<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD Journal Publishing DTD v2.3 20070202//EN" "journalpublishing.dtd">
<article article-type="research-article" dtd-version="2.3" xml:lang="EN" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Phys.</journal-id>
<journal-title>Frontiers in Physics</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Phys.</abbrev-journal-title>
<issn pub-type="epub">2296-424X</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="publisher-id">1249771</article-id>
<article-id pub-id-type="doi">10.3389/fphy.2023.1249771</article-id>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Physics</subject>
<subj-group>
<subject>Original Research</subject>
</subj-group>
</subj-group>
</article-categories>
<title-group>
<article-title>Low-field MRI at high magnetic field instability and inhomogeneity conditions</article-title>
<alt-title alt-title-type="left-running-head">Rodriguez et al.</alt-title>
<alt-title alt-title-type="right-running-head">
<ext-link ext-link-type="uri" xlink:href="https://doi.org/10.3389/fphy.2023.1249771">10.3389/fphy.2023.1249771</ext-link>
</alt-title>
</title-group>
<contrib-group>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Rodriguez</surname>
<given-names>Gonzalo G.</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="corresp" rid="c001">&#x2a;</xref>
<xref ref-type="fn" rid="fn1">
<sup>&#x2020;</sup>
</xref>
<uri xlink:href="https://loop.frontiersin.org/people/2424900/overview"/>
</contrib>
<contrib contrib-type="author">
<name>
<surname>Sch&#xfc;rrer</surname>
<given-names>Clemar A.</given-names>
</name>
<xref ref-type="aff" rid="aff2">
<sup>2</sup>
</xref>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name>
<surname>Anoardo</surname>
<given-names>Esteban</given-names>
</name>
<xref ref-type="aff" rid="aff1">
<sup>1</sup>
</xref>
<xref ref-type="corresp" rid="c001">&#x2a;</xref>
<uri xlink:href="https://loop.frontiersin.org/people/1334173/overview"/>
</contrib>
</contrib-group>
<aff id="aff1">
<sup>1</sup>
<institution>Laboratorio de Relaxometr&#xed;a y T&#xe9;cnicas Especiales (LaRTE)</institution>, <institution>FaMAF&#x2013;Universidad Nacional de C&#xf3;rdoba (UNC) &#x26; IFEG-CONICET</institution>, <addr-line>C&#xf3;rdoba</addr-line>, <country>Argentina</country>
</aff>
<aff id="aff2">
<sup>2</sup>
<institution>Centro de Investigaci&#x00F3;n y Transferencia en Metrolog&#x00ED;a (CEMETRO)</institution>, <institution>Facultad Regional C&#xf3;rdoba y FaMAF-UNC</institution>, <institution>Universidad Tecnol&#xf3;gica Nacional</institution>, <addr-line>C&#xf3;rdoba</addr-line>, <country>Argentina</country>
</aff>
<author-notes>
<fn fn-type="edited-by">
<p>
<bold>Edited by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/1915698/overview">Fangrong Zong</ext-link>, Beijing University of Posts and Telecommunications (BUPT), China</p>
</fn>
<fn fn-type="edited-by">
<p>
<bold>Reviewed by:</bold> <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/126201/overview">Per Magnelind</ext-link>, Los Alamos National Laboratory (DOE), United States</p>
<p>
<ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/1838231/overview">Feng Deng</ext-link>, Research Institute of Petroleum Exploration and Development (RIPED), China</p>
</fn>
<corresp id="c001">&#x2a;Correspondence: Esteban Anoardo, <email>eanoardo@unc.edu.ar</email>; Gonzalo G. Rodriguez, <email>gonzalogabriel.rodriguez@mpinat.mpg.de</email>
</corresp>
<fn fn-type="present-address" id="fn1">
<label>
<sup>&#x2020;</sup>
</label>
<p>
<bold>Present address:</bold> Gonzalo G. Rodriguez, NMR Signal Enhancement Group, Max Planck Institute for Multidisciplinary Sciences, G&#xf6;ttingen, Germany</p>
</fn>
</author-notes>
<pub-date pub-type="epub">
<day>02</day>
<month>11</month>
<year>2023</year>
</pub-date>
<pub-date pub-type="collection">
<year>2023</year>
</pub-date>
<volume>11</volume>
<elocation-id>1249771</elocation-id>
<history>
<date date-type="received">
<day>29</day>
<month>06</month>
<year>2023</year>
</date>
<date date-type="accepted">
<day>18</day>
<month>10</month>
<year>2023</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#xa9; 2023 Rodriguez, Sch&#xfc;rrer and Anoardo.</copyright-statement>
<copyright-year>2023</copyright-year>
<copyright-holder>Rodriguez, Sch&#xfc;rrer and Anoardo</copyright-holder>
<license xlink:href="http://creativecommons.org/licenses/by/4.0/">
<p>This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</p>
</license>
</permissions>
<abstract>
<p>Understanding the effects of the magnetic field time instabilities in magnetic resonance imaging (MRI) is fundamental for the success of portable and low-cost MRI hardware based on electromagnets. In this work we propose a magnetic field model that considers the field instability in addition to the inhomogeneity. We have successfully validated the model on signals acquired with a commercial NMR instrument. It was used to simulate the image defects due to different types of instability for both the spin-echo and the gradient-echo sequences. We have considered both random field fluctuations, and an instability having a dominant harmonic component. Strategies are suggested to minimize the artifacts generated by these instabilities. Images were acquired using a home-made MRI relaxometer to show the consistency of the analysis.</p>
</abstract>
<kwd-group>
<kwd>low-field</kwd>
<kwd>field instability</kwd>
<kwd>MRI</kwd>
<kwd>electromagnets</kwd>
<kwd>field inhomogeneity</kwd>
</kwd-group>
<contract-sponsor id="cn001">Consejo Nacional de Investigaciones Cient&#xed;ficas y T&#xe9;cnicas<named-content content-type="fundref-id">10.13039/501100002923</named-content>
</contract-sponsor>
<custom-meta-wrap>
<custom-meta>
<meta-name>section-at-acceptance</meta-name>
<meta-value>Medical Physics and Imaging</meta-value>
</custom-meta>
</custom-meta-wrap>
</article-meta>
</front>
<body>
<sec id="s1">
<title>1 Introduction</title>
<p>The development of low field MRI instruments (&#x3c;0.2&#xa0;T) is growing drastically [<xref ref-type="bibr" rid="B1">1</xref>&#x2013;<xref ref-type="bibr" rid="B3">3</xref>] motivated by the possibility of cost reductions [<xref ref-type="bibr" rid="B4">4</xref>, <xref ref-type="bibr" rid="B5">5</xref>], portability [<xref ref-type="bibr" rid="B6">6</xref>&#x2013;<xref ref-type="bibr" rid="B10">10</xref>], and contrast-enhancement [<xref ref-type="bibr" rid="B11">11</xref>]. However, critical points to consider are the lower signal to noise ratio (SNR) [<xref ref-type="bibr" rid="B12">12</xref>], magnetic field homogeneity [<xref ref-type="bibr" rid="B9">9</xref>], and magnetic field stability [<xref ref-type="bibr" rid="B13">13</xref>]. These limitations may strongly affect the image quality by introducing biasing and artifacts. Therefore, the success of low field scanners strongly depends on our ability to minimize or compensate these undesired effects.</p>
<p>In MRI experiments the SNR degrades with the magnetic flux density (<italic>B</italic>
<sub>0</sub>) as <italic>B</italic>
<sub>0</sub>
<sup>
<italic>x</italic>
</sup>, with an exponent <italic>x</italic> ranging from 1.65 to 1.75 [<xref ref-type="bibr" rid="B12">12</xref>, <xref ref-type="bibr" rid="B14">14</xref>]. In consequence, the SNR working conditions for low-field MRI can be much poorer that those usually available at high-field MRI. However, the SNR can be improved by different means, from a simple signal averaging to different hardware and computational contraptions. Some explored hardware solutions include, among other, hyperpolarization schemes [<xref ref-type="bibr" rid="B11">11</xref>], magnetic field-cycling technology [<xref ref-type="bibr" rid="B3">3</xref>, <xref ref-type="bibr" rid="B13">13</xref>, <xref ref-type="bibr" rid="B15">15</xref>], single or multiple receiver channels using superconducting quantum interference devices (SQUID) [<xref ref-type="bibr" rid="B16">16</xref>&#x2013;<xref ref-type="bibr" rid="B18">18</xref>] and coupling to external resonators or magnetic lenses for signal enhancement [<xref ref-type="bibr" rid="B19">19</xref>&#x2013;<xref ref-type="bibr" rid="B21">21</xref>]. Specially designed pulse sequences are also considered [<xref ref-type="bibr" rid="B22">22</xref>, <xref ref-type="bibr" rid="B23">23</xref>], while image pos-processing can be used to artificially enhance the image SNR [<xref ref-type="bibr" rid="B24">24</xref>, <xref ref-type="bibr" rid="B25">25</xref>].</p>
<p>Against the natural lower SNR of the acquired signal at low magnetic field conditions, some advantages of the methodology are compensating this fact. Spin-lattice (<italic>T</italic>
<sub>1</sub>) relaxation times are usually much shorter at low magnetic fields [<xref ref-type="bibr" rid="B26">26</xref>], thus allowing a higher number of scans (signal acquisitions) per unit of time. The specific absorption rate (SAR) is proportional to the square of the magnetic field. Consequently, at low magnetic fields, pulse sequences with high flip angles or increased duty-cycles are favored. In addition, the effects of superparamagnetic nanoparticles in modulating the spin relaxation are very relevant at low magnetic fields [<xref ref-type="bibr" rid="B27">27</xref>], thus allowing an efficient contrast enhancement [<xref ref-type="bibr" rid="B7">7</xref>, <xref ref-type="bibr" rid="B28">28</xref>]. It is opportune to mention that the SNR of the image is related to the contrast-to-noise ratio (CNR), which can be increased by a proper use of contrast agents [<xref ref-type="bibr" rid="B29">29</xref>] and/or mathematically manipulated through computational post processing of the acquired image [<xref ref-type="bibr" rid="B30">30</xref>, <xref ref-type="bibr" rid="B31">31</xref>].</p>
<p>Compact magnet designs and imperfect mechanical assembling are usually associated with highly inhomogeneous magnetic fields, with inhomogeneities orders of magnitude bigger than superconducting magnets [<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B13">13</xref>]. Notwithstanding this feature, it has been shown that MRI images with reasonable quality are achievable under high inhomogeneities [<xref ref-type="bibr" rid="B32">32</xref>, <xref ref-type="bibr" rid="B33">33</xref>], including inhomogeneities higher than 1,000&#xa0;ppm [<xref ref-type="bibr" rid="B34">34</xref>, <xref ref-type="bibr" rid="B35">35</xref>]. Image distortions due to magnetic field inhomogeneities can be compensated by different methods [<xref ref-type="bibr" rid="B36">36</xref>&#x2013;<xref ref-type="bibr" rid="B38">38</xref>].</p>
<p>The use of non-cryogenic magnets usually implies dealing with magnetic field instability. This feature is usually an important point to consider if electromagnets are chosen. However, the implementation of electromagnets in MRI prototypes has shown promising results [<xref ref-type="bibr" rid="B2">2</xref>, <xref ref-type="bibr" rid="B3">3</xref>, <xref ref-type="bibr" rid="B39">39</xref>, <xref ref-type="bibr" rid="B40">40</xref>]. This approach is attractive since it allows switching the magnetic field [<xref ref-type="bibr" rid="B41">41</xref>, <xref ref-type="bibr" rid="B42">42</xref>], then enabling the acquisition of images with contrasts unavailable at fixed magnetic fields [<xref ref-type="bibr" rid="B43">43</xref>, <xref ref-type="bibr" rid="B44">44</xref>]. Recent examples are the combination of field cycling MRI with superparamagnetic nanoparticles [<xref ref-type="bibr" rid="B45">45</xref>] and proton double irradiation [<xref ref-type="bibr" rid="B46">46</xref>].</p>
<p>Despite the high potential associated with the use of electromagnets in low-field MRI, the study of magnetic field instability-induced artifacts is a topic that has hardly been considered in the literature. In the context of MRI, this point is usually skipped although some mentions can be found in hardware-related papers [<xref ref-type="bibr" rid="B47">47</xref>, <xref ref-type="bibr" rid="B48">48</xref>]. In addition, the main strategies for artifact correction due to low field stability are post-processing methods [<xref ref-type="bibr" rid="B37">37</xref>, <xref ref-type="bibr" rid="B49">49</xref>], and most of them were focused on mitigating motion artifacts at high magnetic fields [<xref ref-type="bibr" rid="B50">50</xref>, <xref ref-type="bibr" rid="B51">51</xref>]. Therefore, the practically nonexistent studies on this area represent a limitation for the development of the new generation of low field MRI scanners based on electromagnets.</p>
