<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Publishing DTD v1.3 20210610//EN" "JATS-journalpublishing1-3-mathml3.dtd">
<article article-type="research-article" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:ali="http://www.niso.org/schemas/ali/1.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" dtd-version="1.3" xml:lang="EN">
<front>
<journal-meta>
<journal-id journal-id-type="publisher-id">Front. Med. Technol.</journal-id><journal-title-group>
<journal-title>Frontiers in Medical Technology</journal-title>
<abbrev-journal-title abbrev-type="pubmed">Front. Med. Technol.</abbrev-journal-title></journal-title-group>
<issn pub-type="epub">2673-3129</issn>
<publisher>
<publisher-name>Frontiers Media S.A.</publisher-name>
</publisher>
</journal-meta>
<article-meta>
<article-id pub-id-type="doi">10.3389/fmedt.2025.1748577</article-id>
<article-version article-version-type="Version of Record" vocab="NISO-RP-8-2008"/>
<article-categories>
<subj-group subj-group-type="heading">
<subject>Original Research</subject>
</subj-group>
</article-categories>
<title-group>
<article-title>Ventricular suction detection algorithm designed for ventricular assist devices</article-title>
</title-group>
<contrib-group>
<contrib contrib-type="author" equal-contrib="yes"><name><surname>Wu</surname><given-names>Yijiao</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="author-notes" rid="an1"><sup>&#x2020;</sup></xref><uri xlink:href="https://loop.frontiersin.org/people/3283493/overview"/><role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; original draft" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-original-draft/">Writing &#x2013; original draft</role></contrib>
<contrib contrib-type="author" equal-contrib="yes"><name><surname>Yang</surname><given-names>Yuzhuo</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="author-notes" rid="an1"><sup>&#x2020;</sup></xref><role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &#x0026; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x0026; editing</role></contrib>
<contrib contrib-type="author"><name><surname>Pan</surname><given-names>Xudong</given-names></name>
<xref ref-type="aff" rid="aff2"><sup>2</sup></xref><role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &#x0026; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x0026; editing</role></contrib>
<contrib contrib-type="author" corresp="yes"><name><surname>Yu</surname><given-names>Shunzhou</given-names></name>
<xref ref-type="aff" rid="aff1"><sup>1</sup></xref>
<xref ref-type="corresp" rid="cor1">&#x002A;</xref><role vocab="credit" vocab-identifier="https://credit.niso.org/" vocab-term="Writing &#x2013; review &#x0026; editing" vocab-term-identifier="https://credit.niso.org/contributor-roles/writing-review-editing/">Writing &#x2013; review &#x0026; editing</role></contrib>
</contrib-group>
<aff id="aff1"><label>1</label><institution>ShenZhen Core Technology Co., Ltd.</institution>, <city>ShenZhen</city>, <country country="cn">China</country></aff>
<aff id="aff2"><label>2</label><institution>School of Mechatronics Engineering, Harbin Institute of Technology</institution>, <city>Harbin</city>, <country country="cn">China</country></aff>
<author-notes>
<corresp id="cor1"><label>&#x002A;</label><bold>Correspondence:</bold> Shunzhou Yu <email xlink:href="mailto:yushunzhou@coretechmed.com">yushunzhou@coretechmed.com</email></corresp>
<fn fn-type="equal" id="an1"><label>&#x2020;</label><p>These authors have contributed equally to this work</p></fn>
</author-notes>
<pub-date publication-format="electronic" date-type="pub" iso-8601-date="2026-01-13"><day>13</day><month>01</month><year>2026</year></pub-date>
<pub-date publication-format="electronic" date-type="collection"><year>2025</year></pub-date>
<volume>7</volume><elocation-id>1748577</elocation-id>
<history>
<date date-type="received"><day>18</day><month>11</month><year>2025</year></date>
<date date-type="rev-recd"><day>15</day><month>12</month><year>2025</year></date>
<date date-type="accepted"><day>23</day><month>12</month><year>2025</year></date>
</history>
<permissions>
<copyright-statement>&#x00A9; 2026 Wu, Yang, Pan and Yu.</copyright-statement>
<copyright-year>2026</copyright-year><copyright-holder>Wu, Yang, Pan and Yu</copyright-holder><license><ali:license_ref start_date="2026-01-13">https://creativecommons.org/licenses/by/4.0/</ali:license_ref><license-p>This is an open-access article distributed under the terms of the <ext-link ext-link-type="uri" xlink:href="https://creativecommons.org/licenses/by/4.0/">Creative Commons Attribution License (CC BY)</ext-link>. The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.</license-p></license>
</permissions>
<abstract><sec><title>Background</title>
<p>Ventricular assist devices (VADs) are an effective treatment for end-stage heart failure and can significantly improve patients&#x0027; quality of life. However, when the rotational speed of the VAD does not match the intraventricular blood volume, ventricular suction may occur. Severe suction can lead to ventricular collapse, making accurate and real-time suction detection critically important.</p>
</sec><sec><title>Methods</title>
<p>Two statistical features and two frequency-domain features were extracted from the pump flow signal to build a classification and regression tree (CART) model. Additionally, a secondary decision-making process was applied using a time-domain threshold.</p>
</sec><sec><title>Results</title>
<p>The proposed method was validated using both <italic>in vivo</italic> and <italic>in vitro</italic> experimental data. Experimental results show that, compared to existing suction detection techniques, the proposed approach not only reduces computational complexity but also achieves higher detection accuracy and enhanced algorithmic stability.</p>
</sec><sec><title>Conclusions</title>
<p>The proposed method provides a more efficient and reliable solution for real-time ventricular suction detection, which is crucial for the safe operation of VADs in clinical settings.</p>
</sec>
</abstract>
<kwd-group>
<kwd>classification model</kwd>
<kwd>suction</kwd>
<kwd>suction detection</kwd>
<kwd>ventricular assist device</kwd>
<kwd>ventricular suction</kwd>
</kwd-group><funding-group><funding-statement>The author(s) declared that financial support was received for this work and/or its publication. This research was funded by the Key Technology Research and Development Project for the Miniaturized Centrifugal Magnetic Levitation Artificial Heart, grant number KOTD20210811090222040 and The APC was funded by ShenZhen Core Technology CO.</funding-statement></funding-group><counts>
<fig-count count="9"/>
<table-count count="3"/><equation-count count="9"/><ref-count count="18"/><page-count count="11"/><word-count count="54864"/></counts><custom-meta-group><custom-meta><meta-name>section-at-acceptance</meta-name><meta-value>Medtech Data Analytics</meta-value></custom-meta></custom-meta-group>
</article-meta>
</front>
<body><sec id="s1" sec-type="intro"><label>1</label><title>Introduction</title>
<p>Ventricular assist device (VAD), as an important mechanical circulatory support therapy, provides essential cardiac support for heart failure patients in the context of a shortage of heart donors, significantly improving patient survival rates and quality of life (<xref ref-type="bibr" rid="B1">1</xref>, <xref ref-type="bibr" rid="B2">2</xref>). However, the management of a VAD in clinical practice still faces several challenges. There are several conditions that can lead to ventricular suction: pericardial tamponade increases the pressure outside the heart, limiting the expansion of the ventricles (<xref ref-type="bibr" rid="B3">3</xref>), which leads to an increase in ventricular pressure and subsequently triggers suction; Abnormal positioning of the inflow cannula can also lead to suction: left ventricular apical aneurysm, extensive calcification at the left ventricular apex, and scars or patches from previous surgeries such as ventricular aneurysmectomy may prevent the inflow cannula from being properly positioned against the mitral valve. Ventricular suction is one such severe and potentially fatal complication, which may lead to ventricular collapse and even threaten the patient&#x0027;s life (<xref ref-type="bibr" rid="B4">4</xref>, <xref ref-type="bibr" rid="B5">5</xref>). Although many existing studies focus on preventing ventricular suction through active control methods, most of these approaches rely on invasive sensors (<xref ref-type="bibr" rid="B6">6</xref>, <xref ref-type="bibr" rid="B7">7</xref>). The long-term reliability of invasive sensors in a VAD faces challenges, including potential risks such as drift, which hinders the application of these active control methods in practical products. Therefore, developing an accurate, timely, and non-invasive suction detection method is of great significance.</p>
