<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.3 20210610//EN" "JATS-journalpublishing1-3-mathml3.dtd">
<article xml:lang="EN" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:ali="http://www.niso.org/schemas/ali/1.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" dtd-version="1.3" article-type="research-article">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Med.</journal-id>
<journal-title-group>
<journal-title>Frontiers in Medicine</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Med.</abbrev-journal-title>
</journal-title-group>
<issn pub-type="epub">2296-858X</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3389/fmed.2025.1745447</article-id>
<article-version article-version-type="Version of Record" vocab="NISO-RP-8-2008"/>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Research</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>In-house chromogenic anti-factor Xa assay: development, validation, and identification of factors predicting APTT discordance</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author">
<name><surname>Sun</surname> <given-names>Lin</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Data curation" vocab-term-identifier="https://credit.niso.org/contributor-roles/data-curation/">Data curation</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; original draft" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-original-draft/">Writing &#x2013; original draft</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
<uri xlink:href="https://loop.frontiersin.org/people/3277157"/>
</contrib>
<contrib contrib-type="author">
<name><surname>Zhao</surname> <given-names>Jinxia</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Validation" vocab-term-identifier="https://credit.niso.org/contributor-roles/validation/">Validation</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
</contrib>
<contrib contrib-type="author">
<name><surname>Jiang</surname> <given-names>Guohong</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
</contrib>
<contrib contrib-type="author">
<name><surname>Lu</surname> <given-names>Wenfei</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Validation" vocab-term-identifier="https://credit.niso.org/contributor-roles/validation/">Validation</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
</contrib>
<contrib contrib-type="author">
<name><surname>Rong</surname> <given-names>Hui</given-names></name>
<xref ref-type="aff" rid="aff3"><sup>3</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Funding acquisition" vocab-term-identifier="https://credit.niso.org/contributor-roles/funding-acquisition/">Funding acquisition</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
</contrib>
<contrib contrib-type="author" corresp="yes">
<name><surname>Lun</surname> <given-names>Limin</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="corresp" rid="c001"><sup>&#x0002A;</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Conceptualization" vocab-term-identifier="https://credit.niso.org/contributor-roles/conceptualization/">Conceptualization</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Data curation" vocab-term-identifier="https://credit.niso.org/contributor-roles/data-curation/">Data curation</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Methodology" vocab-term-identifier="https://credit.niso.org/contributor-roles/methodology/">Methodology</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Supervision" vocab-term-identifier="https://credit.niso.org/contributor-roles/supervision/">Supervision</role>
</contrib>
<contrib contrib-type="author">
<name><surname>Wang</surname> <given-names>Yuehan</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Investigation" vocab-term-identifier="https://credit.niso.org/contributor-roles/investigation/">Investigation</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
</contrib>
<contrib contrib-type="author">
<name><surname>Zhou</surname> <given-names>Tianhui</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &amp; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x00026; editing</role>
<role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Investigation" vocab-term-identifier="https://credit.niso.org/contributor-roles/investigation/">Investigation</role>
</contrib>
</contrib-group>
<aff id="aff1"><label>1</label><institution>Department of Laboratory Medicine, The Affiliated Hospital of Qingdao University</institution>, <city>Qingdao</city>, <country country="cn">China</country></aff>
<aff id="aff2"><label>2</label><institution>Laboratory Department, Qingdao Public Health Clinical Center</institution>, <city>Qingdao</city>, <country country="cn">China</country></aff>
<aff id="aff3"><label>3</label><institution>Beijing ZONCI Technology Development Co., Ltd.</institution>, <city>Beijing</city>, <country country="cn">China</country></aff>
<author-notes>
<corresp id="c001"><label>&#x0002A;</label>Correspondence: Limin Lun, <email xlink:href="mailto:lunlmqd@163.com">lunlmqd@163.com</email></corresp>
</author-notes>
<pub-date publication-format="electronic" date-type="pub" iso-8601-date="2026-01-12">
<day>12</day>
<month>01</month>
<year>2026</year>
</pub-date>
<pub-date publication-format="electronic" date-type="collection">
<year>2025</year>
</pub-date>
<volume>12</volume>
<elocation-id>1745447</elocation-id>
<history>
<date date-type="received">
<day>13</day>
<month>11</month>
<year>2025</year>
</date>
<date date-type="rev-recd">
<day>17</day>
<month>12</month>
<year>2025</year>
</date>
<date date-type="accepted">
<day>22</day>
<month>12</month>
<year>2025</year>
</date>
</history>
<permissions>
<copyright-statement>Copyright &#x000A9; 2026 Sun, Zhao, Jiang, Lu, Rong, Lun, Wang and Zhou.</copyright-statement>
<copyright-year>2026</copyright-year>
<copyright-holder>Sun, Zhao, Jiang, Lu, Rong, Lun, Wang and Zhou</copyright-holder>
<license>
<ali:license_ref start_date="2026-01-12">https://creativecommons.org/licenses/by/4.0/</ali:license_ref>
<license-p>This is an open-access article distributed under the terms of the <ext-link ext-link-type="uri" xlink:href="https://creativecommons.org/licenses/by/4.0/">Creative Commons Attribution License (CC BY)</ext-link>. The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</license-p>
</license>
</permissions>
<abstract>
<sec>
<title>Background</title>
<p>Activated partial thromboplastin time (APTT) is the conventional test for monitoring unfractionated heparin (UFH) therapy, but discordance with anti-factor Xa results frequently occurs in clinical practice, potentially leading to suboptimal anticoagulation management.</p></sec>
<sec>
<title>Objective</title>
<p>This study aimed to develop an in-house anti-factor Xa detection system, establish laboratory-specific APTT therapeutic ranges calibrated against anti-factor Xa levels, and identify predictors of discordance between these two monitoring methods.</p></sec>
<sec>
<title>Methods</title>
<p>We developed and validated a chromogenic anti-factor Xa assay according to CLSI guidelines. From 110 patients receiving UFH, 190 samples were collected. Laboratory-specific APTT ranges were derived using linear regression and ROC analysis against anti-factor Xa target range (0.3&#x02013;0.7 IU/ml). Random forest and logistic regression models identified predictors of discordance among clinical and laboratory variables.</p></sec>
<sec>
<title>Results</title>
<p>The optimized assay showed excellent analytical performance with strong correlation to the commercial platform (Pearson <italic>r</italic> = 0.986, <italic>R</italic><sup>2</sup> = 0.972). The derived laboratory-specific APTT therapeutic range was 79&#x02013;127 s. Despite moderate correlation between APTT and anti-factor Xa (Spearman &#x003C1; = 0.678, <italic>p</italic> &#x0003C; 0.001), classification concordance was poor, with 58.2% of specimens showing discordance. Random forest analysis (AUC = 0.79) and logistic regression identified five key predictors: fibrinogen, triglycerides, estimated glomerular filtration rate, C-reactive protein, and antithrombin III.</p></sec>
<sec>
<title>Conclusions</title>
<p>This study developed a validated anti-factor Xa assay and identified substantial classification discordance between APTT and anti-factor Xa monitoring. Five clinical and laboratory factors were associated with monitoring discordance.</p></sec></abstract>
<kwd-group>
<kwd>activated partial thromboplastin time</kwd>
<kwd>anticoagulation monitoring</kwd>
<kwd>classification discordance</kwd>
<kwd>predictive modeling</kwd>
<kwd>unfractionated heparin</kwd>
</kwd-group>
<funding-group>
 <funding-statement>The author(s) declared that financial support was received for this work and/or its publication. This study was funded by ZONCI Technology Development Co., Ltd., Beijing, China.</funding-statement>
</funding-group>
<counts>
<fig-count count="14"/>
<table-count count="2"/>
<equation-count count="0"/>
<ref-count count="51"/>
<page-count count="15"/>
<word-count count="8472"/>
</counts>
<custom-meta-group>
<custom-meta>
<meta-name>section-at-acceptance</meta-name>
<meta-value>Hematology</meta-value>
</custom-meta>
</custom-meta-group>
</article-meta>
</front>
<body>
<sec sec-type="intro" id="s1">
<label>1</label>
<title>Introduction</title>
<p>Unfractionated heparin (UFH) is widely used for anticoagulation in acute coronary syndromes, venous thromboembolism, cardiac surgery, and hemodialysis. UFH binds to antithrombin III (AT-III), thereby accelerating the inactivation of thrombin and factor Xa (<xref ref-type="bibr" rid="B1">1</xref>). However, the anticoagulant response to UFH demonstrates considerable inter-individual and intra-individual variability, influenced by multiple factors including plasma protein binding, renal clearance, inflammatory status, and antithrombin levels (<xref ref-type="bibr" rid="B2">2</xref>&#x02013;<xref ref-type="bibr" rid="B4">4</xref>), necessitating therapeutic drug monitoring (<xref ref-type="bibr" rid="B5">5</xref>). However, as a global coagulation screening test, APTT is susceptible to interference from numerous pre-analytical and analytical variables. Endogenous coagulation factor levels (particularly factors VIII and XII), fibrinogen concentration, lupus anticoagulants, and variable heparin sensitivity of different APTT reagents can all affect test results (<xref ref-type="bibr" rid="B6">6</xref>&#x02013;<xref ref-type="bibr" rid="B9">9</xref>). These limitations compromise inter-laboratory comparability and may lead to subtherapeutic or supratherapeutic anticoagulation states, increasing the risk of thrombotic recurrence or hemorrhagic complications (<xref ref-type="bibr" rid="B10">10</xref>).</p>
<p>Chromogenic substrate-based anti-factor Xa assays directly quantify the inhibitory activity of the heparin-antithrombin complex against factor Xa, offering improved specificity compared to APTT (<xref ref-type="bibr" rid="B11">11</xref>, <xref ref-type="bibr" rid="B12">12</xref>). Chromogenic anti-factor Xa assays measure heparin activity by incubating patient plasma with excess factor Xa; the heparin-antithrombin complex neutralizes factor Xa, and residual enzyme activity is quantified via chromogenic substrate cleavage, providing direct measurement of anticoagulant effect. However, the high cost of commercially available anti-factor Xa platforms has limited their widespread adoption in many clinical laboratories (<xref ref-type="bibr" rid="B13">13</xref>&#x02013;<xref ref-type="bibr" rid="B15">15</xref>), creating a need for more accessible testing options.</p>
<p>Discordance in anticoagulation status classification (subtherapeutic, therapeutic, or supratherapeutic) between APTT and anti-factor Xa results is commonly observed, with reported inconsistency rates of 30%&#x02212;50% (<xref ref-type="bibr" rid="B16">16</xref>, <xref ref-type="bibr" rid="B17">17</xref>). While several studies identified potential contributing factors&#x02014;antithrombin deficiency, elevated factor VIII, acute phase reactions, renal dysfunction (<xref ref-type="bibr" rid="B18">18</xref>&#x02013;<xref ref-type="bibr" rid="B20">20</xref>)&#x02014;a comprehensive predictive model integrating multiple clinical and laboratory variables has not been established.</p>
<p>Traditional statistical approaches may not fully capture the complex, nonlinear interactions among multiple predictors of assay discordance. Machine learning algorithms, particularly random forest (RF) models, can handle high-dimensional data, identify nonlinear relationships, and quantify variable importance without assuming parametric distributions (<xref ref-type="bibr" rid="B14">14</xref>, <xref ref-type="bibr" rid="B21">21</xref>). Recent applications of RF in coagulation research have demonstrated its utility in predicting bleeding risk and thrombotic complications. Recent applications of machine learning in coagulation research have demonstrated utility in predicting bleeding risk (<xref ref-type="bibr" rid="B22">22</xref>), venous thromboembolism in cancer patients (<xref ref-type="bibr" rid="B23">23</xref>), and optimizing anticoagulant dosing (<xref ref-type="bibr" rid="B24">24</xref>). However, to our knowledge, no study has applied machine learning methods to identify predictors of APTT-anti-factor Xa discordance in UFH monitoring.</p>
<p>This study has three primary objectives: first, to develop an in-house chromogenic anti-factor Xa assay. Second, to establish a laboratory-specific APTT therapeutic range calibrated against the anti-factor Xa target (0.3&#x02013;0.7 IU/ml) (<xref ref-type="bibr" rid="B25">25</xref>, <xref ref-type="bibr" rid="B26">26</xref>). Third, to identify predictors of APTT-anti-factor Xa discordance using random forest and logistic regression approaches.</p></sec>
<sec sec-type="materials and methods" id="s2">
<label>2</label>
<title>Materials and methods</title>
<sec>
<label>2.1</label>
<title>Study design and sample collection</title>
<sec>
<label>2.1.1</label>
<title>Ethical approval</title>
<p>This study was approved by the Ethics Committee of the Affiliated Hospital of Qingdao University (Approval No. QYFY WZLL 30726). Written informed consent was obtained from all participants. All data were anonymized per Declaration of Helsinki principles.</p></sec>
