sec>
<title><italic>Ex vivo</italic> connection</title>
<p>The intracardiac signal such as QRS wave is characterized by low amplitude (&#x0007E;mV) and high impedance (&#x0007E;k&#x003A9;). Therefore, modern ICD sensing circuitry needs high input impedance (&#x0007E;M&#x003A9;) to minimize the distortion of the QRS wave, so that the signal sampling is not significantly affected by trivial increases in load resistance (Myers et al., <xref ref-type="bibr" rid="B15">1978</xref>). In this study, we evaluated R waves by <italic>ex vivo</italic> ICD connections, which produced similar patterns of ventricular sensing as implanted ICDs (data not shown). This suggests that ventricular sensing could be evaluated precisely by an <italic>ex vivo</italic> connection method before the ICD was implanted.</p>
</sec>
<sec>
<title>R wave and T wave amplitude</title>
<p>In the sensing circuitry of ICD, both R and T waves are ventricular signals, so it is reasonable to measure R wave and T wave amplitude on ICD printouts. In our study, T wave amplitude post sensing was lower than after pacing in both Fortify and Teligen ICDs. This result was consistent with higher T waves after pacing than after sensing on surface EKGs. Interestingly, T wave amplitudes showed no difference between Fortify and Teligen ICDs, although the R wave amplitudes differed.</p>
</sec>
<sec>
<title>TWOS and TWSM</title>
<p>In general clinical practice, ICD sensitivity is often programed to default values unless TWOS occurs. This &#x0201C;wait-and-see&#x0201D; strategy creates risks of TWOS and inappropriate therapy in ICD patients, especially in those at high risk of TWOS. In our study, 9/32 (28%) patients had episodes of TWOS at the maximum sensitivity settings. In these nine patients, default values could only provide a TWSM of &#x0003C;2 (i.e., 200%). When T wave amplitude increased for various clinical reasons, TWOS might occur. Sensing TWOS can cause inappropriate therapy, while pacing TWOS created a risk for inappropriate bradycardia and insufficient pacing in pacing-dependent patients, such as those with long QT syndrome and dilated cardiomyopathy (Iijima et al., <xref ref-type="bibr" rid="B3">2011</xref>; Korantzopoulos et al., <xref ref-type="bibr" rid="B7">2013</xref>). In our study, a dilated cardiomyopathy patient with a Consulta ICD showed TWOS post biventricular pacing during regular follow-up at 3 months. The TWOS resulted in a decreased pacing proportion. The situation was resolved by decreasing the ventricular sensitivity to 0.45 mV.</p>
<p>Further analysis showed that all nine episodes of TWOS occurred in the Epic ICDs and the two Medtronic ICDs. TWOS in the Epic ICD was caused by high T wave amplitude. It is noteworthy that the T wave amplitudes of the same patient were 2.4&#x02013;3.8 mV by the Epic ICD, and 0.74 mV by the Fortify ICD, which might reflect technical improvement in the sensing circuitry of ICDs. In contrast, TWOS episodes in Medtronic ICDs were caused by low sensing thresholds rather than high T wave amplitudes. In Medtronic ICDs, sensing thresholds after sensing and pacing are limited to 8&#x0007E;10 and 4.5 times of the maximum sensitivity, respectively. At the default sensing threshold of 0.3 mV, the ventricular sensing thresholds are not more than 3 mV and 1.35 mV, respectively. This algorithm ensures high sensitivity to ventricular signals, but leads to a lower TWSM, especially in patients with high R wave amplitudes (Swerdlow and Friedman, <xref ref-type="bibr" rid="B26">2005</xref>). Recently, a new TWOS rejection algorithm was employed in Medtronic ICDs, but its effectiveness in patients with a high risk of TWOS is still uncertain (Cao et al., <xref ref-type="bibr" rid="B2">2012</xref>).</p>