<p>Many scholars have previously conducted research on ventricular suction recognition methods, with the primary differences lying in the selection of classification models and suction features. The classification model, serving as a decision-making tool, is responsible for determining whether a suction event occurs based on the features. Suction features, as input data to the classification model, significantly influence the model&#x0027;s performance and accuracy. Therefore, the classification model and suction features must work in close coordination to ensure the accuracy and reliability of suction detection. With the continuous development of classification model research, current classification methods have become well-established. To further improve the accuracy of suction detection methods, it is necessary to focus on feature selection, opting for features with higher discriminative power and greater computational efficiency.</p>
<p>Studies (<xref ref-type="bibr" rid="B8">8</xref>, <xref ref-type="bibr" rid="B9">9</xref>) used the same suction features, which span time-domain, frequency-domain, and time-frequency domain characteristics. While the time-frequency and frequency-domain features can capture detailed signal information from multiple perspectives and provide a deeper analysis of suction events, their high computational complexity makes them unsuitable for real-time applications. Under the constraints of low power consumption and compact size in VAD control systems, these features may cause response delays, impacting the timely detection and feedback of ventricular suction.</p>
<p>Study (<xref ref-type="bibr" rid="B10">10</xref>) proposed a time-domain suction detection index based on the pump flow derivative. Studies (<xref ref-type="bibr" rid="B11">11</xref>, <xref ref-type="bibr" rid="B12">12</xref>) extracted time-domain features from signals such as pump flow and rotational speed for suction detection. While time-domain features are computationally simple, relying solely on them often struggles to address the complex variations caused by individual differences among patients. Variations in cardiac function among different patients result in distinct patterns of ventricular suction signals. A single time-domain feature may not effectively eliminate this interference, leading to poor model performance in individualized patients and inaccurate suction event detection.</p>
<p>In addition to the aforementioned points, current research on sensorless physiological control for ventricular assist devices has also investigated suction detection. Study (<xref ref-type="bibr" rid="B13">13</xref>) predicts the preload value based on a Convolutional Neural Network (CNN), and then considers suction to have occurred when the preload value is 0. This method addresses the core nature of the suction problem by using preload for detection; however, running a deep Convolutional Neural Network requires relatively high computational resources and processing power.</p>
</sec>
<sec id="s2" sec-type="methods"><label>2</label><title>Materials and methods</title>
<sec id="s2a"><label>2.1</label><title>Data acquisition and suction definition</title>
<p>The patient data in this study consists of a single case, where suction occurred during pVAD support due to insufficient volume. Clinical imaging revealed intermittent blockage of the inlet window. The sampling frequency used is 50&#x2005;Hz. The inclusion criterion for suction is a sudden drop in pump flow and ultrasound image. Based on this criterion, manual labeling was performed by clinical medical personnel.</p>
<p>The definitions of suction and non-suction states in this study are shown in <xref ref-type="fig" rid="F1">Figure&#x00A0;1</xref>. As seen in the figure, suction events are marked by a significant decrease in flow rate compared to the non-suction state (<xref ref-type="bibr" rid="B4">4</xref>). The pump flow data used in this study is divided into two parts: patient data and test data. The patient data was selected by medical researchers from the pump operation data of actual VAD patients, including both suction and non-suction state data. The test data consists of data from both an <italic>in vitro</italic> circulation simulation platform and <italic>in vivo</italic> data from animal experiments.</p>
<fig id="F1" position="float"><label>Figure&#x00A0;1</label>
<caption><p>Typical VAD pump flow: stable Non-suction (top) and irregular suction (bottom).</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g001.tif"><alt-text content-type="machine-generated">Two line graphs show pump flow over time in milliliters per minute. The top graph, labeled \"pump flow (no suction),\" shows a consistent flow ranging from 3500 to 4500 ml/min across 100 seconds. The bottom graph, labeled \"pump flow (suction),\" displays variable flow with spikes and drops ranging from nearly 0 to 5000 ml/min, also over 100 seconds.</alt-text>
</graphic>
</fig>
</sec>
<sec id="s2b"><label>2.2</label><title>Algorithmic procedure</title>
<p><xref ref-type="fig" rid="F2">Figure&#x00A0;2</xref> presents the flowchart of the suction detection algorithm used in this study. The algorithm mainly consists of two components: CART-based classification and time-domain threshold-based refinement.</p>
<fig id="F2" position="float"><label>Figure&#x00A0;2</label>
<caption><p>Flowchart of the suction detection algorithm.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g002.tif"><alt-text content-type="machine-generated">Flowchart depicting pump flow feature extraction and suction detection. It uses frequency, statistical, and time domain features, followed by CART training. The process leads to preliminary suction detection, distinguished by thresholds determining suction presence.</alt-text>
</graphic>
</fig>
<p>The CART model is used for preliminary suction identification of the pump flow data, detecting anomalies. Then, thresholding is applied for further classification to identify pump flow signals with flow rate drop characteristics. If state A is identified as suction by the CART model, but its <italic>Z</italic>-value is within the safe range (<italic>Z</italic>&#x2009;&#x003C;&#x2009;0.25), it indicates that the characteristic feature of suction&#x2014;sharp flow decrease&#x2014;is not present. Therefore, the identification result for state A is classified as non-suction. If state A is identified as non-suction by the CART model, it suggests that the frequency-domain behavior is stable and there are no anomalies in the statistical distribution of the data, thus it is considered a non-suction state.</p>
<p>Next, this paper will introduce how to implement the four suction features and the decision threshold, and demonstrate their effectiveness in distinguishing between suction and non-suction states from both theoretical and practical perspectives.</p>
</sec>
<sec id="s2c"><label>2.3</label><title>Suction features and decision threshold</title>
<p>First, the continuous pump flow time-series signal needs to be segmented into appropriately sized segments using a sliding time window. The longer the window length, the more feature information can be extracted, thereby improving the accuracy of detection. However, a longer window also leads to higher latency, which affects real-time performance. Since the normal cardiac cycle lasts between 0.2 and 2&#x2005;s, there can be at least 2.5 cardiac cycles within a 5-s window. Therefore, in this study, the length of each sliding window is set to 5&#x2005;s, with a step size of 1&#x2005;s, balancing both accuracy and real-time performance.</p>
<p>To avoid potential temporal leakage, the data partitioning was conducted before any window segmentation. The patient dataset consists of a single continuous recording of approximately 1&#x2005;h and 30&#x2005;min. The raw time series was first divided into two non-overlapping segments: 60&#x0025; of the data was used to generate the training dataset, and the remaining 40&#x0025; was used to generate the testing dataset. Sliding-window segmentation (5-s window, 1-s step) was then applied separately within each of these two partitions.</p>