<p>In reference 37 we proposed an algorithm to correct instabilities and inhomogeneities for Cartesian spin-echo acquisitions. The method consists in a two step sequential algorithm considering both random field fluctuations and field inhomogeneity corrections, neglecting any potential correlation between them. In this work we made an exhaustive analysis of the role of magnetic field stability in the context of MRI. We discuss the relationship between magnetic field stability and homogeneity in the context of both the NMR and MRI signal. The presented model allows observing that spin dephasing due to magnetic field instability is proportional to the field strength at which the imaging sequence is applied. This fact represents a key feature supporting the use of electromagnets in low-field MRI technology.</p>
<p>We present a simple mathematical model for an inhomogeneous and unstable magnetic field, and performed simulations using this model for the spin-echo and the gradient-echo pulse sequences [<xref ref-type="bibr" rid="B52">52</xref>]. For the simulations, we considered two types of instabilities, one completely random and the case where a dominant monochromatic harmonic component is present. NMR experiments were done to test and validate our model and simulations. Experimental images were acquired in a resistive homemade MRI relaxometer with different echo times, in order to characterize the image degradation due to instability. Finally, strategies are discussed to minimize the undesirable effects induced by magnetic field instability. This work is important in view of a potential reduction of costs associated to the involved hardware.</p>
</sec>
<sec sec-type="materials|methods" id="s2">
<title>2 Materials and methods</title>
<sec id="s2-1">
<title>2.1 Magnetic field model</title>
<p>In this section we consider a magnetic field model for an electromagnet (switched or not) that is fed by a current source. In this frame, we consider a realistic picture in which the magnet homogeneity and the current stability are not perfect. However, we will neglect a potential time dependence of the magnetic field homogeneity.</p>
<p>The magnetic flux density at a given position <bold>r</bold> (<italic>x</italic>,<italic>y,z</italic>), <bold>B</bold> (<bold>r</bold>,<italic>t</italic>), can be factorized as a product of two well defined independent functions. The first one, <bold>C</bold>(<bold>r</bold>) is associated with the spatial dependence of the generated magnetic field, which essentially depends on the magnet geometry. This functional also includes magnet machining and assembly imperfections that degrade the generated field homogeneity. The second function <italic>I</italic>(<italic>t</italic>) is the current in the circuit. If we define the mean temporal value <italic>I</italic>
<sub>0</sub> of the current then, <italic>I</italic>(<italic>t</italic>) &#x3d; <italic>I</italic>
<sub>0</sub>&#x2b;<italic>&#x394;I</italic>(<italic>t</italic>). <italic>&#x394;I</italic>(<italic>t</italic>) is associated with the temporal instability of the current, which is mainly attributed to the quality and performance of the control and power electronics of the magnet current power supply. In the same way, <bold>C</bold>(<bold>r</bold>) can be expressed as the sum of a constant <bold>C</bold>
<sub>
<bold>0</bold>
</sub> <italic>&#x3d;</italic> <bold>C</bold> (<bold>0</bold>) (where the position <bold>0</bold> &#x3d; (0,0,0) is located at the geometric center of the magnet), plus a position-dependent term <italic>&#x394;</italic>
<bold>C</bold>(<bold>r</bold>). The proposed factorization allows us to clearly separate the spatial inhomogeneities from the temporal instability (described by &#x394;<bold>C</bold>(<bold>r</bold>) and <italic>&#x394;I</italic>(t) respectively) in the factors that describe the evolution of the phase during the image acquisition (see Eq. <xref ref-type="disp-formula" rid="e4">(4)</xref>). Through these considerations, it is possible to express the magnetic field as it is shown in Eq. <xref ref-type="disp-formula" rid="e1">(1)</xref>:<disp-formula id="e1">
<mml:math id="m1">
<mml:mrow>
<mml:mi mathvariant="bold">B</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mn mathvariant="bold">0</mml:mn>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:mo>&#x394;</mml:mo>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
<label>(1)</label>
</disp-formula>
</p>
<p>Applying the distributive property of the product, <bold>B</bold> (<bold>r</bold>,<italic>t</italic>) can be expressed as:<disp-formula id="e2">
<mml:math id="m2">
<mml:mrow>
<mml:mi mathvariant="bold">B</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:msub>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:mo>&#x394;</mml:mo>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:mo>&#x394;</mml:mo>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
<label>(2)</label>
</disp-formula>
</p>
<p>The first term of Eq. <xref ref-type="disp-formula" rid="e2">2</xref> can be associated with the main magnetic field value <bold>B</bold>
<sub>0</sub>. The second term can be interpreted as the magnetic field inhomogeneity while the third term as the magnetic field instability. The last term is typically much smaller than the others and it will be neglected. Therefore, neglecting the last term, the magnetic field can be approximated as:<disp-formula id="e3">
<mml:math id="m3">
<mml:mrow>
<mml:mi mathvariant="bold">B</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2248;</mml:mo>
<mml:msub>
<mml:mi mathvariant="bold">B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mn>1</mml:mn>
<mml:mo>&#x2b;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi mathvariant="bold">C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mo>&#x2b;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
<label>(3)</label>
</disp-formula>
</p>
<p>The fact that <bold>C</bold> and <italic>I</italic> have been considered as independent variables is an approximation. The model does not consider variations in the geometry of the magnet during the experiment. This condition is fulfilled in most cases, although it can be violated in extreme situations. Nevertheless, this formalization is adequate to describe and interpret the effects of magnetic field instability in several NMR and MRI experiments.</p>
<sec id="s2-1-1">
<title>2.1.1 Free induction decay signal under magnetic field instability and inhomogeneity</title>
<p>As the free induction decay (FID) is the simplest experiment in NMR, it is an appropriate test for our magnetic field model. Furthermore, if the model is accurate, it can be useful to understand the relationship between the magnetic field inhomogeneity and stability in complex experiments.</p>
<p>The signal is modeled without considering relaxation effects as the magnetic field inhomogeneity and instability are considered to be dominant. That means that the signal loss attributed to <italic>T</italic>
<sub>1</sub> and <italic>T</italic>
<sub>2</sub> is much smaller than the loss induced by the magnetic field deviations. The defacing during the RF pulses is neglected. This is equivalent to assuming that the RF pulses are perfect and infinitely short. Defining <italic>&#x3c1;</italic>(<bold>r</bold>) as the spin density of the sample and &#x3b3; the proton gyromagnetic ratio, the FID signal can be modeled as:<disp-formula id="e4">
<mml:math id="m4">
<mml:mrow>
<mml:mi>s</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mo>&#x222d;</mml:mo>
<mml:mi>&#x3c1;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>z</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:msup>
<mml:mi mathvariant="italic">exp</mml:mi>
<mml:mrow>
<mml:mo>&#x2212;</mml:mo>
<mml:mi mathvariant="normal">&#x3b3;</mml:mi>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold-italic">x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi mathvariant="bold-italic">y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi mathvariant="bold-italic">z</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>t</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mn>0</mml:mn>
<mml:mi>t</mml:mi>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:msup>
<mml:mo>&#x2061;</mml:mo>
<mml:mi>d</mml:mi>
<mml:mi>x</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>y</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>z</mml:mi>
</mml:mrow>
</mml:math>
<label>(4)</label>
</disp-formula>
</p>
<p>In Eq. <xref ref-type="disp-formula" rid="e4">(4)</xref>, we integrate over the whole space (the integrals limits go from -&#x221e; to &#x221e;), and the dimensions of the sample are contained in <inline-formula id="inf1">
<mml:math id="m5">
<mml:mrow>
<mml:mi>&#x3c1;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>z</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
</inline-formula>. Eq. <xref ref-type="disp-formula" rid="e4">(4)</xref> suggests that the magnetic field instability does not modify the signal amplitude, but it generates frequency variations and phase shifts accumulations.</p>
<p>According to this scheme, if the following condition is satisfied:<disp-formula id="e5">
<mml:math id="m6">
<mml:mrow>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi mathvariant="bold">r</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi mathvariant="normal">C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mo>&#x226b;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mn>1</mml:mn>
</mml:mrow>
<mml:mrow>
<mml:mi mathvariant="normal">t</mml:mi>
</mml:mrow>
</mml:mfrac>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mn>0</mml:mn>
<mml:mi mathvariant="normal">t</mml:mi>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi mathvariant="normal">I</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi mathvariant="normal">t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi mathvariant="normal">I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi mathvariant="normal">d</mml:mi>
<mml:mover accent="true">
<mml:mrow>
<mml:mi mathvariant="normal">t</mml:mi>
<mml:mo>,</mml:mo>
</mml:mrow>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
<label>(5)</label>
</disp-formula>then the NMR signal will show a certain immunity to the magnetic field instability. This means that the effects of the instability on the NMR signal will be less relevant if the inhomogeneity is dominant over the instability. In other words, as the frequency spectrum of the instability is constrained within the bandwidth covered by the field inhomogeneity, the NMR signal gains immunity against magnetic field fluctuations (periodic and random noise). This fact constitutes a key-feature for gaining signal stability against magnetic field or current instability through controlled homogeneity degradation. In <xref ref-type="sec" rid="s2-3-1">section 2.3.1</xref> we will present experimental evidence that verify this theoretical speculation.</p>
</sec>
<sec id="s2-1-2">
<title>2.1.2 MRI signal under magnetic field instability and inhomogeneity</title>
<p>Now we extend the model described on Eq. <xref ref-type="disp-formula" rid="e3">3</xref> to the context of 2D MRI. Here we consider signals for spin-echo (SEs) and gradient-echo (GEs) sequences [<xref ref-type="bibr" rid="B52">52</xref>]. The mathematical expressions correspond to the pulse sequences shown in <xref ref-type="fig" rid="F1">Figure 1</xref>. The gray rectangles represent the (ideal-perfect) RF pulses, <italic>AD</italic> the acquisition delay (time between the last RF pulse and the beginning of the acquisition window), and <italic>AT</italic> the acquisition time. In the spin echo sequence, <italic>ET</italic> coincides with the center of the acquisition window, that is, <italic>ET/</italic>2 <italic>&#x3d; AD &#x2b; AT/</italic>2.</p>
<fig id="F1" position="float">
<label>FIGURE 1</label>
<caption>
<p>pulse sequences used for the simulations. <italic>ET</italic>: echo-time, <italic>AD</italic>: acquisition delay, <italic>AT</italic>: acquisition time, and <inline-formula id="inf2">
<mml:math id="m7">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
</inline-formula> evolution time. The gray rectangles represent the RF pulses. The defacing during the RF pulses was not considered, which is equivalent to assuming that the pulses are perfect and as Dirac pulse (RF duration &#x3c;&#x3c; <italic>ET</italic>). Thus, the echo will be centered at <italic>ET</italic> in the spin echo sequences.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g001.tif"/>
</fig>
<p>Hence, the expression for an acquired signal with the 2D spin echo sequence is:<disp-formula id="e6">
<mml:math id="m8">
<mml:mrow>
<mml:msub>
<mml:mi>s</mml:mi>
<mml:mrow>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:munderover>
<mml:mo>&#x222d;</mml:mo>
<mml:mrow>
<mml:msub>
<mml:mi>z</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x2212;</mml:mo>
<mml:mo>&#x2206;</mml:mo>
<mml:mi>z</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
<mml:mrow>
<mml:msub>
<mml:mi>z</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:mo>&#x2206;</mml:mo>
<mml:mi>z</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