<p>Subcutaneous ICDs (S-ICD), as an alternative to the transvenous ICD, are employed in recent years. The most common cause of inappropriate shocks in S-ICD was TWOS (Aydin et al., <xref ref-type="bibr" rid="B1">2012</xref>; Olde Nordkamp et al., <xref ref-type="bibr" rid="B16">2012</xref>; Kobe et al., <xref ref-type="bibr" rid="B5">2013</xref>). Although, there are some methods to predict TWOS (Wilson et al., <xref ref-type="bibr" rid="B29">2016</xref>) and resolve these problems (Kooiman et al., <xref ref-type="bibr" rid="B6">2014</xref>), evaluation of TWSM during the procedure of ICD implantation seems to be a better choice to reduce TWOS in advance. However, the sensing signal in S-ICD is more similar to surface EKG rather than intracardiac signal in transvenous ICD. The validity of the hypothesis still needs further verification.</p>
</sec>
<sec>
<title>Limitations</title>
<p>Only six ICD models are used to evaluate signal sensing. Some new types of ICD (e.g., Protecta and Evera of Medtronic and ICDs of Biotronik) are not tested because they are not available during our study. Even though paired comparisons employed in this study have adequate statistical power, 32 patients is a small sample population. T wave and TWSM were not fully evaluated under different heart rhythm or under different sympathetic active drugs (e.g., isoproterenol). Post-pacing T wave sensing is not fully tested for various reasons. Furthermore, R wave and T wave amplitudes may significantly change according to postural and emotional status, and should be further evaluated. The heterogeneity of the patients and lack of control group were the biggest drawback of our study. Further studies should be designed to clarify these issues.</p>
</sec>
</sec>
<sec sec-type="conclusions" id="s5">
<title>Conclusion</title>
<p>We first propose the index of TWSM during ICD implantation as a potentially efficient predictor for TWOS. Evaluation of TWSM may help to reduce TWOS episode in patients with high risk. Further studies are warranted to validate this predictor and its potential in clinical practice.</p>
</sec>
<sec id="s6">
<title>Author contributions</title>
<p>DH, GF, YS, and HM. design the project. YS, JGao, XS, YY, and HH. coordinated and collected the clinical evaluations. DH, CJ, YMX, and YZX organized and summarized the data. YZ, GL, and JGuo. analyzed the data and made the figures and table. DH, YS, and JGao wrote the manuscript. All co-authors contributed to editing and proving of manuscript.</p>
<sec>
<title>Conflict of interest statement</title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
</sec>
</body>
<back>
<ack><p>We thank technical stuffs from St. Jude Medical, Inc., Medtronic, Inc., and Boston Scientific, Inc. for providing the ICD manuals. We appreciate Dr. Roderick Tung from University of Chicago for his help in the manuscript revision.</p>
</ack>
<sec sec-type="supplementary-material" id="s7">
<title>Supplementary material</title>
<p>The Supplementary Material for this article can be found online at: <ext-link ext-link-type="uri" xlink:href="http://journal.frontiersin.org/article/10.3389/fphys.2017.00659/full#supplementary-material">http://journal.frontiersin.org/article/10.3389/fphys.2017.00659/full#supplementary-material</ext-link></p>
<supplementary-material xlink:href="Table1.docx" id="SM1" mimetype="application/vnd.openxmlformats-officedocument.wordprocessingml.document" xmlns:xlink="http://www.w3.org/1999/xlink"/>
</sec>
<ref-list>
<title>References</title>
<ref id="B1">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Aydin</surname> <given-names>A.</given-names></name> <name><surname>Hartel</surname> <given-names>F.</given-names></name> <name><surname>Schluter</surname> <given-names>M.</given-names></name> <name><surname>Butter</surname> <given-names>C.</given-names></name> <name><surname>Kobe</surname> <given-names>J.</given-names></name> <name><surname>Seifert</surname> <given-names>M.</given-names></name> <etal/></person-group>. (<year>2012</year>). <article-title>Shock efficacy of subcutaneous implantable cardioverter-defibrillator for prevention of sudden cardiac death: initial multicenter experience</article-title>. <source>Circ. Arrhythm. Electrophysiol.</source> <volume>5</volume>, <fpage>913</fpage>&#x02013;<lpage>919</lpage>. <pub-id pub-id-type="doi">10.1161/CIRCEP.112.973339</pub-id><pub-id pub-id-type="pmid">22923274</pub-id></citation></ref>