<p>This paper designs four suction features (SF) for the CART model based on the characteristic of a significant flow decrease during suction events. Among these, SF1 and SF2 are statistical features, while SF3 and SF4 are frequency-domain features.<disp-formula id="disp-formula1"><label>(1)</label><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM1"><mml:mi>S</mml:mi><mml:mi>F</mml:mi><mml:mn>1</mml:mn><mml:mspace width="0.25em"/><mml:mo>=</mml:mo><mml:mspace width="0.25em"/><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>E</mml:mi><mml:mo stretchy="false">[</mml:mo><mml:mrow><mml:msup><mml:mrow><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mi>i</mml:mi><mml:mo>&#x2212;</mml:mo><mml:mspace width="0.25em"/><mml:mi>m</mml:mi><mml:mi>e</mml:mi><mml:mi>a</mml:mi><mml:mi>n</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mn>3</mml:mn></mml:msup></mml:mrow><mml:mo stretchy="false">]</mml:mo></mml:mrow><mml:mrow><mml:msup><mml:mi>&#x03B4;</mml:mi><mml:mn>3</mml:mn></mml:msup></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math></disp-formula></p>
<p>The skewness (<xref ref-type="disp-formula" rid="disp-formula1">Equation 1</xref>) of the pump flow (SF1) can be used to measure the asymmetry of the data.<disp-formula id="disp-formula2"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM2"><mml:mi>S</mml:mi><mml:mi>F</mml:mi><mml:mn>2</mml:mn><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mn>1</mml:mn><mml:mi>n</mml:mi></mml:mfrac></mml:mrow><mml:munderover><mml:mo movablelimits="false">&#x2211;</mml:mo><mml:mrow><mml:mi>i</mml:mi><mml:mo>=</mml:mo><mml:mn>1</mml:mn></mml:mrow><mml:mi>n</mml:mi></mml:munderover><mml:msup><mml:mrow><mml:mrow><mml:mo>(</mml:mo><mml:mrow><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mi>i</mml:mi><mml:mo>&#x2212;</mml:mo><mml:mi>m</mml:mi><mml:mi>e</mml:mi><mml:mi>a</mml:mi><mml:mi>n</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mi>s</mml:mi></mml:mfrac></mml:mrow></mml:mstyle></mml:mrow><mml:mo>)</mml:mo></mml:mrow></mml:mrow><mml:mn>4</mml:mn></mml:msup><mml:mo>&#x2212;</mml:mo><mml:mn>3</mml:mn></mml:mstyle></mml:math><label>(2)</label></disp-formula>Kurtosis (<xref ref-type="disp-formula" rid="disp-formula2">Equation 2</xref>) (SF2) describes the distribution shape of the flow data. When suction occurs, the pump flow suddenly decreases, causing a change in the tail of the data distribution. From a probability distribution perspective, the data corresponding to the sharp flow drop becomes more deviated from the mean flow, making the distribution more peaked. As a result, the kurtosis value increases compared to the non-suction state.<disp-formula id="disp-formula3"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM3"><mml:mi>S</mml:mi><mml:mi>F</mml:mi><mml:mn>3</mml:mn><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:msub><mml:mo movablelimits="false">&#x2211;</mml:mo><mml:mi>f</mml:mi></mml:msub><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mo>&#x002A;</mml:mo><mml:msup><mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mi>f</mml:mi><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo></mml:mrow><mml:mn>2</mml:mn></mml:msup></mml:mrow></mml:mrow><mml:mrow><mml:msub><mml:mo movablelimits="false">&#x2211;</mml:mo><mml:mi>f</mml:mi></mml:msub><mml:mrow><mml:msup><mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mi>f</mml:mi><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo></mml:mrow><mml:mn>2</mml:mn></mml:msup></mml:mrow></mml:mrow></mml:mfrac></mml:mrow><mml:mo>&#x2212;</mml:mo><mml:mi>f</mml:mi><mml:mi>c</mml:mi></mml:mstyle></mml:math><label>(3)</label></disp-formula>The spectral centroid represents the weighted average frequency of the signal&#x0027;s spectrum, reflecting the position of the &#x201C;center of mass&#x201D; of the spectral energy distribution. For the pump flow signal (PF), the spectral centroid in its Fourier-transformed spectrum PF(f) can be used to measure the concentration of frequency components in the signal. The heart rate fc represents the primary frequency component during normal pumping. In the absence of suction, the spectral centroid is close to the heart rate, indicating that the main energy of the signal&#x0027;s spectrum is concentrated around the heart rate. However, during suction, due to abnormal fluctuations in the pump flow, the spectral centroid undergoes a noticeable frequency shift. To quantitatively describe this shift, the difference between the spectral centroid and the heart rate (<xref ref-type="disp-formula" rid="disp-formula3">Equation 3</xref>) (SF3) is used as an indicator of the &#x201C;center of mass&#x201D; displacement. This difference effectively reflects the frequency changes caused by ventricular suction.<disp-formula id="disp-formula4"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM4"><mml:mi>p</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mi>i</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:msup><mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mi>i</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo></mml:mrow><mml:mn>2</mml:mn></mml:msup></mml:mrow><mml:mrow><mml:msubsup><mml:mo movablelimits="false">&#x2211;</mml:mo><mml:mrow><mml:mi>i</mml:mi><mml:mo>=</mml:mo><mml:mn>1</mml:mn></mml:mrow><mml:mi>N</mml:mi></mml:msubsup><mml:mrow><mml:msup><mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mi>i</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:mrow><mml:mo fence="false" stretchy="false">|</mml:mo></mml:mrow><mml:mn>2</mml:mn></mml:msup></mml:mrow></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math><label>(4)</label></disp-formula><disp-formula id="disp-formula5"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM5"><mml:mi>S</mml:mi><mml:mi>F</mml:mi><mml:mn>4</mml:mn><mml:mo>=</mml:mo><mml:mo>&#x2212;</mml:mo><mml:munderover><mml:mo movablelimits="false">&#x2211;</mml:mo><mml:mrow><mml:mi>i</mml:mi><mml:mo>=</mml:mo><mml:mn>1</mml:mn></mml:mrow><mml:mi>N</mml:mi></mml:munderover><mml:mrow><mml:mspace width=".1em"/><mml:mi>p</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mi>i</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo><mml:mi>l</mml:mi><mml:mi>o</mml:mi><mml:mi>g</mml:mi><mml:mn>2</mml:mn></mml:mrow><mml:mi>p</mml:mi><mml:mo stretchy="false">(</mml:mo><mml:mrow><mml:mspace width=".1em"/><mml:mi>f</mml:mi><mml:mi>i</mml:mi></mml:mrow><mml:mo stretchy="false">)</mml:mo></mml:math><label>(5)</label></disp-formula>Spectral entropy (<xref ref-type="disp-formula" rid="disp-formula4">Equations 4</xref>&#x2013;<xref ref-type="disp-formula" rid="disp-formula5">5</xref>) (SF4) is used to measure the complexity of the signal&#x0027;s spectrum, reflecting the uniformity of the energy distribution within the spectrum. P(fi) represents the probability distribution of the pump flow signal at frequency fi. A higher spectral entropy value indicates more complexity in the frequency components of the pump flow signal. The occurrence of a suction event expands the signal&#x0027;s spectrum, increasing the proportion of energy from non-heart-rate components in the flow signal, which in turn leads to an increase in the spectral entropy value.<disp-formula id="disp-formula6"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM6"><mml:mi>Z</mml:mi><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>n</mml:mi><mml:mi>o</mml:mi><mml:mi>r</mml:mi><mml:mi>m</mml:mi><mml:mi>a</mml:mi><mml:mi>l</mml:mi><mml:mi mathvariant="normal">&#x005F;</mml:mi><mml:mi>v</mml:mi><mml:mi>a</mml:mi><mml:mi>l</mml:mi><mml:mi>l</mml:mi><mml:mi>e</mml:mi><mml:mi>y</mml:mi><mml:mo>&#x2212;</mml:mo><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mi mathvariant="normal">&#x005F;</mml:mi><mml:mi>m</mml:mi><mml:mi>i</mml:mi><mml:mi>n</mml:mi></mml:mrow><mml:mrow><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mi mathvariant="normal">&#x005F;</mml:mi><mml:mi>m</mml:mi><mml:mi>a</mml:mi><mml:mi>x</mml:mi><mml:mo>&#x2212;</mml:mo><mml:mi>P</mml:mi><mml:mi>F</mml:mi><mml:mi mathvariant="normal">&#x005F;</mml:mi><mml:mi>m</mml:mi><mml:mi>i</mml:mi><mml:mi>n</mml:mi></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math><label>(6)</label></disp-formula>The decision threshold (<xref ref-type="disp-formula" rid="disp-formula6">Equation 6</xref>) (Z) represents the degree of flow reduction and serves as an intuitive indicator. In the formula, normal_valley represents the flow valley value in the non-suction state, and PF_min represents the minimum pump flow value within the window. When suction has not occurred, normal_valley and PF_min have the same physical meaning and are numerically very close. During suction, PF_min represents the flow value at the suction point.</p>
<p>To scientifically and rationally determine the threshold Z, this paper proposes a universal calculation method based on the VAD pressure-flow (P-Q) curve. The aim is to provide a threshold determination method that is universally applicable to different VADs.</p>