</mml:munderover>
<mml:mi>&#x3c1;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>z</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:msup>
<mml:mi mathvariant="italic">exp</mml:mi>
<mml:mrow>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>i</mml:mi>
<mml:mi mathvariant="normal">&#x3b3;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>y</mml:mi>
<mml:mtext>&#x2009;</mml:mtext>
<mml:msub>
<mml:mi>&#x3c4;</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>r</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>x</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:msubsup>
<mml:mo>&#x394;</mml:mo>
<mml:mi>B</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
<mml:mo>&#x2212;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mn>0</mml:mn>
<mml:mrow>
<mml:mi mathvariant="italic">ET</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
</mml:msubsup>
<mml:mo>&#x394;</mml:mo>
<mml:mi>B</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:msup>
<mml:mo>&#x2061;</mml:mo>
<mml:mi>d</mml:mi>
<mml:mi>x</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>y</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>z</mml:mi>
<mml:mo>,</mml:mo>
</mml:mrow>
</mml:math>
<label>(6)</label>
</disp-formula>where <italic>&#x394;B</italic> (<italic>x,y,t</italic>) <italic>&#x3d; B</italic> (<italic>x,y,t</italic>)<italic>&#x2010;B</italic>
<sub>0</sub>, <italic>G</italic>
<sub>p</sub> the phase gradient intensity and <italic>&#x3c4;</italic>
<sub>p</sub> the phase gradient pulse time. <italic>G</italic>
<sub>r</sub> is the read gradient intensity, and <italic>t</italic> &#x2208; [0<italic>, AT</italic>]. <italic>z</italic> is the slice selection direction, <italic>&#x394;z</italic> is the slice size and <italic>z</italic>
<sub>0</sub> the center of the slice. This equation does not consider the effects that instability could generate during the slice selection process. Under this approximation, the dependence on <italic>z</italic> can be avoided by considering the projection of the slice in <italic>z</italic>
<sub>0</sub>. The times associated with the readout pulses were defined as <italic>AT/</italic>2 before the &#x3c0; pulse, and <italic>AT</italic> during the signal acquisition. Using the field model of Eq. <xref ref-type="disp-formula" rid="e3">3</xref> and Eq. <xref ref-type="disp-formula" rid="e6">6</xref> can be rewritten as:<disp-formula id="e7">
<mml:math id="m9">
<mml:mrow>
<mml:msub>
<mml:mi>s</mml:mi>
<mml:mrow>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mo>&#x222c;</mml:mo>
<mml:mi>&#x3c1;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:msub>
<mml:mi>z</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:msup>
<mml:mi mathvariant="italic">exp</mml:mi>
<mml:mrow>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>i</mml:mi>
<mml:mi mathvariant="normal">&#x3b3;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>y</mml:mi>
<mml:mtext>&#x2009;</mml:mtext>
<mml:msub>
<mml:mi>&#x3c4;</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>p</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>r</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>x</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>r</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:msup>
<mml:mo>&#x2061;</mml:mo>
<mml:mi>d</mml:mi>
<mml:mi>x</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
</mml:mrow>
</mml:math>
<label>(7)</label>
</disp-formula>where <inline-formula id="inf3">
<mml:math id="m10">
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>p</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mo>&#x3d;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mrow>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mrow>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
<mml:mo>&#x2212;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mn>0</mml:mn>
<mml:mrow>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mrow>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
</inline-formula> and <inline-formula id="inf4">
<mml:math id="m11">
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>r</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>s</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mrow>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mrow>
<mml:mi>E</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
</mml:mrow>
<mml:mn>2</mml:mn>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:msubsup>
</mml:mrow>
<mml:mrow>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
<mml:mo>.</mml:mo>
</mml:mrow>
</mml:math>
</inline-formula>
</p>
<p>According to these equations, the accumulated dephasing (non-refocused spins) due to the magnetic field instability at the echo time is <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>p-se</sub>. This term can be considered time-independent and is the equivalent to the phase error proposed in reference [<xref ref-type="bibr" rid="B48">48</xref>]. Therefore, <italic>I</italic>
<sub>p-se</sub> generates defects along the phase encoding direction. Correspondingly, <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>r-se</sub> is the dephasing generated by the field instability during the signal acquisition. It is time-dependent and affects the read encoding direction (in similarity with the approach made in [<xref ref-type="bibr" rid="B47">47</xref>]). It is important to notice that both terms (<italic>I</italic>
<sub>p-se</sub> and <italic>I</italic>
<sub>r-se</sub>) are multiplied by <italic>B</italic>
<sub>0</sub>, which means that the signal is more sensitive to the instability at high fields.</p>
<p>On the other hand, the signal in a 2D gradient echo sequence can be written as:<disp-formula id="e8">
<mml:math id="m12">
<mml:mrow>
<mml:msub>
<mml:mi>s</mml:mi>
<mml:mrow>
<mml:mi>g</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>,</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:mo>&#x222c;</mml:mo>
<mml:mi>&#x3c1;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
<mml:msub>
<mml:mi>z</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:msup>
<mml:mi>e</mml:mi>
<mml:mrow>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>i</mml:mi>
<mml:mi mathvariant="normal">&#x3b3;</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>y</mml:mi>
<mml:mtext>&#x2009;</mml:mtext>
<mml:msub>
<mml:mi>&#x3c4;</mml:mi>
<mml:mi>p</mml:mi>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>p</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>g</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>D</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:msub>
<mml:mi>G</mml:mi>
<mml:mi>r</mml:mi>
</mml:msub>
<mml:mtext>&#x2009;</mml:mtext>
<mml:mi>x</mml:mi>
<mml:msub>
<mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>B</mml:mi>
</mml:mrow>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mfrac>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
</mml:mrow>
<mml:mn>2</mml:mn>
</mml:mfrac>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>C</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>x</mml:mi>
<mml:mo>,</mml:mo>
<mml:mi>y</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>C</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mo>&#x2b;</mml:mo>
<mml:msub>
<mml:mi>B</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>r</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>g</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:msup>
<mml:mi>d</mml:mi>
<mml:mi>x</mml:mi>
<mml:mi>d</mml:mi>
<mml:mi>y</mml:mi>
<mml:mo>,</mml:mo>
</mml:mrow>
</mml:math>
<label>(8)</label>
</disp-formula>where <inline-formula id="inf5">
<mml:math id="m13">
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>p</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>g</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mo>&#x3d;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mn>0</mml:mn>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>D</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
</inline-formula> and <inline-formula id="inf6">
<mml:math id="m14">
<mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mi>r</mml:mi>
<mml:mo>&#x2212;</mml:mo>
<mml:mi>g</mml:mi>
<mml:mi>e</mml:mi>
</mml:mrow>
</mml:msub>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>&#x3d;</mml:mo>
<mml:msubsup>
<mml:mo>&#x222b;</mml:mo>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>D</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>A</mml:mi>
<mml:mi>T</mml:mi>
<mml:mo>/</mml:mo>
<mml:mn>2</mml:mn>
</mml:mrow>
<mml:mrow>
<mml:mi>A</mml:mi>
<mml:mi>D</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:msubsup>
<mml:mfrac>
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
</mml:mfrac>
<mml:mi>d</mml:mi>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
</inline-formula>.</p>
<p>Here, <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>p-ge</sub> is the accumulated dephasing due to the magnetic field instability in the middle of the acquisition window. In contrast with the spin-echo sequence, the integral has the same sign at any instant. It can be considered time-independent and, as a consequence, generates defects along the phase encoding direction. In contrast, the dephasing generated by the instability during the signal acquisition is <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>r-ge</sub>. It is time-dependent and affects the readout encoding direction. It is important to note that in this case the magnetic field inhomogeneity generates artifacts in both encoding directions. This is a key-difference between both sequences: in the spin-echo sequence the inhomogeneity only affects the readout encoding direction. In contrast, in the gradient-echo sequence the inhomogeneity is never refocused, thus increasing the undesired effects in the images. As well as for the SE, both terms (<italic>I</italic>
<sub>p-ge</sub> and <italic>I</italic>
<sub>r-ge</sub>) are multiplied by <italic>B</italic>
<sub>0</sub>. That is, the signal is more sensitive to the instability at high fields.</p>
</sec>
</sec>
<sec id="s2-2">
<title>2.2 Simulations</title>
<p>In order to analyze the described effects on the signals we have simulated instabilities with two different characteristics: random fluctuations of the magnetic field, and including a dominant frequency component. These two cases represent the main sources of instabilities usually present in the associated instrumentation. Either random noise originated in the electronics (thermal noise, shot, flicker and transit time noises originated in semiconductor devices, and other), or unfiltered low-frequency components (typical harmonics from the power line or originated in the power electronics), and/or high-frequency radio signals that resonate, couples or interferes with the instrument electronics.</p>
<p>For both cases, &#x394;C (<italic>x,y</italic>) was simulated using a random surface rsgeng2D (<ext-link ext-link-type="uri" xlink:href="http://www.mysimlabs.com/">http://www.mysimlabs.com</ext-link>). This function generates a 2D matrix that was previously tested in diverse areas [<xref ref-type="bibr" rid="B53">53</xref>, <xref ref-type="bibr" rid="B54">54</xref>]. The algorithm allows for changing the correlation length between matrix components in both directions (rows and columns). &#x394;C (<italic>x,y</italic>) has the same matrix size as the image, and each element corresponds to a position in the space. As the inhomogeneity is expected to be smooth, the correlation parameter was fixed at 0.5 (it can vary from 0 to 1) in both directions.</p>
<p>The degradation of the image quality was calculated as a function of the <italic>ET</italic> for both cases. In order to simplify calculations, the slice selection process was neglected, and the images were simulated as 2D projection images. As mentioned above, <inline-formula id="inf7">
<mml:math id="m15">
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
</inline-formula> was simulated in two different ways: random and sinusoidal. These two modalities are detailed in <xref ref-type="sec" rid="s2-2-1">Subsections 2.2.1</xref> and <xref ref-type="sec" rid="s2-2-2">2.2.2</xref>.</p>
<p>For all the simulations, we fixed the maximum gradient amplitudes at 45&#xa0;mT/m. The main magnetic field was fixed at 0.125 T, the maximum inhomogeneity was fixed at 1,400&#xa0;ppm (representative of the measured value within the volume of interest, VOI, in our experiments) and the instability at 220&#xa0;ppm (with a simulation timescale that is consistent with the acquired signal duration), in accordance with the characteristics of our homemade prototype [<xref ref-type="bibr" rid="B35">35</xref>]. The simulations were performed without adding noise, that is, the whole signal comes from the sample. In addition, the gradient fields were assumed to be perfectly linear, and concomitant fields were not considered.</p>