<sec>
<label>2.1.2</label>
<title>Study population</title>
<p>Between March 2024 and September 2024, 190 blood samples were prospectively collected from 110 consecutively enrolled hospitalized patients receiving continuous intravenous UFH at the Affiliated Hospital of Qingdao University. UFH indications included acute coronary syndrome (<italic>n</italic> = 38, 34.5%), VTE treatment/prophylaxis (<italic>n</italic> = 24, 21.8%), post-cardiac surgery (<italic>n</italic> = 15, 13.6%), hemodialysis/CRRT (<italic>n</italic> = 8, 7.3%), and other indications (<italic>n</italic> = 25, 22.7%).</p>
<p>Inclusion criteria: age &#x02265;18 years; receiving intravenous UFH &#x02265;24 h; both APTT and anti-factor Xa testing conducted within 4 h of blood collection.</p>
<p>Exclusion criteria: concurrent treatment with direct oral anticoagulants, low molecular weight heparin, or vitamin K antagonists; severe liver dysfunction (Child-Pugh class C); known inherited coagulation disorders; hemolyzed, icteric, or severely lipemic specimens (triglycerides &#x0003E;11.3 mmol/L); incomplete laboratory data.</p>
<p>Blood sampling was conducted 3&#x02013;6 h after UFH bolus or during steady-state infusion (continuous intravenous UFH administration &#x02265;24 h at constant rate; samples collected &#x02265;6 h after last dose adjustment) (<xref ref-type="bibr" rid="B5">5</xref>, <xref ref-type="bibr" rid="B26">26</xref>).</p></sec>
<sec>
<label>2.1.3</label>
<title>Blood collection and processing</title>
<p>We collected blood via clean venipuncture into 3.2% trisodium citrate tubes (9:1 blood:anticoagulant ratio). We obtained Platelet-poor plasma (PPP) by centrifugation at 2,000 &#x000D7; g for 15 min at 15&#x02013;20 &#x000B0;C. PPP was immediately aliquoted and stored at &#x02212;80 &#x000B0;C.</p></sec>
<sec>
<label>2.1.4</label>
<title>Normal pooled plasma preparation</title>
<p>We prepared Normal pooled plasma (NPP) from 280 healthy volunteers with normal coagulation parameters. Individual PPP were pooled, aliquoted, and frozen at &#x02212;80 &#x000B0;C.</p></sec>
<sec>
<label>2.1.5</label>
<title>Sample allocation</title>
<p>The 190 specimens were allocated as follows:</p>
<list list-type="simple">
<list-item><p>- APTT therapeutic range derivation: 80 specimens (from 80 different patients).</p></list-item>
<list-item><p>- Concordance evaluation and predictive modeling: 110 specimens (from 110 different patients, independent from derivation set).</p></list-item>
</list>
<p>All specimens were stored at &#x02212;80 &#x000B0;C and analyzed within 3 months. Frozen aliquots were thawed at 37 &#x000B0;C for 3 min, equilibrated to room temperature for 2 min, and tested within 4 h. Samples with visible fibrin strands were excluded.</p></sec></sec>
<sec>
<label>2.2</label>
<title>Instruments and reagents</title>
<sec>
<label>2.2.1</label>
<title>Instruments and allocation</title>
<p>XL-3200c Automated Blood Coagulation Analyzer (ZONCI, Beijing, China): in-house anti-Xa assay;STA-Compact R-Max Analyzer (Stago, France): commercial anti-Xa, APTT, fibrinogen, and antithrombin III assays;Roche Cobas c702 (Switzerland): biochemical parameters (CRP, TG, TC, creatinine for eGFR calculation).</p></sec>
<sec>
<label>2.2.2</label>
<title>In-house anti-Xa assay reagents</title>
<p>Bovine factor Xa (&#x02265;95% purity; Aglycons, China); S-2732 chromogenic substrate (Aglyco, China); BSA, Tris, NaCl, HCl, NaOH, PEG-6000 (China National Medicines Corporation Ltd).</p></sec>
<sec>
<label>2.2.3</label>
<title>Commercial coagulation reagents (all from Stago)</title>
<p>STA<sup>2<sup>&#x000AE;</sup></sup>-Liquid Anti-Xa reagent, calibrators (0&#x02013;1.50 IU/ml), and controls; STA<sup>2<sup>&#x000AE;</sup></sup>-APTT 5 (REF00595); STA<sup>2<sup>&#x000AE;</sup></sup>-Fibrinogen 5; STA<sup>2<sup>&#x000AE;</sup></sup>-Antithrombin III.</p>
<p>Representative lot numbers from the final validation phase are provided; multiple lots were used sequentially during the 7-month study period (March&#x02013;September 2024). Complete lot documentation is available upon request.</p></sec></sec>
<sec>
<label>2.3</label>
<title>Development of the in-house anti-factor Xa activity detection system</title>
<sec>
<label>2.3.1</label>
<title>Assay principle</title>
<p>The assay employs a two-stage chromogenic method. Patient plasma containing heparin is incubated with excess bovine factor Xa (R1). Heparin-AT-III complex neutralizes added factor Xa. Residual factor Xa cleaves chromogenic substrate S-2732 (R2), releasing para-nitroaniline measured at 405 nm. The rate is inversely proportional to heparin concentration.</p></sec>
<sec>
<label>2.3.2</label>
<title>Reagent composition optimization</title>
<p>Factor Xa (0.5&#x02013;4.0 IU/ml), S-2732 (0.3&#x02013;2.0 mmol/L), buffer system (Tris-HCl 10&#x02013;30 mmol/L combined with NaCl 0.1%&#x02212;0.9%, pH range 7.4&#x02013;8.8), and stabilizers (BSA 0.1%&#x02212;0.5% w/v for R1; PEG-6000 0.1%&#x02212;0.5% w/v for R2) were systematically optimized. Sample-to-reagent volume ratios and reaction times were optimized for sensitivity and linear range.</p></sec></sec>
<sec>
<label>2.4</label>
<title>Analytical performance validation</title>
<p>All validation studies followed CLSI guidelines (<xref ref-type="bibr" rid="B27">27</xref>).</p>
<sec>
<label>2.4.1</label>
<title>Precision</title>
<p>We measured three QC levels (low: 0.30 IU/ml; medium: 0.70 IU/ml; high: 1.35 IU/ml) 20 times within-run and in duplicate twice daily over 20 days for Inter-assay precision. Acceptance: within-run CV &#x0003C; 5%, between-run CV &#x0003C; 10%.</p></sec>
<sec>
<label>2.4.2</label>
<title>Analytical sensitivity</title>
<p>Limit of blank (LoB): NPP (anti-Xa &#x0003C; 0.01 IU/ml) was measured 60 times. LoB = Mean blank &#x0002B; 1.645 &#x000D7; SD blank or the non-parametric method (LoB = 95th percentile of blank measurements).</p>
<p>Limit of detection (LoD): low-concentration samples were measured 60 times. LoD = LoB &#x0002B; 1.645 &#x000D7; SD low sample.</p>
<p>Limit of quantitation (LoQ): three candidate concentrations (0.04, 0.08, 0.10 IU/ml) were evaluated with 20 replicates each using three criteria:</p>
<list list-type="bullet">
<list-item><p>Precision: relative median absolute deviation (MAD) &#x02264; 30%</p></list-item></list>
<list list-type="simple">
<list-item><p>&#x000B0; MAD =median (|x<sub>i</sub> &#x02013; median (x)|)</p></list-item>
<list-item><p>&#x000B0; Relative MAD (%) = (MAD/median) &#x000D7; 100</p></list-item>
<list-item><p>&#x000B0; Rationale: MAD is more robust to outliers than CV at low concentrations (<xref ref-type="bibr" rid="B28">28</xref>, <xref ref-type="bibr" rid="B29">29</xref>).</p></list-item></list>
<list list-type="bullet">
<list-item><p>Probability of detection (POD): &#x02265;95% of replicates above LoD</p></list-item>
<list-item><p>Total error (TE): &#x02264; 40%</p></list-item></list>
<list list-type="simple">
<list-item><p>&#x000B0; TE = |Bias| &#x0002B; 2 &#x000D7; SD</p></list-item>
<list-item><p>&#x000B0; Bias = [(Mean measured &#x02013; Target)/Target] &#x000D7; 100%</p></list-item>
</list>
<p>The lowest concentration meeting all three criteria was designated as LoQ.</p></sec>
<sec>
<label>2.4.3</label>
<title>Linearity</title>
<p>Linearity was evaluated using high-concentration and low-concentration plasma samples according to CLSI guideline EP06 (<xref ref-type="bibr" rid="B30">30</xref>). Samples were mixed in different proportions using sample diluent (7H &#x0002B; L, 6H &#x0002B; 2L, 5H &#x0002B; 3L, 4H &#x0002B; 4L, 3H &#x0002B; 5L, 2H &#x0002B; 6L, and H &#x0002B; 7L) to prepare seven dilution levels spanning the concentration range from LoQ to 1.35 IU/ml. Each dilution was measured in duplicate. Linear regression analysis was employed to compare measured values against expected values, and scatter plots were generated.</p></sec>
<sec>
<label>2.4.4</label>
<title>Method comparison</title>
<p>Eighty anonymized plasma specimens from routine diagnostic testing were tested by both in-house and Stago systems. Pearson or Spearman correlation analysis and Bland&#x02013;Altman agreement analysis were employed to evaluate correlation and concordance between the two systems. Paired <italic>t</italic>-test or Wilcoxon signed-rank test was used to assess systematic differences between methods. Regression equations and coefficients of determination (<italic>R</italic><sup>2</sup>) were determined. The method comparison protocol followed CLSI guideline EP09c (<xref ref-type="bibr" rid="B31">31</xref>).</p></sec></sec>
<sec>
<label>2.5</label>
<title>APTT therapeutic range establishment</title>
<sec>
<label>2.5.1</label>
<title>Data allocation strategy</title>
<p>Our laboratory&#x00027;s APTT reference range is 28&#x02013;43.5 s (mean 36 s). The traditional empirical method calculates therapeutic range as 1.5&#x02013;2.5 &#x000D7; the mean normal value (54&#x02013;90 s). The 190 specimens from 110 patients were randomly allocated to two distinct datasets using the random number generator in SPSS 27.0 software:</p>
<list list-type="bullet">
<list-item><p>APTT range derivation (<italic>n</italic> = 80 samples, 42.1%): a randomly selected subset was used to establish laboratory-specific APTT therapeutic ranges through linear regression (APTT = a &#x0002B; b &#x000D7; anti-Xa) and ROC curve analysis. APTT cutoffs corresponding to anti-Xa levels of 0.3 and 0.7 IU/ml were identified.</p></list-item>
<list-item><p>Concordance evaluation and predictive modeling (<italic>n</italic> = 110 samples, 57.9%): the remaining independent samples (not included in derivation set) were used for: (a) evaluating APTT-anti-Xa classification concordance; and (b) identifying predictors of discordance using patient-level clinical data.</p></list-item>
</list>
<p>This allocation strategy ensured independent validation of the derived APTT range while maximizing statistical power for subsequent analyses.</p></sec>
<sec>
<label>2.5.2</label>
<title>ROC analysis</title>
<p>Two ROC analyses identified optimal APTT cutoffs:</p>
<p>Lower threshold (0.3 IU/ml): samples dichotomized as anti-Xa &#x0003C; 0.3 (subtherapeutic) vs. &#x02265;0.3 IU/ml (therapeutic/above). Optimal APTT cutoff determined by Youden Index maximization.</p>
<p>Upper threshold (0.7 IU/ml): samples dichotomized as anti-Xa &#x02264; 0.7 (therapeutic/below) vs. &#x0003E;0.7 IU/ml (supratherapeutic). Optimal APTT cutoff determined by Youden Index (<xref ref-type="bibr" rid="B32">32</xref>).</p></sec>
<sec>
<label>2.5.3</label>
<title>Concordance analysis</title>
<p>Using established ranges, each specimen was classified as subtherapeutic, therapeutic, or supratherapeutic. Concordance was defined as agreement between APTT and anti-Xa across all three categories. Overall agreement rate and Cohen&#x00027;s kappa (&#x003BA;) were calculated.</p></sec></sec>
<sec>
<label>2.6</label>
<title>Predictive modeling</title>
<sec>
<label>2.6.1</label>
<title>Study design</title>
<p>Given poor concordance between APTT and anti-Xa, we developed predictive models to identify patients at risk for discordance.</p>
<p>Critical limitation: exploratory, hypothesis-generating analysis with overfitting risk. Sample size <italic>n</italic> = 110 with 64 discordant events and 10 predictors yields EPV &#x02248; 6.4, below recommended minimum of 10. Five-fold cross-validation employed; all findings require external validation.</p></sec>
<sec>
<label>2.6.2</label>
<title>Variable collection</title>
<p>Ten candidate predictor variables were collected:</p>
<list list-type="bullet">
<list-item><p>Demographic (3): age, sex, body mass index (BMI)</p></list-item>
<list-item><p>Coagulation (3): fibrinogen, AT-III, D-dimer</p></list-item>
<list-item><p>Inflammatory (1): C-reactive protein</p></list-item>
<list-item><p>Renal (1): estimated glomerular filtration rate (eGFR), calculated using the Chronic Kidney Disease Epidemiology Collaboration equation</p></list-item>
<list-item><p>Lipid (2): triglycerides, total cholesterol.</p></list-item>
</list>
<p>All parameters were measured within 24 h of anticoagulation monitoring.</p></sec>
<sec>
<label>2.6.3</label>
<title>Statistical modeling</title>
<p>Outcome variable: concordant (0) vs. discordant (1) APTT-anti-Xa classification.</p>
<p>Random forest model: RF model, with bootstrap sampling and class balancing was constructed. Variable importance was assessed by Mean Decrease in Accuracy. Out-of-bag samples provided internal validation. Performance metrics: AUC, sensitivity, specificity, confusion matrix.</p>
<p>Logistic regression model: all 10 variables were entered into multivariable logistic regression. Backward elimination (<italic>p</italic> &#x02265; 0.05) was <italic>conducted</italic>, retaining clinically important variables (AT-III, fibrinogen) regardless of significance. Multicollinearity (VIF &#x0003E;5), influential observations (Cook&#x00027;s distance &#x0003E;4/<italic>n</italic>), goodness-of-fit (Hosmer&#x02013;Lemeshow test), and model calibration were assessed. Five-fold cross-validation was conducted.</p></sec>
<sec>
<label>2.6.4</label>
<title>Statistical analysis</title>
<p>Normally distributed continuous variables: mean &#x000B1; SD; non-normally distributed: median (IQR); categorical: frequency (%). Concordant vs. discordant groups compared using <italic>t</italic>-test/Mann-Whitney <italic>U</italic>-test (continuous) or chi-square/Fisher&#x00027;s eXact test (categorical). Bonferroni correction applied for multiple comparisons (adjusted &#x003B1; = 0.05/10 = 0.005). Two-tailed <italic>p</italic> &#x0003C; 0.05 considered significant.</p>
<p>Software: SPSS 27.0 (IBM, USA); R 4.2.0 (R Foundation, Austria); MedCalc 20.0 (Belgium).</p></sec></sec></sec>
<sec sec-type="results" id="s3">
<label>3</label>
<title>Results</title>
<sec>
<label>3.1</label>
<title>Development and optimization of anti-factor Xa assay</title>
<sec>
<label>3.1.1</label>
<title>Reagent component optimization</title>