<p>&#x201C;Normal_valley&#x201D; refers to the flow valley value under non-suction conditions, derived from the flow valley pattern observed within a 5-s sliding window. In suction events, most of the valleys correspond to non-suction diastole, which is referred to as &#x201C;Normal_valley&#x201D;. The normal diastolic pressure difference ranges from 50 to 70&#x2005;mmHg, while in suction, it increases by 10&#x2005;mmHg (because the suction alarm must start at a ventricular pressure of 0&#x2005;mmHg rather than complete ventricular collapse, the ventricular pressure during suction is 5&#x2013;10&#x2005;mmHg lower than the typical diastolic ventricular pressure. To reduce the risk of false alarms, the diastolic pressure difference in the suction state is defined to be 10&#x2005;mmHg higher than in the normal non-suction state) (<xref ref-type="bibr" rid="B14">14</xref>). By using P-Q curves from our products and others, such as the Impella 2.5, we can determine values for normal_Valley, PF_min, and PF_max. Sensitivity analysis shows that a <italic>Z</italic>-value of 0.25 offers the optimal Youden Index for our VAD, striking the best balance between sensitivity and specificity. Setting a higher or lower threshold may either miss suction events or trigger false alarms. For other devices, it is strongly recommended that Z be determined based on the specific P-Q curves within the clinically recommended range of rotational speeds.</p>
<p>Suction occurs at high speeds, where the pump generates extreme negative pressure, leading to ventricular collapse. <italic>Z</italic>&#x2009;&#x003D;&#x2009;0.25 is the optimal threshold, balancing sensitivity and specificity. A higher or lower threshold risks missing suction events or causing false alarms.</p>
<p>Since the threshold Z is essentially a time-domain feature, this study takes into account the local minimum characteristic of normal_valley and the noise-induced spike interference when extracting the threshold Z. The mode of the local minimum values within the window is used as the normal valley value. This approach ultimately enables accurate identification of the normal valley value, as shown in <xref ref-type="fig" rid="F3">Figure&#x00A0;3</xref>.</p>
<fig id="F3" position="float"><label>Figure&#x00A0;3</label>
<caption><p>Determination of the normal valley through mode calculation of local Minima.the horizontal axis represents the number of sampling points, which can be understood as a window consisting of 500 sampling points.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g003.tif"><alt-text content-type="machine-generated">Line graph showing pump flow in milliliters per minute over 500 sample points. Red circles indicate local minima, a cyan circle highlights the pump flow minimum, and a magenta circle marks a normal valley.</alt-text>
</graphic>
</fig>
<p>Before training the CART model, the aforementioned features were extracted from both suction and non-suction data. It was found that the four suction features and the threshold Z exhibited a high degree of distinction between suction and non-suction states, as reflected in the numerical values, as shown in <xref ref-type="fig" rid="F4">Figure&#x00A0;4</xref>.</p>
<fig id="F4" position="float"><label>Figure&#x00A0;4</label>
<caption><p>Effectiveness validation of four features and decision threshold Z for suction detection.SF1 represents skewness. After suction occurs, due to the extreme low values, SF1 shifts negatively. SF2 represents kurtosis, which increases significantly during suction. SF3 represents the shift of the spectral centroid relative to the heart rate. After suction, the energy is no longer solely concentrated at the heart rate frequency, and the shift in all the figures is quite apparent. SF4 represents spectral entropy, which increases during suction.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g004.tif"><alt-text content-type="machine-generated">Graph showing five plots. The top plot depicts \"Pump Flow\" in milliliters per minute over 150 seconds, with fluctuations. Below, four subplots labeled SF1 to SF4 and Z, show data variation across a window count of thirty. Each subplot displays unique trend patterns, with spikes and dips at different intervals.</alt-text>
</graphic>
</fig>
</sec>
</sec>
<sec id="s3" sec-type="results"><label>3</label><title>Results</title>
<p>The results section evaluates the performance of the ventricular suction detection method from two perspectives. Section <xref ref-type="sec" rid="s3a">3.1</xref> discusses the performance of the &#x201C;CART model&#x201D; on the test set, while Section <xref ref-type="sec" rid="s3b">3.2</xref> presents the performance of the &#x201C;CART model&#x2009;&#x002B;&#x2009;decision threshold&#x201D; method on the same test set.</p>
<sec id="s3a"><label>3.1</label><title>CART model</title>
<p>In this study, a sliding window of 5&#x2005;s in length with a 1-s step size was used. A total of 2,625 suction state flow data segments and 1,907 non-suction state segments were obtained from the <italic>in-vivo</italic> pump operation data (patient data) of VADs. The data was divided into training and testing sets with a 60&#x0025;/40&#x0025; ratio. Suction features SF1-SF4 were extracted from the patient data, and the feature set was labeled, with non-suction states labeled as &#x201C;1&#x201D; and suction states labeled as &#x201C;&#x2212;1&#x201D;.</p>
<p>To train the classification model, a CART model (<xref ref-type="bibr" rid="B18">18</xref>) was used. The model was configured with a maximum tree depth of 3 and a minimum leaf size of 20. Cost-complexity pruning was applied, with MATLAB automatically selecting the optimal <italic>&#x03B1;</italic> value to balance model complexity and performance. The CART model was trained using the training set, and its accuracy and effectiveness were evaluated using a confusion matrix (<xref ref-type="bibr" rid="B15">15</xref>) and the following five evaluation metrics.</p>
<p>The feature importance scores calculated using the Gini impurity decrease method are also included and presented in descending order, as shown in <xref ref-type="table" rid="T1">Table&#x00A0;1</xref>.<disp-formula id="disp-formula7"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM7"><mml:mi>S</mml:mi><mml:mi>e</mml:mi><mml:mi>n</mml:mi><mml:mi>s</mml:mi><mml:mi>i</mml:mi><mml:mi>t</mml:mi><mml:mi>i</mml:mi><mml:mi>v</mml:mi><mml:mi>i</mml:mi><mml:mi>t</mml:mi><mml:mi>y</mml:mi><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>T</mml:mi><mml:mi>P</mml:mi></mml:mrow><mml:mrow><mml:mi>T</mml:mi><mml:mi>P</mml:mi><mml:mo>+</mml:mo><mml:mi>F</mml:mi><mml:mi>N</mml:mi></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math><label>(7)</label></disp-formula><disp-formula id="disp-formula8"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM8"><mml:mi>S</mml:mi><mml:mi>p</mml:mi><mml:mi>e</mml:mi><mml:mi>c</mml:mi><mml:mi>i</mml:mi><mml:mi>f</mml:mi><mml:mi>i</mml:mi><mml:mi>c</mml:mi><mml:mi>i</mml:mi><mml:mi>t</mml:mi><mml:mi>y</mml:mi><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>T</mml:mi><mml:mi>N</mml:mi></mml:mrow><mml:mrow><mml:mi>T</mml:mi><mml:mi>N</mml:mi><mml:mo>+</mml:mo><mml:mi>F</mml:mi><mml:mi>P</mml:mi></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math><label>(8)</label></disp-formula><disp-formula id="disp-formula9"><mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML" id="DM9"><mml:mi>A</mml:mi><mml:mi>c</mml:mi><mml:mi>c</mml:mi><mml:mi>u</mml:mi><mml:mi>r</mml:mi><mml:mi>a</mml:mi><mml:mi>c</mml:mi><mml:mi>y</mml:mi><mml:mo>=</mml:mo><mml:mstyle displaystyle="true" scriptlevel="0"><mml:mrow><mml:mfrac><mml:mrow><mml:mi>T</mml:mi><mml:mi>P</mml:mi><mml:mo>+</mml:mo><mml:mi>T</mml:mi><mml:mi>N</mml:mi></mml:mrow><mml:mrow><mml:mi>T</mml:mi><mml:mi>P</mml:mi><mml:mo>+</mml:mo><mml:mi>T</mml:mi><mml:mi>N</mml:mi><mml:mo>+</mml:mo><mml:mi>F</mml:mi><mml:mi>P</mml:mi><mml:mo>+</mml:mo><mml:mi>F</mml:mi><mml:mi>N</mml:mi><mml:mtable rowspacing="4pt" columnspacing="1em"><mml:mtr><mml:mtd></mml:mtd></mml:mtr></mml:mtable></mml:mrow></mml:mfrac></mml:mrow></mml:mstyle></mml:math><label>(9)</label></disp-formula></p>
<table-wrap id="T1" position="float"><label>Table&#x00A0;1</label>
<caption><p>Feature importance ranking.</p></caption>
<table>
<colgroup>
<col align="left"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left">Suction features</th>
<th valign="top" align="center">Importance score</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">SF1</td>
<td valign="top" align="center">0.409</td>
</tr>
<tr>
<td valign="top" align="left">SF2</td>
<td valign="top" align="center">0.312</td>
</tr>
<tr>
<td valign="top" align="left">SF3</td>
<td valign="top" align="center">0.125</td>
</tr>
<tr>
<td valign="top" align="left">SF4</td>
<td valign="top" align="center">0.154</td>
</tr>
</tbody>
</table>
</table-wrap>