<ref id="B2">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Cao</surname> <given-names>J.</given-names></name> <name><surname>Gillberg</surname> <given-names>J. M.</given-names></name> <name><surname>Swerdlow</surname> <given-names>C. D.</given-names></name></person-group> (<year>2012</year>). <article-title>A fully automatic, implantable cardioverter-defibrillator algorithm to prevent inappropriate detection of ventricular tachycardia or fibrillation due to T-wave oversensing in spontaneous rhythm</article-title>. <source>Heart Rhythm</source> <volume>9</volume>, <fpage>522</fpage>&#x02013;<lpage>530</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2011.11.023</pub-id><pub-id pub-id-type="pmid">22094074</pub-id></citation></ref>
<ref id="B3">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Iijima</surname> <given-names>K.</given-names></name> <name><surname>Chinushi</surname> <given-names>M.</given-names></name> <name><surname>Hasegawa</surname> <given-names>K.</given-names></name> <name><surname>Izumi</surname> <given-names>D.</given-names></name> <name><surname>Hosaka</surname> <given-names>Y.</given-names></name> <name><surname>Furushima</surname> <given-names>H.</given-names></name> <etal/></person-group>. (<year>2011</year>). <article-title>Inappropriate pacing inhibition triggered by QT prolongation due to T wave oversensing in an ICD recipient presenting with long QT syndrome</article-title>. <source>Intern. Med.</source> <volume>50</volume>, <fpage>1021</fpage>&#x02013;<lpage>1024</lpage>. <pub-id pub-id-type="doi">10.2169/internalmedicine.50.4837</pub-id><pub-id pub-id-type="pmid">21532225</pub-id></citation></ref>
<ref id="B4">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Juan</surname> <given-names>E. A.</given-names></name> <name><surname>Pollack</surname> <given-names>M.</given-names></name></person-group> (<year>2010</year>). <article-title>Phantom shocks in patients with an implantable cardioverter defibrillator</article-title>. <source>J. Emerg. Med.</source> <volume>38</volume>, <fpage>22</fpage>&#x02013;<lpage>24</lpage>. <pub-id pub-id-type="doi">10.1016/j.jemermed.2007.09.036</pub-id><pub-id pub-id-type="pmid">18394854</pub-id></citation></ref>
<ref id="B5">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Kobe</surname> <given-names>J.</given-names></name> <name><surname>Reinke</surname> <given-names>F.</given-names></name> <name><surname>Meyer</surname> <given-names>C.</given-names></name> <name><surname>Shin</surname> <given-names>D. I.</given-names></name> <name><surname>Martens</surname> <given-names>E.</given-names></name> <name><surname>Kaab</surname> <given-names>S.</given-names></name> <etal/></person-group>. (<year>2013</year>). <article-title>Implantation and follow-up of totally subcutaneous versus conventional implantable cardioverter-defibrillators: a multicenter case-control study</article-title>. <source>Heart Rhythm</source> <volume>10</volume>, <fpage>29</fpage>&#x02013;<lpage>36</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2012.09.126</pub-id><pub-id pub-id-type="pmid">23032867</pub-id></citation></ref>
<ref id="B6">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Kooiman</surname> <given-names>K. M.</given-names></name> <name><surname>Knops</surname> <given-names>R. E.</given-names></name> <name><surname>Olde Nordkamp</surname> <given-names>L.</given-names></name> <name><surname>Wilde</surname> <given-names>A. A.</given-names></name> <name><surname>De Groot</surname> <given-names>J. R.</given-names></name></person-group> (<year>2014</year>). <article-title>Inappropriate subcutaneous implantable cardioverter-defibrillator shocks due to T-wave oversensing can be prevented: implications for management</article-title>. <source>Heart Rhythm</source> <volume>11</volume>, <fpage>426</fpage>&#x02013;<lpage>434</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2013.12.007</pub-id><pub-id pub-id-type="pmid">24321235</pub-id></citation></ref>