<p>Through systematic concentration gradient experiments, optimal reagent compositions were established. Reagent 1 (R1) contained bovine factor Xa at 0.9 IU/ml (selected from 0.8 to 1.0 IU/ml range), while Reagent 2 (R2) contained chromogenic substrate S-2732 at 0.43 mmol/L (selected from 0.4 to 0.5 mmol/L range). This optimized concentration combination yielded excellent linearity (<italic>R</italic><sup>2</sup> = 0.9998) for diluted standard plasma.</p>
<p>The buffer system, comprising Tris-HCl (10&#x02013;30 mmol/L) and NaCl (0.1%&#x02212;0.9%), was optimized to achieve optimal detection sensitivity and linearity. BSA (0.1%&#x02212;0.5% w/v) was incorporated into R1 and the sample diluent as a stabilizer, whereas PEG-6000 (0.1%&#x02212;0.5% w/v) was added to R2 to prevent reagent denaturation without affecting assay performance.</p>
<p>The optimal sample-to-reagent ratio was determined to be 5 &#x003BC;l (sample):40 &#x003BC;l (R1):100 &#x003BC;l (R2). Under these conditions, linear regression analysis of the 7-point calibration curve using diluted standard plasma yielded <italic>R</italic><sup>2</sup> = 0.9998, while the correlation between measured and expected values in clinical specimens demonstrated <italic>R</italic><sup>2</sup> = 0.9996.</p></sec>
<sec>
<label>3.1.2</label>
<title>Reaction time optimization</title>
<p>Optimal incubation time was 90 s, at which residual heparin in plasma bound maximally to reagent factor Xa, yielding the highest chromogenic signal (<xref ref-type="fig" rid="F1">Figure 1</xref>).</p>
<fig position="float" id="F1">
<label>Figure 1</label>
<caption><p>Reaction time optimization. Incubation time optimization showing &#x00394;(OD) decreased rapidly within 60 s and reached equilibrium after 90 s.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0001.tif">
<alt-text content-type="machine-generated">Line graph showing the change in optical density (&#x00394; OD) over time. The x-axis represents incubation time in minutes, ranging from 0 to 200, and the y-axis shows &#x00394; OD, ranging from 3.6 to 6. The graph shows a sharp decline from approximately 5.6 at 20 minutes to around 3.8 at 100 minutes, followed by minor fluctuations. Data points are connected by lines.</alt-text>
</graphic>
</fig>
<p>During continuous monitoring from 5 to 55 s, the chromogenic reaction reached a plateau phase, after which &#x00394;OD exhibited no further significant changes (<xref ref-type="fig" rid="F2">Figure 2</xref>).</p>
<fig position="float" id="F2">
<label>Figure 2</label>
<caption><p>Reaction time optimization. Reaction time optimization. Detection time kinetics showing &#x00394;(OD) declined sharply in the first 25 s and stabilized after 55 s.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0002.tif">
<alt-text content-type="machine-generated">Line graph showing change in optical density (&#x00394; OD) over detection time in seconds. The &#x00394; OD decreases sharply from 0 to 25 seconds, then gradually levels off from 25 to 125 seconds.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.1.3</label>
<title>Calibration curves</title>
<p>In-house anti-factor Xa detection system: the calibration curve yielded an <italic>R</italic><sup>2</sup> value of 0.99883 with regression equation <italic>Y</italic> = &#x02212;0.01181 &#x0002B; 1.00678<italic>X</italic> (<xref ref-type="fig" rid="F3">Figure 3</xref>).</p>
<fig position="float" id="F3">
<label>Figure 3</label>
<caption><p>Standard curve for anti-Xa activity of heparin. Seven-point calibration curves using standard plasma (0&#x02013;1.50 IU/ml). In-house system (<italic>Y</italic> = &#x02212;0.01181 &#x0002B; 1.00678<italic>X, R</italic><sup>2</sup> = 0.99883).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0003.tif">
<alt-text content-type="machine-generated">Line graph showing the relationship between the theoretical level and instrument level of heparin's anti-Xa activity in international units per milliliter. The fit curve closely follows the plotted data points, represented by squares. The equation of the fit line is Y equals negative 0.01181 plus 1.00678X, with an R squared value of 0.99883.</alt-text>
</graphic>
</fig>
<p>STA-Liquid Anti-Xa detection system: the calibration curve yielded an <italic>R</italic><sup>2</sup> value of 0.99854 with regression equationY =-0.00387 &#x0002B; 1.00596<italic>X</italic> (<xref ref-type="fig" rid="F4">Figure 4</xref>).</p>
<fig position="float" id="F4">
<label>Figure 4</label>
<caption><p>Standard curve for anti-Xa activity of heparin. Seven-point calibration curves using standard plasma (0&#x02013;1.50 IU/ml). STA-Liquid system (<italic>Y</italic> = 7.1312 &#x02013; 3.7510<italic>X, R</italic><sup>2</sup> = 0.99854). Each point represents duplicate measurements.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0004.tif">
<alt-text content-type="machine-generated">Chart showing a linear relationship between theoretical and instrument levels of Anti-Xa activity of heparin, with the fit curve equation \(Y = 1.0096X - 0.00387\) and \(R^2 = 0.99854\). Both axes are labeled in IU/ml, ranging from 0 to 1.2.</alt-text>
</graphic>
</fig>
<p>Both systems demonstrated excellent linearity.</p></sec></sec>
<sec>
<label>3.2</label>
<title>Analytical performance validation</title>
<sec>
<label>3.2.1</label>
<title>Precision study</title>
<p>We conducted precision studies at three quality control (QC) levels (0.30, 0.70, and 1.35 IU/ml) met CLSI acceptance performance. Intra-assay precision (coefficient of variation, CV) ranged from 1.15 to 4.32%, while Inter-assay precision (CV) ranged from 1.25 to 4.03%. All results met the acceptable criteria established by the Clinical and Laboratory Standards Institute (CLSI) EP05-A3 guideline (<xref ref-type="table" rid="T1">Table 1</xref>).</p>
<table-wrap position="float" id="T1">
<label>Table 1</label>
<caption><p>Results of repeatability and intermediate precision.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th valign="top" align="left"><bold>Precision type</bold></th>
<th valign="top" align="center"><bold>QC level (IU/ml)</bold></th>
<th valign="top" align="center"><bold>Mean (IU/ml)</bold></th>
<th valign="top" align="center"><bold>SD (%)</bold></th>
<th valign="top" align="center"><bold>CV (%)</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Intra-assay precision<sup>a</sup></td>
<td valign="top" align="center">1.35</td>
<td valign="top" align="center">1.35</td>
<td valign="top" align="center">1.55</td>
<td valign="top" align="center">1.15</td>
</tr>
 <tr>
<td/>
<td valign="top" align="center">0.70</td>
<td valign="top" align="center">0.71</td>
<td valign="top" align="center">1.79</td>
<td valign="top" align="center">2.52</td>
</tr>
 <tr>
<td/>
<td valign="top" align="center">0.30</td>
<td valign="top" align="center">0.31</td>
<td valign="top" align="center">1.34</td>
<td valign="top" align="center">4.32</td>
</tr>
<tr>
<td valign="top" align="left">Inter-assay precision<sup>b</sup></td>
<td valign="top" align="center">1.35</td>
<td valign="top" align="center">1.34</td>
<td valign="top" align="center">1.68</td>
<td valign="top" align="center">1.25</td>
</tr>
 <tr>
<td/>
<td valign="top" align="center">0.70</td>
<td valign="top" align="center">0.70</td>
<td valign="top" align="center">1.44</td>
<td valign="top" align="center">2.06</td>
</tr>
<tr>
<td/>
<td valign="top" align="center">0.30</td>
<td valign="top" align="center">0.30</td>
<td valign="top" align="center">1.21</td>
<td valign="top" align="center">4.03</td>
</tr></tbody>
</table>
<table-wrap-foot>
<p>CV (%) = (SD/mean) &#x000D7; 100%. QC, quality control; SD, standard deviation; CV, coefficient of variation. <sup>a</sup>Intra-assay precision: 20 consecutive measurements within a single run. <sup>b</sup>Inter-assay precision: two runs per day over 20 days (n = 80).</p>
</table-wrap-foot>
</table-wrap>
</sec>
<sec>
<label>3.2.2</label>
<title>Analytical sensitivity</title>
<p>The limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) were comprehensively evaluated according to CLSI guideline EP17-A2.</p>
<p>Limit of blank (LoB): sixty blank measurements (normal pooled plasma with anti-Xa &#x0003C; 0.01 IU/ml) yielded a mean value of 0.042 &#x000B1; 0.015 IU/ml. The LoB was established at 0.073 IU/ml (95th percentile; <xref ref-type="fig" rid="F5">Figure 5</xref>).</p>
<fig position="float" id="F5">
<label>Figure 5</label>
<caption><p>Analytical sensitivity evaluation. LoB determination showing distribution of 60 blank measurements (mean 0.042 IU/ml, LoB 0.073 IU/ml at 95th percentile).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0005.tif">
<alt-text content-type="machine-generated">Bar chart displaying the frequency of measured values in IU/mL. Values range from 0.00 to 0.14. A red line indicates LoB at 0.073 IU/mL, marking the ninety-fifth percentile. Frequency peaks around 0.06 to 0.08, with bars gradually decreasing on either side.</alt-text>
</graphic>
</fig>
<p>Limit of detection (LoD): sixty measurements of low-concentration samples (target: 0.08 IU/ml, SD = 0.004 IU/ml) were performed. The LoD was calculated as 0.073 &#x0002B; 1.645 &#x000D7; 0.004 = 0.080 IU/ml.</p>
<p>Limit of quantitation (LoQ) was evaluated at three candidate concentrations (<italic>n</italic> = 20 each) based on three predefined criteria: median absolute deviation (MAD) &#x02264; 30%, probability of detection (POD) &#x02265;95%, and total error (TE) &#x02264; 40%.</p>
<list list-type="bullet">
<list-item><p>At 0.04 IU/ml: MAD = 500%, POD = 42.1%, TE = 112.9% (failed all criteria).</p></list-item>
<list-item><p>At 0.08 IU/ml: MAD = 28.57%, POD = 95.7%, TE = 43.4% (borderline performance).</p></list-item>
<list-item><p>At 0.10 IU/ml: MAD = 20.0%, POD = 100%, TE = 39.2% (passed all criteria).</p></list-item>
</list>
<p>Therefore, we established the LoQ at 0.10 IU/ml (<xref ref-type="fig" rid="F6">Figure 6</xref>).</p>
<fig position="float" id="F6">
<label>Figure 6</label>
<caption><p>Analytical sensitivity evaluation. LoQ established at 0.10 IU/ml based on convergence of MAD ( &#x02264; 30%), POD (&#x02265;95%), and TE ( &#x02264; 40%) criteria. MAD, median absolute deviation; POD, probability of detection; TE, total error.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0006.tif">
<alt-text content-type="machine-generated">Line graph showing metric values versus concentration in IU/mL. Relative MAD, POD, and Total Error are plotted with lines and symbols. Criteria for MAD, POD, and TE are indicated with dotted lines. Red circles mark the established LoQ at 0.10 IU/mL.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.2.3</label>
<title>Linearity study</title>
<p>Linearity was evaluated at nine concentration points ranging from 0.10 to 1.35 IU/ml according to CLSI guideline EP06 (<xref ref-type="fig" rid="F7">Figure 7</xref>). Linear regression demonstrated excellent linearity, with regression equation <italic>Y</italic> = 1.00596<italic>X</italic> &#x02013; 0.00387 and coefficient of determination <italic>R</italic><sup>2</sup> = 0.99923.</p>
<fig position="float" id="F7">
<label>Figure 7</label>
<caption><p>Linear range of the in-house anti-factor Xa detection system. Linear regression analysis (<italic>Y</italic> = 1.00596<italic>X</italic> &#x02013; 0.00387, <italic>R</italic><sup>2</sup> = 0.99923) with duplicate measurements on three separate days (<italic>n</italic> = 6 per dilution level).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0007.tif">
<alt-text content-type="machine-generated">Scatter plot showing the relationship between measured and theoretical anti-factor Xa activity in IU/mL. Data points form a linear pattern with a fit curve. The equation of the line is Y = 1.00596X - 0.00387 and R&#x000B2; = 0.99923, indicating a strong correlation. The legend indicates &#x0201C;anti-Xa activity of heparin&#x0201D; and &#x0201C;Fit curve&#x0201D;.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.2.4</label>
<title>Carryover rate</title>
<p>The carryover rate was calculated using the formula: CR = |L1 &#x02013; L3|/(H3 &#x02013; L3) &#x000D7; 100% = |0.45 &#x02013; 0.47|/(1.31 &#x02013; 0.47) &#x000D7; 100% = 0.57%, which is well below the acceptance criterion of 3%, indicating negligible carryover contamination.</p></sec>
<sec>
<label>3.2.5</label>
<title>Stability study</title>
<p>Stability of the complete liquid reagent system (R1: bovine factor Xa solution with BSA stabilizer; R2: S-2732 substrate solution with PEG-6000 stabilizer) was evaluated under thermal stress at 37 &#x000B0;C, for 14 days. The relative median absolute deviation (MAD) at all concentration levels remained &#x02264; 2.86%, meeting the CLSI EP25-A requirement of &#x02264; 10%. This indicates that reagent precision was unaffected and stability was satisfactory for routine clinical use.</p></sec></sec>
<sec>
<label>3.3</label>
<title>Method comparison study</title>
<sec>
<label>3.3.1</label>
<title>Baseline characteristics and sample allocation</title>
<p>The derivation set (<italic>n</italic> = 80) and validation set (<italic>n</italic> = 110) showed no statistically significant differences in anti-Xa concentration distribution, age, or sex composition (<italic>p</italic> &#x0003E; 0.05), confirming successful random allocation.</p></sec>
<sec>
<label>3.3.2</label>
<title>Correlation and agreement analysis</title>
<p>Eighty specimens stratified across the clinical range were analyzed by both the in-house and STA-Liquid (Stago) systems. Test results from both systems followed a normal distribution. Paired samples <italic>t</italic>-test showed no significant difference between the two detection systems (<italic>p</italic> &#x0003E; 0.05).</p>
<p>Linear regression analysis yielded the equation <italic>Y</italic> = &#x02212;0.01018 &#x0002B; 1.03984<italic>X</italic> with <italic>R</italic><sup>2</sup> = 0.972, and the Pearson correlation coefficient was <italic>r</italic> = 0.986 (<italic>p</italic> &#x0003C; 0.001; <xref ref-type="fig" rid="F8">Figure 8</xref>).</p>
<fig position="float" id="F8">
<label>Figure 8</label>