<p>In this context, TP refers to the samples that truly belong to the positive class and are correctly classified as positive by the model. TN refers to the samples that truly belong to the negative class and are correctly classified as negative by the model. FP refers to the samples that actually belong to the negative class but are incorrectly classified as positive by the model. FN refers to the samples that actually belong to the positive class but are incorrectly classified as negative by the model.</p>
<p><xref ref-type="fig" rid="F5">Figure&#x00A0;5</xref> shows the confusion matrix of the CART model&#x0027;s classification results on the test set, while <xref ref-type="table" rid="T2">Table&#x00A0;2</xref> presents the results of the three evaluation metrics. The CART model achieved a sensitivity (<xref ref-type="disp-formula" rid="disp-formula7">Equation 7</xref>) of 99.81&#x0025; and specificity (<xref ref-type="disp-formula" rid="disp-formula8">Equation 8</xref>) of 98.69&#x0025; on the test set, with an overall accuracy (<xref ref-type="disp-formula" rid="disp-formula9">Equation 9</xref>) of 99.3&#x0025;, indicating that high-confidence suction event detection can be achieved using the four-dimensional flow features SF1-SF4. However, there are still a few misclassifications (FP&#x2009;&#x003D;&#x2009;10, FN&#x2009;&#x003D;&#x2009;2), suggesting that there is room for improvement in the model&#x0027;s discriminative ability.</p>
<fig id="F5" position="float"><label>Figure&#x00A0;5</label>
<caption><p>Confusion matrix of the CART model.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g005.tif"><alt-text content-type="machine-generated">Confusion matrix for a Classification and Regression Tree model. True class -1: 1048 true negatives, 2 false positives. True class 1: 10 false negatives, 753 true positives. Overall accuracies: 99.8% for -1 class, 98.7% for 1 class. Predicted class accuracies: 99.1% negative, 99.7% positive.</alt-text>
</graphic>
</fig>
<table-wrap id="T2" position="float"><label>Table&#x00A0;2</label>
<caption><p>Performance evaluation metrics of the CART model.</p></caption>
<table>
<colgroup>
<col align="left"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left">Evaluation criteria</th>
<th valign="top" align="center">Value</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Sensitivity</td>
<td valign="top" align="center">0.9981</td>
</tr>
<tr>
<td valign="top" align="left">Specificity</td>
<td valign="top" align="center">0.9869</td>
</tr>
<tr>
<td valign="top" align="left">Accuracy</td>
<td valign="top" align="center">0.9934</td>
</tr>
<tr>
<td valign="top" align="left">AUC</td>
<td valign="top" align="center">0.9803</td>
</tr>
<tr>
<td valign="top" align="left">F1 score</td>
<td valign="top" align="center">0.9943</td>
</tr>
</tbody>
</table>
</table-wrap>
</sec>
<sec id="s3b"><label>3.2</label><title>CART model&#x2009;&#x002B;&#x2009;threshold Z</title>
<p>Based on the CART model, the suction detection method with the inclusion of decision threshold Z was applied to the same test set. The evaluation metrics are shown in <xref ref-type="table" rid="T3">Table&#x00A0;3</xref>. Compared to the results using only the CART model, the specificity increased by approximately 1.3&#x0025; and the overall accuracy improved by about 0.55&#x0025; after applying the decision threshold (Z). This indicates that the decision threshold can further enhance the detection capability of suction events.</p>
<table-wrap id="T3" position="float"><label>Table&#x00A0;3</label>
<caption><p>Performance evaluation metrics of the CART model.</p></caption>
<table>
<colgroup>
<col align="left"/>
<col align="center"/>
</colgroup>
<thead>
<tr>
<th valign="top" align="left">Evaluation criteria</th>
<th valign="top" align="center">Value</th>
</tr>
</thead>
<tbody>
<tr>
<td valign="top" align="left">Sensitivity</td>
<td valign="top" align="center">0.9981</td>
</tr>
<tr>
<td valign="top" align="left">Specificity</td>
<td valign="top" align="center">1</td>
</tr>
<tr>
<td valign="top" align="left">Accuracy</td>
<td valign="top" align="center">0.9989</td>
</tr>
</tbody>
</table>
</table-wrap>
</sec>
</sec>
<sec id="s4"><label>4</label><title>Validation</title>
<p>To further validate the real-time performance and accuracy of the algorithm, this study conducted effectiveness verification using both <italic>in vitro</italic> and <italic>in vivo</italic> experimental data.</p>
<sec id="s4a"><label>4.1</label><title><italic>In vitro</italic> experimental results</title>
<p>The <italic>in vitro</italic> experiment used a hydraulic circulation platform to simulate the coupling relationship between the ventricle and the VAD, as shown in <xref ref-type="fig" rid="F6">Figure&#x00A0;6</xref>. In the figure, the cylindrical container is used to simulate the ventricular chamber. On the left side of the ventricular chamber is the pulsatile control device, which simulates the contraction characteristics of a normal heart. On the right side is the VAD model pump, with the pump inlet connected to the ventricular chamber.</p>
<fig id="F6" position="float"><label>Figure&#x00A0;6</label>
<caption><p>Schematic of the <italic>in vitro</italic> experiment.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g006.tif"><alt-text content-type="machine-generated">Diagram showing a volumetric chamber connected to a Pulsatile Generation Device (PGD) on the left and a Ventricular Assist Device (VAD) on the right. The chamber contains liquid, with lines linking all components.</alt-text>
</graphic>
</fig>
<p>Before inducing suction in the experimental setup, we validated the effects of factors such as fluid viscosity, conduit resistance and ventricular pressure on suction phenomena in a blood circulation simulation platform (mock circulation loop). We found that suction was insensitive to these parameters. Additionally, suction induced by preload changes and suction caused by blockage of the pump inlet exhibited the same flow characteristics. Therefore, in the <italic>in vitro</italic> experiment, suction was simulated by blocking the pump inlet (<xref ref-type="bibr" rid="B16">16</xref>). A 5-s sliding window was employed, with a step size of 1&#x2005;s, meaning the window data is updated and features are extracted and classified every second. This rapid window update mechanism and the low computational complexity of the features ensure that the algorithm can instantly detect suction events in real-time.</p>
<p>As shown in <xref ref-type="fig" rid="F7">Figure&#x00A0;7</xref>, in the <italic>in vitro</italic> experimental data of the non-suction state, one window was misclassified as a suction state when only the CART model was used. After the final classification using the threshold Z, it was correctly classified as a non-suction state. In the <italic>in vitro</italic> experimental data of the suction state, the CART model achieved a classification accuracy of 100&#x0025;. After introducing the threshold Z, it enabled the classification of suction severity levels.</p>
<fig id="F7" position="float"><label>Figure&#x00A0;7</label>
<caption><p>Validation of the ventricular suction detection algorithm on <italic>in vitro</italic> experimental data.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g007.tif"><alt-text content-type="machine-generated">Two sets of graphs compare pump flow and analysis results. \n\nTop left: \"Pump Flow(no suction)\" shows a consistent flow rate between 3,500 and 4,500 mL/min over 2,000 seconds. \n\nTop right: \"Pump Flow(suction)\" indicates fluctuating flow rates ranging from 0 to 4,500 mL/min. \n\nMiddle left: \"CART\" remains near zero without significant deviation. \n\nMiddle right: \"CART\" shows a constant value slightly below negative one. \n\nBottom left: \"CART + Threshold Z\" maintains a consistent positive value of one. \n\nBottom right: \"CART + Threshold Z\" displays varied values between negative three and zero with multiple spikes.</alt-text>
</graphic>
</fig>
</sec>
<sec id="s4b"><label>4.2</label><title><italic>In vivo</italic> experimental results</title>
<p>Pump flow data from the non-suction and suction states of the VAD in animal experiments were extracted. The animal experimental data in this study were obtained from experiments conducted on 10 sheep. Although the original purpose of the animal experiment was not to induce suction, factors such as animal blood volume, overly deep insertion of the interventional ventricular assist device, and abnormal positioning of the implanted ventricular assist device, which could induce suction, occurred during the experiment. Therefore, corresponding suction data were extracted from these experiments. Suction detection was performed using both the CART model and the combined algorithm of the CART model with decision threshold. The results are shown in <xref ref-type="fig" rid="F8">Figure&#x00A0;8</xref> (non-suction state) and <xref ref-type="fig" rid="F9">Figure&#x00A0;9</xref> (suction state).</p>