<ref id="B7">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Korantzopoulos</surname> <given-names>P.</given-names></name> <name><surname>Kyrlas</surname> <given-names>K.</given-names></name> <name><surname>Goudevenos</surname> <given-names>J. A.</given-names></name></person-group> (<year>2013</year>). <article-title>Loss of biventricular pacing due to T wave oversensing: case report and review of the literature</article-title>. <source>Int. J. Cardiol.</source> <volume>168</volume>, <fpage>1626</fpage>&#x02013;<lpage>1628</lpage>. <pub-id pub-id-type="doi">10.1016/j.ijcard.2013.01.065</pub-id><pub-id pub-id-type="pmid">23398824</pub-id></citation></ref>
<ref id="B8">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Koul</surname> <given-names>A. K.</given-names></name> <name><surname>Keller</surname> <given-names>S.</given-names></name> <name><surname>Clancy</surname> <given-names>J. F.</given-names></name> <name><surname>Lampert</surname> <given-names>R.</given-names></name> <name><surname>Batsford</surname> <given-names>W. P.</given-names></name> <name><surname>Rosenfeld</surname> <given-names>L. E.</given-names></name></person-group> (<year>2004</year>). <article-title>Hyperkalemia induced T wave oversensing leading to loss of biventricular pacing and inappropriate ICD shocks</article-title>. <source>Pacing Clin. Electrophysiol.</source> <volume>27</volume>, <fpage>681</fpage>&#x02013;<lpage>683</lpage>. <pub-id pub-id-type="doi">10.1111/j.1540-8159.2004.00509.x</pub-id><pub-id pub-id-type="pmid">15125731</pub-id></citation></ref>
<ref id="B9">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Maesato</surname> <given-names>A.</given-names></name> <name><surname>Higa</surname> <given-names>S.</given-names></name> <name><surname>Lin</surname> <given-names>Y. J.</given-names></name> <name><surname>Chinen</surname> <given-names>I.</given-names></name> <name><surname>Ishigaki</surname> <given-names>S.</given-names></name> <name><surname>Yajima</surname> <given-names>M.</given-names></name> <etal/></person-group>. (<year>2011</year>). <article-title>Impact of pacing and high-pass filter settings on ventricular bipolar electrograms in implantable cardioverter defibrillator systems&#x02014;Implication of predictors for inappropriate therapy caused by oversensing of repolarization electrograms</article-title>. <source>Circ. J.</source> <volume>75</volume>, <fpage>2095</fpage>&#x02013;<lpage>2104</lpage>. <pub-id pub-id-type="doi">10.1253/circj.CJ-11-0200</pub-id><pub-id pub-id-type="pmid">21757820</pub-id></citation></ref>
<ref id="B10">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Magnusson</surname> <given-names>P.</given-names></name> <name><surname>Gadler</surname> <given-names>F.</given-names></name> <name><surname>Liv</surname> <given-names>P.</given-names></name> <name><surname>Morner</surname> <given-names>S.</given-names></name></person-group> (<year>2015</year>). <article-title>Hypertrophic cardiomyopathy and implantable defibrillators in sweden: inappropriate shocks and complications requiring surgery</article-title>. <source>J. Cardiovasc. Electrophysiol.</source> <volume>26</volume>, <fpage>1088</fpage>&#x02013;<lpage>1094</lpage>. <pub-id pub-id-type="doi">10.1111/jce.12750</pub-id><pub-id pub-id-type="pmid">26178879</pub-id></citation></ref>