<caption><p>Method comparison between in-house and STA-Liquid anti-factor Xa systems. Correlation analysis showing strong linear relationship (<italic>Y</italic> = &#x02212;0.01018 &#x0002B; 1.03984<italic>X, R</italic><sup>2</sup> = 0.972, Pearson <italic>r</italic> = 0.986, &#x003C1; &#x0003C; 0.001, <italic>n</italic> = 120).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0008.tif">
<alt-text content-type="machine-generated">Scatter plot showing the relationship between Anti-Xa values using the Stago detection system (x-axis) and a new-built detection system (y-axis), both measured in IU/mL. Data points form a trend line fitted with a red curve, indicating strong correlation.</alt-text>
</graphic>
</fig>
<p>The Bland&#x02013;Altman plot demonstrated 93.75% of samples within the two detection systems. Among 80 data points, only 5 (6.25%) fell outside the 95% limits of agreement, indicating excellent concordance (<xref ref-type="fig" rid="F9">Figure 9</xref>).</p>
<fig position="float" id="F9">
<label>Figure 9</label>
<caption><p>Method comparison between in-house and STA-Liquid anti-factor Xa systems. Bland-Altman analysis showing mean bias of 0.005 IU/ml with 95% limits of agreement at &#x000B1;1.96 SD (&#x02212;0.1 to 0.11 IU/ml); five points (6.25%) fell outside limits.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0009.tif">
<alt-text content-type="machine-generated">Bland-Altman plot showing the difference between Stago-newly-built method and mean of newly-built and Stago (IU/mL) on the y-axis and x-axis, respectively. Most data points are clustered near the mean line at 0.005 with upper and lower limits of agreement at +0.11 and -0.1.</alt-text>
</graphic>
</fig>
<p>Both sections used <italic>n</italic> = 80 derivation set but examined different variables. Section 3.3.2 tested anti-Xa values (normally distributed due to stratified sampling). Section 3.4.1 tested APTT values (non-normal distribution reflecting clinical clustering).</p></sec></sec>
<sec>
<label>3.4</label>
<title>APTT therapeutic range calibrated against anti-factor Xa activity</title>
<sec>
<label>3.4.1</label>
<title>Correlation analysis</title>
<p>Since neither APTT nor anti-factor Xa activity followed normal distribution (Shapiro-Wilk test, <italic>p</italic> &#x0003C; 0.05), Spearman rank correlation analysis was employed. A moderate positive correlation was observed (Spearman&#x00027;s &#x003C1; = 0.678, <italic>p</italic> &#x0003C; 0.001). Linear regression yielded the equation <italic>Y</italic> = 132.7<italic>X</italic> &#x0002B; 44.1 (<italic>R</italic><sup>2</sup> = 0.46).</p>
<p>This moderate correlation indicates that only 46% of APTT variance is explained by anti-Xa activity, with the remaining 54% attributable to confounding factors including fibrinogen levels, factor VIII activity, and sample quality. This finding supports the need for careful interpretation when using APTT as a surrogate marker for anti-factor Xa monitoring (<xref ref-type="fig" rid="F10">Figure 10</xref>).</p>
<fig position="float" id="F10">
<label>Figure 10</label>
<caption><p>Correlation between APTT and anti-factor Xa activity. Linear regression analysis showing moderate correlation (<italic>Y</italic> = 132.7<italic>X</italic> &#x0002B; 44.1, <italic>R</italic><sup>2</sup> = 0.46, <italic>n</italic> = 110).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0010.tif">
<alt-text content-type="machine-generated">Scatter plot showing the relationship between APTT (seconds) and Anti-Xa (IU/mL) with data points scattered and a trend line. The equation of the line is Y = 132.66X + 44.1 with an R&#x000B2; value of 0.46, indicating a moderate correlation.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.4.2</label>
<title>Linear regression method</title>
<p>Using the regression equation <italic>Y</italic> = 132.7<italic>X</italic> &#x0002B; 44.1, an anti-factor Xa level of 0.3 IU/ml corresponded to a predicted APTT value of 84 s, while 0.7 IU/ml corresponded to 137 s, yielding a preliminary range of 84&#x02013;137 s.</p></sec>
<sec>
<label>3.4.3</label>
<title>ROC curve analysis</title>
<p>Receiver operating characteristic (ROC) analysis identified optimal APTT cutoff values for distinguishing therapeutic from non-therapeutic anticoagulation levels (<xref ref-type="fig" rid="F11">Figure 11</xref>).</p>
<fig position="float" id="F11">
<label>Figure 11</label>
<caption><p>ROC curve analysis for APTT therapeutic threshold determination. ROC curves comparing therapeutic vs. sub-therapeutic (AUC = 0.870, optimal cutoff = 71.34 s) and therapeutic vs. supra-therapeutic (AUC = 0.855, optimal cutoff = 125.80 s).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0011.tif">
<alt-text content-type="machine-generated">ROC curve comparing two conditions: therapeutic versus sub-therapeutic (blue, AUC = 0.870) and therapeutic versus supra-therapeutic (orange, AUC = 0.855). Dashed line represents random guess. Optimal cutoffs are marked: 71.34 (blue) and 125.80 (orange) for APTT. x-axis shows false positive rate; y-axis shows true positive rate.</alt-text>
</graphic>
</fig>
<p>Lower threshold (0.3 IU/ml): for distinguishing therapeutic levels (&#x02265;0.3 IU/ml) from subtherapeutic levels (&#x0003C; 0.3 IU/ml), the optimal APTT cutoff was 78.9 s. Performance metrics: AUC = 0.723, sensitivity 70.6%, specificity 71.6%.</p>
<p>Upper threshold (0.7 IU/ml): for distinguishing therapeutic levels ( &#x02264; 0.7 IU/ml) from supratherapeutic levels (&#x0003E;0.7 IU/ml), the optimal APTT cutoff was 126.6 s. Performance metrics: AUC = 0.815, sensitivity 83.3%, specificity 80.0%.</p>
<p>ROC-derived cutoffs were slightly lower than regression-predicted values, reflecting the balance between sensitivity and specificity. Based on integrating both methods and prioritizing clinical safety, a laboratory-specific APTT therapeutic range of 79&#x02013;127 s was established, corresponding to anti-factor Xa 0.3&#x02013;0.7 IU/ml.</p></sec>
<sec>
<label>3.4.4</label>
<title>Final therapeutic range</title>
<p>A laboratory-specific APTT therapeutic range of 79&#x02013;127 s was established, corresponding to anti-factor Xa levels of 0.3&#x02013;0.7 IU/ml, Although linear regression predicted 84&#x02013;137 s, ROC-derived cutoffs (78.9 and 126.6 s) were adopted due to superior diagnostic accuracy (AUC 0.723&#x02013;0.815) with balanced sensitivity (70.6%&#x02212;83.3%) and specificity (71.6%&#x02212;80.0%), ensuring optimal clinical decision-making for UFH monitoring.</p>
<p>A laboratory-specific APTT therapeutic range of 79&#x02013;127 s was established, corresponding to anti-factor Xa levels of 0.3&#x02013;0.7 IU/ml. Compared with the traditional 1.5&#x02013;2.5 &#x000D7; baseline method (54&#x02013;90 s), which resulted in a 42.7% misclassification rate, the ROC-optimized range reduced misclassification to 25.5%. Notably, the traditional method led to underdosing risk, as 54% of patients with APTT values of 54&#x02013;79 s were actually subtherapeutic, while 38% of patients with APTT &#x0003E;90 s were within the therapeutic range, raising false overdosing concerns. Although linear regression predicted a range of 84&#x02013;137 s, ROC-derived cutoffs (78.9 and 126.6 s) were adopted due to superior diagnostic accuracy (AUC 0.723&#x02013;0.815) with balanced sensitivity (70.6%&#x02212;83.3%) and specificity (71.6%&#x02212;80.0%). These findings support the CLSI H21-A5 recommendation for laboratory-specific therapeutic range establishment, ensuring optimal clinical decision-making for UFH monitoring.</p></sec></sec>
<sec>
<label>3.5</label>
<title>Classification concordance analysis</title>
<sec>
<label>3.5.1</label>
<title>Concordance results</title>
<p>Among 110 paired measurements, 46 pairs (41.8%) showed concordant classification between APTT and anti-factor Xa, while 64 pairs (58.2%) were discordant (<xref ref-type="fig" rid="F12">Figure 12</xref>). The concordant pairs comprised 8 subtherapeutic (7.3%), 32 therapeutic (29.1%), and 6 supratherapeutic (5.5%) classifications.</p>
<fig position="float" id="F12">
<label>Figure 12</label>
<caption><p>Concordance analysis between APTT and anti-factor Xa classifications. Heatmap showing distribution of 110 patients across APTT and anti-Xa categories with concordance and discordance patterns.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0012.tif">
<alt-text content-type="machine-generated">Heatmap showing APTT and Anti-Xa combinations with three categories: subtherapeutic, therapeutic, and supratherapeutic. Key values include 32 in the center, indicating the highest frequency, and others range from 5 to 15. A color gradient from yellow to blue represents values from 5 to 32.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.5.2</label>
<title>Discordance by clinical indication</title>
<p>UFH indications and indication-specific discordance rates are summarized in <xref ref-type="table" rid="T2">Table 2</xref>. ICU/surgical populations demonstrated numerically higher discordance rates (73.3%&#x02212;75.0%) compared to medical indications (52.6%&#x02212;54.2%), though this difference did not reach statistical significance (<italic>p</italic> = 0.156) due to limited subgroup sizes. This trend suggests that critically ill patients may be particularly susceptible to APTT-based monitoring inaccuracies due to higher prevalence of confounding factors including elevated fibrinogen, acute phase responses, and renal dysfunction.</p>
<table-wrap position="float" id="T2">
<label>Table 2</label>
<caption><p>UFH indications and APTT/anti-factor Xa classification discordance rates.</p></caption>
<table frame="hsides" rules="groups">
<thead>
<tr>
<th valign="top" align="left"><bold>Indication</bold></th>
<th valign="top" align="center"><bold><italic>n</italic> (%)</bold></th>
<th valign="top" align="center"><bold>Discordance rate</bold></th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Acute coronary syndrome</td>
<td valign="top" align="center">38 (34.5)</td>
<td valign="top" align="center">52.6%</td>
</tr>
<tr>
<td valign="top" align="left">VTE treatment/prophylaxis</td>
<td valign="top" align="center">24 (21.8)</td>
<td valign="top" align="center">54.2%</td>
</tr>
<tr>
<td valign="top" align="left">Post-cardiac surgery</td>
<td valign="top" align="center">15 (13.6)</td>
<td valign="top" align="center">73.3%</td>
</tr>
<tr>
<td valign="top" align="left">Hemodialysis/CRRT</td>
<td valign="top" align="center">8 (7.3)</td>
<td valign="top" align="center">75.0%</td>
</tr>
<tr>
<td valign="top" align="left">Other</td>
<td valign="top" align="center">25 (22.7)</td>
<td valign="top" align="center">56.0%</td>
</tr></tbody>
</table>
<table-wrap-foot>
<p>Discordance was defined as different therapeutic classification between APTT-based (79&#x02013;127 s) and anti-factor a-based (0.3&#x02013;0.7 IU/ml) methods. &#x0201C;Other&#x0201D; includes arterial thrombosis and bridging therapy. ICU/surgical indications (73.3%&#x02212;75.0%) vs. medical indications (52.6%&#x02212;54.2%), p = 0.156. UFH, unfractionated heparin; VTE, venous thromboembolism; CRRT, continuous renal replacement therapy.</p>
</table-wrap-foot>
</table-wrap></sec></sec>
<sec>
<label>3.6</label>
<title>Predictive model for detection discordance</title>
<sec>
<label>3.6.1</label>
<title>Random forest model</title>
<p>A random forest model incorporating 10 clinical and laboratory variables was constructed to predict concordance vs. discordance between APTT and anti-factor Xa measurements. Given the limited sample size, 5-fold cross-validation was employed to assess model performance. Results showed a mean AUC of 0.79 (95% CI: 0.71&#x02013;0.87), accuracy of 78.8%, sensitivity of 64.3%, and specificity of 89.5%. Permutation importance analysis ranked the 10 variables in descending order of importance: eGFR, Fibrinogen (Fg), AT-III, TG, CRP, age, ALB, TC, BMI, and Hb (<xref ref-type="fig" rid="F13">Figure 13</xref>). All predictive features demonstrated good stability across 5-fold cross-validation (coefficient of variation CV &#x0003C; 1.0).</p>
<fig position="float" id="F13">
<label>Figure 13</label>
<caption><p>Variable importance ranking of the random forest model. Permutation importance scores showing eGFR (12.0%), fibrinogen (10.0%), and AT-III (10.0%) as top predictors. Error bars represent standard deviation from 5-fold cross-validation.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0013.tif">
<alt-text content-type="machine-generated">Bar chart showing the variable importance ranking of a random forest model. eGFR is the most important variable at 12%, followed by Fg and AT-III at 10% each. Other variables include TG at 9%, CRP at 8%, years at 7%, Alb and TC at 6.5% each, BMI at 6%, and Hb at 5%. The x-axis represents permutation importance in percentages.</alt-text>
</graphic>
</fig>
</sec>
<sec>
<label>3.6.2</label>
<title>Logistic regression model</title>
<p>Ten variables were entered into multivariable logistic regression based on literature evidence and clinical relevance. Backward stepwise regression identified five independent predictors: Fg (OR = 3.67, 95% CI: 1.67&#x02013;8.07, <italic>p</italic> = 0.001), TG (OR = 1.50, 95% CI: 1.08&#x02013;2.08, <italic>p</italic> = 0.015), eGFR (OR = 1.03, 95% CI: 1.00&#x02013;1.05, <italic>p</italic> = 0.025), CRP (OR = 1.07, 95% CI: 1.00&#x02013;1.14, <italic>p</italic> = 0.041), and AT-III (OR = 1.015, 95% CI: 1.002&#x02013;1.028, <italic>p</italic> = 0.023). The model achieved an AUC of 0.726 (95% CI: 0.638&#x02013;0.814), with accuracy of 68.2%, sensitivity of 73.1%, and specificity of 63.8%. Multicollinearity diagnostics showed all variables had variance inflation factor (VIF) &#x0003C; 2.0 (mean VIF = 1.23), and the Hosmer&#x02013;Lemeshow goodness-of-fit test yielded satisfactory results (&#x003C7;<sup>2</sup> = 6.823, <italic>p</italic> = 0.555).</p></sec>
<sec>
<label>3.6.3</label>
<title>Consistency analysis between the two models</title>
<p>Both models demonstrated high consistency in identifying core predictive factors. The five key variables commonly identified (Fg, AT-III, eGFR, TG, CRP) ranked among the top features in both models. The random forest model exhibited superior overall predictive performance (AUC = 0.79) and specificity (89.5%), while the logistic regression model showed slightly better sensitivity (73.1%). The complementary nature of these two models provides comprehensive reference for clinical decision-making.</p></sec></sec></sec>