<fig id="F8" position="float"><label>Figure&#x00A0;8</label>
<caption><p>Suction detection algorithm&#x0027;s identification effect on Non-suction data from <italic>in vivo</italic> animal experiments.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g008.tif"><alt-text content-type="machine-generated">Graph showing pump flow over time with fluctuations between 1,000 and 5,000 milliliters per minute. Confusion matrices display classification model performance: top matrix shows true positives as 122, false positives as 0, false negatives as 4, and true negatives as 0; bottom matrix shows true positives as 126, false positives as 0, false negatives as 0, and true negatives as 0. Pie charts illustrate performance: one shows 97% no suction, 3% suction; another shows 100% no suction.</alt-text>
</graphic>
</fig>
<fig id="F9" position="float"><label>Figure&#x00A0;9</label>
<caption><p>Suction detection algorithm&#x0027;s identification effect on suction data from <italic>in vivo</italic> animal experiments.</p></caption>
<graphic mimetype="image" mime-subtype="tiff" xmlns:xlink="http://www.w3.org/1999/xlink" xlink:href="fmedt-07-1748577-g009.tif"><alt-text content-type="machine-generated">Graphical analysis of a pump flow classification model includes: a line chart of pump flow over time, confusion matrices showing model performance before and after threshold adjustments, classification result plots, and pie charts depicting classification performance. The matrices show true positives, false positives, false negatives, and true negatives. Pie charts illustrate a distribution of classifications, with yellow indicating suction and blue no suction, showing predominant suction results.</alt-text>
</graphic>
</fig>
<p>By observing <xref ref-type="fig" rid="F8">Figure&#x00A0;8</xref>, it can be seen that although suction did not occur, the animal&#x0027;s heart rate showed significant changes. These changes in heart rate affect the frequency-domain features, leading to a misclassification as suction. Therefore, when using only the CART model, the accuracy was 92.37&#x0025;. After applying the decision threshold, the accuracy increased to 100&#x0025;. The decision threshold further analyzes the flow variation amplitude based on frequency and data distribution anomalies, allowing misclassifications caused by heart rate irregularities to be reclassified as normal states, ultimately improving the accuracy of suction detection.</p>
<p>In <xref ref-type="fig" rid="F9">Figure&#x00A0;9</xref>, between 10&#x2005;s and 15&#x2005;s, the window exhibits an unstable heart rate and an abnormal flow waveform, which was identified as suction by the CART model. After applying the decision threshold, it was correctly recognized as a non-suction state, with the accuracy improving to 95.24&#x0025;.</p>
</sec>
</sec>
<sec id="s5" sec-type="discussion"><label>5</label><title>Discussion</title>
<p>The ventricular suction detection algorithm presented in this paper has demonstrated high accuracy and real-time performance in the test set, <italic>in vivo</italic> experiments, and <italic>in vitro</italic> experiments.</p>
<p>In the selection of time-domain features, previous studies on suction detection often used flow derivatives or derivative-related indicators to capture sudden changes in the signal. However, derivative operations are highly sensitive to high-frequency noise, which can lead to significant deviations due to small fluctuations in the signal, increasing the risk of misclassification. To improve the stability of the features, this study did not use traditional derivative features. Instead, it innovatively applied two statistical features: kurtosis and skewness. The former reflects the sharpness of the signal waveform, while the latter measures the direction of the signal&#x0027;s distribution shift. Both features effectively detect suction events without relying on instantaneous change rates. This approach not only reduces the impact of noise interference but also enhances sensitivity to abnormal waveform patterns, significantly improving the model&#x0027;s generalization ability and stability in complex data environments.</p>
<p>In the selection of frequency-domain features, the impact of computational complexity on the performance of real-time detection systems was carefully considered. The two frequency-domain features used in this study&#x2014;spectral centroid shift (SF3) and spectral entropy (SF4)&#x2014;are both calculated based on the frequency spectrum energy distribution function constructed by performing a single Fourier transform on the pump flow signal. SF3 calculates the difference between the weighted average frequency of the energy spectrum and the central frequency, with the core operation being a linear weighted summation, resulting in a computational complexity of approximately O(N). SF4 further computes the entropy value of the spectral energy distribution, with its computational complexity also maintaining an O(N) level.</p>
<p>In contrast, frequency-domain features from existing studies are based on frequency-domain integration, calculating energy ratios. While their complexity is also O(N), the integration limits depend on frequency band divisions, indirectly increasing the computational steps. Furthermore, time-frequency domain features from previous studies require performing a short-time Fourier transform on the flow signal. For a 5-s window, a sliding window needs to be set, with each sliding window requiring the calculation of instantaneous frequency followed by variance computation. This results in an overall complexity of approximately O(MNlogN), where M is the number of sliding windows in the 5-s window, and N is the number of sample points in the window. This significantly increases the computational load, making real-time detection more challenging.</p>
<p>Therefore, by selecting low-complexity frequency-domain features while ensuring high suction detection accuracy, this study effectively enhances the computational efficiency of the model and the system&#x0027;s real-time response capability, offering higher engineering practicality.</p>
<p>To validate real-time performance, we also conducted rigorous physical delay testing in our actual experiments. The GPIO flip signal of the MCU was monitored using a logic analyzer. We defined the system delay as the time from the completion of sensor ADC sampling, through the algorithm, to the detection of anomalies and the issuance of the control command. The measurement results showed that the core algorithm&#x0027;s processing delay is only 50&#x2005;&#x03BC;s, while the total monitoring loop delay, including data communication, is approximately 2&#x2005;ms. Considering that the cardiac cycle typically lasts 1,000&#x2005;ms, and the fluid dynamic response caused by suction occurs on the order of tens of milliseconds, the 2-ms system delay fully meets the clinical real-time control requirements.</p>
<p>In addition, to further improve the reliability of the detection method, the decision threshold (Z) was designed, based on the essential characteristics of suction events, to measure the degree of flow reduction. This helps avoid misclassifications caused by various heart rate abnormalities.</p>
<p>In addition to validating the accuracy and real-time performance of the suction detection algorithm, the study also considered the applicability of the algorithm under different physiological conditions, primarily focusing on left ventricular function, right heart failure, hypotension, and pericardial tamponade. To verify the sensitivity of suction phenomena to these parameters, the study used a mock circulation loop to replicate these conditions. Left and right ventricular compliance were simulated by adjusting the motor stroke, and hypotension was simulated by reducing blood volume. Initially, suction was simulated under baseline conditions for the above parameters, and suction waveforms and flow data were recorded. Subsequently, the parameters were altered one by one, and it was found that the characteristics of suction remained unchanged. This indicates that changes in these physiological conditions do not affect the accuracy of suction detection.</p>
<p>On the other hand, considering that different patients have varying body surface areas and that ejection fraction also differs from person to person, this study also investigates the impact of these parameter variations on suction. Differences in body surface area lead to variations in blood flow, which can cause a shift in the average pump flow. However, since the suction detection in this study is based on relative changes to identify anomalies, this baseline shift does not affect the detection accuracy. As for the ejection fraction, it is similar to left ventricular compliance, and as mentioned earlier, suction phenomena are not sensitive to this parameter.</p>