<p>The simulations were performed in Matlab and processed on an Intel Core i7-8565U CPU with a processor base frequency of 1.8&#xa0;GHz.</p>
<sec id="s2-2-1">
<title>2.2.1 Random instability</title>
<p>
<inline-formula id="inf8">
<mml:math id="m16">
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
</inline-formula> is represented by a 2D matrix associated with the two <italic>k</italic>-space coordinates (the acquisition time <inline-formula id="inf9">
<mml:math id="m17">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:math>
</inline-formula> and the phase-encoding index). Each row is associated with a phase-encoding index. The time difference between matrix components of the same row is <italic>dt &#x3d; AT/N</italic>, where <italic>N</italic> is the number of acquired points. The first value of the 2D matrix is random, and the rest of the values are defined by two correlations parameters. This function was also generated with rsgeng2D but using different correlation parameters. A correlation parameter along the column direction (short-term instability) was set to 0.5 while for the row direction (long-term instability) to 0.005. That means that the stability between acquisitions is less correlated than the stability during each acquisition. This choice turns consistent with the observed behavior during an experiment. We considered 4 cases: perfect image (perfectly homogeneous and stable magnetic field), instability effects only, inhomogeneity effects only, and image simulated under both inhomogeneity and instability effects. The 4 cases were simulated for both GEs and SEs sequences. The GEs simulations were made with <italic>N &#x3d;</italic> 64, <italic>AD &#x3d;</italic> 0.5&#xa0;ms and <italic>AT &#x3d;</italic> 1&#xa0;ms, while the SEs with <italic>N &#x3d;</italic> 64<italic>, ET &#x3d;</italic> 2&#xa0;ms and <italic>AT &#x3d;</italic> 1&#xa0;ms.</p>
<p>To characterize more precisely how the instability affects the images acquired through these two pulse sequences, we have simulated images with stabilities from 50&#xa0;ppm to 1,200&#xa0;ppm. In addition, a quantitative parameter was defined as the sum of the pixels outside of the reference image (ORI). Thus, the ORI is zero for the ideal reference image. If the image becomes distorted or we have ghosting due to magnetic field instability, non-zero signal intensity will appear in the region corresponding to the dark portion of the reference image. In this case the ORI becomes nonzero. Simulations were repeated 32 times for each instability value, and the ORI signal intensity was averaged to avoid any abnormal behavior due to the random instability.</p>
<p>The two bright circles of the reference image represent, for example, two tubes filled with water. This means that the total signal intensity of the image must be independent of the image distortion induced by magnetic field instability. In other words, an image distortion can be thought as a signal spreading outside the region defined by the bright circles, i.e., into the dark region corresponding to the reference image (which emulates the situation of an image acquired with perfect magnetic field homogeneity and stability).</p>
<p>Finally, we analyzed the effect of the echo time duration in the SEs. We fixed the instability to 220&#xa0;ppm, varied the echo time from 1&#xa0;ms to 21&#xa0;ms, and calculated the average signal intensity in the ORI over 32 simulations.</p>
</sec>
<sec id="s2-2-2">
<title>2.2.2 Sinusoidal instability</title>
<p>In this subsection, the instability is simulated as a pure harmonic component where each acquisition starts with a random phase <italic>&#x3d5;</italic>:<disp-formula id="e9">
<mml:math id="m18">
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mi>t</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
<mml:mo>/</mml:mo>
<mml:msub>
<mml:mi>I</mml:mi>
<mml:mn>0</mml:mn>
</mml:msub>
<mml:mo>&#x3d;</mml:mo>
<mml:mi>A</mml:mi>
<mml:mo>&#x2061;</mml:mo>
<mml:mi>sin</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mn>2</mml:mn>
<mml:mi>&#x3c0;</mml:mi>
<mml:mi>f</mml:mi>
<mml:mi>t</mml:mi>
<mml:mo>&#x2b;</mml:mo>
<mml:mi>&#x3d5;</mml:mi>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
<label>(9)</label>
</disp-formula>
</p>
<p>The maximum amplitude <italic>A</italic> was fixed at 220&#xa0;ppm; thus, the instability will be a sinusoidal function oscillating between &#xb1;220&#xa0;ppm. This represents a high instability considering that cryogenic magnets can be stabilized to values lower than 1 part-per-billion [<xref ref-type="bibr" rid="B55">55</xref>].</p>
<p>Simulations were only performed for the SEs. Firstly, we simulated images with instabilities with the main frequency in the range of 0&#x2013;2&#xa0;kHz for three different echo times (<italic>ET</italic> &#x3d; 1&#xa0;ms, 2&#xa0;ms and 4&#xa0;ms), and calculated the mean signal in the ORI. Then, we performed simulations with instabilities with frequencies in the range of 0&#x2013;20&#xa0;kHz and for four echo times (<italic>ET</italic> &#x3d; 1&#xa0;ms, 2&#xa0;ms, 4&#xa0;ms, and 16&#xa0;ms).</p>
<p>Finally, we simulated three images: perfect conditions (reference), and then two images with different echo times (38&#xa0;ms and 40&#xa0;ms) and same instability with a main frequency of 50&#xa0;Hz (typical frequency instability induced by the electric line). We specifically chose 40&#xa0;ms as echo time because <italic>ET</italic>/2 is exactly one frequency period.</p>
</sec>
</sec>
<sec id="s2-3">
<title>2.3 Experiments</title>
<p>We present two experiments to demonstrate the correct behavior of our model and simulations. The first experiment consisted of measuring FID signals under different magnetic field homogeneities. The second one consisted of acquiring 2D fast-field-cycling (FFC) MRI images in our home-made prototype under different echo times. This last experiment allows determining the nature of the dominant instability affecting the images (random noise or dominant frequency). This information turns relevant to delineate the best method to minimize the artifacts.</p>
<sec id="s2-3-1">
<title>2.3.1 NMR signal</title>
<p>The following experiment was aimed to check the predictions suggested by the proposed model. FID signals were acquired with one scan, two scans, and four scans under different magnetic field homogeneities (90&#xa0;ppm, 250&#xa0;ppm, and 370&#xa0;ppm). Hence, if the signals are immune to the instability, they can be accumulated (or averaged) reducing the white noise without signal loss. That is, the only difference related to thenumber of scans must be the SNR.</p>
<p>The experiment was performed in a commercial Stelar Spinmaster FC-2000/C/D (Mede, Italy) relaxometer. The magnetic field homogeneity was changed by relocating the probe (sample) in the magnet away from the central region (maximal homogeneity). The sample was deionized water with 4.5&#xa0;mm of height in a standard 10&#xa0;mm diameter test tube. The pulse sequence was a non-polarized (NP) sequence [<xref ref-type="bibr" rid="B42">42</xref>], with a polarization and detection field of 0.375&#xa0;T (15&#xa0;MHz in terms of proton Larmor frequency). The receptor bandwidth was set to 40&#xa0;kHz and the acquisition window 4.5&#xa0;ms.</p>
</sec>
<sec id="s2-3-2">
<title>2.3.2 2D spin echo images</title>
<p>Images were acquired with different echo times in a homemade MRI-relaxometer with magnetic field instability of 220&#xa0;ppm/s and inhomogeneity of 1,400&#xa0;ppm within a VOI of the order of 35&#xa0;cm<sup>3</sup> [<xref ref-type="bibr" rid="B35">35</xref>]. The instability was measured from the standard deviation of the frequency shifts of the peak value of the Fourier transform (FT) of the echo signal (acquired without gradients, 50 samples acquired consecutively). The pulse sequence was a non-polarized (NP) sequence [<xref ref-type="bibr" rid="B42">42</xref>] with a relaxation and detection field of 0.125&#xa0;T (5&#xa0;MHz in terms of proton Larmor frequency). The image acquisition was based on a spin-echo sequence with hard RF pulses, an irradiation bandwidth of 100&#xa0;kHz andan acquisition window of 1,024&#xa0;&#xb5;s. Six images with echo times from 2&#xa0;ms to 11.1&#xa0;ms were acquired. Images were acquired without slice selection to maximize the SNR. The sample was a 25&#xa0;mM solution of deionized water (18&#xa0;M&#x2126;/cm, milli-Q purification system Osmion 5, Apema, Villa Dominico, Buenos Aires, Argentina) and copper sulfate (Ciccarelli, San Lorenzo, Santa F&#xe9;, Argentina).</p>
</sec>
</sec>
</sec>
<sec sec-type="results" id="s3">
<title>3 Results</title>
<sec id="s3-1">
<title>3.1 Simulations</title>
<sec id="s3-1-1">
<title>3.1.1 Random instability</title>
<p>
<xref ref-type="fig" rid="F2">Figure 2</xref> shows simulated images for a GEs with <italic>N</italic> &#x3d; 64, <italic>AD</italic> &#x3d; 0.5&#xa0;ms and <italic>AT</italic> &#x3d; 1&#xa0;ms. <xref ref-type="fig" rid="F2">Figure 2A</xref> shows the reference image (perfect homogeneity and stability). <xref ref-type="fig" rid="F2">Figure 2B</xref> corresponds to the image affected by both instability and inhomogeneity. <xref ref-type="fig" rid="F2">Figure 2C</xref> the image under effects of inhomogeneity only (<inline-formula id="inf10">
<mml:math id="m19">
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
</inline-formula> <italic>&#x3d;</italic> 0) and <xref ref-type="fig" rid="F2">Figure 2D</xref> the image under effects of instability only (<italic>&#x394;C</italic> (<italic>x,y</italic>) <italic>&#x3d;</italic> 0).</p>
<fig id="F2" position="float">
<label>FIGURE 2</label>
<caption>
<p>Simulated images with gradient-echo sequence. The vertical direction corresponds to the phase encoding direction. <bold>(A)</bold> Reference image, <bold>(B)</bold> Degraded by both inhomogeneity and instability, <bold>(C)</bold> Degraded by inhomogeneity, <bold>(D)</bold> Degraded by instability. <italic>N</italic> &#x3d; 64, <italic>AD</italic> &#x3d; 0.5&#xa0;ms and <italic>AT</italic> &#x3d; 1&#xa0;ms.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g002.tif"/>
</fig>
<p>
<xref ref-type="fig" rid="F3">Figure 3C</xref> shows images simulated for a SEs with <italic>ET</italic> &#x3d; 2&#xa0;ms and <italic>AT</italic> &#x3d; 1&#xa0;ms. As in the previous case, A: reference, B: dual degradation, C: inhomogeneity only and D: instability only.</p>
<fig id="F3" position="float">
<label>FIGURE 3</label>
<caption>
<p>Simulated images with spin-echo sequence. The vertical direction corresponds to phase encoding. <bold>(A)</bold> Reference image, <bold>(B)</bold> Degraded by both inhomogeneity and instability, <bold>(C)</bold> Degraded by inhomogeneity, <bold>(D)</bold> Degraded by instability. <italic>N &#x3d;</italic> 64<italic>, ET &#x3d;</italic> 2&#xa0;ms <italic>and AT &#x3d;</italic> 1&#xa0;ms.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g003.tif"/>
</fig>
<p>
<xref ref-type="fig" rid="F4">Figure 4</xref> shows the ORI signal intensity as a function of the magnetic field instability from 50&#xa0;ppm to 1200&#xa0;ppm. Due to the random nature of the instability, each point in the curve wasobtained by averaging 32 images. The associated uncertainty to each point is the standard deviation of the mean value after considering a coverage factor of 95%. The total intensity of each image was normalized to 1, so the calculated ORI is proportional to the percentage of signal spread outside thebright circles. The ORI signal is zero for the reference image and is thus not shown in <xref ref-type="fig" rid="F4">Figure 4</xref>. For a highly degraded image, where the signal intensity is homogeneously spread along the phase direction, the ORI intensity approaches 0.5. That is, this is the expected maximum for a totally degraded image. In fact, the bright circles occupy 50.7% of the image area along the phase direction. Thus, in the case the net signal intensity of the circles becomes homogeneously spread along the phase direction, the theoretical value of the ORI in this example is 0.493.</p>
<fig id="F4" position="float">
<label>FIGURE 4</label>
<caption>
<p>signal intensity ORI vs. magnetic field instability. The black curve is associated with the GEs and the red with SEs.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g004.tif"/>
</fig>
<p>In <xref ref-type="fig" rid="F4">Figure 4</xref>, the signal intensity ORI in the GEs is greater than in SEs until a stability of 1,000&#xa0;ppm, where the results from both sequences are undistinguishable. The GEs saturate rapidly to values close to 0.5, showing a great sensitivity to instability variations (between 0&#xa0;ppm and 250&#xa0;ppm). On the other hand, the SEs is more robust, saturating at instabilities of 1,000&#xa0;ppm.</p>
<p>