<ref id="B11">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Mark</surname> <given-names>D. B.</given-names></name> <name><surname>Anstrom</surname> <given-names>K. J.</given-names></name> <name><surname>Sun</surname> <given-names>J. L.</given-names></name> <name><surname>Clapp-Channing</surname> <given-names>N. E.</given-names></name> <name><surname>Tsiatis</surname> <given-names>A. A.</given-names></name> <name><surname>Davidson-Ray</surname> <given-names>L.</given-names></name> <etal/></person-group>. (<year>2008</year>). <article-title>Quality of life with defibrillator therapy or amiodarone in heart failure</article-title>. <source>N. Engl. J. Med.</source> <volume>359</volume>, <fpage>999</fpage>&#x02013;<lpage>1008</lpage>. <pub-id pub-id-type="doi">10.1056/NEJMoa0706719</pub-id></citation></ref>
<ref id="B12">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Moss</surname> <given-names>A. J.</given-names></name> <name><surname>Hall</surname> <given-names>W. J.</given-names></name> <name><surname>Cannom</surname> <given-names>D. S.</given-names></name> <name><surname>Daubert</surname> <given-names>J. P.</given-names></name> <name><surname>Higgins</surname> <given-names>S. L.</given-names></name> <name><surname>Klein</surname> <given-names>H.</given-names></name> <etal/></person-group>. (<year>1996</year>). <article-title>Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators</article-title>. <source>N. Engl. J. Med.</source> <volume>335</volume>, <fpage>1933</fpage>&#x02013;<lpage>1940</lpage>. <pub-id pub-id-type="doi">10.1056/NEJM199612263352601</pub-id><pub-id pub-id-type="pmid">8960472</pub-id></citation></ref>
<ref id="B13">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Moss</surname> <given-names>A. J.</given-names></name> <name><surname>Hall</surname> <given-names>W. J.</given-names></name> <name><surname>Cannom</surname> <given-names>D. S.</given-names></name> <name><surname>Klein</surname> <given-names>H.</given-names></name> <name><surname>Brown</surname> <given-names>M. W.</given-names></name> <name><surname>Daubert</surname> <given-names>J. P.</given-names></name> <etal/></person-group>. (<year>2009</year>). <article-title>Cardiac-resynchronization therapy for the prevention of heart-failure events</article-title>. <source>N. Engl. J. Med.</source> <volume>361</volume>, <fpage>1329</fpage>&#x02013;<lpage>1338</lpage>. <pub-id pub-id-type="doi">10.1056/NEJMoa0906431</pub-id><pub-id pub-id-type="pmid">19723701</pub-id></citation></ref>
<ref id="B14">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Moss</surname> <given-names>A. J.</given-names></name> <name><surname>Zareba</surname> <given-names>W.</given-names></name> <name><surname>Hall</surname> <given-names>W. J.</given-names></name> <name><surname>Klein</surname> <given-names>H.</given-names></name> <name><surname>Wilber</surname> <given-names>D. J.</given-names></name> <name><surname>Cannom</surname> <given-names>D. S.</given-names></name> <etal/></person-group>. (<year>2002</year>). <article-title>Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction</article-title>. <source>N. Engl. J. Med.</source> <volume>346</volume>, <fpage>877</fpage>&#x02013;<lpage>883</lpage>. <pub-id pub-id-type="doi">10.1056/NEJMoa013474</pub-id><pub-id pub-id-type="pmid">11907286</pub-id></citation></ref>
<ref id="B15">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Myers</surname> <given-names>G. H.</given-names></name> <name><surname>Kresh</surname> <given-names>Y. M.</given-names></name> <name><surname>Parsonnet</surname> <given-names>V.</given-names></name></person-group> (<year>1978</year>). <article-title>Characteristics of intracardiac electrograms</article-title>. <source>Pacing Clin. Electrophysiol.</source> <volume>1</volume>, <fpage>90</fpage>&#x02013;<lpage>103</lpage>. <pub-id pub-id-type="doi">10.1111/j.1540-8159.1978.tb03445.x</pub-id><pub-id pub-id-type="pmid">83626</pub-id></citation></ref>