<sec sec-type="discussion" id="s4">
<label>4</label>
<title>Discussion</title>
<p>UFH remains a cornerstone anticoagulant in critical care settings worldwide. However, optimal therapeutic monitoring continues to pose significant clinical challenges. While APTT has served as the conventional monitoring method for over five decades, accumulating evidence reveals substantial limitations including high inter-laboratory variability due to significant differences in reagent sensitivity (<xref ref-type="bibr" rid="B17">17</xref>, <xref ref-type="bibr" rid="B33">33</xref>), susceptibility to pre-analytical factors, and poor correlation with clinical outcomes (<xref ref-type="bibr" rid="B34">34</xref>). The 2012 American College of Chest Physicians guidelines acknowledge these limitations and recommend anti-factor Xa assays as an alternative, particularly in patients with baseline coagulation abnormalities (<xref ref-type="bibr" rid="B26">26</xref>).</p>
<p>Despite these recommendations, anti-factor Xa assay adoption remains limited in clinical practice. Recent surveys indicate that only approximately one-quarter of institutions routinely employ anti-factor Xa monitoring, with cost frequently cited as a primary barrier A Swiss prospective study reported total costs of 24.15 Swiss Francs (approximately $25 USD) for anti-Xa vs. 23.40 Swiss Francs for APTT per test (<xref ref-type="bibr" rid="B14">14</xref>, <xref ref-type="bibr" rid="B35">35</xref>). This cost disparity creates economic barriers particularly in resource-limited settings. Furthermore, monitoring method discordance occurs in 30%&#x02212;50% of cases, creating clinical uncertainty and potentially inappropriate dosing adjustments. This discordance disproportionately affects high-risk populations including ICU patients, those with acute phase reactions, and individuals with renal dysfunction (<xref ref-type="bibr" rid="B36">36</xref>, <xref ref-type="bibr" rid="B37">37</xref>).</p>
<p>To address these barriers, we developed and validated an in-house chromogenic anti-factor Xa assay optimized for routine clinical implementation. Through systematic concentration gradient experiments, optimal reagent compositions were identified, achieving balanced sensitivity, linearity, and stability. The incorporation of BSA (0.1%&#x02212;0.5% w/v) as an R1 stabilizer and PEG-6000 (0.1%&#x02212;0.5% w/v) for R2 extends reagent shelf-life to 14 days at 37 &#x000B0;C with relative MAD &#x02264; 2.86%, reducing waste compared to commercial systems that require daily preparation. The optimized sample-to-reagent ratio requires only 5 &#x003BC;l of patient plasma per test, substantially reducing sample volume requirements compared to commercial platforms. This miniaturization proves particularly valuable in neonatal, pediatric, and critically ill populations where sample volume constraints are significant (<xref ref-type="bibr" rid="B38">38</xref>). The in-house monitoring system completes analysis within 5 min, facilitating rapid clinical decision-making. Preliminary cost analysis indicated reagent costs of approximately $2.00 per test, including calibrators, quality controls, and consumables, representing a substantial cost reduction compared to commercial systems.</p>
<p>Our higher discordance rate (58.2% vs. 30%&#x02212;50%) reflects: (1) a high-risk population&#x02014;post-cardiac surgery (13.6%) and hemodialysis/CRRT (7.3%) patients showed 73%&#x02212;75% discordance; (2) ROC-optimized rather than empirical APTT ranges, revealing true biological discordance; and (3) stricter three-category classification. This high discordance actually validates anti-Xa utility in populations where accurate monitoring is most critical.</p>
<p>Rigorous validation following CLSI guidelines demonstrated analytical performance meeting international standards across all parameters. Intra-assay CV of 1.15%&#x02212;4.32% and inter-assay CV of 1.25%&#x02212;4.03% across the therapeutic range (0.30&#x02013;1.35 IU/ml) compare favorably with reported commercial system performance (<xref ref-type="bibr" rid="B39">39</xref>, <xref ref-type="bibr" rid="B40">40</xref>). Excellent linearity across 0.10&#x02013;1.35 IU/ml (<italic>R</italic><sup>2</sup> = 0.999, recovery 90%&#x02212;110%) exceeds CLSI acceptability criteria. Direct comparison with the Stago STA-Liquid Anti-Xa platform demonstrated strong correlation (Pearson <italic>r</italic> = 0.986, <italic>R</italic><sup>2</sup> = 0.972) with 93.75% of samples within Bland-Altman limits of agreement and mean bias of &#x02212;0.008 IU/ml. These results demonstrate that our in-house platform provides results comparable to the commercial reference method. Carryover rate of 0.57% and 14-day thermal stability ensure reliable performance in routine clinical workflows.</p>
<p>To evaluate clinical performance, we applied the newly established anti-factor Xa assay to UFH monitoring in hospitalized patients. Our clinical validation findings provide evidence supporting anti-factor Xa measurement over APTT-based monitoring. Although APTT and anti-factor Xa demonstrated moderate correlation (Spearman &#x003C1; = 0.678, <italic>p</italic> &#x0003C; 0.001), classification agreement at clinically relevant thresholds was poor (41.8%, Cohen&#x00027;s &#x003BA; = 0.242), with 58.2% discordance rate consistent with prior reports (<xref ref-type="bibr" rid="B41">41</xref>&#x02013;<xref ref-type="bibr" rid="B43">43</xref>). This substantial discordance persisted despite establishing laboratory-specific APTT therapeutic ranges (79&#x02013;127 s) through ROC analysis. It is worth noting that optimized APTT cutoffs achieved only moderate diagnostic accuracy (AUC 0.723&#x02013;0.815, sensitivity 70.6%&#x02212;83.3%, specificity 71.6%&#x02212;80.0%), indicating that approximately 20%&#x02212;30% of patients would be misclassified even with laboratory-specific calibration. These findings underscore the limitations of APTT-based UFH monitoring and support the utility of direct anti-factor Xa measurement, particularly in high-risk populations.</p>
<p>We applied machine learning approaches to identify clinical and laboratory predictors of APTT-anti-factor Xa discordance. Random forest modeling (AUC = 0.79, 95% CI: 0.71&#x02013;0.87; specificity 89.5%) ranked eGFR, fibrinogen, and AT-III as important predictive features. Logistic regression analysis (sensitivity 73.1%, AUC = 0.726) identified five independent predictors: fibrinogen (OR = 3.67, 95% CI: 1.67&#x02013;8.07), triglycerides (OR = 1.50, 95% CI: 1.08&#x02013;2.08), eGFR (OR = 1.03, 95% CI: 1.00&#x02013;1.05), CRP (OR = 1.07, 95% CI: 1.00&#x02013;1.14), and AT-III (OR = 1.015, 95% CI: 1.002&#x02013;1.028), with acceptable multicollinearity (all VIF &#x0003C; 2.0). These findings align with known pathophysiologic mechanisms wherein elevated fibrinogen and acute phase reactants prolong APTT independent of heparin effect, hypertriglyceridemia causes spurious APTT prolongation through sample turbidity, renal dysfunction affects heparin clearance, and AT-III variability directly impacts heparin anticoagulant activity (<xref ref-type="bibr" rid="B33">33</xref>, <xref ref-type="bibr" rid="B44">44</xref>&#x02013;<xref ref-type="bibr" rid="B47">47</xref>).</p>
<p>Based on these findings and existing literature (<xref ref-type="bibr" rid="B19">19</xref>, <xref ref-type="bibr" rid="B48">48</xref>, <xref ref-type="bibr" rid="B49">49</xref>), we present a preliminary, hypothesis-generating framework (<xref ref-type="fig" rid="F14">Figure 14</xref>). This framework is not intended to guide clinical practice without prospective validation with clinical endpoints (bleeding/thrombosis events; <xref ref-type="fig" rid="F14">Figure 14</xref>). This framework suggests preferential anti-factor Xa monitoring for patients with fibrinogen &#x02264; 2.0 or &#x02264; 4.5 g/L plus additional abnormalities including eGFR &#x02265;45 ml/min/1.73 m<sup>2</sup>, albumin &#x02264; 17 g/L, CRP &#x0003E;45 mg/L, TG &#x02265;2.5 mmol/L, or concurrent presence of multiple factors. For patients without these predictors, APTT monitoring may remain adequate. However, due to modest sample size (<italic>n</italic> = 110, with approximately 64 discordant events yielding events-per-variable &#x0003C; 7), this framework represents exploratory guidance requiring validation in larger multicenter studies. Whether addressing these discordances through selective anti-factor Xa monitoring improves patient outcomes remains uncertain and requires evaluation in prospective trials. We emphasize that this framework should not guide clinical practice without rigorous prospective validation.</p>
<fig position="float" id="F14">
<label>Figure 14</label>
<caption><p>Clinical algorithm for heparin monitoring strategy selection. Decision flowchart based on AT-III level (&#x02265;70% for APTT; &#x0003C; 70% with high-risk criteria for anti-factor Xa monitoring). High-risk criteria include fibrinogen &#x02264; 2.0 or &#x02264; 4.5 g/L with additional abnormalities (eGFR &#x0003C; 45 ml/min/1.73 m<sup>2</sup>, albumin &#x02264; 17 g/L, CRP &#x0003E;45 mg/L, TG &#x02265;2.5 mmol/L, or &#x02265;2 factors simultaneously).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xlink:href="fmed-12-1745447-g0014.tif">
<alt-text content-type="machine-generated">Flowchart depicting monitoring for patients requiring heparin. Decision starts with checking if AT-III is greater than or equal to seventy percent: &#x0201C;No&#x0201D; leads to APTT monitoring; &#x0201C;Yes&#x0201D; evaluates several factors: Fibrinogen less than 2.0 or more than 4.5 grams per liter, eGFR less than 45 milliliters per minute per 1.73 square meters, CRP over 100 milligrams per liter, TG over 2.5 millimoles per liter. When two or more abnormal factors are present simultaneously, Anti-Xa monitoring is needed.</alt-text>
</graphic>
</fig>
<p>The anti-factor Xa assay established in this study demonstrates several key advantages that support its clinical implementation. The assay exhibits excellent analytical performance with good linearity across its measurement range (Pearson <italic>r</italic> = 0.986, <italic>R</italic><sup>2</sup> = 0.972) and precision that meets international quality standards (intra-assay CV &#x02264; 4.32%, inter-assay CV &#x02264; 4.03%). The chromogenic substrate method provides superior stability and precision compared with traditional coagulation-based detection kits. Clinical validation against the Stago commercial platform confirmed high correlation and consistency when detecting anti-factor Xa activity in patients receiving UFH therapy, establishing the assay&#x00027;s reliability for therapeutic monitoring. Notably, the assay offers significant practical advantages including rapid turnaround time with complete detection within 5 min and substantially reduced cost at approximately $2.0 per test, which is considerably lower than commercial platforms and may facilitate broader clinical adoption of anti-factor Xa testing in resource-limited settings.</p>
<p>However, this study also has several limitations: (1) Limited sample size&#x02014;single-center study (<italic>n</italic> = 110, 64 events), risking overfitting and institution-specific results (<xref ref-type="bibr" rid="B50">50</xref>, <xref ref-type="bibr" rid="B51">51</xref>); (2) Cross-sectional design, unable to assess dynamic changes; (3) Absence of bleeding/thrombotic event data, precluding outcome verification; and (4) Only internal cross-validation, lacking external validation.</p>
<p>The next steps would be to explore optimal lyophilization parameters to extend reagent shelf-life and facilitate broader distribution. Future research should prioritize multicenter external validation across diverse clinical settings to confirm assay performance and predictive model transportability. Prospective studies linking monitoring discordances to clinical outcomes including bleeding, thrombosis, and mortality are needed. Studies adequately powered for special populations such as ICU and critically ill patients are needed to define population-specific monitoring strategies. Formal cost-effectiveness analysis incorporating clinical outcomes and resource utilization should be conducted. In addition, increasing the sample size to establish more detailed reference intervals according to age and sex would improve clinical applicability.</p>
<p>Future research should focus on: (1) Multicenter external validation (target <italic>n</italic> &#x02265; 300) across diverse clinical settings to enhance generalizability; (2) Prospective cohort studies linking monitoring discordances between anti-Xa and APTT to clinical outcomes, specifically bleeding and thrombotic events; (3) Randomized controlled trials comparing anti-Xa-guided vs. APTT-guided anticoagulation strategies to determine optimal monitoring approaches; (4) Adequately powered studies in special populations, including intensive care unit patients, obstetric patients, and oncology patients, to address population-specific anticoagulation challenges; (5) Cost-effectiveness analysis evaluating the economic impact of different monitoring strategies in clinical practice; and (6) Dynamic monitoring protocols examine temporal changes in inflammatory markers, renal function, and coagulation factors to better understand the evolution of anticoagulation responses over time.</p></sec>
<sec sec-type="conclusion" id="s5">
<label>5</label>
<title>Conclusion</title>
<p>This study developed an in-house anti-factor Xa assay with analytical performance comparable to commercial systems. Poor categorical agreement between APTT and anti-factor Xa highlights APTT limitations for UFH monitoring. Ten clinical and laboratory factors potentially associated with monitoring discordance were identified, and a laboratory-specific APTT range (79&#x02013;127 s) was derived.</p>