<xref ref-type="fig" rid="F5">Figure 5</xref> shows the effect of the echo time duration in the SEs for a fixed instability to 220&#xa0;ppm, within an echo time range from 1&#xa0;ms to 21&#xa0;ms. As in the previous case, the average ORI signal intensity results from 32 simulations. This plot suggests that short echo times are preferable to avoid image degradation due to instability.</p>
<fig id="F5" position="float">
<label>FIGURE 5</label>
<caption>
<p>ORI signal intensity as a function of <italic>ET</italic> in images simulated through SEs, for a fixed instability to 220&#xa0;ppm.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g005.tif"/>
</fig>
</sec>
<sec id="s3-1-2">
<title>3.1.2 Sinusoidal instability</title>
<p>
<xref ref-type="fig" rid="F6">Figure 6</xref> shows the ORI signal intensity dependence on the main frequency of the instability in the range from 0 to 2&#xa0;kHz for three different echo times (<italic>ET</italic> &#x3d; 1&#xa0;ms, 2&#xa0;ms, and 4&#xa0;ms). As can be observed, when the frequency associated with <italic>ET</italic>/2 is proportional to the instability frequency, the ORI signal intensity becomes minimized. In other words, when an integer number of cycles fits <italic>ET</italic>/2, each term of <italic>I</italic>
<sub>r-se</sub> is null, and, consequently, the ORI signal is minimal (see <xref ref-type="fig" rid="F7">Figure 7</xref>). Furthermore, each minimum has an associated bandwidth, showing that the pulse sequences work as a filter. In addition, the bandwidth increases at higher frequencies (see <xref ref-type="fig" rid="F6">Figure 6</xref>, <italic>ET</italic> &#x3d; 4&#xa0;ms). However, the ORI is not equal to 0 because <italic>I</italic>
<sub>r-se</sub> is always nonzero. <xref ref-type="fig" rid="F8">Figure 8</xref> shows how critical is the choice of the echo time is SEs experiment to avoid ghosting in the image due to monochromatic instability.</p>
<fig id="F6" position="float">
<label>FIGURE 6</label>
<caption>
<p>Signal intensity ORI as function of the instability frequency. Images simulated for SEs. Left: <italic>ET</italic> &#x3d; 1&#xa0;ms, middle: <italic>ET</italic> &#x3d; 2&#xa0;ms and right: <italic>ET</italic> &#x3d; 4&#xa0;ms. The uncertainty of each point is not shown for a clear interpretation of the curve&#x2019;s behaviors, but it is represented in the variability of the dataset.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g006.tif"/>
</fig>
<fig id="F7" position="float">
<label>FIGURE 7</label>
<caption>
<p>Signal intensity ORI as function of the instability frequency. Images simulated for SEs. Top left: <italic>ET</italic> &#x3d; 1&#xa0;ms, top right: <italic>ET</italic> &#x3d; 2&#xa0;ms, bottom left: <italic>ET</italic> &#x3d; 4&#xa0;ms and bottom right: <italic>ET</italic> &#x3d; 16&#xa0;ms.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g007.tif"/>
</fig>
<fig id="F8" position="float">
<label>FIGURE 8</label>
<caption>
<p>Simulated images with SEs. <bold>(A)</bold> Reference image. <bold>(B)</bold> Image with frequency instability of 50&#xa0;Hz and <italic>ET</italic> &#x3d; 38&#xa0;ms. <bold>(C)</bold> Image with frequency stability of 50&#xa0;Hz and <italic>ET</italic> &#x3d; 40&#xa0;ms.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g008.tif"/>
</fig>
</sec>
</sec>
<sec id="s3-2">
<title>3.2 Experiments</title>
<sec id="s3-2-1">
<title>3.2.1 NMR signal</title>
<p>
<xref ref-type="fig" rid="F9">Figure 9</xref> shows the NMR signal magnitudes and its Fourier Transform (FT). The FID signals magnitude acquired with 1, 2, and 4 scans are indistinguishable, independently of the magnetic field inhomogeneity, for a stability of 35&#xa0;ppm. This is consistent withequation 4, and shows that the signal magnitude is unaltered by the magnetic field instability. However, the increase of the magnetic field inhomogeneity leads to a decrease of the signal amplitude and length. On the other hand, the FT is sensitive to phase changes in the NMR signal, which means that it is not immune to magnetic field instability. The FT of the signals acquired with an inhomogeneity of 90&#xa0;ppm shows that there is a high variability between the spectra obtained after 1, 2, and 4 scans, even for a field stability of 35&#xa0;ppm. At 250&#xa0;ppm of inhomogeneity, the differences between the signals are smaller, while for 370&#xa0;ppm are practically indistinguishable. Consequently, this simple experiment evidences a direct relationship between the instability immunity and the magnetic field inhomogeneity in an FID experiment. Furthermore, it agrees with Eq. <xref ref-type="disp-formula" rid="e5">(5)</xref>, experimentally proving that the model is accurate to describe and understand the influence of magnetic field instability and inhomogeneity in NMR experiments.</p>
<fig id="F9" position="float">
<label>FIGURE 9</label>
<caption>
<p>NMR signal amplitudes (top) and its FT (bottom) acquired under 90&#xa0;ppm (left), 250&#xa0;ppm (middle) and 370&#xa0;ppm (right) with 1 scan (black line), 2 scans (red line), and 4 scans (blue line). The mean value of the field instability was 35&#xa0;ppm.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g009.tif"/>
</fig>
</sec>
<sec id="s3-2-2">
<title>3.2.2 2D MRI acquisition</title>
<p>
<xref ref-type="fig" rid="F10">Figure 10</xref> shows the image acquired in our MRI relaxometer. The artifacts increase drastically with the echo time suggesting that the dominant component of the instability is originated in random fluctuations.</p>
<fig id="F10" position="float">
<label>FIGURE 10</label>
<caption>
<p>Images acquired with different echo times. Image parameters: size 64 &#xd7; 64, <italic>G</italic>
<sub>
<italic>r</italic>
</sub> &#x3d; 54.9&#xa0;mT/m, <italic>G</italic>
<sub>
<italic>p</italic>
</sub> &#x3d; 71.3&#xa0;mT/m, 2 scans and 1 dummy.</p>
</caption>
<graphic xlink:href="fphy-11-1249771-g010.tif"/>
</fig>
</sec>
</sec>
</sec>
<sec sec-type="discussion" id="s4">
<title>4 Discussion</title>
<p>In this manuscript we have presented a simple model to describe the effects of the magnetic field inhomogeneity and instability in MRI. The model could successfully explain the observed behavior of FID signals acquired with different inhomogeneities in a commercial relaxometer. A key feature of the presented experiments and simulations is that a correlation between the magnetic field inhomogeneity and the signal immunity due to the instability can be established. However, this behavior does not replicate in MRI, where a more complex situation holds. The pulse gradients and the <italic>k</italic>-space sampling broke this correlation due to the creation of a second dimension.</p>
<p>It is worth noting that spins dephasing accumulation in both phase and readout encoding directions, <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>p-se</sub> and <italic>&#x3b3;B</italic>
<sub>0</sub>
<italic>I</italic>
<sub>r-se</sub>, are proportional to the magnetic field strength <italic>B</italic>
<sub>
<italic>0</italic>
</sub>. This feature suggests a higher natural immunity to magnetic field instability at low-field conditions, thus representing an additional motivation for pre-polarized MRI schemes. That is, magnetization boosted-up with a polarization <italic>B</italic>
<sub>
<italic>0</italic>
</sub> pulse, followed by an MRI sequence applied at a lower field. In any case the result holds valid for fixed-field MRI systems operating at low field conditions. On the other side, this result prevents the use of electromagnets for magnetic field strengths were, due to associated costs and technical difficulties, they would not be the best option anyway.</p>
<p>As it is possible to observe in <xref ref-type="fig" rid="F2">Figures 2</xref>, <xref ref-type="fig" rid="F3">3</xref>, simulated images for the GEs are highly degraded, while the images with SEs show a higher robustness to magnetic field instabilities. The main difference relies on the effect of the magnetic field inhomogeneity. The GEs generate artifacts along both encoding directions, leading to frequency shifts. Moreover, the dephasing induced by the inhomogeneity is accumulated all the time during GEs, while in the SEs, the magnetization is re-focused due to the presence of a &#x3c0; pulse [<xref ref-type="bibr" rid="B56">56</xref>]. Nonetheless, in the GEs the defects are also highlighted in the case of images degraded only by the stability. The GEs accumulate the dephasing of the low-frequency instabilities, an undesirable effect that can be avoided by using the SEs.</p>
<p>The results from <xref ref-type="fig" rid="F4">Figure 4</xref> are consistent with the simulations showed in <xref ref-type="fig" rid="F2">Figures 2</xref>, <xref ref-type="fig" rid="F3">3</xref>, highlighting the importance of the low-frequency re-focusing. Therefore, the spin-echo sequence is the recommended sequence to minimize instability-induced image distortions and ghosting.</p>
<p>Instabilities with different characteristics were described. The first one considers the instability as a random fluctuation, where the magnetic field between scans is less correlated than during a single scan. The second one approximates the instability as a monochromatic harmonic fluctuation with random phase between scans. Different strategies are to be considered to minimize the effects induced by these two kinds of instability. While for random instability the best strategy relies on minimizing the echo time, for a monochromatic instability, the best strategy consists of selecting an echo time in which a complete cycle of the instability is contained in <italic>ET</italic>/2. As expected, a realistic experimental situation may have both contributions. Therefore, a compromise relationship is presented, that is, the echo time that minimize random fluctuations should fit a proper value in order to cancel a dominant monochromatic component, if any.</p>
<p>Experimental images with different echo times were acquired in our MRI relaxometer, showing a strong correlation between the ET and the image degradation. This result shows that in the moment of the experiment, the instability of the magnetic field of the machine was mainly of random nature. A concluding remark from these results is that short echo times seem to be preferable in case of random magnetic field fluctuations.</p>
<p>In this work, we have shown that the study of the nature of the instability is a fundamental point to find the best strategy to minimize instability artifacts. Furthermore, if a pick-up coil or sensors are implemented to measure <inline-formula id="inf11">
<mml:math id="m20">
<mml:mrow>
<mml:mo>&#x394;</mml:mo>
<mml:mi>I</mml:mi>
<mml:mrow>
<mml:mfenced open="(" close=")" separators="|">
<mml:mrow>
<mml:mover accent="true">
<mml:mi>t</mml:mi>
<mml:mo>&#x223c;</mml:mo>
</mml:mover>
</mml:mrow>
</mml:mfenced>
</mml:mrow>
</mml:mrow>
</mml:math>
</inline-formula> during the MRI experiment, the phase shifts generated by the instability could be potentially corrected by a post-processing algorithm using the presented model. This approach would be similar to proposed methods aimed to minimize the noise in MRI images acquired without shielding [<xref ref-type="bibr" rid="B57">57</xref>, <xref ref-type="bibr" rid="B58">58</xref>]. Nonetheless, this strategy implies a complexity increase of both hardware and software, while it is not effective for image correction due to magnetic field instability. However, these days efficient post-processing methods can be successfully implemented at a negligible computational cost [<xref ref-type="bibr" rid="B37">37</xref>, <xref ref-type="bibr" rid="B49">49</xref>].</p>
</sec>
</body>
<back>
<sec sec-type="data-availability" id="s5">
<title>Data availability statement </title>
<p>The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.</p>
</sec>
<sec id="s6">
<title>Author contributions </title>
<p>GR performed the simulations, realized experimental work and wrote the manuscript. CS conceptualized the method and wrote the manuscript. EA conceptualized the method, realized experimental work and wrote the manuscript. All authors contributed to the article and approved the submitted version.</p>
</sec>
<sec id="s7">
<title>Funding </title>
<p>The author(s) declare financial support was received for the research, authorship, and/or publication of this article. This work was funded by FONCYT project number PICT-2017&#x2013;2195 and CONICET PIP 2021&#x2013;2013 11220200102521CO. Support from Secyt-UNC is also acknowledged.</p>
</sec>
<ack>
<p>GR acknowledges CONICET for the postdoctoral fellowship during this work.</p>
</ack>
<sec sec-type="COI-statement" id="s8">
<title>Conflict of interest </title>
<p>The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="disclaimer" id="s9">
<title>Publisher&#x2019;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<ref-list>