<ref id="B16">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Olde Nordkamp</surname> <given-names>L. R.</given-names></name> <name><surname>Dabiri Abkenari</surname> <given-names>L.</given-names></name> <name><surname>Boersma</surname> <given-names>L. V.</given-names></name> <name><surname>Maass</surname> <given-names>A. H.</given-names></name> <name><surname>De Groot</surname> <given-names>J. R.</given-names></name> <name><surname>Van Oostrom</surname> <given-names>A. J.</given-names></name> <etal/></person-group>. (<year>2012</year>). <article-title>The entirely subcutaneous implantable cardioverter-defibrillator: initial clinical experience in a large Dutch cohort</article-title>. <source>J. Am. Coll. Cardiol</source> <volume>60</volume>, <fpage>1933</fpage>&#x02013;<lpage>1939</lpage>. <pub-id pub-id-type="doi">10.1016/j.jacc.2012.06.053</pub-id><pub-id pub-id-type="pmid">23062537</pub-id></citation></ref>
<ref id="B17">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Olde Nordkamp</surname> <given-names>L. R.</given-names></name> <name><surname>Postema</surname> <given-names>P. G.</given-names></name> <name><surname>Knops</surname> <given-names>R. E.</given-names></name> <name><surname>Van Dijk</surname> <given-names>N.</given-names></name> <name><surname>Limpens</surname> <given-names>J.</given-names></name> <name><surname>Wilde</surname> <given-names>A. A.</given-names></name> <etal/></person-group>. (<year>2016</year>). <article-title>Implantable cardioverter-defibrillator harm in young patients with inherited arrhythmia syndromes: a systematic review and meta-analysis of inappropriate shocks and complications</article-title>. <source>Heart Rhythm</source> <volume>13</volume>, <fpage>443</fpage>&#x02013;<lpage>454</lpage>. <pub-id pub-id-type="doi">10.1016/j.hrthm.2015.09.010</pub-id><pub-id pub-id-type="pmid">26385533</pub-id></citation></ref>
<ref id="B18">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Otmani</surname> <given-names>A.</given-names></name> <name><surname>Rey</surname> <given-names>J. L.</given-names></name> <name><surname>Leborgne</surname> <given-names>L.</given-names></name></person-group> (<year>2008</year>). <article-title>T-wave oversensing during exercise one year after cardioverter defibrillator implantation for Brugada syndrome</article-title>. <source>Arch. Cardiovasc. Dis.</source> <volume>101</volume>, <fpage>292</fpage>&#x02013;<lpage>294</lpage>. <pub-id pub-id-type="doi">10.1016/S1875-2136(08)73708-2</pub-id><pub-id pub-id-type="pmid">18654108</pub-id></citation></ref>
<ref id="B19">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Powell</surname> <given-names>B. D.</given-names></name> <name><surname>Saxon</surname> <given-names>L. A.</given-names></name> <name><surname>Boehmer</surname> <given-names>J. P.</given-names></name> <name><surname>Day</surname> <given-names>J. D.</given-names></name> <name><surname>Gilliam</surname> <given-names>F. R.</given-names> <suffix>III.</suffix></name> <name><surname>Heidenreich</surname> <given-names>P. A.</given-names></name> <etal/></person-group>. (<year>2013</year>). <article-title>Survival after shock therapy in implantable cardioverter-defibrillator and cardiac resynchronization therapy-defibrillator recipients according to rhythm shocked. The altitude survival by rhythm study</article-title>. <source>J. Am. Coll. Cardiol.</source> <volume>62</volume>, <fpage>1674</fpage>&#x02013;<lpage>1679</lpage>. <pub-id pub-id-type="doi">10.1016/j.jacc.2013.04.083</pub-id><pub-id pub-id-type="pmid">23810882</pub-id></citation></ref>
<ref id="B20">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sacher</surname> <given-names>F.</given-names></name> <name><surname>Probst</surname> <given-names>V.</given-names></name> <name><surname>Iesaka</surname> <given-names>Y.</given-names></name> <name><surname>Jacon</surname> <given-names>P.</given-names></name> <name><surname>Laborderie</surname> <given-names>J.</given-names></name> <name><surname>Mizon-Gerard</surname> <given-names>F.</given-names></name> <etal/></person-group>. (<year>2006</year>). <article-title>Outcome after implantation of a cardioverter-defibrillator in patients with Brugada syndrome: a multicenter study</article-title>. <source>Circulation</source> <volume>114</volume>, <fpage>2317</fpage>&#x02013;<lpage>2324</lpage>. <pub-id pub-id-type="doi">10.1161/CIRCULATIONAHA.106.628537</pub-id><pub-id pub-id-type="pmid">17116772</pub-id></citation></ref>