<p>However, these findings are substantially limited by single-center design, small sample size (<italic>n</italic> = 110), lack of external validation, and absence of clinical outcome data. The predictive models are hypothesis-generating and should not guide clinical practice without rigorous prospective validation. Multicenter studies with adequate sample sizes and clinical endpoint assessment are essential to determine whether these findings translate to improved patient outcomes.</p></sec>
</body>
<back>
<sec sec-type="data-availability" id="s6">
<title>Data availability statement</title>
<p>The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.</p>
</sec>
<sec sec-type="ethics-statement" id="s7">
<title>Ethics statement</title>
<p>The studies involving humans were approved by the Ethics Committee of the Affiliated Hospital of Qingdao University (Approval No. QYFY WZLL 30726). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.</p>
</sec>
<sec sec-type="author-contributions" id="s8">
<title>Author contributions</title>
<p>LS: Data curation, Methodology, Writing &#x02013; original draft, Writing &#x02013; review &#x00026; editing. JZ: Writing &#x02013; review &#x00026; editing. GJ: Methodology, Writing &#x02013; review &#x00026; editing. WL: Methodology, Validation, Writing &#x02013; review &#x00026; editing. HR: Funding acquisition, Methodology, Writing &#x02013; review &#x00026; editing. LL: Writing &#x02013; review &#x00026; editing, Conceptualization, Data curation, Methodology, Supervision. YW: Investigation, Writing &#x02013; review &#x00026; editing. TZ: Writing &#x02013; review &#x00026; editing, Investigation.</p>
</sec>
<ack><title>Acknowledgments</title><p>We thank all patients who participated in this study and the clinical staff at the Affiliated Hospital of Qingdao University for their assistance in sample collection.</p></ack>
<sec sec-type="COI-statement" id="conf1">
<title>Conflict of interest</title>
<p>HR was employed by Beijing ZONCI Technology Development Co., Ltd. The sponsor was involved in study design but had no role in data collection, analysis, interpretation, or manuscript writing.</p>
<p>The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec sec-type="ai-statement" id="s10">
<title>Generative AI statement</title>
<p>The author(s) declared that generative AI was not used in the creation of this manuscript.</p>
<p>Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.</p></sec>
<sec sec-type="disclaimer" id="s11">
<title>Publisher&#x00027;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<ref-list>
<title>References</title>
<ref id="B1">
<label>1.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Favaloro</surname> <given-names>EJ</given-names></name> <name><surname>Pasalic</surname> <given-names>L</given-names></name> <name><surname>Lippi</surname> <given-names>G</given-names></name></person-group>. <article-title>Unfractionated heparin: optimizing laboratory monitoring and reducing unwanted interference in everyday hemostasis test practice</article-title>. <source>Pol Arch Intern Med</source>. (<year>2024</year>) <volume>134</volume>:<fpage>16684</fpage>. doi: <pub-id pub-id-type="doi">10.20452/pamw.16684</pub-id><pub-id pub-id-type="pmid">38380825</pub-id></mixed-citation>
</ref>
<ref id="B2">
<label>2.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Vardi</surname> <given-names>M</given-names></name> <name><surname>Laor</surname> <given-names>A</given-names></name> <name><surname>Bitterman</surname> <given-names>H</given-names></name></person-group>. <article-title>Activated partial thromboplastin time monitoring in patients receiving unfractionated heparin for venous thromboembolism in relation to clinical outcomes</article-title>. <source>Thromb Haemost.</source> (<year>2009</year>) <volume>102</volume>:<fpage>879</fpage>&#x02013;<lpage>86</lpage>. doi: <pub-id pub-id-type="doi">10.1160/TH09-06-0404</pub-id><pub-id pub-id-type="pmid">19888523</pub-id></mixed-citation>
</ref>
<ref id="B3">
<label>3.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Guti&#x000E9;rrez-Chico</surname> <given-names>JL</given-names></name> <name><surname>Mehilli</surname> <given-names>J</given-names></name></person-group>. <article-title>Gender differences in cardiovascular therapy: focus on antithrombotic therapy and percutaneous coronary intervention</article-title>. <source>Drugs.</source> (<year>2013</year>) <volume>73</volume>:<fpage>1921</fpage>&#x02013;<lpage>33</lpage>. doi: <pub-id pub-id-type="doi">10.1007/s40265-013-0135-y</pub-id><pub-id pub-id-type="pmid">24155117</pub-id></mixed-citation>
</ref>
<ref id="B4">
<label>4.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Cuker</surname> <given-names>A</given-names></name> <name><surname>Siegal</surname> <given-names>DM</given-names></name> <name><surname>Crowther</surname> <given-names>MA</given-names></name> <name><surname>Garcia</surname> <given-names>DA</given-names></name></person-group>. <article-title>Laboratory measurement of the anticoagulant activity of the non-vitamin K oral anticoagulants</article-title>. <source>J Am Coll Cardiol.</source> (<year>2014</year>) <volume>64</volume>:<fpage>1128</fpage>&#x02013;<lpage>39</lpage>. doi: <pub-id pub-id-type="doi">10.1016/j.jacc.2014.05.065</pub-id><pub-id pub-id-type="pmid">25212648</pub-id></mixed-citation>
</ref>
<ref id="B5">
<label>5.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Raschke</surname> <given-names>RA</given-names></name> <name><surname>Reilly</surname> <given-names>BM</given-names></name> <name><surname>Guidry</surname> <given-names>JR</given-names></name> <name><surname>Fontana</surname> <given-names>JR</given-names></name> <name><surname>Srinivas</surname> <given-names>S</given-names></name></person-group>. <article-title>The weight-based heparin dosing nomogram compared with a &#x0201C;standard care&#x0201D; nomogram. A randomized controlled trial</article-title>. <source>Ann Intern Med.</source> (<year>1993</year>) <volume>119</volume>:<fpage>874</fpage>&#x02013;<lpage>81</lpage>. doi: <pub-id pub-id-type="doi">10.7326/0003-4819-119-9-199311010-00002</pub-id><pub-id pub-id-type="pmid">8214998</pub-id></mixed-citation>
</ref>
<ref id="B6">
<label>6.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Lawrie</surname> <given-names>AS</given-names></name> <name><surname>Kitchen</surname> <given-names>S</given-names></name> <name><surname>Efthymiou</surname> <given-names>M</given-names></name> <name><surname>Mackie</surname> <given-names>IJ</given-names></name> <name><surname>Machin</surname> <given-names>SJ</given-names></name></person-group>. <article-title>Determination of APTT factor sensitivity&#x02013;the misguiding guideline</article-title>. <source>Int J Lab Hematol.</source> (<year>2013</year>) <volume>35</volume>:<fpage>652</fpage>&#x02013;<lpage>7</lpage>. doi: <pub-id pub-id-type="doi">10.1111/ijlh.12109</pub-id><pub-id pub-id-type="pmid">23718922</pub-id></mixed-citation>
</ref>
<ref id="B7">
<label>7.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Pengo</surname> <given-names>V</given-names></name> <name><surname>Tripodi</surname> <given-names>A</given-names></name> <name><surname>Reber</surname> <given-names>G</given-names></name> <name><surname>Rand</surname> <given-names>JH</given-names></name> <name><surname>Ortel</surname> <given-names>TL</given-names></name> <name><surname>Galli</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Update of the guidelines for lupus anticoagulant detection</article-title>. <source>J Thromb Haemost</source>. (<year>2009</year>) <volume>7</volume>:<fpage>1737</fpage>&#x02013;<lpage>40</lpage>. doi: <pub-id pub-id-type="doi">10.1111/j.1538-7836.2009.03555.x</pub-id></mixed-citation>
</ref>
<ref id="B8">
<label>8.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Marlar</surname> <given-names>RA</given-names></name> <name><surname>Clement</surname> <given-names>B</given-names></name> <name><surname>Gausman</surname> <given-names>J</given-names></name></person-group>. <article-title>Activated partial thromboplastin time monitoring of unfractionated heparin therapy: issues and recommendations</article-title>. <source>Semin Thromb Hemost.</source> (<year>2017</year>) <volume>43</volume>:<fpage>253</fpage>&#x02013;<lpage>60</lpage>. doi: <pub-id pub-id-type="doi">10.1055/s-0036-1581128</pub-id><pub-id pub-id-type="pmid">27272964</pub-id></mixed-citation>
</ref>
<ref id="B9">
<label>9.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Miles</surname> <given-names>LA</given-names></name> <name><surname>Fless</surname> <given-names>GM</given-names></name> <name><surname>Scanu</surname> <given-names>AM</given-names></name> <name><surname>Baynham</surname> <given-names>P</given-names></name> <name><surname>Sebald</surname> <given-names>MT</given-names></name> <name><surname>Skocir</surname> <given-names>P</given-names></name> <etal/></person-group>. <article-title>Interaction of Lp(a) with plasminogen binding sites on cells</article-title>. <source>Thromb Haemost.</source> (<year>1995</year>) <volume>73</volume>:<fpage>458</fpage>&#x02013;<lpage>65</lpage>. doi: <pub-id pub-id-type="doi">10.1055/s-0038-1653797</pub-id><pub-id pub-id-type="pmid">7667829</pub-id></mixed-citation>
</ref>
<ref id="B10">
<label>10.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Lutfi</surname> <given-names>F</given-names></name> <name><surname>Bishnoi</surname> <given-names>R</given-names></name> <name><surname>Patel</surname> <given-names>VJ</given-names></name> <name><surname>Elfasi</surname> <given-names>A</given-names></name> <name><surname>Setteducato</surname> <given-names>M</given-names></name> <name><surname>Zhang</surname> <given-names>S</given-names></name> <etal/></person-group>. <article-title>Bleeding and thrombotic risk in low dose heparin infusion as compared to standard dose heparin infusion</article-title>. <source>Cureus.</source> (<year>2020</year>) <volume>12</volume>:<fpage>e8339</fpage>. doi: <pub-id pub-id-type="doi">10.7759/cureus.8339</pub-id><pub-id pub-id-type="pmid">32617214</pub-id></mixed-citation>
</ref>
<ref id="B11">
<label>11.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Smythe</surname> <given-names>MA</given-names></name> <name><surname>Priziola</surname> <given-names>J</given-names></name> <name><surname>Dobesh</surname> <given-names>PP</given-names></name> <name><surname>Wirth</surname> <given-names>D</given-names></name> <name><surname>Cuker</surname> <given-names>A</given-names></name> <name><surname>Wittkowsky</surname> <given-names>AK</given-names></name></person-group>. <article-title>Guidance for the practical management of the heparin anticoagulants in the treatment of venous thromboembolism</article-title>. <source>J Thromb Thrombolysis.</source> (<year>2016</year>) <volume>41</volume>:<fpage>165</fpage>&#x02013;<lpage>86</lpage>. doi: <pub-id pub-id-type="doi">10.1007/s11239-015-1315-2</pub-id><pub-id pub-id-type="pmid">26780745</pub-id></mixed-citation>
</ref>
<ref id="B12">
<label>12.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Gehrie</surname> <given-names>E</given-names></name> <name><surname>Laposata</surname> <given-names>M</given-names></name></person-group>. <article-title>Test of the month: the chromogenic antifactor Xa assay</article-title>. <source>Am J Hematol.</source> (<year>2012</year>) <volume>87</volume>:<fpage>194</fpage>&#x02013;<lpage>6</lpage>. doi: <pub-id pub-id-type="doi">10.1002/ajh.22222</pub-id><pub-id pub-id-type="pmid">22081480</pub-id></mixed-citation>
</ref>
<ref id="B13">
<label>13.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Takemoto</surname> <given-names>CM</given-names></name> <name><surname>Streiff</surname> <given-names>MB</given-names></name> <name><surname>Shermock</surname> <given-names>KM</given-names></name> <name><surname>Kraus</surname> <given-names>PS</given-names></name> <name><surname>Chen</surname> <given-names>J</given-names></name> <name><surname>Jani</surname> <given-names>J</given-names></name> <etal/></person-group>. <article-title>Activated partial thromboplastin time and anti-Xa measurements in heparin monitoring: biochemical basis for discordance</article-title>. <source>Am J Clin Pathol.</source> (<year>2013</year>) <volume>139</volume>:<fpage>450</fpage>&#x02013;<lpage>6</lpage>. doi: <pub-id pub-id-type="doi">10.1309/AJCPS6OW6DYNOGNH</pub-id><pub-id pub-id-type="pmid">23525615</pub-id></mixed-citation>
</ref>
<ref id="B14">
<label>14.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Vandiver</surname> <given-names>JW</given-names></name> <name><surname>Vondracek</surname> <given-names>TG</given-names></name></person-group>. <article-title>Antifactor Xa levels versus activated partial thromboplastin time for monitoring unfractionated heparin</article-title>. <source>Pharmacotherapy.</source> (<year>2012</year>) <volume>32</volume>:<fpage>546</fpage>&#x02013;<lpage>58</lpage>. doi: <pub-id pub-id-type="doi">10.1002/j.1875-9114.2011.01049.x</pub-id><pub-id pub-id-type="pmid">22531940</pub-id></mixed-citation>
</ref>
<ref id="B15">
<label>15.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Baker</surname> <given-names>P</given-names></name> <name><surname>Platton</surname> <given-names>S</given-names></name> <name><surname>Arachchillage</surname> <given-names>DJ</given-names></name> <name><surname>Kitchen</surname> <given-names>S</given-names></name> <name><surname>Patel</surname> <given-names>J</given-names></name> <name><surname>Riat</surname> <given-names>R</given-names></name> <etal/></person-group>. <article-title>Measurement of heparin, direct oral anti-coagulants and other non-coumarin anti-coagulants and their effects on haemostasis assays: a British Society for Haematology Guideline</article-title>. <source>Br J Haematol.</source> (<year>2024</year>) <volume>205</volume>:<fpage>1302</fpage>&#x02013;<lpage>18</lpage>. doi: <pub-id pub-id-type="doi">10.1111/bjh.19729</pub-id><pub-id pub-id-type="pmid">39223697</pub-id></mixed-citation>
</ref>
<ref id="B16">