<title>References</title>
<ref id="B1">
<label>1.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Marques</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Simonis</surname>
<given-names>FFJ</given-names>
</name>
<name>
<surname>Webb</surname>
<given-names>AG</given-names>
</name>
</person-group>. <article-title>Low-field MRI: an MR physics perspective</article-title>. <source>J Magn Reson Imaging</source> (<year>2019</year>) <volume>49</volume>:<fpage>1528</fpage>&#x2013;<lpage>42</lpage>. <pub-id pub-id-type="doi">10.1002/jmri.26637</pub-id>
</citation>
</ref>
<ref id="B2">
<label>2.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sarracanie</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Salameh</surname>
<given-names>N</given-names>
</name>
</person-group>. <article-title>Low-field MRI: how low can we go? A fresh view on an old debate</article-title>. <source>Front Phys</source> (<year>2020</year>) <volume>8</volume>:<fpage>1</fpage>&#x2013;<lpage>14</lpage>. <pub-id pub-id-type="doi">10.3389/fphy.2020.00172</pub-id>
</citation>
</ref>
<ref id="B3">
<label>3.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Rodriguez</surname>
<given-names>GG</given-names>
</name>
</person-group>. <article-title>New challenges and opportunities for low-field MRI</article-title>. <source>J Magn Reson Open</source> (<year>2023</year>) <volume>14-15</volume>:<fpage>100086</fpage>. <pub-id pub-id-type="doi">10.1016/j.jmro.2022.100086</pub-id>
</citation>
</ref>
<ref id="B4">
<label>4.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>O&#x2019;Reilly</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Teeuwisse</surname>
<given-names>WM</given-names>
</name>
<name>
<surname>de Gans</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Koolstra</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Webb</surname>
<given-names>AG</given-names>
</name>
</person-group>. <article-title>
<italic>In vivo</italic> 3D brain and extremity MRI at 50 mT using a permanent magnet Halbach array</article-title>. <source>Magn Reson Med</source> (<year>2020</year>) <volume>85</volume>:<fpage>495</fpage>&#x2013;<lpage>505</lpage>. <pub-id pub-id-type="doi">10.1002/mrm.28396</pub-id>
</citation>
</ref>
<ref id="B5">
<label>5.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sarracanie</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Lapierre</surname>
<given-names>C</given-names>
</name>
<name>
<surname>Salameh</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Waddington</surname>
<given-names>DEJ</given-names>
</name>
<name>
<surname>Witzel</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Rosen</surname>
<given-names>MS</given-names>
</name>
</person-group>. <article-title>Low-cost high-performance MRI</article-title>. <source>Sci Rep</source> (<year>2015</year>) <volume>5</volume>:<fpage>15177</fpage>&#x2013;<lpage>9</lpage>. <pub-id pub-id-type="doi">10.1038/srep15177</pub-id>
</citation>
</ref>
<ref id="B6">
<label>6.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Nakagomi</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Kajiwara</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Matsuzaki</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Tanabe</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Hoshiai</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Okamoto</surname>
<given-names>Y</given-names>
</name>
<etal/>
</person-group> <article-title>Development of a small car-mounted magnetic resonance imaging system for human elbows using a 0.2 T permanent magnet</article-title>. <source>J Magn Reson</source> (<year>2019</year>) <volume>304</volume>:<fpage>1</fpage>&#x2013;<lpage>6</lpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2019.04.017</pub-id>
</citation>
</ref>
<ref id="B7">
<label>7.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Waddington</surname>
<given-names>DEJ</given-names>
</name>
<name>
<surname>Boele</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Maschmeyer</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Kuncic</surname>
<given-names>Z</given-names>
</name>
<name>
<surname>Rosen</surname>
<given-names>MS</given-names>
</name>
</person-group>. <article-title>High-sensitivity <italic>in vivo</italic> contrast for ultra-low field magnetic resonance imaging using superparamagnetic iron oxide nanoparticles</article-title>. <source>Sci Adv</source> (<year>2020</year>) <volume>6</volume>:<fpage>eabb0998</fpage>&#x2013;<lpage>10</lpage>. <pub-id pub-id-type="doi">10.1126/sciadv.abb0998</pub-id>
</citation>
</ref>
<ref id="B8">
<label>8.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Wald</surname>
<given-names>LL</given-names>
</name>
<name>
<surname>McDaniel</surname>
<given-names>PC</given-names>
</name>
<name>
<surname>Witzel</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Stockmann</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Cooley</surname>
<given-names>CZ</given-names>
</name>
</person-group>. <article-title>Low-cost and portable MRI</article-title>. <source>J Magn Reson Imaging</source> (<year>2019</year>) <volume>52</volume>:<fpage>686</fpage>&#x2013;<lpage>96</lpage>. <pub-id pub-id-type="doi">10.1002/jmri.26942</pub-id>
</citation>
</ref>
<ref id="B9">
<label>9.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Cooley</surname>
<given-names>CZ</given-names>
</name>
<name>
<surname>McDaniel</surname>
<given-names>PC</given-names>
</name>
<name>
<surname>Stockmann</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Srinivas</surname>
<given-names>SA</given-names>
</name>
<name>
<surname>Cauley</surname>
<given-names>SF</given-names>
</name>
<name>
<surname>&#x15a;liwiak</surname>
<given-names>M</given-names>
</name>
<etal/>
</person-group> <article-title>A portable scanner for magnetic resonance imaging of the brain</article-title>. <source>Nat Biomed Eng</source> (<year>2020</year>) <volume>5</volume>:<fpage>229</fpage>&#x2013;<lpage>39</lpage>. <pub-id pub-id-type="doi">10.1038/s41551-020-00641-5</pub-id>
</citation>
</ref>
<ref id="B10">
<label>10.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Guallart-Naval</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Algar&#xed;n</surname>
<given-names>JM</given-names>
</name>
<name>
<surname>Pellicer-Guridi</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Galve</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Vives-Gilabert</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Bosch</surname>
<given-names>R</given-names>
</name>
<etal/>
</person-group> <article-title>Portable magnetic resonance imaging of patients indoors, outdoors and at home</article-title>. <source>Sci Rep</source> (<year>2022</year>) <volume>12</volume>:<fpage>13147</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-022-17472-w</pub-id>
</citation>
</ref>
<ref id="B11">
<label>11.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Coffey</surname>
<given-names>AM</given-names>
</name>
<name>
<surname>Truong</surname>
<given-names>ML</given-names>
</name>
<name>
<surname>Chekmenev</surname>
<given-names>EY</given-names>
</name>
</person-group>. <article-title>Low-field MRI can be more sensitive than high-field MRI</article-title>. <source>J Magn Reson</source> (<year>2013</year>) <volume>237</volume>:<fpage>169</fpage>&#x2013;<lpage>74</lpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2013.10.013</pub-id>
</citation>
</ref>
<ref id="B12">
<label>12.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hoult</surname>
<given-names>DI</given-names>
</name>
<name>
<surname>Richards</surname>
<given-names>RE</given-names>
</name>
</person-group>. <article-title>The signal-to-noise ratio of the nuclear magnetic resonance experiment</article-title>. <source>J Magn Reson</source> (<year>1976</year>) <volume>24</volume>:<fpage>71</fpage>&#x2013;<lpage>85</lpage>. <pub-id pub-id-type="doi">10.1016/0022-2364(76)90233-X</pub-id>
</citation>
</ref>
<ref id="B13">
<label>13.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Broche</surname>
<given-names>LM</given-names>
</name>
<name>
<surname>Ross</surname>
<given-names>PJ</given-names>
</name>
<name>
<surname>Davies</surname>
<given-names>GR</given-names>
</name>
<name>
<surname>MacLeod</surname>
<given-names>MJ</given-names>
</name>
<name>
<surname>Lurie</surname>
<given-names>DJ</given-names>
</name>
</person-group>. <article-title>A whole-body Fast Field-Cycling scanner for clinical molecular imaging studies</article-title>. <source>Sci Rep</source> (<year>2019</year>) <volume>9</volume>:<fpage>10402</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-019-46648-0</pub-id>
</citation>
</ref>
<ref id="B14">
<label>14.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Edelstein</surname>
<given-names>WA</given-names>
</name>
<name>
<surname>Glover</surname>
<given-names>GH</given-names>
</name>
<name>
<surname>Hardy</surname>
<given-names>CJ</given-names>
</name>
<name>
<surname>Redington</surname>
<given-names>RW</given-names>
</name>
</person-group>. <article-title>The intrinsic signal-to-noise ratio in NMR imaging</article-title>. <source>Magn Reson Med</source> (<year>1986</year>) <volume>3</volume>:<fpage>604</fpage>&#x2013;<lpage>18</lpage>. <pub-id pub-id-type="doi">10.1002/mrm.1910030413</pub-id>
</citation>
</ref>
<ref id="B15">
<label>15.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Matter</surname>
<given-names>NI</given-names>
</name>
<name>
<surname>Scott</surname>
<given-names>GC</given-names>
</name>
<name>
<surname>Grafendorfer</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Macovski</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Conolly</surname>
<given-names>SM</given-names>
</name>
</person-group>. <article-title>Rapid polarizing field cycling in magnetic resonance imaging</article-title>. <source>IEEE Trans Med Imag</source> (<year>2006</year>) <volume>25</volume>:<fpage>84</fpage>&#x2013;<lpage>93</lpage>. <pub-id pub-id-type="doi">10.1109/tmi.2005.861014</pub-id>
</citation>
</ref>
<ref id="B16">
<label>16.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>McDermott</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Lee</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Haken</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Trabesinger</surname>
<given-names>AH</given-names>
</name>
<name>
<surname>Pines</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Clarke</surname>
<given-names>J</given-names>
</name>
</person-group>. <article-title>Microtesla MRI with a superconducting quantum interference device</article-title>. <source>PNAS</source> (<year>2004</year>) <volume>21</volume>:<fpage>7857</fpage>&#x2013;<lpage>61</lpage>. <pub-id pub-id-type="doi">10.1073/pnas.0402382101</pub-id>
</citation>
</ref>
<ref id="B17">
<label>17.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Dong</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Zhang</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Krause</surname>
<given-names>H-J</given-names>
</name>
<name>
<surname>Xie</surname>
<given-names>X</given-names>
</name>
<name>
<surname>Offenh&#xe4;usser</surname>
<given-names>A</given-names>
</name>
</person-group>. <article-title>Low field MRI detection with tuned HTS SQUID magnetometer</article-title>. <source>IEEE Trans Appl Supercon</source> (<year>2011</year>) <volume>21</volume>:<fpage>509</fpage>&#x2013;<lpage>13</lpage>. <pub-id pub-id-type="doi">10.1109/tasc.2010.2091713</pub-id>
</citation>
</ref>
<ref id="B18">
<label>18.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Demachi</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Hayashi</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Adachi</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Tanabe</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Tanaka</surname>
<given-names>S</given-names>
</name>
</person-group>. <article-title>T<sub>1</sub>-weighted image by ultra-low field SQUID-MRI</article-title>. <source>IEEE Trans Appl Supercon</source> (<year>2019</year>) <volume>29</volume>:<fpage>1</fpage>&#x2013;<lpage>5</lpage>. <pub-id pub-id-type="doi">10.1109/TASC.2019.2902772</pub-id>
</citation>
</ref>
<ref id="B19">
<label>19.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Suefke</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Liebisch</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Bl&#xfc;mich</surname>
<given-names>B</given-names>
</name>
</person-group>. <article-title>External high-quality-factor resonator tunes up nuclear magnetic resonance</article-title>. <source>Nat Phys</source> (<year>2015</year>) <volume>11</volume>:<fpage>767</fpage>&#x2013;<lpage>71</lpage>. <pub-id pub-id-type="doi">10.1038/nphys3382</pub-id>
</citation>
</ref>
<ref id="B20">