<p>There are still some points that need to be further discussed regarding suction detection: the solutions for suction (<xref ref-type="bibr" rid="B17">17</xref>) and the application of the suction detection algorithm in right ventricular assist devices. Currently, after detecting ventricular suction, it is generally addressed by reducing the pump speed. Regarding the applicability of the suction detection algorithm in right ventricular suction detection in this study, from the perspective of the algorithm&#x0027;s features, the four suction characteristics and the threshold Z are designed based on the sudden decrease in pump flow when suction occurs. When suction happens in the right heart, the pump flow exhibits a similar pattern of change. Since these features analyze the relative changes in pump flow, they can also be applied to the detection of right heart suction. This flow-change-based approach enables the algorithm to identify fluctuations in flow during right heart suction, ensuring its applicability and effectiveness across different parts of the heart.</p>
</sec>
<sec id="s6" sec-type="conclusions"><label>6</label><title>Conclusions</title>
<p>Based on the time-domain and frequency-domain characteristics of pump flow during ventricular suction, this paper proposes an algorithm combining the CART model with decision threshold, and validates the algorithm through both <italic>in vitro</italic> and <italic>in vivo</italic> experiments. Innovatively, skewness, kurtosis, spectral entropy, and spectral centroid shift are selected as features to construct the CART model, with the flow reduction amplitude used as the decision threshold. This approach enhances the real-time performance, accuracy, and engineering applicability of ventricular suction detection, providing useful references for VAD product design.</p>
</sec>
</body>
<back>
<sec id="s8" sec-type="data-availability"><title>Data availability statement</title>
<p>The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.</p>
</sec>
<sec id="s9" sec-type="ethics-statement"><title>Ethics statement</title>
<p>The animal study was approved by Silver Snake (Guangzhou) Medical Technology Co. The study was conducted in accordance with the local legislation and institutional requirements.</p>
</sec>
<sec id="s10" sec-type="author-contributions"><title>Author contributions</title>
<p>YW: Writing &#x2013; original draft. YY: Writing &#x2013; review &#x0026; editing. XP: Writing &#x2013; review &#x0026; editing. SY: Writing &#x2013; review &#x0026; editing.</p>
</sec>
<sec id="s12" sec-type="COI-statement"><title>Conflict of interest</title>
<p>YW, YY and SY were employed by ShenZhen Core Technology Co., Ltd.</p>
<p>The remaining author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.</p>
</sec>
<sec id="s13" sec-type="ai-statement"><title>Generative AI statement</title>
<p>The author(s) declared that generative AI was used in the creation of this manuscript. During the preparation of this manuscript/study, the author(s) used ChatGpt for the purposes of translating the content. The authors have reviewed and edited the output and take full responsibility for the content of this publication.</p>
<p>Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.</p>
</sec>
<sec id="s14" sec-type="disclaimer"><title>Publisher&#x0027;s note</title>
<p>All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.</p>
</sec>
<ref-list><title>References</title>
<ref id="B1"><label>1.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Meki</surname> <given-names>M</given-names></name> <name><surname>Wang</surname> <given-names>Y</given-names></name> <name><surname>Sethu</surname> <given-names>P</given-names></name> <name><surname>Ghazal</surname> <given-names>M</given-names></name> <name><surname>El-Baz</surname> <given-names>A</given-names></name> <name><surname>Giridharan</surname> <given-names>G</given-names></name></person-group>. <article-title>A sensorless rotational speed-based control system for continuous flow left ventricular assist devices</article-title>. <source>IEEE Trans Biomed Eng</source>. (<year>2020</year>) <volume>67</volume>(<issue>4</issue>):<fpage>1050</fpage>&#x2013;<lpage>60</lpage>. <pub-id pub-id-type="doi">10.1109/TBME.2019.2928826</pub-id><pub-id pub-id-type="pmid">31329102</pub-id></mixed-citation></ref>
<ref id="B2"><label>2.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Takatani</surname> <given-names>S</given-names></name></person-group>. <article-title>Progress of rotary blood pumps</article-title>. <source>Artif Organs</source>. (<year>2006</year>) <volume>30</volume>(<issue>5</issue>):<fpage>317</fpage>&#x2013;<lpage>21</lpage>. <pub-id pub-id-type="doi">10.1111/j.1525-1594.2006.00220.x</pub-id><pub-id pub-id-type="pmid">16683947</pub-id></mixed-citation></ref>
<ref id="B3"><label>3.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Hoit</surname> <given-names>BD</given-names></name></person-group>. <article-title>Pericardial effusion and cardiac tamponade pathophysiology and new approaches to treatment</article-title>. <source>Curr Cardiol Rep</source>. (<year>2023</year>) <volume>25</volume>(<issue>9</issue>):<fpage>1003</fpage>&#x2013;<lpage>14</lpage>. <pub-id pub-id-type="doi">10.1007/s11886-023-01920-8</pub-id><pub-id pub-id-type="pmid">37515704</pub-id></mixed-citation></ref>
<ref id="B4"><label>4.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Reesink</surname> <given-names>K</given-names></name> <name><surname>Dekker</surname> <given-names>A</given-names></name> <name><surname>Van der Nagel</surname> <given-names>T</given-names></name> <name><surname>Beghi</surname> <given-names>C</given-names></name> <name><surname>Leonardi</surname> <given-names>F</given-names></name> <name><surname>Botti</surname> <given-names>P</given-names></name><etal/></person-group> <article-title>Suction due to left ventricular assist: implications for device control and management</article-title>. <source>Artif Organs</source>. (<year>2007</year>) <volume>31</volume>:<fpage>542</fpage>&#x2013;<lpage>9</lpage>. <pub-id pub-id-type="doi">10.1111/j.1525-1594.2007.00420.x</pub-id><pub-id pub-id-type="pmid">17584479</pub-id></mixed-citation></ref>
<ref id="B5"><label>5.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Gross</surname> <given-names>C</given-names></name> <name><surname>Schima</surname> <given-names>H</given-names></name> <name><surname>Schl&#x00F6;glhofer</surname> <given-names>T</given-names></name> <name><surname>Dimitrov</surname> <given-names>K</given-names></name> <name><surname>Maw</surname> <given-names>M</given-names></name> <name><surname>Riebandt</surname> <given-names>J</given-names></name><etal/></person-group> <article-title>Continuous LVAD monitoring reveals high suction rates in clinically stable outpatients</article-title>. <source>Artif Organs</source>. (<year>2020</year>) <volume>44</volume>(<issue>7</issue>):<fpage>E251</fpage>&#x2013;<lpage>62</lpage>. <pub-id pub-id-type="doi">10.1111/aor.13638</pub-id><pub-id pub-id-type="pmid">31945201</pub-id></mixed-citation></ref>
<ref id="B6"><label>6.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Gregory</surname> <given-names>SD</given-names></name> <name><surname>Stevens</surname> <given-names>MC</given-names></name> <name><surname>Pauls</surname> <given-names>JP</given-names></name> <name><surname>Schummy</surname> <given-names>E</given-names></name> <name><surname>Diab</surname> <given-names>S</given-names></name> <name><surname>Thomson</surname> <given-names>B</given-names></name><etal/></person-group> <article-title><italic>In Vivo</italic> evaluation of physiological control systems</article-title>. <source>Artif Organs</source>. (<year>2016</year>) <volume>40</volume>:<fpage>894</fpage>&#x2013;<lpage>903</lpage>. <pub-id pub-id-type="doi">10.1111/aor.12654</pub-id><pub-id pub-id-type="pmid">26748566</pub-id></mixed-citation></ref>
<ref id="B7"><label>7.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Santos</surname> <given-names>BJ</given-names></name> <name><surname>Cestari</surname> <given-names>IA</given-names></name></person-group>. <article-title>Intelligent reflux and suction detection system for ventricular assist devices: in silico study</article-title>. <source>Res Biomed Eng</source>. (<year>2025</year>) <volume>41</volume>(<issue>1</issue>):<fpage>1</fpage>&#x2013;<lpage>23</lpage>. <pub-id pub-id-type="doi">10.1007/s42600-024-00393-0</pub-id></mixed-citation></ref>