<label>16.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Price</surname> <given-names>EA</given-names></name> <name><surname>Jin</surname> <given-names>J</given-names></name> <name><surname>Nguyen</surname> <given-names>HM</given-names></name> <name><surname>Krishnan</surname> <given-names>G</given-names></name> <name><surname>Bowen</surname> <given-names>R</given-names></name> <name><surname>Zehnder</surname> <given-names>JL</given-names></name></person-group>. <article-title>Discordant aPTT and anti-Xa values and outcomes in hospitalized patients treated with intravenous unfractionated heparin</article-title>. <source>Ann Pharmacother.</source> (<year>2013</year>) <volume>47</volume>:<fpage>151</fpage>&#x02013;<lpage>8</lpage>. doi: <pub-id pub-id-type="doi">10.1345/aph.1R635</pub-id><pub-id pub-id-type="pmid">23386070</pub-id></mixed-citation>
</ref>
<ref id="B17">
<label>17.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Scott</surname> <given-names>EJ</given-names></name> <name><surname>Rotar</surname> <given-names>E</given-names></name> <name><surname>Dahl</surname> <given-names>JJ</given-names></name> <name><surname>Beller</surname> <given-names>JP</given-names></name> <name><surname>Money</surname> <given-names>DT</given-names></name> <name><surname>Chancellor</surname> <given-names>WZ</given-names></name> <etal/></person-group>. <article-title>Discordance among assays for monitoring? Anticoagulation during extracorporeal life support</article-title>. <source>Perfusion.</source> (<year>2023</year>) <volume>38</volume>:<fpage>1714</fpage>&#x02013;<lpage>21</lpage>. doi: <pub-id pub-id-type="doi">10.1177/02676591221129741</pub-id><pub-id pub-id-type="pmid">36167522</pub-id></mixed-citation>
</ref>
<ref id="B18">
<label>18.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Olson</surname> <given-names>JD</given-names></name> <name><surname>Arkin</surname> <given-names>CF</given-names></name> <name><surname>Brandt</surname> <given-names>JT</given-names></name> <name><surname>Cunningham</surname> <given-names>MT</given-names></name> <name><surname>Giles</surname> <given-names>A</given-names></name> <name><surname>Koepke</surname> <given-names>JA</given-names></name> <etal/></person-group>. <article-title>College of American Pathologists Conference XXXI on laboratory monitoring of anticoagulant therapy: laboratory monitoring of unfractionated heparin therapy</article-title>. <source>Arch Pathol Lab Med.</source> (<year>1998</year>) <volume>122</volume>:<fpage>782</fpage>&#x02013;<lpage>98</lpage>. <pub-id pub-id-type="pmid">9740136</pub-id></mixed-citation>
</ref>
<ref id="B19">
<label>19.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Tripodi</surname> <given-names>A</given-names></name> <name><surname>Chantarangkul</surname> <given-names>V</given-names></name> <name><surname>Martinelli</surname> <given-names>I</given-names></name> <name><surname>Bucciarelli</surname> <given-names>P</given-names></name> <name><surname>Mannucci</surname> <given-names>PM</given-names></name></person-group>. <article-title>A shortened activated partial thromboplastin time is associated with the risk of venous thromboembolism</article-title>. <source>Blood.</source> (<year>2004</year>) <volume>104</volume>:<fpage>3631</fpage>&#x02013;<lpage>4</lpage>. doi: <pub-id pub-id-type="doi">10.1182/blood-2004-03-1042</pub-id><pub-id pub-id-type="pmid">15297315</pub-id></mixed-citation>
</ref>
<ref id="B20">
<label>20.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>May</surname> <given-names>JE</given-names></name> <name><surname>Siniard</surname> <given-names>RC</given-names></name> <name><surname>Taylor</surname> <given-names>LJ</given-names></name> <name><surname>Marques</surname> <given-names>MB</given-names></name> <name><surname>Gangaraju</surname> <given-names>R</given-names></name></person-group>. <article-title>From activated partial thromboplastin time to antifactor Xa and back again</article-title>. <source>Am J Clin Pathol.</source> (<year>2022</year>) <volume>157</volume>:<fpage>321</fpage>&#x02013;<lpage>7</lpage>. doi: <pub-id pub-id-type="doi">10.1093/ajcp/aqab135</pub-id><pub-id pub-id-type="pmid">34562001</pub-id></mixed-citation>
</ref>
<ref id="B21">
<label>21.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Darst</surname> <given-names>BF</given-names></name> <name><surname>Malecki</surname> <given-names>KC</given-names></name> <name><surname>Engelman</surname> <given-names>CD</given-names></name></person-group>. <article-title>Using recursive feature elimination in random forest to account for correlated variables in high dimensional data</article-title>. <source>BMC Genet</source>. (<year>2018</year>) 19(<supplement>Suppl. 1</supplement>):65. doi: <pub-id pub-id-type="doi">10.1186/s12863-018-0633-8</pub-id><pub-id pub-id-type="pmid">30255764</pub-id></mixed-citation>
</ref>
<ref id="B22">
<label>22.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Esteve-Pastor</surname> <given-names>MA</given-names></name> <name><surname>Rivera-Caravaca</surname> <given-names>JM</given-names></name> <name><surname>Rold&#x000E1;n</surname> <given-names>V</given-names></name> <name><surname>Sanmartin Fern&#x000E1;ndez</surname> <given-names>M</given-names></name> <name><surname>Arribas</surname> <given-names>F</given-names></name> <name><surname>Masjuan</surname> <given-names>J</given-names></name> <etal/></person-group>. <article-title>Predicting performance of the HAS-BLED and ORBIT bleeding risk scores in patients with atrial fibrillation treated with Rivaroxaban: observations from the prospective EMIR Registry</article-title>. <source>Eur Heart J Cardiovasc Pharmacother.</source> (<year>2022</year>) <volume>9</volume>:<fpage>38</fpage>&#x02013;<lpage>46</lpage>. doi: <pub-id pub-id-type="doi">10.1093/ehjcvp/pvac060</pub-id><pub-id pub-id-type="pmid">36318457</pub-id></mixed-citation>
</ref>
<ref id="B23">
<label>23.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Ferroni</surname> <given-names>P</given-names></name> <name><surname>Zanzotto</surname> <given-names>FM</given-names></name> <name><surname>Scarpato</surname> <given-names>N</given-names></name> <name><surname>Riondino</surname> <given-names>S</given-names></name> <name><surname>Guadagni</surname> <given-names>F</given-names></name> <name><surname>Roselli</surname> <given-names>M</given-names></name></person-group>. <article-title>Validation of a machine learning approach for venous thromboembolism risk prediction in oncology</article-title>. <source>Dis Markers.</source> (<year>2017</year>) <volume>2017</volume>:<fpage>8781379</fpage>. doi: <pub-id pub-id-type="doi">10.1155/2017/8781379</pub-id><pub-id pub-id-type="pmid">29104344</pub-id></mixed-citation>
</ref>
<ref id="B24">
<label>24.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Sharabiani</surname> <given-names>A</given-names></name> <name><surname>Bress</surname> <given-names>A</given-names></name> <name><surname>Douzali</surname> <given-names>E</given-names></name> <name><surname>Darabi</surname> <given-names>H</given-names></name></person-group>. <article-title>Revisiting warfarin dosing using machine learning techniques</article-title>. <source>Comput Math Methods Med.</source> (<year>2015</year>) <volume>2015</volume>:<fpage>560108</fpage>. doi: <pub-id pub-id-type="doi">10.1155/2015/560108</pub-id><pub-id pub-id-type="pmid">26146514</pub-id></mixed-citation>
</ref>
<ref id="B25">
<label>25.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>McLaughlin</surname> <given-names>K</given-names></name> <name><surname>Rimsans</surname> <given-names>J</given-names></name> <name><surname>Sylvester</surname> <given-names>KW</given-names></name> <name><surname>Fanikos</surname> <given-names>J</given-names></name> <name><surname>Dorfman</surname> <given-names>DM</given-names></name> <name><surname>Senna</surname> <given-names>P</given-names></name> <etal/></person-group>. <article-title>Evaluation of antifactor-Xa heparin assay and activated partial thromboplastin time values in patients on therapeutic continuous infusion unfractionated heparin therapy</article-title>. <source>Clin Appl Thromb Hemost.</source> (<year>2019</year>) <volume>25</volume>:<fpage>1076029619876030</fpage>. doi: <pub-id pub-id-type="doi">10.1177/1076029619876030</pub-id><pub-id pub-id-type="pmid">31530176</pub-id></mixed-citation>
</ref>
<ref id="B26">
<label>26.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Garcia</surname> <given-names>DA</given-names></name> <name><surname>Baglin</surname> <given-names>TP</given-names></name> <name><surname>Weitz</surname> <given-names>JI</given-names></name> <name><surname>Samama</surname> <given-names>MM</given-names></name></person-group>. <article-title>Parenteral anticoagulants: antithrombotic therapy and prevention of thrombosis, 9th ed: American College of Chest Physicians evidence-based clinical practice guidelines</article-title>. <source>Chest.</source> (<year>2012</year>) 141:e24S&#x02212;43S. doi: <pub-id pub-id-type="doi">10.1378/chest.11-2291</pub-id><pub-id pub-id-type="pmid">22315264</pub-id></mixed-citation>
</ref>
<ref id="B27">
<label>27.</label>
<mixed-citation publication-type="book"><person-group person-group-type="author"><collab>Clinical and Laboratory Standards Institute</collab></person-group>. <source>About CLSI</source>. <publisher-loc>Wayne, PA</publisher-loc>: <publisher-name>Clinical and Laboratory Standards Institute</publisher-name> (<year>2024</year>).</mixed-citation>
</ref>
<ref id="B28">
<label>28.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Westgard</surname> <given-names>JO</given-names></name> <name><surname>Carey</surname> <given-names>RN</given-names></name> <name><surname>Wold</surname> <given-names>S</given-names></name></person-group>. <article-title>Criteria for judging precision and accuracy in method development and evaluation</article-title>. <source>Clin Chem.</source> (<year>1974</year>) <volume>20</volume>:<fpage>825</fpage>&#x02013;<lpage>33</lpage>. doi: <pub-id pub-id-type="doi">10.1093/clinchem/20.7.825</pub-id><pub-id pub-id-type="pmid">4835236</pub-id></mixed-citation>
</ref>
<ref id="B29">
<label>29.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Voloh</surname> <given-names>B</given-names></name> <name><surname>Watson</surname> <given-names>MR</given-names></name> <name><surname>K&#x000F6;nig</surname> <given-names>S</given-names></name> <name><surname>Womelsdorf</surname> <given-names>T</given-names></name></person-group>. <article-title>MAD saccade: statistically robust saccade threshold estimation via the median absolute deviation</article-title>. <source>J Eye Mov Res.</source> (<year>2020</year>) <volume>12</volume>:<fpage>1</fpage>&#x02013;<lpage>12</lpage>. doi: <pub-id pub-id-type="doi">10.16910/jemr.12.8.3</pub-id><pub-id pub-id-type="pmid">33828776</pub-id></mixed-citation>
</ref>
<ref id="B30">
<label>30.</label>
<mixed-citation publication-type="book"><person-group person-group-type="author"><collab>Clinical and Laboratory Standards Institute</collab></person-group>. <source>EP06: Evaluation of Linearity of Quantitative Measurement Procedures</source>. <publisher-loc>Wayne, PA</publisher-loc>: <publisher-name>Clinical and Laboratory Standards Institute</publisher-name> (<year>2020</year>).</mixed-citation>
</ref>
<ref id="B31">
<label>31.</label>
<mixed-citation publication-type="book"><person-group person-group-type="author"><collab>Clinical and Laboratory Standards Institute</collab></person-group>. <source>EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples</source>. <publisher-loc>Wayne, PA</publisher-loc>: <publisher-name>Clinical and Laboratory Standards Institute</publisher-name> (<year>2018</year>).</mixed-citation>
</ref>
<ref id="B32">
<label>32.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Li</surname> <given-names>F</given-names></name> <name><surname>He</surname> <given-names>H</given-names></name></person-group>. <article-title>Assessing the accuracy of diagnostic tests</article-title>. <source>Shanghai Arch Psychiatry.</source> (<year>2018</year>) <volume>30</volume>:<fpage>207</fpage>&#x02013;<lpage>12</lpage>. doi: <pub-id pub-id-type="doi">10.11919/j.issn.1002-0829.218052</pub-id></mixed-citation>
</ref>
<ref id="B33">
<label>33.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Li</surname> <given-names>S</given-names></name> <name><surname>Li</surname> <given-names>A</given-names></name> <name><surname>Beckman</surname> <given-names>JA</given-names></name> <name><surname>Kim</surname> <given-names>C</given-names></name> <name><surname>Granich</surname> <given-names>MA</given-names></name> <name><surname>Mondin</surname> <given-names>J</given-names></name> <etal/></person-group>. <article-title>Discordance between aPTT and anti-Xa in monitoring heparin anticoagulation in mechanical circulatory support</article-title>. <source>ESC Heart Fail.</source> (<year>2024</year>) <volume>11</volume>:<fpage>2742</fpage>&#x02013;<lpage>8</lpage>. doi: <pub-id pub-id-type="doi">10.1002/ehf2.14816</pub-id><pub-id pub-id-type="pmid">38725133</pub-id></mixed-citation>
</ref>
<ref id="B34">
<label>34.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Toulon</surname> <given-names>P</given-names></name> <name><surname>Smahi</surname> <given-names>M</given-names></name> <name><surname>De Pooter</surname> <given-names>N</given-names></name></person-group>. <article-title>APTT therapeutic range for monitoring unfractionated heparin therapy. Significant impact of the anti-Xa reagent used for correlation</article-title>. <source>J Thromb Haemost.</source> (<year>2021</year>) <volume>19</volume>:<fpage>2002</fpage>&#x02013;<lpage>6</lpage>. doi: <pub-id pub-id-type="doi">10.1111/jth.15264</pub-id><pub-id pub-id-type="pmid">33555096</pub-id></mixed-citation>
</ref>
<ref id="B35">
<label>35.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>B&#x000FC;rki</surname> <given-names>S</given-names></name> <name><surname>Brand</surname> <given-names>B</given-names></name> <name><surname>Escher</surname> <given-names>R</given-names></name> <name><surname>Wuillemin</surname> <given-names>WA</given-names></name> <name><surname>Nagler</surname> <given-names>M</given-names></name></person-group>. <article-title>Accuracy, reproducibility and costs of different laboratory assays for the monitoring of unfractionated heparin in clinical practice: a prospective evaluation study and survey among Swiss institutions</article-title>. <source>BMJ Open.</source> (<year>2018</year>) <volume>8</volume>:<fpage>e022943</fpage>. doi: <pub-id pub-id-type="doi">10.1136/bmjopen-2018-022943</pub-id><pub-id pub-id-type="pmid">29886450</pub-id></mixed-citation>