<label>20.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Jouda</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Kamberger</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Leupold</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Spengler</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Henning</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Gruschke</surname>
<given-names>O</given-names>
</name>
<etal/>
</person-group> <article-title>A comparison of Lenz lenses and LC resonators for NMR signal enhancement</article-title>. <source>Concepts Magn Reson B</source> (<year>2018</year>) <volume>47b</volume>:<fpage>e21357</fpage>. <pub-id pub-id-type="doi">10.1002/cmr.b.21357</pub-id>
</citation>
</ref>
<ref id="B21">
<label>21.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Zhang</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Guo</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Kong</surname>
<given-names>X</given-names>
</name>
<name>
<surname>Zeng</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Yin</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Wu</surname>
<given-names>J</given-names>
</name>
<etal/>
</person-group> <article-title>Improving local SNR of a single-channel 54.6 mT MRI system using additional LC-resonator</article-title>. <source>J Magn Reson</source> (<year>2022</year>) <volume>339</volume>:<fpage>107215</fpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2022.107215</pub-id>
</citation>
</ref>
<ref id="B22">
<label>22.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Sarracanie</surname>
<given-names>M</given-names>
</name>
</person-group>. <article-title>Fast quantitative low-field magnetic resonance imaging WithOPTIMUM&#x2014;optimized magnetic resonance fingerprinting using a stationary steady-state cartesian approach and accelerated acquisition schedules</article-title>. <source>Invest Radiol</source> (<year>2022</year>) <volume>57</volume>:<fpage>263</fpage>&#x2013;<lpage>71</lpage>. <pub-id pub-id-type="doi">10.1097/RLI.0000000000000836</pub-id>
</citation>
</ref>
<ref id="B23">
<label>23.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Lehmkuhl</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Fleischer</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Lohmann</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Rosen</surname>
<given-names>MS</given-names>
</name>
<name>
<surname>Chekmenev</surname>
<given-names>EY</given-names>
</name>
<name>
<surname>Adams</surname>
<given-names>A</given-names>
</name>
<etal/>
</person-group> <article-title>RASER MRI: magnetic resonance images formed spontaneously exploiting cooperative nonlinear interaction</article-title>. <source>Sci Adv</source> (<year>2022</year>) <volume>8</volume>:<fpage>eabp8483</fpage>. <pub-id pub-id-type="doi">10.1126/sciadv.abp8483</pub-id>
</citation>
</ref>
<ref id="B24">
<label>24.</label>
<citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname>Aja-Fern&#xe1;ndez</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Vegas-S&#xe1;nchez-Ferrero</surname>
<given-names>G</given-names>
</name>
</person-group>. <article-title>The problem of noise in MRI</article-title>. In: <source>Statistical analysis of noise in MRI</source>. <publisher-loc>Berlin, Germany</publisher-loc>: <publisher-name>Springer</publisher-name> (<year>2016</year>). <pub-id pub-id-type="doi">10.1007/978-3-319-39934-8_1</pub-id>
</citation>
</ref>
<ref id="B25">
<label>25.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Koonjoo</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Zhu</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Bagnall</surname>
<given-names>GC</given-names>
</name>
<name>
<surname>Bhutto</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Rosen</surname>
<given-names>MS</given-names>
</name>
</person-group>. <article-title>Boosting the signal-to-noise of low-field MRI with deep learning image reconstruction</article-title>. <source>Sci Rep</source> (<year>2021</year>) <volume>11</volume>:<fpage>8248</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-021-87482-7</pub-id>
</citation>
</ref>
<ref id="B26">
<label>26.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Araya</surname>
<given-names>YT</given-names>
</name>
<name>
<surname>Mart&#xed;nez-Santiesteban</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Handler</surname>
<given-names>WB</given-names>
</name>
<name>
<surname>Harris</surname>
<given-names>CT</given-names>
</name>
<name>
<surname>Chronik</surname>
<given-names>BA</given-names>
</name>
<name>
<surname>Scholl</surname>
<given-names>TJ</given-names>
</name>
</person-group>. <article-title>Nuclear magnetic relaxation dispersion of murine tissue for development of T<sub>1</sub> (R<sub>1</sub>) dispersion contrast imaging</article-title>. <source>NMR Biomed</source> (<year>2017</year>) <volume>30</volume>:<fpage>e3789</fpage>. <pub-id pub-id-type="doi">10.1002/nbm.3789</pub-id>
</citation>
</ref>
<ref id="B27">
<label>27.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Gossuin</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Gillis</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Hocq</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Vuong</surname>
<given-names>QL</given-names>
</name>
<name>
<surname>Roch</surname>
<given-names>A</given-names>
</name>
</person-group>. <article-title>Magnetic resonance relaxation properties of superparamagnetic particles</article-title>. <source>Wiley Interdiscip Rev Nanomedicine Nanobiotechnology</source> (<year>2009</year>) <volume>1</volume>:<fpage>299</fpage>&#x2013;<lpage>310</lpage>. <pub-id pub-id-type="doi">10.1002/wnan.36</pub-id>
</citation>
</ref>
<ref id="B28">
<label>28.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Yin</surname>
<given-names>X</given-names>
</name>
<name>
<surname>Russek</surname>
<given-names>SE</given-names>
</name>
<name>
<surname>Zabow</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Sun</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Mohapatra</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Keenan</surname>
<given-names>KE</given-names>
</name>
<etal/>
</person-group> <article-title>Large T<sub>1</sub> contrast enhancement using superparamagnetic nanoparticles in ultra-low field MRI</article-title>. <source>Sci Rep</source> (<year>2018</year>) <volume>8</volume>:<fpage>11863</fpage>. <pub-id pub-id-type="doi">10.1038/s41598-018-30264-5</pub-id>
</citation>
</ref>
<ref id="B29">
<label>29.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hendrick</surname>
<given-names>RE</given-names>
</name>
<name>
<surname>Haacke</surname>
<given-names>EM</given-names>
</name>
</person-group>. <article-title>Basic physics of MR contrast agents and maximization of image contrast</article-title>. <source>J Magn Reson Imaging</source> (<year>1993</year>) <volume>3</volume>:<fpage>137</fpage>&#x2013;<lpage>48</lpage>. <pub-id pub-id-type="doi">10.1002/jmri.1880030126</pub-id>
</citation>
</ref>
<ref id="B30">
<label>30.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rabbani</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Nezafat</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Gazor</surname>
<given-names>S</given-names>
</name>
</person-group>. <article-title>Wavelet-domain medical image denoising using bivariate Laplacian mixture model</article-title>. <source>IEEE Trans Biomed Eng</source> (<year>2009</year>) <volume>56</volume>:<fpage>2826</fpage>&#x2013;<lpage>37</lpage>. <pub-id pub-id-type="doi">10.1109/TBME.2009.2028876</pub-id>
</citation>
</ref>
<ref id="B31">
<label>31.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ippoliti</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Adams</surname>
<given-names>LC</given-names>
</name>
<name>
<surname>Winfried</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Hamm</surname>
<given-names>B</given-names>
</name>
<name>
<surname>Spincemaille</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>Y</given-names>
</name>
<etal/>
</person-group> <article-title>Quantitative susceptibility mapping across two clinical field strengths:contrast-to-noise ratio enhancement at 1.5T</article-title>. <source>J Magn Reson Imaging</source> (<year>2018</year>) <volume>48</volume>:<fpage>1410</fpage>&#x2013;<lpage>20</lpage>. <pub-id pub-id-type="doi">10.1002/jmri.26045</pub-id>
</citation>
</ref>
<ref id="B32">
<label>32.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Epstein</surname>
<given-names>CL</given-names>
</name>
</person-group>. <article-title>Magnetic resonance imaging in inhomogeneous fields</article-title>. <source>Inverse Probl</source> (<year>2004</year>) <volume>20</volume>:<fpage>753</fpage>&#x2013;<lpage>80</lpage>. <pub-id pub-id-type="doi">10.1088/0266-5611/20/3/007</pub-id>
</citation>
</ref>
<ref id="B33">
<label>33.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kim</surname>
<given-names>JK</given-names>
</name>
<name>
<surname>Plewes</surname>
<given-names>DB</given-names>
</name>
<name>
<surname>Henkelman</surname>
<given-names>RM</given-names>
</name>
</person-group>. <article-title>Magnetic resonance imaging in an inhomogeneous magnetic field</article-title>. <source>US Patent</source> (<year>1994</year>):<fpage>5309101</fpage>.</citation>
</ref>
<ref id="B34">
<label>34.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Perlo</surname>
<given-names>J</given-names>
</name>
<name>
<surname>Casanova</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Bl&#xfc;mich</surname>
<given-names>B</given-names>
</name>
</person-group>. <article-title>3D imaging with a single-sided sensor: an open tomograph</article-title>. <source>J Magn Reson</source> (<year>2004</year>) <volume>166</volume>:<fpage>228</fpage>&#x2013;<lpage>35</lpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2003.10.018</pub-id>
</citation>
</ref>
<ref id="B35">
<label>35.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Romero</surname>
<given-names>JA</given-names>
</name>
<name>
<surname>Rodriguez</surname>
<given-names>GG</given-names>
</name>
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
</person-group>. <article-title>A fast field-cycling MRI relaxometer for physical contrasts design and pre-clinical studies in small animals</article-title>. <source>J Magn Reson</source> (<year>2020</year>) <volume>311</volume>:<fpage>106682</fpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2019.106682</pub-id>
</citation>
</ref>
<ref id="B36">
<label>36.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Vovk</surname>
<given-names>U</given-names>
</name>
<name>
<surname>Pernu&#x161;</surname>
<given-names>F</given-names>
</name>
<name>
<surname>Likar</surname>
<given-names>B</given-names>
</name>
</person-group>. <article-title>A review of methods for correction of intensity inhomogeneity in MRI</article-title>. <source>IEEE Trans Med Imaging</source> (<year>2007</year>) <volume>26</volume>:<fpage>405</fpage>&#x2013;<lpage>21</lpage>. <pub-id pub-id-type="doi">10.1109/TMI.2006.891486</pub-id>
</citation>
</ref>
<ref id="B37">
<label>37.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rodriguez</surname>
<given-names>GG</given-names>
</name>
<name>
<surname>Salvatori</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
</person-group>. <article-title>Dual k-space and image-space post-processing for field-cycling MRI under low magnetic field stability and homogeneity conditions</article-title>. <source>Magn Reson Imag</source> (<year>2022</year>) <volume>87</volume>:<fpage>157</fpage>&#x2013;<lpage>68</lpage>. <pub-id pub-id-type="doi">10.1016/j.mri.2022.01.008</pub-id>
</citation>
</ref>
<ref id="B38">
<label>38.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Koolstra</surname>
<given-names>K</given-names>
</name>
<name>
<surname>O&#x2019;Reilly</surname>
<given-names>T</given-names>
</name>
<name>
<surname>B&#xf6;rnert</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Webb</surname>
<given-names>A</given-names>
</name>
</person-group>. <article-title>Image distortion correction for MRI in low field permanent magnet systems with strong B<sub>0</sub> inhomogeneity and gradient field nonlinearities</article-title>. <source>MAGMA</source> (<year>2021</year>) <volume>34</volume>:<fpage>631</fpage>&#x2013;<lpage>42</lpage>. <pub-id pub-id-type="doi">10.1007/s10334-021-00907-2</pub-id>
</citation>
</ref>
<ref id="B39">
<label>39.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Lother</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Schiff</surname>
<given-names>SJ</given-names>
</name>
<name>
<surname>Neuberger</surname>
<given-names>T</given-names>
</name>
<name>
<surname>Jakob</surname>
<given-names>PM</given-names>
</name>
<name>
<surname>Fidler</surname>
<given-names>F</given-names>
</name>