<ref id="B8"><label>8.</label><mixed-citation publication-type="book"><person-group person-group-type="author"><name><surname>Ferreira</surname> <given-names>A</given-names></name> <name><surname>Chen</surname> <given-names>S</given-names></name> <name><surname>Simaan</surname> <given-names>MA</given-names></name> <name><surname>Boston</surname> <given-names>JR</given-names></name> <name><surname>Antaki</surname> <given-names>JF</given-names></name></person-group>. <article-title>A discriminant-analysis-based suction detection system for rotary blood pumps</article-title>. In: <source>2006 International Conference of the IEEE Engineering in Medicine and Biology Society</source>. <publisher-loc>New York, NY, USA</publisher-loc> (<year>2006</year>). p. <fpage>5382</fpage>&#x2013;<lpage>5</lpage>. <pub-id pub-id-type="doi">10.1109/IEMBS.2006.260621</pub-id></mixed-citation></ref>
<ref id="B9"><label>9.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Wang</surname> <given-names>Y</given-names></name> <name><surname>Simaan</surname> <given-names>MA</given-names></name></person-group>. <article-title>A suction detection system for rotary blood pumps based on the Lagrangian support vector machine algorithm</article-title>. <source>IEEE J Biomed Health Inform</source>. (<year>2013</year>) <volume>17</volume>(<issue>3</issue>):<fpage>654</fpage>&#x2013;<lpage>63</lpage>. <pub-id pub-id-type="doi">10.1109/TITB.2012.2228877</pub-id><pub-id pub-id-type="pmid">23192602</pub-id></mixed-citation></ref>
<ref id="B10"><label>10.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Liang</surname> <given-names>L</given-names></name> <name><surname>Meki</surname> <given-names>M</given-names></name> <name><surname>Wang</surname> <given-names>W</given-names></name> <name><surname>Sethu</surname> <given-names>P</given-names></name> <name><surname>El-Baz</surname> <given-names>A</given-names></name> <name><surname>Giridharan</surname> <given-names>GA</given-names></name><etal/></person-group> <article-title>A suction index based control system for rotary blood pumps</article-title>. <source>Biomed Signal Process Control</source>. (<year>2020</year>) <volume>62</volume>:<fpage>102057</fpage>. <pub-id pub-id-type="doi">10.1016/j.bspc.2020.102057</pub-id></mixed-citation></ref>
<ref id="B11"><label>11.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Maw</surname> <given-names>M</given-names></name> <name><surname>Gross</surname> <given-names>C</given-names></name> <name><surname>Schl&#x00F6;glhofer</surname> <given-names>T</given-names></name> <name><surname>Dimitrov</surname> <given-names>K</given-names></name> <name><surname>Zimpfer</surname> <given-names>D</given-names></name> <name><surname>Moscato</surname> <given-names>F</given-names></name><etal/></person-group> <article-title>Development of suction detection algorithms for a left ventricular assist device from patient data</article-title>. <source>Biomed Signal Process Control</source>. (<year>2021</year>) <volume>69</volume>:<fpage>102910</fpage>. <pub-id pub-id-type="doi">10.1016/j.bspc.2021.102910</pub-id></mixed-citation></ref>
<ref id="B12"><label>12.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Ng</surname> <given-names>S-C</given-names></name> <name><surname>Lim</surname> <given-names>E</given-names></name> <name><surname>Mason</surname> <given-names>DG</given-names></name> <name><surname>Avolio</surname> <given-names>AP</given-names></name> <name><surname>Lovell</surname> <given-names>NH</given-names></name></person-group>. <article-title>Evaluation of suction detection</article-title>. <source>Artif Organs</source>. (<year>2013</year>) <volume>37</volume>:<fpage>E145</fpage>&#x2013;<lpage>54</lpage>. <pub-id pub-id-type="doi">10.1111/aor.12079</pub-id><pub-id pub-id-type="pmid">23635073</pub-id></mixed-citation></ref>
<ref id="B13"><label>13.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Fetanat</surname> <given-names>M</given-names></name> <name><surname>Stevens</surname> <given-names>M</given-names></name> <name><surname>Hayward</surname> <given-names>C</given-names></name> <name><surname>Lovell</surname> <given-names>NH</given-names></name></person-group>. <article-title>A sensorless control system for an implantable heart pump using a real-time deep convolutional neural network</article-title>. <source>IEEE Transactions on Biomedical Engineering</source>. (<year>2021</year>) <volume>68</volume>(<issue>10</issue>):<fpage>3029</fpage>&#x2013;<lpage>38</lpage>. <pub-id pub-id-type="doi">10.1109/TBME.2021.3061405</pub-id><pub-id pub-id-type="pmid">33621164</pub-id></mixed-citation></ref>
<ref id="B14"><label>14.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Zein</surname> <given-names>R</given-names></name> <name><surname>Patel</surname> <given-names>C</given-names></name> <name><surname>Mercado-Alamo</surname> <given-names>A</given-names></name> <name><surname>Schreiber</surname> <given-names>T</given-names></name> <name><surname>Kaki</surname> <given-names>A</given-names></name></person-group>. <article-title>A review of the impella devices</article-title>. <source>Interv Cardiol</source>. (<year>2022</year>) <volume>17</volume>:<fpage>e05</fpage>. <pub-id pub-id-type="doi">10.15420/icr.2021.11</pub-id><pub-id pub-id-type="pmid">35474971</pub-id></mixed-citation></ref>
<ref id="B15"><label>15.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Visa</surname> <given-names>S</given-names></name> <name><surname>Ramsay</surname> <given-names>B</given-names></name> <name><surname>Ralescu</surname> <given-names>AL</given-names></name> <name><surname>Van Der Knaap</surname> <given-names>E</given-names></name></person-group>. <article-title>Confusion matrix-based feature selection</article-title>. <source>Maics</source>. (<year>2011</year>) <volume>710</volume>(<issue>1</issue>):<fpage>120</fpage>&#x2013;<lpage>7</lpage>.</mixed-citation></ref>
<ref id="B16"><label>16.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Rocchi</surname> <given-names>M</given-names></name> <name><surname>Gross</surname> <given-names>C</given-names></name> <name><surname>Moscato</surname> <given-names>F</given-names></name> <name><surname>Schl&#x00F6;glhofer</surname> <given-names>T</given-names></name> <name><surname>Meyns</surname> <given-names>B</given-names></name> <name><surname>Fresiello</surname> <given-names>L</given-names></name></person-group>. <article-title>An <italic>in vitro</italic> model to study suction events by a ventricular assist device: validation with clinical data</article-title>. <source>Front. Physiol</source>. (<year>2023</year>) <volume>14</volume>:<fpage>1155032</fpage>. <pub-id pub-id-type="doi">10.3389/fphys.2023.1155032</pub-id><pub-id pub-id-type="pmid">37560156</pub-id></mixed-citation></ref>
<ref id="B17"><label>17.</label><mixed-citation publication-type="journal"><person-group person-group-type="author"><name><surname>Maw</surname> <given-names>M</given-names></name> <name><surname>Schl&#x00F6;glhofer</surname> <given-names>T</given-names></name> <name><surname>Marko</surname> <given-names>C</given-names></name> <name><surname>Aigner</surname> <given-names>P</given-names></name> <name><surname>Gross</surname> <given-names>C</given-names></name> <name><surname>Widhalm</surname> <given-names>G</given-names></name><etal/></person-group> <article-title>A sensorless modular multiobjective control algorithm for left ventricular assist devices: a clinical pilot study</article-title>. <source>Front. Cardiovasc. Med</source>. (<year>2022</year>) <volume>9</volume>:<fpage>888269</fpage>. <pub-id pub-id-type="doi">10.3389/fcvm.2022.888269</pub-id><pub-id pub-id-type="pmid">35548436</pub-id></mixed-citation></ref>
<ref id="B18"><label>18.</label><mixed-citation publication-type="book"><person-group person-group-type="author"><name><surname>Breiman</surname> <given-names>L</given-names></name> <name><surname>Friedman</surname> <given-names>J</given-names></name> <name><surname>Olshen</surname> <given-names>RA</given-names></name> <name><surname>Stone</surname> <given-names>CJ</given-names></name></person-group>. <source>Classification and Regression Trees</source>. Boca Raton, FL: <publisher-name>Chapman and Hall/CRC</publisher-name> (<year>2017</year>).</mixed-citation></ref></ref-list>
<fn-group>
<fn id="n1" fn-type="custom" custom-type="edited-by"><p>Edited by: <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/996909/overview">Kunal Pal</ext-link>, National Institute of Technology Rourkela, India</p></fn>
<fn id="n2" fn-type="custom" custom-type="reviewed-by"><p>Reviewed by: <ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/1399562/overview">Tahir Yagdi</ext-link>, EGE University, T&#x00FC;rkiye</p>
<p><ext-link ext-link-type="uri" xlink:href="https://loop.frontiersin.org/people/3290450/overview">Mohamed Bounouib</ext-link>, &#x00C9;cole Nationale Sup&#x00E9;ieure d&#x0027;Arts et M&#x00E9;tiers Rabat, Morocco</p></fn>
</fn-group>
</back>
</article>