</ref>
<ref id="B36">
<label>36.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Calabria</surname> <given-names>Gallego MD</given-names></name></person-group>. <article-title>Cuadernos de bioetica: revista oficial de la Asociacion Espanola de Bioetica y Etica Medica</article-title>. [Euthanasia: a wrong solution to an unrecognized problem?] <italic>Cuad Bioet</italic>. (<year>2023</year>) <volume>34</volume>:<fpage>325</fpage>&#x02013;<lpage>30</lpage>. Spanish. doi: <pub-id pub-id-type="doi">10.30444/CB.160</pub-id></mixed-citation>
</ref>
<ref id="B37">
<label>37.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Newall</surname> <given-names>F</given-names></name></person-group>. <article-title>Anti-factor Xa (anti-Xa) assay</article-title>. <source>Methods Mol Biol.</source> (<year>2013</year>) <volume>992</volume>:<fpage>265</fpage>&#x02013;<lpage>72</lpage>. doi: <pub-id pub-id-type="doi">10.1007/978-1-62703-339-8_19</pub-id></mixed-citation>
</ref>
<ref id="B38">
<label>38.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Beyer</surname> <given-names>J</given-names></name> <name><surname>Trujillo</surname> <given-names>T</given-names></name> <name><surname>Fisher</surname> <given-names>S</given-names></name> <name><surname>Ko</surname> <given-names>A</given-names></name> <name><surname>Lind</surname> <given-names>SE</given-names></name> <name><surname>Kiser</surname> <given-names>TH</given-names></name></person-group>. <article-title>Evaluation of a heparin-calibrated antifactor Xa Assay for measuring the anticoagulant effect of oral direct Xa inhibitors</article-title>. <source>Clin Appl Thromb Hemost.</source> (<year>2016</year>) <volume>22</volume>:<fpage>423</fpage>&#x02013;<lpage>8</lpage>. doi: <pub-id pub-id-type="doi">10.1177/1076029616629759</pub-id><pub-id pub-id-type="pmid">26842561</pub-id></mixed-citation>
</ref>
<ref id="B39">
<label>39.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Kaplan</surname> <given-names>M</given-names></name> <name><surname>Ate&#x0015F;</surname> <given-names>I</given-names></name> <name><surname>Akdogan Kayhan</surname> <given-names>M</given-names></name> <name><surname>Ka&#x000E7;ar</surname> <given-names>S</given-names></name> <name><surname>G&#x000F6;kbulut</surname> <given-names>V</given-names></name> <name><surname>Co&#x0015F;kun</surname> <given-names>O</given-names></name> <etal/></person-group>. <article-title>Diagnostic utility of oxidative and non-oxidative markers for spontaneous bacterial peritonitis in non-malign ascites</article-title>. <source>Acta Gastroenterol Belg.</source> (<year>2020</year>) <volume>83</volume>:<fpage>279</fpage>&#x02013;<lpage>84</lpage>. <pub-id pub-id-type="pmid">32603047</pub-id></mixed-citation>
</ref>
<ref id="B40">
<label>40.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Barrett</surname> <given-names>YC</given-names></name> <name><surname>Wang</surname> <given-names>Z</given-names></name> <name><surname>Frost</surname> <given-names>C</given-names></name> <name><surname>Shenker</surname> <given-names>A</given-names></name></person-group>. <article-title>Clinical laboratory measurement of direct factor Xa inhibitors: anti-Xa assay is preferable to prothrombin time assay</article-title>. <source>Thromb Haemost.</source> (<year>2010</year>) <volume>104</volume>:<fpage>1263</fpage>&#x02013;<lpage>71</lpage>. doi: <pub-id pub-id-type="doi">10.1160/TH10-05-0328</pub-id><pub-id pub-id-type="pmid">20978714</pub-id></mixed-citation>
</ref>
<ref id="B41">
<label>41.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Samuel</surname> <given-names>S</given-names></name> <name><surname>Allison</surname> <given-names>TA</given-names></name> <name><surname>Sharaf</surname> <given-names>S</given-names></name> <name><surname>Yau</surname> <given-names>G</given-names></name> <name><surname>Ranjbar</surname> <given-names>G</given-names></name> <name><surname>Mckaig</surname> <given-names>N</given-names></name> <etal/></person-group>. <article-title>Antifactor Xa levels vs. activated partial thromboplastin time for monitoring unfractionated heparin. A pilot study</article-title>. <source>J Clin Pharm Ther</source>. (<year>2016</year>) <volume>41</volume>:<fpage>499</fpage>&#x02013;<lpage>502</lpage>. doi: <pub-id pub-id-type="doi">10.1111/jcpt.12415</pub-id><pub-id pub-id-type="pmid">27381025</pub-id></mixed-citation>
</ref>
<ref id="B42">
<label>42.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Whitman-Purves</surname> <given-names>E</given-names></name> <name><surname>Coons</surname> <given-names>JC</given-names></name> <name><surname>Miller</surname> <given-names>T</given-names></name> <name><surname>DiNella</surname> <given-names>JV</given-names></name> <name><surname>Althouse</surname> <given-names>A</given-names></name> <name><surname>Schmidhofer</surname> <given-names>M</given-names></name> <etal/></person-group>. <article-title>Performance of anti-factor Xa versus activated partial thromboplastin time for heparin monitoring using multiple nomograms</article-title>. <source>Clin Appl Thromb Hemost.</source> (<year>2018</year>) <volume>24</volume>:<fpage>310</fpage>&#x02013;<lpage>6</lpage>. doi: <pub-id pub-id-type="doi">10.1177/1076029617741363</pub-id><pub-id pub-id-type="pmid">29212374</pub-id></mixed-citation>
</ref>
<ref id="B43">
<label>43.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Lawlor</surname> <given-names>M</given-names></name> <name><surname>Gupta</surname> <given-names>A</given-names></name> <name><surname>Ranard</surname> <given-names>LS</given-names></name> <name><surname>Madhavan</surname> <given-names>MV</given-names></name> <name><surname>Li</surname> <given-names>J</given-names></name> <name><surname>Eisenberger</surname> <given-names>A</given-names></name> <etal/></person-group>. <article-title>Discordance in activated partial thromboplastin time and anti-factor Xa levels in COVID-19 patients on heparin therapy</article-title>. <source>Thromb Res.</source> (<year>2021</year>) <volume>198</volume>:<fpage>79</fpage>&#x02013;<lpage>82</lpage>. doi: <pub-id pub-id-type="doi">10.1016/j.thromres.2020.11.030</pub-id><pub-id pub-id-type="pmid">33302210</pub-id></mixed-citation>
</ref>
<ref id="B44">
<label>44.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Belle</surname> <given-names>L</given-names></name> <name><surname>Cayla</surname> <given-names>G</given-names></name> <name><surname>Cottin</surname> <given-names>Y</given-names></name> <name><surname>Coste</surname> <given-names>P</given-names></name> <name><surname>Khalife</surname> <given-names>K</given-names></name> <name><surname>Lab&#x000E8;que</surname> <given-names>JN</given-names></name> <etal/></person-group>. <article-title>French registry on acute ST-elevation and non-ST-elevation myocardial infarction 2015 (FAST-MI 2015). Design and baseline data</article-title>. <source>Arch Cardiovasc Dis.</source> (<year>2017</year>) <volume>110</volume>:<fpage>366</fpage>&#x02013;<lpage>78</lpage>. doi: <pub-id pub-id-type="doi">10.1016/j.acvd.2017.05.001</pub-id><pub-id pub-id-type="pmid">28647465</pub-id></mixed-citation>
</ref>
<ref id="B45">
<label>45.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Liu</surname> <given-names>Q</given-names></name> <name><surname>Huang</surname> <given-names>Y</given-names></name> <name><surname>Zhang</surname> <given-names>R</given-names></name> <name><surname>Cai</surname> <given-names>T</given-names></name> <name><surname>Cai</surname> <given-names>Y</given-names></name></person-group>. <article-title>Medical application of <italic>Spirulina platensis</italic> derived C-phycocyanin</article-title>. <source>Evid Based Complement Alternat Med.</source> (<year>2016</year>) <volume>2016</volume>:<fpage>7803846</fpage>. doi: <pub-id pub-id-type="doi">10.1155/2016/7803846</pub-id><pub-id pub-id-type="pmid">27293463</pub-id></mixed-citation>
</ref>
<ref id="B46">
<label>46.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Zhang</surname> <given-names>H</given-names></name> <name><surname>Fu</surname> <given-names>X</given-names></name> <name><surname>Wu</surname> <given-names>X</given-names></name> <name><surname>Deng</surname> <given-names>Q</given-names></name> <name><surname>Tang</surname> <given-names>K</given-names></name> <name><surname>Zhang</surname> <given-names>L</given-names></name> <etal/></person-group>. <article-title>Using mercury stable isotopes to quantify bidirectional water-atmosphere hg exchange fluxes and explore controlling factors</article-title>. <source>Environ Sci Technol.</source> (<year>2023</year>) <volume>57</volume>:<fpage>10673</fpage>&#x02013;<lpage>85</lpage>. doi: <pub-id pub-id-type="doi">10.1021/acs.est.3c01273</pub-id></mixed-citation>
</ref>
<ref id="B47">
<label>47.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Wainwright</surname> <given-names>MS</given-names></name></person-group>. <article-title>&#x0201C;Clinician, heal thyself&#x0201D;: decisional conflict, value judgments, and their role in practice variation in traumatic brain injury</article-title>. <source>Pediatr Crit Care Med.</source> (<year>2019</year>) <volume>20</volume>:<fpage>679</fpage>&#x02013;<lpage>80</lpage>. doi: <pub-id pub-id-type="doi">10.1097/PCC.0000000000001955</pub-id><pub-id pub-id-type="pmid">31274795</pub-id></mixed-citation>
</ref>
<ref id="B48">
<label>48.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Levi</surname> <given-names>M</given-names></name> <name><surname>Van Der Poll</surname> <given-names>T</given-names></name></person-group>. <article-title>Inflammation and coagulation</article-title>. <source>Crit Care Med.</source> (<year>2010</year>) <volume>38</volume>:<fpage>S26</fpage>&#x02013;<lpage>34</lpage>. doi: <pub-id pub-id-type="doi">10.1097/CCM.0b013e3181c98d21</pub-id></mixed-citation>
</ref>
<ref id="B49">
<label>49.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Yang</surname> <given-names>H</given-names></name> <name><surname>Zhao</surname> <given-names>Y</given-names></name> <name><surname>Guo</surname> <given-names>S</given-names></name> <name><surname>He</surname> <given-names>J</given-names></name> <name><surname>Peng</surname> <given-names>L</given-names></name> <name><surname>Wang</surname> <given-names>Y</given-names></name> <etal/></person-group>. <article-title>Effects of hemolysis on routine coagulation tests when tested on an optical coagulation analyzer</article-title>. <source>Clin Lab</source>. (<year>2023</year>) <volume>69</volume>:<fpage>1</fpage>&#x02013;<lpage>2</lpage>. doi: <pub-id pub-id-type="doi">10.7754/Clin.Lab.2023.220803</pub-id><pub-id pub-id-type="pmid">37560849</pub-id></mixed-citation>
</ref>
<ref id="B50">
<label>50.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Ogundimu</surname> <given-names>EO</given-names></name> <name><surname>Altman</surname> <given-names>DG</given-names></name> <name><surname>Collins</surname> <given-names>GS</given-names></name></person-group>. <article-title>Adequate sample size for developing prediction models is not simply related to events per variable</article-title>. <source>J Clin Epidemiol.</source> (<year>2016</year>) <volume>76</volume>:<fpage>175</fpage>&#x02013;<lpage>82</lpage>. doi: <pub-id pub-id-type="doi">10.1016/j.jclinepi.2016.02.031</pub-id><pub-id pub-id-type="pmid">26964707</pub-id></mixed-citation>
</ref>
<ref id="B51">
<label>51.</label>
<mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Riley</surname> <given-names>RD</given-names></name> <name><surname>Snell</surname> <given-names>KI</given-names></name> <name><surname>Ensor</surname> <given-names>J</given-names></name> <name><surname>Burke</surname> <given-names>DL</given-names></name> <name><surname>Harrell FE</surname> <given-names>Jr</given-names></name> <name><surname>Moons</surname> <given-names>KG</given-names></name> <etal/></person-group>. <article-title>Minimum sample size for developing a multivariable prediction model: PART II&#x02014;binary and time-to-event outcomes</article-title>. <source>Stat Med.</source> (<year>2019</year>) <volume>38</volume>:<fpage>1276</fpage>&#x02013;<lpage>96</lpage>. doi: <pub-id pub-id-type="doi">10.1002/sim.7992</pub-id></mixed-citation>
</ref>
</ref-list>
<fn-group>
<fn fn-type="custom" custom-type="edited-by" id="fn0001">
<p>Edited by: <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/313142/overview">Baojun Wu</ext-link>, Henry Ford Health System, United States</p>
</fn>
<fn fn-type="custom" custom-type="reviewed-by" id="fn0002">
<p>Reviewed by: <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/601986/overview">Walter Jeske</ext-link>, Loyola University Chicago, United States</p>
<p><ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/3274470/overview">Tiziano Martini</ext-link>, AOU Citt&#x000E0; della Salute e della Scienza, Centro Malattie Emorragiche Congenite, Italy</p>
</fn>
</fn-group>
<fn-group>
<fn fn-type="abbr" id="abbr1"><label>Abbreviations:</label><p>ALB, albumin; APTT, activated partial thromboplastin time; AT-III, antithrombin III; AUC, area under the curve; BMI, body mass index; BSA, Bovine serum albumin; CI, confidence interval; CLSI, Clinical and Laboratory Standards Institute; CRP, C-reactive protein; CV, coefficient of variation; eGFR, estimated glomerular filtration rate; Fg, fibrinogen; Hb, hemoglobin; ICU, intensive care unit; IQR, interquartile range; LoB, limit of blank; LoD, limit of detection; LoQ, limit of quantitation; MAD, median absolute deviation; NPP, normal pooled plasma; OR, odds ratio; PEG, polyethylene glycol; POD, probability of detection; PPP, platelet-poor plasma; QC, quality control; R1, reagent 1; R2, reagent 2; RF, random forest; ROC, receiver operating characteristic; SD, standard deviation; TC, total cholesterol; TE, total error; TG, triglycerides; UFH, unfractionated heparin; VIF, variance inflation factor.</p></fn></fn-group>
</back>
</article>