</person-group>. <article-title>Design of a mobile, homogeneous, and efficient electromagnet with a large field of view for neonatal low-field MRI</article-title>. <source>Magn Reson Mater Phys Biol. Med.</source> (<year>2016</year>) <volume>29</volume>:<fpage>691</fpage>&#x2013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.1007/s10334-016-0525-8</pub-id>
</citation>
</ref>
<ref id="B40">
<label>40.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Michal</surname>
<given-names>CA</given-names>
</name>
</person-group>. <article-title>Low-cost low-field NMR and MRI: instrumentation and applications</article-title>. <source>J Magn Reson</source> (<year>2020</year>) <volume>319</volume>:<fpage>106800</fpage>. <pub-id pub-id-type="doi">10.1016/j.jmr.2020.106800</pub-id>
</citation>
</ref>
<ref id="B41">
<label>41.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Kimmich</surname>
<given-names>R</given-names>
</name>
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
</person-group>. <article-title>Field-cycling NMR relaxometry</article-title>. <source>Prog NMR Sepctrosc</source> (<year>2004</year>) <volume>44</volume>:<fpage>257</fpage>&#x2013;<lpage>320</lpage>. <pub-id pub-id-type="doi">10.1016/j.pnmrs.2004.03.002</pub-id>
</citation>
</ref>
<ref id="B42">
<label>42.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
<name>
<surname>Galli</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Ferrante</surname>
<given-names>G</given-names>
</name>
</person-group>. <article-title>Fast-field-cycling NMR: applications and instrumentation</article-title>. <source>Appl Magn Reson</source> (<year>2001</year>) <volume>20</volume>:<fpage>365</fpage>&#x2013;<lpage>404</lpage>. <pub-id pub-id-type="doi">10.1007/BF03162287</pub-id>
</citation>
</ref>
<ref id="B43">
<label>43.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Lurie</surname>
<given-names>DJ</given-names>
</name>
<name>
<surname>Hutchison</surname>
<given-names>JMS</given-names>
</name>
<name>
<surname>Bell</surname>
<given-names>LH</given-names>
</name>
<name>
<surname>Nicholson</surname>
<given-names>I</given-names>
</name>
<name>
<surname>Bussell</surname>
<given-names>DM</given-names>
</name>
<name>
<surname>Mallard</surname>
<given-names>JR</given-names>
</name>
</person-group>. <article-title>Field-cycled proton-electron double-resonance imaging of free radicals in large aqueous samples</article-title>. <source>J Magn Reson</source> (<year>1989</year>) <volume>84</volume>:<fpage>431</fpage>&#x2013;<lpage>7</lpage>. <pub-id pub-id-type="doi">10.1016/0022-2364(89)90392-2</pub-id>
</citation>
</ref>
<ref id="B44">
<label>44.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ungersma</surname>
<given-names>SE</given-names>
</name>
<name>
<surname>Matter</surname>
<given-names>NI</given-names>
</name>
<name>
<surname>Hardy</surname>
<given-names>JW</given-names>
</name>
<name>
<surname>Venook</surname>
<given-names>RD</given-names>
</name>
<name>
<surname>Macovski</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Conolly</surname>
<given-names>SM</given-names>
</name>
<etal/>
</person-group> <article-title>Magnetic resonance imaging with T<sub>1</sub> dispersion contrast</article-title>. <source>Magn Reson Med</source> (<year>2006</year>) <volume>55</volume>:<fpage>1362</fpage>&#x2013;<lpage>71</lpage>. <pub-id pub-id-type="doi">10.1002/mrm.20910</pub-id>
</citation>
</ref>
<ref id="B45">
<label>45.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rodriguez</surname>
<given-names>GG</given-names>
</name>
<name>
<surname>Erro</surname>
<given-names>EM</given-names>
</name>
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
</person-group>. <article-title>Fast iron oxide-induced low-field magnetic resonance imaging</article-title>. <source>J Phys D: Appl Phys</source> (<year>2020</year>) <volume>54</volume>:<fpage>025003</fpage>. <pub-id pub-id-type="doi">10.1088/1361-6463/abbe4d</pub-id>
</citation>
</ref>
<ref id="B46">
<label>46.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Rodriguez</surname>
<given-names>GG</given-names>
</name>
<name>
<surname>Anoardo</surname>
<given-names>E</given-names>
</name>
</person-group>. <article-title>Proton double irradiation field-cycling nuclear magnetic resonance imaging: testing new concepts and calibration methods</article-title>. <source>IEEE Trans Instrum Meas</source> (<year>2021</year>) <volume>70</volume>:<fpage>1</fpage>&#x2013;<lpage>8</lpage>. <pub-id pub-id-type="doi">10.1109/TIM.2020.3037304</pub-id>
</citation>
</ref>
<ref id="B47">
<label>47.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Morgan</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Conolly</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Scott</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Macovski</surname>
<given-names>A</given-names>
</name>
</person-group>. <article-title>A readout magnet for prepolarized MRI</article-title>. <source>Magn Reson Med</source> (<year>1996</year>) <volume>36</volume>:<fpage>527</fpage>&#x2013;<lpage>36</lpage>. <pub-id pub-id-type="doi">10.1002/mrm.1910360405</pub-id>
</citation>
</ref>
<ref id="B48">
<label>48.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>B&#xf6;denler</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Rochefort</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Ross</surname>
<given-names>PJ</given-names>
</name>
<name>
<surname>Chanet</surname>
<given-names>N</given-names>
</name>
<name>
<surname>Guillot</surname>
<given-names>G</given-names>
</name>
<name>
<surname>Davies</surname>
<given-names>GR</given-names>
</name>
<etal/>
</person-group> <article-title>Comparison of fast field-cycling magnetic resonance imaging methods and future perspectives</article-title>. <source>Mol Phys</source> (<year>2019</year>) <volume>117</volume>:<fpage>832</fpage>&#x2013;<lpage>48</lpage>. <pub-id pub-id-type="doi">10.1080/00268976.2018.1557349</pub-id>
</citation>
</ref>
<ref id="B49">
<label>49.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Broche</surname>
<given-names>LM</given-names>
</name>
<name>
<surname>Ross</surname>
<given-names>PJ</given-names>
</name>
<name>
<surname>Davies</surname>
<given-names>GR</given-names>
</name>
<name>
<surname>Lurie</surname>
<given-names>DJ</given-names>
</name>
</person-group>. <article-title>Simple algorithm for the correction of MRI image artefacts due to random phase fluctuations</article-title>. <source>Magn Reson Imaging</source> (<year>2017</year>) <volume>44</volume>:<fpage>55</fpage>&#x2013;<lpage>9</lpage>. <pub-id pub-id-type="doi">10.1016/j.mri.2017.07.023</pub-id>
</citation>
</ref>
<ref id="B50">
<label>50.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Atkinson</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Hill</surname>
<given-names>DLG</given-names>
</name>
<name>
<surname>Stoyle</surname>
<given-names>PNR</given-names>
</name>
<name>
<surname>Summers</surname>
<given-names>PE</given-names>
</name>
<name>
<surname>Keevil</surname>
<given-names>SF</given-names>
</name>
</person-group>. <article-title>Automatic correction of motion artifacts in magnetic resonance images using an entropy focus criterion</article-title>. <source>IEEE Trans Med Imaging</source> (<year>1997</year>) <volume>16</volume>:<fpage>903</fpage>&#x2013;<lpage>10</lpage>. <pub-id pub-id-type="doi">10.1109/42.650886</pub-id>
</citation>
</ref>
<ref id="B51">
<label>51.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Atkinson</surname>
<given-names>D</given-names>
</name>
<name>
<surname>Hill</surname>
<given-names>DLG</given-names>
</name>
<name>
<surname>Stoyle</surname>
<given-names>PNR</given-names>
</name>
<name>
<surname>Summers</surname>
<given-names>PE</given-names>
</name>
<name>
<surname>Clare</surname>
<given-names>S</given-names>
</name>
<name>
<surname>Bowtell</surname>
<given-names>R</given-names>
</name>
<etal/>
</person-group> <article-title>Automatic compensation of motion artifacts in MRI</article-title>. <source>Magn Reson Med</source> (<year>1999</year>) <volume>41</volume>:<fpage>163</fpage>&#x2013;<lpage>70</lpage>. <pub-id pub-id-type="doi">10.1002/(SICI)1522-2594(199901)41:1&#x3c;163::AID-MRM23&#x3e;3.0.CO;2-9</pub-id>
</citation>
</ref>
<ref id="B52">
<label>52.</label>
<citation citation-type="book">
<person-group person-group-type="author">
<name>
<surname>Bernstein</surname>
<given-names>MA</given-names>
</name>
<name>
<surname>King</surname>
<given-names>KF</given-names>
</name>
<name>
<surname>Joe Zhou</surname>
<given-names>X</given-names>
</name>
</person-group>. <source>Handbook of MRI pulse sequences</source>. <publisher-loc>Amsterdam</publisher-loc>: <publisher-name>Elsevier Academic Press</publisher-name> (<year>2007</year>). <pub-id pub-id-type="doi">10.4314/bcse.v21i3.21211</pub-id>
</citation>
</ref>
<ref id="B53">
<label>53.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Javili</surname>
<given-names>A</given-names>
</name>
<name>
<surname>McBride</surname>
<given-names>A</given-names>
</name>
<name>
<surname>Steinmann</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Reddy</surname>
<given-names>BD</given-names>
</name>
</person-group>. <article-title>A unified computational framework for bulk and surface elasticity theory: a curvilinear-coordinate-based finite element methodology</article-title>. <source>Comput Mech</source> (<year>2014</year>) <volume>54</volume>:<fpage>745</fpage>&#x2013;<lpage>62</lpage>. <pub-id pub-id-type="doi">10.1007/s00466-014-1030-4</pub-id>
</citation>
</ref>
<ref id="B54">
<label>54.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Ye</surname>
<given-names>M</given-names>
</name>
<name>
<surname>Feng</surname>
<given-names>P</given-names>
</name>
<name>
<surname>Li</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Wang</surname>
<given-names>D</given-names>
</name>
<name>
<surname>He</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Cui</surname>
<given-names>W</given-names>
</name>
</person-group>. <article-title>The total secondary electron yield of a conductive random rough surface</article-title>. <source>J Appl Phys</source> (<year>2019</year>) <volume>125</volume>:<fpage>043301</fpage>. <pub-id pub-id-type="doi">10.1063/1.5023769</pub-id>
</citation>
</ref>
<ref id="B55">
<label>55.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Takeda</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Maeda</surname>
<given-names>H</given-names>
</name>
<name>
<surname>Ohki</surname>
<given-names>K</given-names>
</name>
<name>
<surname>Yanagisawa</surname>
<given-names>Y</given-names>
</name>
</person-group>. <article-title>Review of the temporal stability of themagnetic field for ultra-high field superconducting magnets with aparticular focus on superconducting joints between HTS conductors</article-title>. <source>Supercond Sci Technol</source> (<year>2022</year>) <volume>35</volume>:<fpage>043002</fpage>. <pub-id pub-id-type="doi">10.1088/1361-6668/ac5645</pub-id>
</citation>
</ref>
<ref id="B56">
<label>56.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Hahn</surname>
<given-names>EL</given-names>
</name>
</person-group>. <article-title>Spin echoes</article-title>. <source>Phys Rev</source> (<year>1950</year>) <volume>80</volume>:<fpage>580</fpage>&#x2013;<lpage>94</lpage>. <pub-id pub-id-type="doi">10.1103/PhysRev.80.580</pub-id>
</citation>
</ref>
<ref id="B57">
<label>57.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Liu</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Leon</surname>
<given-names>ATL</given-names>
</name>
<name>
<surname>Zhao</surname>
<given-names>Y</given-names>
</name>
<name>
<surname>Xiao</surname>
<given-names>L</given-names>
</name>
<name>
<surname>Mak</surname>
<given-names>HKF</given-names>
</name>
<name>
<surname>Tsang</surname>
<given-names>ACO</given-names>
</name>
<etal/>
</person-group> <article-title>A low-cost and shielding-free ultra-low-field brain MRI scanner</article-title>. <source>Nat Commun</source> (<year>2021</year>) <volume>12</volume>:<fpage>7238</fpage>. <pub-id pub-id-type="doi">10.1038/s41467-021-27317-1</pub-id>
</citation>
</ref>
<ref id="B58">
<label>58.</label>
<citation citation-type="journal">
<person-group person-group-type="author">
<name>
<surname>Srinivas</surname>
<given-names>SA</given-names>
</name>
<name>
<surname>Cauley</surname>
<given-names>SF</given-names>
</name>
<name>
<surname>Stockmann</surname>
<given-names>JP</given-names>
</name>
<name>
<surname>Sappo</surname>
<given-names>CR</given-names>
</name>
<name>
<surname>Vaughn</surname>
<given-names>CE</given-names>
</name>
<name>
<surname>Wald</surname>
<given-names>LL</given-names>
</name>
<etal/>
</person-group> <article-title>External dynamic interference estimation and removal (EDITER) for low field MRI</article-title>. <source>Magn Reson Med</source> (<year>2021</year>) <volume>87</volume>:<fpage>614</fpage>&#x2013;<lpage>28</lpage>. <pub-id pub-id-type="doi">10.1002/mrm.28992</pub-id>
</citation>
</ref>
</ref-list>
</back>
</article>