<ref id="B21">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sacher</surname> <given-names>F.</given-names></name> <name><surname>Probst</surname> <given-names>V.</given-names></name> <name><surname>Maury</surname> <given-names>P.</given-names></name> <name><surname>Babuty</surname> <given-names>D.</given-names></name> <name><surname>Mansourati</surname> <given-names>J.</given-names></name> <name><surname>Komatsu</surname> <given-names>Y.</given-names></name> <etal/></person-group>. (<year>2013</year>). <article-title>Outcome after implantation of a cardioverter-defibrillator in patients with Brugada syndrome: a multicenter study-part 2</article-title>. <source>Circulation</source> <volume>128</volume>, <fpage>1739</fpage>&#x02013;<lpage>1747</lpage>. <pub-id pub-id-type="doi">10.1161/CIRCULATIONAHA.113.001941</pub-id><pub-id pub-id-type="pmid">23995538</pub-id></citation></ref>
<ref id="B22">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Saxonhouse</surname> <given-names>S. J.</given-names></name> <name><surname>Conti</surname> <given-names>J. B.</given-names></name> <name><surname>Curtis</surname> <given-names>A. B.</given-names></name></person-group> (<year>2005</year>). <article-title>Current of injury predicts adequate active lead fixation in permanent pacemaker/defibrillation leads</article-title>. <source>J. Am. Coll. Cardiol.</source> <volume>45</volume>, <fpage>412</fpage>&#x02013;<lpage>417</lpage>. <pub-id pub-id-type="doi">10.1016/j.jacc.2004.10.045</pub-id><pub-id pub-id-type="pmid">15680721</pub-id></citation></ref>
<ref id="B23">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Schimpf</surname> <given-names>R.</given-names></name> <name><surname>Wolpert</surname> <given-names>C.</given-names></name> <name><surname>Bianchi</surname> <given-names>F.</given-names></name> <name><surname>Giustetto</surname> <given-names>C.</given-names></name> <name><surname>Gaita</surname> <given-names>F.</given-names></name> <name><surname>Bauersfeld</surname> <given-names>U.</given-names></name> <etal/></person-group>. (<year>2003</year>). <article-title>Congenital short QT syndrome and implantable cardioverter defibrillator treatment: inherent risk for inappropriate shock delivery</article-title>. <source>J. Cardiovasc. Electrophysiol.</source> <volume>14</volume>, <fpage>1273</fpage>&#x02013;<lpage>1277</lpage>. <pub-id pub-id-type="doi">10.1046/j.1540-8167.2003.03278.x</pub-id><pub-id pub-id-type="pmid">14678099</pub-id></citation></ref>
<ref id="B24">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sun</surname> <given-names>Y.</given-names></name> <name><surname>Quan</surname> <given-names>X. Q.</given-names></name> <name><surname>Fromme</surname> <given-names>S.</given-names></name> <name><surname>Cox</surname> <given-names>R. H.</given-names></name> <name><surname>Zhang</surname> <given-names>P.</given-names></name> <name><surname>Zhang</surname> <given-names>L.</given-names></name> <etal/></person-group>. (<year>2011</year>). <article-title>A novel mutation in the KCNH2 gene associated with short QT syndrome</article-title>. <source>J. Mol. Cell Cardiol.</source> <volume>50</volume>, <fpage>433</fpage>&#x02013;<lpage>441</lpage>. <pub-id pub-id-type="doi">10.1016/j.yjmcc.2010.11.017</pub-id><pub-id pub-id-type="pmid">21130771</pub-id></citation></ref>
<ref id="B25">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Sun</surname> <given-names>Y.</given-names></name> <name><surname>Zhang</surname> <given-names>P.</given-names></name> <name><surname>Li</surname> <given-names>X.</given-names></name> <name><surname>Guo</surname> <given-names>J.</given-names></name></person-group> (<year>2010</year>). <article-title>Inappropriate ICD discharge due to T-wave oversensing in a patient with short QT syndrome</article-title>. <source>Pacing Clin. Electrophysiol.</source> <volume>33</volume>, <fpage>113</fpage>&#x02013;<lpage>116</lpage>. <pub-id pub-id-type="doi">10.1111/j.1540-8159.2009.02538.x</pub-id><pub-id pub-id-type="pmid">19793372</pub-id></citation></ref>
<ref id="B26">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Swerdlow</surname> <given-names>C. D.</given-names></name> <name><surname>Friedman</surname> <given-names>P. A.</given-names></name></person-group> (<year>2005</year>). <article-title>Advanced ICD troubleshooting: part, I</article-title>. <source>Pacing Clin. Electrophysiol.</source> <volume>28</volume>, <fpage>1322</fpage>&#x02013;<lpage>1346</lpage>. <pub-id pub-id-type="doi">10.1111/j.1540-8159.2005.00275.x</pub-id><pub-id pub-id-type="pmid">16403166</pub-id></citation></ref>
<ref id="B27">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>van Rees</surname> <given-names>J. B.</given-names></name> <name><surname>Borleffs</surname> <given-names>C. J.</given-names></name> <name><surname>De Bie</surname> <given-names>M. K.</given-names></name> <name><surname>Stijnen</surname> <given-names>T.</given-names></name> <name><surname>Van Erven</surname> <given-names>L.</given-names></name> <name><surname>Bax</surname> <given-names>J. J.</given-names></name> <etal/></person-group>. (<year>2011</year>). <article-title>Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors, and impact on mortality</article-title>. <source>J. Am. Coll. Cardiol.</source> <volume>57</volume>, <fpage>556</fpage>&#x02013;<lpage>562</lpage>. <pub-id pub-id-type="doi">10.1016/j.jacc.2010.06.059</pub-id><pub-id pub-id-type="pmid">21272746</pub-id></citation></ref>
<ref id="B28">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Weretka</surname> <given-names>S.</given-names></name> <name><surname>Michaelsen</surname> <given-names>J.</given-names></name> <name><surname>Becker</surname> <given-names>R.</given-names></name> <name><surname>Karle</surname> <given-names>C. A.</given-names></name> <name><surname>Voss</surname> <given-names>F.</given-names></name> <name><surname>Hilbel</surname> <given-names>T.</given-names></name> <etal/></person-group>. (<year>2003</year>). <article-title>Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems</article-title>. <source>Pacing Clin. Electrophysiol.</source> <volume>26</volume>, <fpage>65</fpage>&#x02013;<lpage>70</lpage>. <pub-id pub-id-type="doi">10.1046/j.1460-9592.2003.00152.x</pub-id><pub-id pub-id-type="pmid">12685142</pub-id></citation></ref>
<ref id="B29">
<citation citation-type="journal"><person-group person-group-type="author"><name><surname>Wilson</surname> <given-names>D. G.</given-names></name> <name><surname>Leventigiannis</surname> <given-names>G.</given-names></name> <name><surname>Barr</surname> <given-names>C.</given-names></name> <name><surname>Morgan</surname> <given-names>J. M.</given-names></name></person-group> (<year>2016</year>). <article-title>ECG predictors of T wave oversensing in subcutaneous implantable cardioverter defibrillators</article-title>. <source>Int. J. Cardiol.</source> <volume>220</volume>, <fpage>27</fpage>&#x02013;<lpage>31</lpage>. <pub-id pub-id-type="doi">10.1016/j.ijcard.2016.06.128</pub-id><pub-id pub-id-type="pmid">27372039</pub-id></citation></ref>
</ref-list>
<fn-group>
<fn fn-type="financial-disclosure"><p><bold>Funding.</bold> This study was supported by the National Natural Science Foundation of China (No. 81100067; No. 81670304) and the Medicine &#x00026; Health Platform Program of Zhejiang (No. 2014RCA011) and Natural Science Foundation of Zhejiang Province (No. LY16H020002). The project was supported by Zhengshu Medical Elite grant.</p>